SLAC Magnetic Measurements Date: 11-17-2022 Time: 16:52:39 Project: FACET-II Magnet Type: Undulator Magnet Name: LHU-0003 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched wire Run Number: 9 Measurement Device: Stretched Wire Operator: YL Comment: Parameters: Measurement Type = I12XY VS X VS GAP I1X dy = 0.5000 mm I1Y dx = 0.5000 mm I2X dy = 1.0000 mm I2Y dx = 1.0000 mm Wire # Turns = 1 Preamplifier Gain = 10000.00 Background Field Integrals: I1X = 10.0 (uTm) I1Y = -29.0 (uTm) I2X = 5.0 (uTm2) I2Y = -13.0 (uTm2) Subtracted from results Field Integrals: Gap (mm) X0 (mm) Y0 (mm) I1x (uTm) I1y (uTm) I2x (uTm^2) I2y (uTm^2) ---------- ---------- ---------- ---------------------- ---------------------- ---------------------- ---------------------- 40.000 0.000 0.000 10.717+- 1.284 16.529+- 1.549 -7.768+- 0.712 -13.673+- 0.078