SLAC Magnetic Measurements Date: 08-31-2026 Time: 08:56:59 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_15081 Dataset Number: 6 Run Purpose: Background Run Type: Run Number: 1 Measurement Device: Stretched Wire Operator: YL Comment: Final; after last PS 08:57:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:58:00 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:58:01 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:58:03 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:58:04 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:58:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:58:53 Integrated Voltage Measurement... 08:59:02 Voltage Offset Correction: v_off_ini = 2224.449 uV, v_off_fin = 2219.677 uV 08:59:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -81.662 uVs 08:59:02 Integrated Voltage Measurement... 08:59:11 Voltage Offset Correction: v_off_ini = 2215.591 uV, v_off_fin = 2218.082 uV 08:59:11 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.802 uVs 08:59:11 Integrated Voltage Measurement... 08:59:20 Voltage Offset Correction: v_off_ini = 2219.578 uV, v_off_fin = 2222.002 uV 08:59:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.065 uVs 08:59:20 Integrated Voltage Measurement... 08:59:29 Voltage Offset Correction: v_off_ini = 2226.362 uV, v_off_fin = 2217.748 uV 08:59:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.359 uVs 08:59:29 Integrated Voltage Measurement... 08:59:38 Voltage Offset Correction: v_off_ini = 2232.391 uV, v_off_fin = 2219.436 uV 08:59:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.303 uVs 08:59:38 Integrated Voltage Measurement... 08:59:47 Voltage Offset Correction: v_off_ini = 2201.344 uV, v_off_fin = 2213.171 uV 08:59:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.787 uVs 08:59:51 I1X Average and RMS of Measurements VT = -90.996 +- 10.708 uVs I1X = 18.200 +- 2.141 uTm 08:59:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:59:58 Integrated Voltage Measurement... 09:00:07 Voltage Offset Correction: v_off_ini = 2181.116 uV, v_off_fin = 2177.924 uV 09:00:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -228.125 uVs 09:00:07 Integrated Voltage Measurement... 09:00:16 Voltage Offset Correction: v_off_ini = 2195.154 uV, v_off_fin = 2206.255 uV 09:00:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 227.737 uVs 09:00:16 Integrated Voltage Measurement... 09:00:25 Voltage Offset Correction: v_off_ini = 2214.063 uV, v_off_fin = 2192.695 uV 09:00:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.185 uVs 09:00:25 Integrated Voltage Measurement... 09:00:34 Voltage Offset Correction: v_off_ini = 2227.846 uV, v_off_fin = 2213.280 uV 09:00:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 213.818 uVs 09:00:34 Integrated Voltage Measurement... 09:00:42 Voltage Offset Correction: v_off_ini = 2226.599 uV, v_off_fin = 2214.603 uV 09:00:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -227.086 uVs 09:00:43 Integrated Voltage Measurement... 09:00:51 Voltage Offset Correction: v_off_ini = 2223.368 uV, v_off_fin = 2214.700 uV 09:00:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.748 uVs 09:00:56 I1Y Average and RMS of Measurements VT = -221.616 +- 4.480 uVs I1Y = -44.323 +- 0.896 uTm 09:01:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:01:03 Integrated Voltage Measurement... 09:01:12 Voltage Offset Correction: v_off_ini = 2192.610 uV, v_off_fin = 2198.012 uV 09:01:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -73.217 uVs 09:01:12 Integrated Voltage Measurement... 09:01:21 Voltage Offset Correction: v_off_ini = 2206.924 uV, v_off_fin = 2212.464 uV 09:01:21 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 89.975 uVs 09:01:21 Integrated Voltage Measurement... 09:01:30 Voltage Offset Correction: v_off_ini = 2196.938 uV, v_off_fin = 2172.249 uV 09:01:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.856 uVs 09:01:30 Integrated Voltage Measurement... 09:01:39 Voltage Offset Correction: v_off_ini = 2175.569 uV, v_off_fin = 2181.742 uV 09:01:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.331 uVs 09:01:39 Integrated Voltage Measurement... 09:01:48 Voltage Offset Correction: v_off_ini = 2172.261 uV, v_off_fin = 2180.346 uV 09:01:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.730 uVs 09:01:48 Integrated Voltage Measurement... 09:01:57 Voltage Offset Correction: v_off_ini = 2180.409 uV, v_off_fin = 2192.502 uV 09:01:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.009 uVs 09:02:00 I2X Average and RMS of Measurements VT = -76.186 +- 7.649 uVs I2X = 7.009 +- 0.704 uTm^2 09:02:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:02:08 Integrated Voltage Measurement... 09:02:16 Voltage Offset Correction: v_off_ini = 2172.483 uV, v_off_fin = 2183.736 uV 09:02:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.428 uVs 09:02:17 Integrated Voltage Measurement... 09:02:26 Voltage Offset Correction: v_off_ini = 2177.749 uV, v_off_fin = 2185.986 uV 09:02:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.262 uVs 09:02:26 Integrated Voltage Measurement... 09:02:35 Voltage Offset Correction: v_off_ini = 2192.726 uV, v_off_fin = 2172.453 uV 09:02:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.216 uVs 09:02:35 Integrated Voltage Measurement... 09:02:44 Voltage Offset Correction: v_off_ini = 2162.231 uV, v_off_fin = 2158.225 uV 09:02:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.409 uVs 09:02:44 Integrated Voltage Measurement... 09:02:53 Voltage Offset Correction: v_off_ini = 2158.726 uV, v_off_fin = 2159.126 uV 09:02:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -221.608 uVs 09:02:53 Integrated Voltage Measurement... 09:03:02 Voltage Offset Correction: v_off_ini = 2151.485 uV, v_off_fin = 2147.039 uV 09:03:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.107 uVs 09:03:04 I2Y Average and RMS of Measurements VT = -216.505 +- 4.533 uVs I2Y = -19.918 +- 0.417 uTm^2 09:03:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:03:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:03:34 Integrated Voltage Measurement... 09:03:43 Voltage Offset Correction: v_off_ini = 2160.500 uV, v_off_fin = 2149.659 uV 09:03:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -89.594 uVs 09:03:43 Integrated Voltage Measurement... 09:03:52 Voltage Offset Correction: v_off_ini = 2141.306 uV, v_off_fin = 2127.245 uV 09:03:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.456 uVs 09:03:52 Integrated Voltage Measurement... 09:04:01 Voltage Offset Correction: v_off_ini = 2113.009 uV, v_off_fin = 2124.316 uV 09:04:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -77.962 uVs 09:04:01 Integrated Voltage Measurement... 09:04:10 Voltage Offset Correction: v_off_ini = 2154.801 uV, v_off_fin = 2139.623 uV 09:04:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.785 uVs 09:04:10 Integrated Voltage Measurement... 09:04:19 Voltage Offset Correction: v_off_ini = 2140.745 uV, v_off_fin = 2133.728 uV 09:04:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.492 uVs 09:04:19 Integrated Voltage Measurement... 09:04:28 Voltage Offset Correction: v_off_ini = 2129.289 uV, v_off_fin = 2166.637 uV 09:04:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.145 uVs 09:04:32 I1X Average and RMS of Measurements VT = -86.072 +- 1.520 uVs I1X = 17.215 +- 0.305 uTm 09:04:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:04:39 Integrated Voltage Measurement... 09:04:48 Voltage Offset Correction: v_off_ini = 2160.988 uV, v_off_fin = 2155.239 uV 09:04:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -226.750 uVs 09:04:48 Integrated Voltage Measurement... 09:04:57 Voltage Offset Correction: v_off_ini = 2178.720 uV, v_off_fin = 2167.384 uV 09:04:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 206.104 uVs 09:04:57 Integrated Voltage Measurement... 09:05:06 Voltage Offset Correction: v_off_ini = 2159.095 uV, v_off_fin = 2159.378 uV 09:05:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.192 uVs 09:05:06 Integrated Voltage Measurement... 09:05:15 Voltage Offset Correction: v_off_ini = 2163.884 uV, v_off_fin = 2179.075 uV 09:05:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.352 uVs 09:05:15 Integrated Voltage Measurement... 09:05:24 Voltage Offset Correction: v_off_ini = 2150.671 uV, v_off_fin = 2138.199 uV 09:05:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.785 uVs 09:05:24 Integrated Voltage Measurement... 09:05:33 Voltage Offset Correction: v_off_ini = 2150.203 uV, v_off_fin = 2153.687 uV 09:05:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.361 uVs 09:05:37 I1Y Average and RMS of Measurements VT = -213.757 +- 5.749 uVs I1Y = -42.751 +- 1.150 uTm 09:05:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:05:45 Integrated Voltage Measurement... 09:05:53 Voltage Offset Correction: v_off_ini = 2148.928 uV, v_off_fin = 2155.429 uV 09:05:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.616 uVs 09:05:54 Integrated Voltage Measurement... 09:06:03 Voltage Offset Correction: v_off_ini = 2152.001 uV, v_off_fin = 2168.525 uV 09:06:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.115 uVs 09:06:03 Integrated Voltage Measurement... 09:06:12 Voltage Offset Correction: v_off_ini = 2159.767 uV, v_off_fin = 2164.108 uV 09:06:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.331 uVs 09:06:12 Integrated Voltage Measurement... 09:06:21 Voltage Offset Correction: v_off_ini = 2165.481 uV, v_off_fin = 2165.393 uV 09:06:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.851 uVs 09:06:21 Integrated Voltage Measurement... 09:06:30 Voltage Offset Correction: v_off_ini = 2163.621 uV, v_off_fin = 2153.846 uV 09:06:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.005 uVs 09:06:30 Integrated Voltage Measurement... 09:06:39 Voltage Offset Correction: v_off_ini = 2174.842 uV, v_off_fin = 2186.681 uV 09:06:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.727 uVs 09:06:41 I2X Average and RMS of Measurements VT = -85.441 +- 7.693 uVs I2X = 7.861 +- 0.708 uTm^2 09:06:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:06:49 Integrated Voltage Measurement... 09:06:58 Voltage Offset Correction: v_off_ini = 2154.496 uV, v_off_fin = 2146.274 uV 09:06:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -212.817 uVs 09:06:58 Integrated Voltage Measurement... 09:07:07 Voltage Offset Correction: v_off_ini = 2151.681 uV, v_off_fin = 2164.155 uV 09:07:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 212.749 uVs 09:07:07 Integrated Voltage Measurement... 09:07:16 Voltage Offset Correction: v_off_ini = 2155.326 uV, v_off_fin = 2154.192 uV 09:07:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.902 uVs 09:07:16 Integrated Voltage Measurement... 09:07:25 Voltage Offset Correction: v_off_ini = 2185.438 uV, v_off_fin = 2193.656 uV 09:07:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 221.696 uVs 09:07:25 Integrated Voltage Measurement... 09:07:34 Voltage Offset Correction: v_off_ini = 2160.353 uV, v_off_fin = 2165.980 uV 09:07:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -199.303 uVs 09:07:34 Integrated Voltage Measurement... 09:07:43 Voltage Offset Correction: v_off_ini = 2176.496 uV, v_off_fin = 2187.091 uV 09:07:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.451 uVs 09:07:45 I2Y Average and RMS of Measurements VT = -213.820 +- 7.760 uVs I2Y = -19.671 +- 0.714 uTm^2 09:07:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm