SLAC Magnetic Measurements Date: 01-07-2025 Time: 09:21:29 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_10086 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 3 Measurement Device: On Axis Operator: YL Comment: 0.25mm to 0.375mm 09:21:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:21:39 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:21:40 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:21:42 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:21:43 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:22:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:22:17 Integrated Voltage Measurement... 09:22:26 Voltage Offset Correction: v_off_ini = -66.719 uV, v_off_fin = -83.773 uV 09:22:26 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -110.271 uVs 09:22:26 Integrated Voltage Measurement... 09:22:35 Voltage Offset Correction: v_off_ini = -73.065 uV, v_off_fin = -77.915 uV 09:22:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.329 uVs 09:22:35 Integrated Voltage Measurement... 09:22:44 Voltage Offset Correction: v_off_ini = -51.999 uV, v_off_fin = -44.544 uV 09:22:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.025 uVs 09:22:44 Integrated Voltage Measurement... 09:22:53 Voltage Offset Correction: v_off_ini = -53.224 uV, v_off_fin = -58.377 uV 09:22:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.569 uVs 09:22:53 Integrated Voltage Measurement... 09:23:02 Voltage Offset Correction: v_off_ini = -56.655 uV, v_off_fin = -61.792 uV 09:23:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.442 uVs 09:23:02 Integrated Voltage Measurement... 09:23:11 Voltage Offset Correction: v_off_ini = -58.620 uV, v_off_fin = -46.792 uV 09:23:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.723 uVs 09:23:15 I1X Average and RMS of Measurements VT = -89.393 +- 3.968 uVs I1X = 17.880 +- 0.795 uTm 09:23:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:23:22 Integrated Voltage Measurement... 09:23:31 Voltage Offset Correction: v_off_ini = -31.329 uV, v_off_fin = -45.546 uV 09:23:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -166.348 uVs 09:23:31 Integrated Voltage Measurement... 09:23:40 Voltage Offset Correction: v_off_ini = -33.324 uV, v_off_fin = -27.515 uV 09:23:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 180.857 uVs 09:23:40 Integrated Voltage Measurement... 09:23:49 Voltage Offset Correction: v_off_ini = -33.191 uV, v_off_fin = -28.161 uV 09:23:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -152.817 uVs 09:23:49 Integrated Voltage Measurement... 09:23:58 Voltage Offset Correction: v_off_ini = -23.171 uV, v_off_fin = -20.288 uV 09:23:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.224 uVs 09:23:58 Integrated Voltage Measurement... 09:24:07 Voltage Offset Correction: v_off_ini = -8.878 uV, v_off_fin = -4.348 uV 09:24:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -166.849 uVs 09:24:07 Integrated Voltage Measurement... 09:24:16 Voltage Offset Correction: v_off_ini = -20.893 uV, v_off_fin = -2.373 uV 09:24:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.543 uVs 09:24:21 I1Y Average and RMS of Measurements VT = -167.440 +- 6.163 uVs I1Y = -33.489 +- 1.233 uTm 09:24:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:24:28 Integrated Voltage Measurement... 09:24:37 Voltage Offset Correction: v_off_ini = -24.590 uV, v_off_fin = -26.258 uV 09:24:37 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -73.351 uVs 09:24:37 Integrated Voltage Measurement... 09:24:46 Voltage Offset Correction: v_off_ini = -10.910 uV, v_off_fin = 7.686 uV 09:24:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.561 uVs 09:24:46 Integrated Voltage Measurement... 09:24:55 Voltage Offset Correction: v_off_ini = -6.522 uV, v_off_fin = -17.752 uV 09:24:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.976 uVs 09:24:55 Integrated Voltage Measurement... 09:25:04 Voltage Offset Correction: v_off_ini = 7.010 uV, v_off_fin = 10.302 uV 09:25:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.813 uVs 09:25:04 Integrated Voltage Measurement... 09:25:13 Voltage Offset Correction: v_off_ini = -4.912 uV, v_off_fin = -4.859 uV 09:25:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.697 uVs 09:25:13 Integrated Voltage Measurement... 09:25:22 Voltage Offset Correction: v_off_ini = 17.335 uV, v_off_fin = 33.007 uV 09:25:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.587 uVs 09:25:25 I2X Average and RMS of Measurements VT = -89.997 +- 9.349 uVs I2X = 8.280 +- 0.860 uTm^2