UFx.py version 1.6 from 07/16/2018
Page 1 of 2
Created by user vjeran on 07/17/2018 14:41:00
SLAC Moving Stretched Wire Magnetic Field Measurement Summary
| Magnet ID : | Calibration |
| Directory : | S:\magdata\LCLS-II\Phase Shifter\Calibration\126Gcm_-By\Final Results\Stretched Wire |
| Date : | 06-26-2018 |
| Time : | 11:59:47 - 13:04:45 |
| Operator : | VV |
| Project : | DUMMY |
| Device Type : | Phase Shifter |
| Run Number : | 1 |
| Data Type : | |
| Run Comment : | rewired; 126 Gauss*cm magnet (By negative) |
| gap | I1Y | I2Y | I1X | I2X |
|
- |
-132.2 4.2 (5) |
-27.7 1.5 (5) |
0.3 2.9 (5) |
-5.4 4.5 (5) |
| integral | gap | -02.00;+00.00 | -01.00;+00.00 | +00.00;+00.00 | +01.00;+00.00 | +02.00;+00.00 |
| I1Y |
- |
-136.1 4.5 (3) |
-128.0 4.2 (3) |
-129.4 4.6 (3) |
-134.4 7.3 (3) |
-137.4 2.4 (3) |
| I2Y |
- |
-28.4 2.0 (3) |
-30.5 0.2 (3) |
-27.4 2.7 (3) |
-28.5 1.4 (3) |
-27.0 1.5 (3) |
| I1X |
- |
-1.1 3.3 (3) |
0.1 6.6 (3) |
-2.9 4.1 (3) |
-1.6 8.1 (3) |
-12.2 15.3 (3) |
| I2X |
- |
0.0 1.1 (3) |
-3.0 2.5 (3) |
0.2 3.9 (3) |
-0.5 3.5 (3) |
0.2 1.5 (3) |
| integral | gap | +00.00;-02.00 | +00.00;-01.00 | +00.00;+00.00 | +00.00;+01.00 | +00.00;+02.00 |
| I1Y |
- |
-129.0 5.7 (3) |
-130.9 2.5 (3) |
-134.2 5.9 (3) |
-105.9 48.8 (3) |
-126.9 5.5 (3) |
| I2Y |
- |
-23.0 1.6 (3) |
-26.3 3.1 (3) |
-28.7 1.6 (3) |
-10.2 18.2 (3) |
6.2 27.5 (3) |
| I1X |
- |
2.2 5.3 (3) |
-3.9 5.5 (3) |
0.9 5.1 (3) |
-13.2 8.1 (3) |
-1.6 2.2 (3) |
| I2X |
- |
-0.4 2.0 (3) |
-0.1 2.4 (3) |
-0.0 1.1 (3) |
-5.2 4.8 (3) |
-0.4 3.5 (3) |
UFx.py version 1.6 from 07/16/2018
Page 2 of 2
Created by user vjeran on 07/17/2018 14:41:00
SLAC Moving Stretched Wire Magnetic Field Measurement Summary
| Magnet ID : | Calibration |
| Directory : | S:\magdata\LCLS-II\Phase Shifter\Calibration\126Gcm_-By\Final Results\Stretched Wire |