Coil Constant for 26DQB0_25q_main using magnet LCLS II Quad 4070 New Calibrated Coil Constant = 348047.11684, File coilpar.ru (Old) Coil Constant = 365879.37800 New Constant/Old Constant = 0.9513, Old Coil Radius = 0.0029972, New Radius = 0.0023972, Number of coils = 1 (Only change radius if Number of coils = 1!) Wire Strength = 0.714099 T, Coil Strength = 0.750686 T, Magnet Type is Quadrupole at Imag = 6.000 for coil. Number of Coil Turns = 1 Coil Constant Calculated from radius, turns and magnet order = 222637.6185 (Should be within 5% of Calibrated Coil Constant)