SLAC Magnetic Measurements Date: 01-07-2025 Time: 09:00:10 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_10086 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 2 Measurement Device: On Axis Operator: YL Comment: Shim 0.5mm replaced by shim 0.25mm 09:01:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:01:11 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:01:13 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:01:14 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:01:16 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:02:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:02:07 Integrated Voltage Measurement... 09:02:16 Voltage Offset Correction: v_off_ini = -97.810 uV, v_off_fin = -94.635 uV 09:02:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -106.675 uVs 09:02:16 Integrated Voltage Measurement... 09:02:25 Voltage Offset Correction: v_off_ini = -97.338 uV, v_off_fin = -98.638 uV 09:02:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.228 uVs 09:02:25 Integrated Voltage Measurement... 09:02:34 Voltage Offset Correction: v_off_ini = -109.092 uV, v_off_fin = -96.653 uV 09:02:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -80.222 uVs 09:02:34 Integrated Voltage Measurement... 09:02:43 Voltage Offset Correction: v_off_ini = -98.060 uV, v_off_fin = -96.658 uV 09:02:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.467 uVs 09:02:43 Integrated Voltage Measurement... 09:02:52 Voltage Offset Correction: v_off_ini = -103.841 uV, v_off_fin = -89.188 uV 09:02:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -84.943 uVs 09:02:52 Integrated Voltage Measurement... 09:03:01 Voltage Offset Correction: v_off_ini = -73.656 uV, v_off_fin = -82.649 uV 09:03:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.076 uVs 09:03:05 I1X Average and RMS of Measurements VT = -82.435 +- 10.414 uVs I1X = 16.488 +- 2.083 uTm 09:03:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:03:12 Integrated Voltage Measurement... 09:03:21 Voltage Offset Correction: v_off_ini = -84.140 uV, v_off_fin = -102.949 uV 09:03:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -55.228 uVs 09:03:21 Integrated Voltage Measurement... 09:03:30 Voltage Offset Correction: v_off_ini = -73.770 uV, v_off_fin = -55.817 uV 09:03:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 23.747 uVs 09:03:30 Integrated Voltage Measurement... 09:03:39 Voltage Offset Correction: v_off_ini = -71.310 uV, v_off_fin = -81.862 uV 09:03:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -21.292 uVs 09:03:39 Integrated Voltage Measurement... 09:03:48 Voltage Offset Correction: v_off_ini = -92.755 uV, v_off_fin = -89.630 uV 09:03:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 24.289 uVs 09:03:48 Integrated Voltage Measurement... 09:03:57 Voltage Offset Correction: v_off_ini = -75.640 uV, v_off_fin = -86.543 uV 09:03:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -37.031 uVs 09:03:57 Integrated Voltage Measurement... 09:04:06 Voltage Offset Correction: v_off_ini = -98.743 uV, v_off_fin = -93.011 uV 09:04:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 28.284 uVs 09:04:10 I1Y Average and RMS of Measurements VT = -31.645 +- 6.854 uVs I1Y = -6.329 +- 1.371 uTm 09:04:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:04:18 Integrated Voltage Measurement... 09:04:27 Voltage Offset Correction: v_off_ini = -101.779 uV, v_off_fin = -112.780 uV 09:04:27 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -80.836 uVs 09:04:27 Integrated Voltage Measurement... 09:04:36 Voltage Offset Correction: v_off_ini = -91.585 uV, v_off_fin = -91.041 uV 09:04:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.243 uVs 09:04:36 Integrated Voltage Measurement... 09:04:45 Voltage Offset Correction: v_off_ini = -71.302 uV, v_off_fin = -99.545 uV 09:04:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.205 uVs 09:04:45 Integrated Voltage Measurement... 09:04:54 Voltage Offset Correction: v_off_ini = -87.749 uV, v_off_fin = -93.229 uV 09:04:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.184 uVs 09:04:54 Integrated Voltage Measurement... 09:05:03 Voltage Offset Correction: v_off_ini = -89.056 uV, v_off_fin = -106.500 uV 09:05:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.728 uVs 09:05:03 Integrated Voltage Measurement... 09:05:12 Voltage Offset Correction: v_off_ini = -78.352 uV, v_off_fin = -83.617 uV 09:05:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.591 uVs 09:05:14 I2X Average and RMS of Measurements VT = -81.131 +- 4.760 uVs I2X = 7.464 +- 0.438 uTm^2 09:05:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:05:22 Integrated Voltage Measurement... 09:05:31 Voltage Offset Correction: v_off_ini = -69.451 uV, v_off_fin = -83.209 uV 09:05:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -104.122 uVs 09:05:31 Integrated Voltage Measurement... 09:05:40 Voltage Offset Correction: v_off_ini = -69.467 uV, v_off_fin = -86.930 uV 09:05:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 83.889 uVs 09:05:40 Integrated Voltage Measurement... 09:05:49 Voltage Offset Correction: v_off_ini = -87.425 uV, v_off_fin = -103.650 uV 09:05:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -92.117 uVs 09:05:49 Integrated Voltage Measurement... 09:05:58 Voltage Offset Correction: v_off_ini = -110.836 uV, v_off_fin = -90.336 uV 09:05:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 87.100 uVs 09:05:58 Integrated Voltage Measurement... 09:06:07 Voltage Offset Correction: v_off_ini = -77.043 uV, v_off_fin = -42.775 uV 09:06:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -87.811 uVs 09:06:07 Integrated Voltage Measurement... 09:06:16 Voltage Offset Correction: v_off_ini = -66.011 uV, v_off_fin = -79.481 uV 09:06:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 102.514 uVs 09:06:19 I2Y Average and RMS of Measurements VT = -92.926 +- 2.393 uVs I2Y = -8.549 +- 0.220 uTm^2