SLAC Magnetic Measurements Date: 11-22-2022 Time: 14:34:29 Project: FACET-II Magnet Type: Undulator Magnet Name: LHU-0003 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched wire Run Number: 26 Measurement Device: Stretched Wire Operator: YL Comment: Measurement vs. X; K=1.17 14:35:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:35:09 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.000500 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:35:10 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.000500 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:35:12 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.000500 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:35:13 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.000500 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:37:15 Wire Position: X0 = -15.0000 mm, Y0 = 0.0000 mm 14:37:19 Integrated Voltage Measurement... 14:37:28 Voltage Offset Correction: v_off_ini = 152.723 uV, v_off_fin = 179.282 uV 14:37:28 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 392.480 uVs 14:37:28 Integrated Voltage Measurement... 14:37:37 Voltage Offset Correction: v_off_ini = 177.130 uV, v_off_fin = 153.959 uV 14:37:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -402.480 uVs 14:37:37 Integrated Voltage Measurement... 14:37:47 Voltage Offset Correction: v_off_ini = 163.951 uV, v_off_fin = 201.070 uV 14:37:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 405.819 uVs 14:37:47 Integrated Voltage Measurement... 14:37:56 Voltage Offset Correction: v_off_ini = 172.600 uV, v_off_fin = 156.167 uV 14:37:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -382.253 uVs 14:37:56 Integrated Voltage Measurement... 14:38:05 Voltage Offset Correction: v_off_ini = 181.786 uV, v_off_fin = 198.471 uV 14:38:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 402.994 uVs 14:38:05 Integrated Voltage Measurement... 14:38:14 Voltage Offset Correction: v_off_ini = 161.570 uV, v_off_fin = 141.121 uV 14:38:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -403.942 uVs 14:38:18 I1X Average and RMS of Measurements VT = 398.328 +- 3.897 uVs I1X = -79.670 +- 0.779 uTm 14:38:21 Wire Position: X0 = -15.0000 mm, Y0 = 0.0000 mm 14:38:25 Integrated Voltage Measurement... 14:38:34 Voltage Offset Correction: v_off_ini = 169.119 uV, v_off_fin = 125.142 uV 14:38:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -517.051 uVs 14:38:34 Integrated Voltage Measurement... 14:38:43 Voltage Offset Correction: v_off_ini = 170.385 uV, v_off_fin = 203.943 uV 14:38:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 534.353 uVs 14:38:43 Integrated Voltage Measurement... 14:38:52 Voltage Offset Correction: v_off_ini = 156.550 uV, v_off_fin = 135.918 uV 14:38:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -517.072 uVs 14:38:52 Integrated Voltage Measurement... 14:39:02 Voltage Offset Correction: v_off_ini = 174.475 uV, v_off_fin = 223.216 uV 14:39:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 518.462 uVs 14:39:02 Integrated Voltage Measurement... 14:39:11 Voltage Offset Correction: v_off_ini = 182.220 uV, v_off_fin = 146.305 uV 14:39:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -529.717 uVs 14:39:11 Integrated Voltage Measurement... 14:39:20 Voltage Offset Correction: v_off_ini = 170.106 uV, v_off_fin = 215.915 uV 14:39:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 496.710 uVs 14:39:24 I1Y Average and RMS of Measurements VT = -518.894 +- 5.160 uVs I1Y = -103.779 +- 1.032 uTm 14:39:29 Wire Position: X0 = -15.0000 mm, Y0 = 0.0000 mm 14:39:32 Integrated Voltage Measurement... 14:39:42 Voltage Offset Correction: v_off_ini = 173.594 uV, v_off_fin = 179.338 uV 14:39:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 385.938 uVs 14:39:42 Integrated Voltage Measurement... 14:39:51 Voltage Offset Correction: v_off_ini = 162.886 uV, v_off_fin = 154.205 uV 14:39:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -413.395 uVs 14:39:51 Integrated Voltage Measurement... 14:40:01 Voltage Offset Correction: v_off_ini = 163.939 uV, v_off_fin = 158.376 uV 14:40:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 439.905 uVs 14:40:01 Integrated Voltage Measurement... 14:40:10 Voltage Offset Correction: v_off_ini = 176.496 uV, v_off_fin = 169.521 uV 14:40:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -424.910 uVs 14:40:10 Integrated Voltage Measurement... 14:40:20 Voltage Offset Correction: v_off_ini = 171.886 uV, v_off_fin = 157.299 uV 14:40:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 403.293 uVs 14:40:20 Integrated Voltage Measurement... 14:40:29 Voltage Offset Correction: v_off_ini = 153.819 uV, v_off_fin = 155.462 uV 14:40:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -384.465 uVs 14:40:32 I2X Average and RMS of Measurements VT = 408.651 +- 16.964 uVs I2X = -37.596 +- 1.561 uTm^2 14:40:37 Wire Position: X0 = -15.0000 mm, Y0 = 0.0000 mm 14:40:40 Integrated Voltage Measurement... 14:40:49 Voltage Offset Correction: v_off_ini = 177.715 uV, v_off_fin = 162.831 uV 14:40:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -481.524 uVs 14:40:49 Integrated Voltage Measurement... 14:40:59 Voltage Offset Correction: v_off_ini = 151.912 uV, v_off_fin = 138.224 uV 14:40:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 467.873 uVs 14:40:59 Integrated Voltage Measurement... 14:41:08 Voltage Offset Correction: v_off_ini = 158.269 uV, v_off_fin = 125.925 uV 14:41:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -509.643 uVs 14:41:08 Integrated Voltage Measurement... 14:41:18 Voltage Offset Correction: v_off_ini = 151.891 uV, v_off_fin = 163.946 uV 14:41:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 468.744 uVs 14:41:18 Integrated Voltage Measurement... 14:41:27 Voltage Offset Correction: v_off_ini = 133.038 uV, v_off_fin = 138.177 uV 14:41:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -469.805 uVs 14:41:27 Integrated Voltage Measurement... 14:41:37 Voltage Offset Correction: v_off_ini = 142.558 uV, v_off_fin = 160.745 uV 14:41:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 445.147 uVs 14:41:39 I2Y Average and RMS of Measurements VT = -473.789 +- 12.964 uVs I2Y = -43.589 +- 1.193 uTm^2 14:44:16 Wire Position: X0 = -10.0000 mm, Y0 = 0.0000 mm 14:44:20 Integrated Voltage Measurement... 14:44:30 Voltage Offset Correction: v_off_ini = 174.918 uV, v_off_fin = 180.990 uV 14:44:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 233.908 uVs 14:44:30 Integrated Voltage Measurement... 14:44:39 Voltage Offset Correction: v_off_ini = 155.897 uV, v_off_fin = 135.136 uV 14:44:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.078 uVs 14:44:39 Integrated Voltage Measurement... 14:44:48 Voltage Offset Correction: v_off_ini = 170.508 uV, v_off_fin = 181.038 uV 14:44:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.882 uVs 14:44:48 Integrated Voltage Measurement... 14:44:57 Voltage Offset Correction: v_off_ini = 157.136 uV, v_off_fin = 146.556 uV 14:44:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -243.575 uVs 14:44:57 Integrated Voltage Measurement... 14:45:06 Voltage Offset Correction: v_off_ini = 158.539 uV, v_off_fin = 184.361 uV 14:45:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.497 uVs 14:45:06 Integrated Voltage Measurement... 14:45:15 Voltage Offset Correction: v_off_ini = 167.729 uV, v_off_fin = 142.505 uV 14:45:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.068 uVs 14:45:20 I1X Average and RMS of Measurements VT = 242.501 +- 4.452 uVs I1X = -48.501 +- 0.890 uTm 14:45:22 Wire Position: X0 = -10.0000 mm, Y0 = 0.0000 mm 14:45:26 Integrated Voltage Measurement... 14:45:36 Voltage Offset Correction: v_off_ini = 167.307 uV, v_off_fin = 131.547 uV 14:45:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -377.577 uVs 14:45:36 Integrated Voltage Measurement... 14:45:45 Voltage Offset Correction: v_off_ini = 176.828 uV, v_off_fin = 218.820 uV 14:45:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 352.675 uVs 14:45:45 Integrated Voltage Measurement... 14:45:54 Voltage Offset Correction: v_off_ini = 160.921 uV, v_off_fin = 138.585 uV 14:45:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -345.134 uVs 14:45:54 Integrated Voltage Measurement... 14:46:03 Voltage Offset Correction: v_off_ini = 176.071 uV, v_off_fin = 186.009 uV 14:46:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 359.580 uVs 14:46:03 Integrated Voltage Measurement... 14:46:12 Voltage Offset Correction: v_off_ini = 169.977 uV, v_off_fin = 143.743 uV 14:46:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -354.162 uVs 14:46:12 Integrated Voltage Measurement... 14:46:21 Voltage Offset Correction: v_off_ini = 160.694 uV, v_off_fin = 208.340 uV 14:46:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 361.124 uVs 14:46:26 I1Y Average and RMS of Measurements VT = -358.375 +- 5.239 uVs I1Y = -71.675 +- 1.048 uTm 14:46:31 Wire Position: X0 = -10.0000 mm, Y0 = 0.0000 mm 14:46:34 Integrated Voltage Measurement... 14:46:43 Voltage Offset Correction: v_off_ini = 174.154 uV, v_off_fin = 163.115 uV 14:46:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 247.855 uVs 14:46:43 Integrated Voltage Measurement... 14:46:53 Voltage Offset Correction: v_off_ini = 155.839 uV, v_off_fin = 169.773 uV 14:46:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.325 uVs 14:46:53 Integrated Voltage Measurement... 14:47:02 Voltage Offset Correction: v_off_ini = 176.886 uV, v_off_fin = 164.662 uV 14:47:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.001 uVs 14:47:02 Integrated Voltage Measurement... 14:47:12 Voltage Offset Correction: v_off_ini = 174.371 uV, v_off_fin = 172.487 uV 14:47:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.167 uVs 14:47:12 Integrated Voltage Measurement... 14:47:21 Voltage Offset Correction: v_off_ini = 161.673 uV, v_off_fin = 170.621 uV 14:47:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.966 uVs 14:47:21 Integrated Voltage Measurement... 14:47:31 Voltage Offset Correction: v_off_ini = 155.168 uV, v_off_fin = 168.680 uV 14:47:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.903 uVs 14:47:33 I2X Average and RMS of Measurements VT = 248.869 +- 10.129 uVs I2X = -22.896 +- 0.932 uTm^2 14:47:38 Wire Position: X0 = -10.0000 mm, Y0 = 0.0000 mm 14:47:41 Integrated Voltage Measurement... 14:47:50 Voltage Offset Correction: v_off_ini = 173.098 uV, v_off_fin = 165.752 uV 14:47:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -341.583 uVs 14:47:50 Integrated Voltage Measurement... 14:48:00 Voltage Offset Correction: v_off_ini = 179.702 uV, v_off_fin = 156.603 uV 14:48:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 354.922 uVs 14:48:00 Integrated Voltage Measurement... 14:48:09 Voltage Offset Correction: v_off_ini = 164.305 uV, v_off_fin = 162.480 uV 14:48:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -338.014 uVs 14:48:09 Integrated Voltage Measurement... 14:48:19 Voltage Offset Correction: v_off_ini = 165.870 uV, v_off_fin = 170.566 uV 14:48:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.967 uVs 14:48:19 Integrated Voltage Measurement... 14:48:28 Voltage Offset Correction: v_off_ini = 139.137 uV, v_off_fin = 150.704 uV 14:48:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.104 uVs 14:48:28 Integrated Voltage Measurement... 14:48:38 Voltage Offset Correction: v_off_ini = 147.249 uV, v_off_fin = 148.993 uV 14:48:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 312.953 uVs 14:48:41 I2Y Average and RMS of Measurements VT = -328.090 +- 16.223 uVs I2Y = -30.184 +- 1.493 uTm^2 14:49:31 Wire Position: X0 = -5.0000 mm, Y0 = 0.0000 mm 14:49:35 Integrated Voltage Measurement... 14:49:44 Voltage Offset Correction: v_off_ini = 164.705 uV, v_off_fin = 164.455 uV 14:49:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 56.607 uVs 14:49:44 Integrated Voltage Measurement... 14:49:54 Voltage Offset Correction: v_off_ini = 165.642 uV, v_off_fin = 181.146 uV 14:49:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -59.064 uVs 14:49:54 Integrated Voltage Measurement... 14:50:03 Voltage Offset Correction: v_off_ini = 179.820 uV, v_off_fin = 193.216 uV 14:50:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 38.432 uVs 14:50:03 Integrated Voltage Measurement... 14:50:12 Voltage Offset Correction: v_off_ini = 199.652 uV, v_off_fin = 176.572 uV 14:50:12 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -67.247 uVs 14:50:12 Integrated Voltage Measurement... 14:50:21 Voltage Offset Correction: v_off_ini = 181.172 uV, v_off_fin = 191.067 uV 14:50:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 45.787 uVs 14:50:21 Integrated Voltage Measurement... 14:50:30 Voltage Offset Correction: v_off_ini = 187.132 uV, v_off_fin = 194.596 uV 14:50:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -71.356 uVs 14:50:34 I1X Average and RMS of Measurements VT = 56.416 +- 2.546 uVs I1X = -11.284 +- 0.509 uTm 14:50:37 Wire Position: X0 = -5.0000 mm, Y0 = 0.0000 mm 14:50:41 Integrated Voltage Measurement... 14:50:50 Voltage Offset Correction: v_off_ini = 176.819 uV, v_off_fin = 160.478 uV 14:50:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.030 uVs 14:50:50 Integrated Voltage Measurement... 14:51:00 Voltage Offset Correction: v_off_ini = 186.774 uV, v_off_fin = 196.894 uV 14:51:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 249.120 uVs 14:51:00 Integrated Voltage Measurement... 14:51:09 Voltage Offset Correction: v_off_ini = 177.474 uV, v_off_fin = 169.482 uV 14:51:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.016 uVs 14:51:09 Integrated Voltage Measurement... 14:51:18 Voltage Offset Correction: v_off_ini = 208.181 uV, v_off_fin = 210.539 uV 14:51:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.504 uVs 14:51:18 Integrated Voltage Measurement... 14:51:27 Voltage Offset Correction: v_off_ini = 192.463 uV, v_off_fin = 177.364 uV 14:51:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -239.148 uVs 14:51:27 Integrated Voltage Measurement... 14:51:36 Voltage Offset Correction: v_off_ini = 184.666 uV, v_off_fin = 226.227 uV 14:51:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 269.443 uVs 14:51:40 I1Y Average and RMS of Measurements VT = -252.210 +- 1.650 uVs I1Y = -50.442 +- 0.330 uTm 14:51:46 Wire Position: X0 = -5.0000 mm, Y0 = 0.0000 mm 14:51:48 Integrated Voltage Measurement... 14:51:58 Voltage Offset Correction: v_off_ini = 194.372 uV, v_off_fin = 197.091 uV 14:51:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 48.473 uVs 14:51:58 Integrated Voltage Measurement... 14:52:07 Voltage Offset Correction: v_off_ini = 195.698 uV, v_off_fin = 176.282 uV 14:52:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -28.759 uVs 14:52:07 Integrated Voltage Measurement... 14:52:17 Voltage Offset Correction: v_off_ini = 163.195 uV, v_off_fin = 175.719 uV 14:52:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 78.334 uVs 14:52:17 Integrated Voltage Measurement... 14:52:26 Voltage Offset Correction: v_off_ini = 169.980 uV, v_off_fin = 186.592 uV 14:52:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -37.329 uVs 14:52:26 Integrated Voltage Measurement... 14:52:36 Voltage Offset Correction: v_off_ini = 181.847 uV, v_off_fin = 153.621 uV 14:52:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 31.855 uVs 14:52:36 Integrated Voltage Measurement... 14:52:45 Voltage Offset Correction: v_off_ini = 181.811 uV, v_off_fin = 179.390 uV 14:52:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -66.507 uVs 14:52:48 I2X Average and RMS of Measurements VT = 48.543 +- 7.858 uVs I2X = -4.466 +- 0.723 uTm^2 14:52:53 Wire Position: X0 = -5.0000 mm, Y0 = 0.0000 mm 14:52:56 Integrated Voltage Measurement... 14:53:05 Voltage Offset Correction: v_off_ini = 173.284 uV, v_off_fin = 175.288 uV 14:53:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -191.319 uVs 14:53:05 Integrated Voltage Measurement... 14:53:15 Voltage Offset Correction: v_off_ini = 175.657 uV, v_off_fin = 194.395 uV 14:53:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 175.976 uVs 14:53:15 Integrated Voltage Measurement... 14:53:24 Voltage Offset Correction: v_off_ini = 192.589 uV, v_off_fin = 187.377 uV 14:53:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.576 uVs 14:53:24 Integrated Voltage Measurement... 14:53:34 Voltage Offset Correction: v_off_ini = 173.299 uV, v_off_fin = 156.777 uV 14:53:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 199.376 uVs 14:53:34 Integrated Voltage Measurement... 14:53:43 Voltage Offset Correction: v_off_ini = 157.506 uV, v_off_fin = 148.652 uV 14:53:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -216.034 uVs 14:53:43 Integrated Voltage Measurement... 14:53:53 Voltage Offset Correction: v_off_ini = 163.581 uV, v_off_fin = 179.848 uV 14:53:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.413 uVs 14:53:55 I2Y Average and RMS of Measurements VT = -201.116 +- 12.537 uVs I2Y = -18.503 +- 1.153 uTm^2 14:54:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:54:51 Integrated Voltage Measurement... 14:55:00 Voltage Offset Correction: v_off_ini = 214.144 uV, v_off_fin = 207.340 uV 14:55:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -44.702 uVs 14:55:01 Integrated Voltage Measurement... 14:55:10 Voltage Offset Correction: v_off_ini = 195.269 uV, v_off_fin = 209.723 uV 14:55:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 31.458 uVs 14:55:10 Integrated Voltage Measurement... 14:55:19 Voltage Offset Correction: v_off_ini = 217.632 uV, v_off_fin = 201.832 uV 14:55:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -39.351 uVs 14:55:19 Integrated Voltage Measurement... 14:55:28 Voltage Offset Correction: v_off_ini = 214.701 uV, v_off_fin = 212.383 uV 14:55:28 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 26.334 uVs 14:55:28 Integrated Voltage Measurement... 14:55:37 Voltage Offset Correction: v_off_ini = 209.556 uV, v_off_fin = 201.277 uV 14:55:37 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -27.726 uVs 14:55:37 Integrated Voltage Measurement... 14:55:46 Voltage Offset Correction: v_off_ini = 200.278 uV, v_off_fin = 194.760 uV 14:55:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 18.952 uVs 14:55:51 I1X Average and RMS of Measurements VT = -31.420 +- 6.101 uVs I1X = 6.284 +- 1.220 uTm 14:55:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:55:57 Integrated Voltage Measurement... 14:56:07 Voltage Offset Correction: v_off_ini = 198.070 uV, v_off_fin = 166.870 uV 14:56:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -168.038 uVs 14:56:07 Integrated Voltage Measurement... 14:56:16 Voltage Offset Correction: v_off_ini = 198.793 uV, v_off_fin = 203.400 uV 14:56:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 153.038 uVs 14:56:16 Integrated Voltage Measurement... 14:56:25 Voltage Offset Correction: v_off_ini = 194.773 uV, v_off_fin = 173.254 uV 14:56:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -149.811 uVs 14:56:25 Integrated Voltage Measurement... 14:56:34 Voltage Offset Correction: v_off_ini = 197.671 uV, v_off_fin = 199.472 uV 14:56:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.542 uVs 14:56:34 Integrated Voltage Measurement... 14:56:43 Voltage Offset Correction: v_off_ini = 174.117 uV, v_off_fin = 162.581 uV 14:56:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -147.237 uVs 14:56:43 Integrated Voltage Measurement... 14:56:52 Voltage Offset Correction: v_off_ini = 181.465 uV, v_off_fin = 204.805 uV 14:56:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.329 uVs 14:56:57 I1Y Average and RMS of Measurements VT = -158.833 +- 1.552 uVs I1Y = -31.767 +- 0.310 uTm 14:57:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:57:04 Integrated Voltage Measurement... 14:57:14 Voltage Offset Correction: v_off_ini = 185.319 uV, v_off_fin = 172.989 uV 14:57:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.721 uVs 14:57:14 Integrated Voltage Measurement... 14:57:23 Voltage Offset Correction: v_off_ini = 187.457 uV, v_off_fin = 187.317 uV 14:57:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.699 uVs 14:57:24 Integrated Voltage Measurement... 14:57:33 Voltage Offset Correction: v_off_ini = 183.491 uV, v_off_fin = 192.143 uV 14:57:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -33.869 uVs 14:57:33 Integrated Voltage Measurement... 14:57:43 Voltage Offset Correction: v_off_ini = 183.893 uV, v_off_fin = 181.151 uV 14:57:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.078 uVs 14:57:43 Integrated Voltage Measurement... 14:57:52 Voltage Offset Correction: v_off_ini = 191.479 uV, v_off_fin = 196.991 uV 14:57:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.579 uVs 14:57:52 Integrated Voltage Measurement... 14:58:02 Voltage Offset Correction: v_off_ini = 194.907 uV, v_off_fin = 204.711 uV 14:58:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.150 uVs 14:58:04 I2X Average and RMS of Measurements VT = -55.183 +- 4.499 uVs I2X = 5.077 +- 0.414 uTm^2 14:58:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:58:12 Integrated Voltage Measurement... 14:58:21 Voltage Offset Correction: v_off_ini = 179.618 uV, v_off_fin = 178.677 uV 14:58:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -118.574 uVs 14:58:21 Integrated Voltage Measurement... 14:58:31 Voltage Offset Correction: v_off_ini = 178.929 uV, v_off_fin = 180.229 uV 14:58:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 138.181 uVs 14:58:31 Integrated Voltage Measurement... 14:58:41 Voltage Offset Correction: v_off_ini = 182.750 uV, v_off_fin = 183.226 uV 14:58:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -117.477 uVs 14:58:41 Integrated Voltage Measurement... 14:58:50 Voltage Offset Correction: v_off_ini = 200.111 uV, v_off_fin = 199.707 uV 14:58:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 116.290 uVs 14:58:50 Integrated Voltage Measurement... 14:59:00 Voltage Offset Correction: v_off_ini = 189.718 uV, v_off_fin = 188.802 uV 14:59:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -104.431 uVs 14:59:00 Integrated Voltage Measurement... 14:59:09 Voltage Offset Correction: v_off_ini = 182.780 uV, v_off_fin = 178.124 uV 14:59:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 122.253 uVs 14:59:12 I2Y Average and RMS of Measurements VT = -119.534 +- 6.418 uVs I2Y = -10.997 +- 0.590 uTm^2 14:59:58 Wire Position: X0 = 5.0000 mm, Y0 = 0.0000 mm 15:00:03 Integrated Voltage Measurement... 15:00:12 Voltage Offset Correction: v_off_ini = 205.453 uV, v_off_fin = 210.330 uV 15:00:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 60.062 uVs 15:00:12 Integrated Voltage Measurement... 15:00:21 Voltage Offset Correction: v_off_ini = 207.056 uV, v_off_fin = 201.798 uV 15:00:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -31.080 uVs 15:00:21 Integrated Voltage Measurement... 15:00:30 Voltage Offset Correction: v_off_ini = 186.733 uV, v_off_fin = 188.784 uV 15:00:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 33.593 uVs 15:00:30 Integrated Voltage Measurement... 15:00:39 Voltage Offset Correction: v_off_ini = 208.808 uV, v_off_fin = 212.803 uV 15:00:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -8.515 uVs 15:00:39 Integrated Voltage Measurement... 15:00:48 Voltage Offset Correction: v_off_ini = 188.812 uV, v_off_fin = 202.066 uV 15:00:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 50.111 uVs 15:00:48 Integrated Voltage Measurement... 15:00:57 Voltage Offset Correction: v_off_ini = 176.813 uV, v_off_fin = 189.379 uV 15:00:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -28.605 uVs 15:01:02 I1X Average and RMS of Measurements VT = 35.328 +- 10.407 uVs I1X = -7.066 +- 2.082 uTm 15:01:04 Wire Position: X0 = 5.0000 mm, Y0 = 0.0000 mm 15:01:09 Integrated Voltage Measurement... 15:01:18 Voltage Offset Correction: v_off_ini = 197.661 uV, v_off_fin = 183.532 uV 15:01:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -85.985 uVs 15:01:18 Integrated Voltage Measurement... 15:01:27 Voltage Offset Correction: v_off_ini = 192.626 uV, v_off_fin = 209.006 uV 15:01:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 62.116 uVs 15:01:27 Integrated Voltage Measurement... 15:01:36 Voltage Offset Correction: v_off_ini = 176.270 uV, v_off_fin = 158.172 uV 15:01:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -79.653 uVs 15:01:36 Integrated Voltage Measurement... 15:01:45 Voltage Offset Correction: v_off_ini = 189.068 uV, v_off_fin = 205.226 uV 15:01:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 66.823 uVs 15:01:45 Integrated Voltage Measurement... 15:01:54 Voltage Offset Correction: v_off_ini = 186.127 uV, v_off_fin = 194.720 uV 15:01:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -72.694 uVs 15:01:54 Integrated Voltage Measurement... 15:02:03 Voltage Offset Correction: v_off_ini = 175.878 uV, v_off_fin = 189.596 uV 15:02:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 69.520 uVs 15:02:08 I1Y Average and RMS of Measurements VT = -72.798 +- 1.241 uVs I1Y = -14.560 +- 0.248 uTm 15:02:13 Wire Position: X0 = 5.0000 mm, Y0 = 0.0000 mm 15:02:16 Integrated Voltage Measurement... 15:02:25 Voltage Offset Correction: v_off_ini = 181.516 uV, v_off_fin = 176.668 uV 15:02:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -35.765 uVs 15:02:25 Integrated Voltage Measurement... 15:02:35 Voltage Offset Correction: v_off_ini = 171.257 uV, v_off_fin = 188.001 uV 15:02:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.865 uVs 15:02:35 Integrated Voltage Measurement... 15:02:44 Voltage Offset Correction: v_off_ini = 180.195 uV, v_off_fin = 196.735 uV 15:02:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.157 uVs 15:02:44 Integrated Voltage Measurement... 15:02:54 Voltage Offset Correction: v_off_ini = 187.659 uV, v_off_fin = 181.453 uV 15:02:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 29.480 uVs 15:02:54 Integrated Voltage Measurement... 15:03:03 Voltage Offset Correction: v_off_ini = 185.485 uV, v_off_fin = 200.320 uV 15:03:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.640 uVs 15:03:03 Integrated Voltage Measurement... 15:03:13 Voltage Offset Correction: v_off_ini = 185.688 uV, v_off_fin = 186.870 uV 15:03:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 19.390 uVs 15:03:15 I2X Average and RMS of Measurements VT = -33.216 +- 11.384 uVs I2X = 3.056 +- 1.047 uTm^2 15:03:20 Wire Position: X0 = 5.0000 mm, Y0 = 0.0000 mm 15:03:23 Integrated Voltage Measurement... 15:03:33 Voltage Offset Correction: v_off_ini = 178.674 uV, v_off_fin = 168.854 uV 15:03:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -5.156 uVs 15:03:33 Integrated Voltage Measurement... 15:03:42 Voltage Offset Correction: v_off_ini = 178.898 uV, v_off_fin = 190.623 uV 15:03:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 36.194 uVs 15:03:42 Integrated Voltage Measurement... 15:03:52 Voltage Offset Correction: v_off_ini = 164.995 uV, v_off_fin = 173.450 uV 15:03:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.252 uVs 15:03:52 Integrated Voltage Measurement... 15:04:01 Voltage Offset Correction: v_off_ini = 173.924 uV, v_off_fin = 177.945 uV 15:04:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 34.012 uVs 15:04:01 Integrated Voltage Measurement... 15:04:11 Voltage Offset Correction: v_off_ini = 161.271 uV, v_off_fin = 181.460 uV 15:04:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 8.069 uVs 15:04:11 Integrated Voltage Measurement... 15:04:20 Voltage Offset Correction: v_off_ini = 192.658 uV, v_off_fin = 187.868 uV 15:04:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.051 uVs 15:04:23 I2Y Average and RMS of Measurements VT = -0.498 +- 14.833 uVs I2Y = -0.046 +- 1.365 uTm^2 15:05:03 Wire Position: X0 = 10.0000 mm, Y0 = 0.0000 mm 15:05:07 Integrated Voltage Measurement... 15:05:16 Voltage Offset Correction: v_off_ini = 175.098 uV, v_off_fin = 185.922 uV 15:05:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 181.122 uVs 15:05:16 Integrated Voltage Measurement... 15:05:25 Voltage Offset Correction: v_off_ini = 168.501 uV, v_off_fin = 149.875 uV 15:05:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.955 uVs 15:05:25 Integrated Voltage Measurement... 15:05:34 Voltage Offset Correction: v_off_ini = 159.033 uV, v_off_fin = 172.550 uV 15:05:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.045 uVs 15:05:34 Integrated Voltage Measurement... 15:05:43 Voltage Offset Correction: v_off_ini = 164.247 uV, v_off_fin = 155.055 uV 15:05:43 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.620 uVs 15:05:44 Integrated Voltage Measurement... 15:05:53 Voltage Offset Correction: v_off_ini = 157.622 uV, v_off_fin = 206.932 uV 15:05:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 208.020 uVs 15:05:53 Integrated Voltage Measurement... 15:06:02 Voltage Offset Correction: v_off_ini = 177.132 uV, v_off_fin = 155.754 uV 15:06:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -187.581 uVs 15:06:06 I1X Average and RMS of Measurements VT = 191.557 +- 4.475 uVs I1X = -38.314 +- 0.896 uTm 15:06:09 Wire Position: X0 = 10.0000 mm, Y0 = 0.0000 mm 15:06:13 Integrated Voltage Measurement... 15:06:22 Voltage Offset Correction: v_off_ini = 169.106 uV, v_off_fin = 169.182 uV 15:06:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 13.577 uVs 15:06:22 Integrated Voltage Measurement... 15:06:31 Voltage Offset Correction: v_off_ini = 162.049 uV, v_off_fin = 178.486 uV 15:06:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -26.832 uVs 15:06:31 Integrated Voltage Measurement... 15:06:40 Voltage Offset Correction: v_off_ini = 159.955 uV, v_off_fin = 161.011 uV 15:06:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 20.610 uVs 15:06:40 Integrated Voltage Measurement... 15:06:49 Voltage Offset Correction: v_off_ini = 155.835 uV, v_off_fin = 164.298 uV 15:06:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -17.968 uVs 15:06:50 Integrated Voltage Measurement... 15:06:59 Voltage Offset Correction: v_off_ini = 153.236 uV, v_off_fin = 155.658 uV 15:06:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 21.782 uVs 15:06:59 Integrated Voltage Measurement... 15:07:08 Voltage Offset Correction: v_off_ini = 171.346 uV, v_off_fin = 165.739 uV 15:07:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -7.606 uVs 15:07:12 I1Y Average and RMS of Measurements VT = 18.063 +- 2.411 uVs I1Y = 3.613 +- 0.482 uTm 15:07:17 Wire Position: X0 = 10.0000 mm, Y0 = 0.0000 mm 15:07:20 Integrated Voltage Measurement... 15:07:29 Voltage Offset Correction: v_off_ini = 145.732 uV, v_off_fin = 155.594 uV 15:07:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 47.707 uVs 15:07:30 Integrated Voltage Measurement... 15:07:39 Voltage Offset Correction: v_off_ini = 162.809 uV, v_off_fin = 164.887 uV 15:07:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -56.890 uVs 15:07:39 Integrated Voltage Measurement... 15:07:48 Voltage Offset Correction: v_off_ini = 160.697 uV, v_off_fin = 150.216 uV 15:07:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 88.076 uVs 15:07:49 Integrated Voltage Measurement... 15:07:58 Voltage Offset Correction: v_off_ini = 165.672 uV, v_off_fin = 172.253 uV 15:07:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -108.118 uVs 15:07:58 Integrated Voltage Measurement... 15:08:08 Voltage Offset Correction: v_off_ini = 164.955 uV, v_off_fin = 166.360 uV 15:08:08 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 99.499 uVs 15:08:08 Integrated Voltage Measurement... 15:08:17 Voltage Offset Correction: v_off_ini = 162.545 uV, v_off_fin = 149.643 uV 15:08:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -85.683 uVs 15:08:20 I2X Average and RMS of Measurements VT = 80.995 +- 20.416 uVs I2X = -7.452 +- 1.878 uTm^2 15:08:25 Wire Position: X0 = 10.0000 mm, Y0 = 0.0000 mm 15:08:27 Integrated Voltage Measurement... 15:08:37 Voltage Offset Correction: v_off_ini = 143.402 uV, v_off_fin = 159.277 uV 15:08:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 147.016 uVs 15:08:37 Integrated Voltage Measurement... 15:08:47 Voltage Offset Correction: v_off_ini = 160.914 uV, v_off_fin = 146.990 uV 15:08:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.835 uVs 15:08:47 Integrated Voltage Measurement... 15:08:56 Voltage Offset Correction: v_off_ini = 163.975 uV, v_off_fin = 173.647 uV 15:08:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.101 uVs 15:08:56 Integrated Voltage Measurement... 15:09:06 Voltage Offset Correction: v_off_ini = 154.752 uV, v_off_fin = 158.971 uV 15:09:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.849 uVs 15:09:06 Integrated Voltage Measurement... 15:09:15 Voltage Offset Correction: v_off_ini = 163.329 uV, v_off_fin = 158.934 uV 15:09:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.763 uVs 15:09:15 Integrated Voltage Measurement... 15:09:25 Voltage Offset Correction: v_off_ini = 149.734 uV, v_off_fin = 167.533 uV 15:09:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.602 uVs 15:09:27 I2Y Average and RMS of Measurements VT = 103.694 +- 13.821 uVs I2Y = 9.540 +- 1.272 uTm^2 15:10:11 Wire Position: X0 = 15.0000 mm, Y0 = 0.0000 mm 15:10:15 Integrated Voltage Measurement... 15:10:25 Voltage Offset Correction: v_off_ini = 185.623 uV, v_off_fin = 206.150 uV 15:10:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 311.044 uVs 15:10:25 Integrated Voltage Measurement... 15:10:34 Voltage Offset Correction: v_off_ini = 167.475 uV, v_off_fin = 148.165 uV 15:10:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -323.521 uVs 15:10:34 Integrated Voltage Measurement... 15:10:43 Voltage Offset Correction: v_off_ini = 177.837 uV, v_off_fin = 206.457 uV 15:10:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 319.680 uVs 15:10:43 Integrated Voltage Measurement... 15:10:52 Voltage Offset Correction: v_off_ini = 203.234 uV, v_off_fin = 179.743 uV 15:10:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -307.212 uVs 15:10:52 Integrated Voltage Measurement... 15:11:01 Voltage Offset Correction: v_off_ini = 201.693 uV, v_off_fin = 218.633 uV 15:11:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 317.619 uVs 15:11:01 Integrated Voltage Measurement... 15:11:10 Voltage Offset Correction: v_off_ini = 197.242 uV, v_off_fin = 176.471 uV 15:11:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -326.797 uVs 15:11:15 I1X Average and RMS of Measurements VT = 317.646 +- 3.586 uVs I1X = -63.530 +- 0.717 uTm 15:11:17 Wire Position: X0 = 15.0000 mm, Y0 = 0.0000 mm 15:11:22 Integrated Voltage Measurement... 15:11:31 Voltage Offset Correction: v_off_ini = 206.960 uV, v_off_fin = 210.047 uV 15:11:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.022 uVs 15:11:31 Integrated Voltage Measurement... 15:11:40 Voltage Offset Correction: v_off_ini = 182.496 uV, v_off_fin = 183.011 uV 15:11:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.645 uVs 15:11:40 Integrated Voltage Measurement... 15:11:49 Voltage Offset Correction: v_off_ini = 187.062 uV, v_off_fin = 204.678 uV 15:11:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.327 uVs 15:11:49 Integrated Voltage Measurement... 15:11:58 Voltage Offset Correction: v_off_ini = 182.452 uV, v_off_fin = 187.391 uV 15:11:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -90.849 uVs 15:11:58 Integrated Voltage Measurement... 15:12:07 Voltage Offset Correction: v_off_ini = 206.077 uV, v_off_fin = 205.035 uV 15:12:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.993 uVs 15:12:07 Integrated Voltage Measurement... 15:12:16 Voltage Offset Correction: v_off_ini = 195.887 uV, v_off_fin = 180.211 uV 15:12:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.761 uVs 15:12:21 I1Y Average and RMS of Measurements VT = 78.433 +- 1.105 uVs I1Y = 15.687 +- 0.221 uTm 15:12:26 Wire Position: X0 = 15.0000 mm, Y0 = 0.0000 mm 15:12:29 Integrated Voltage Measurement... 15:12:38 Voltage Offset Correction: v_off_ini = 167.558 uV, v_off_fin = 188.461 uV 15:12:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 176.347 uVs 15:12:38 Integrated Voltage Measurement... 15:12:48 Voltage Offset Correction: v_off_ini = 187.135 uV, v_off_fin = 189.723 uV 15:12:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -155.461 uVs 15:12:48 Integrated Voltage Measurement... 15:12:57 Voltage Offset Correction: v_off_ini = 176.361 uV, v_off_fin = 205.194 uV 15:12:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 152.576 uVs 15:12:57 Integrated Voltage Measurement... 15:13:07 Voltage Offset Correction: v_off_ini = 190.395 uV, v_off_fin = 178.474 uV 15:13:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.584 uVs 15:13:07 Integrated Voltage Measurement... 15:13:16 Voltage Offset Correction: v_off_ini = 169.204 uV, v_off_fin = 167.513 uV 15:13:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 128.524 uVs 15:13:16 Integrated Voltage Measurement... 15:13:26 Voltage Offset Correction: v_off_ini = 182.530 uV, v_off_fin = 175.169 uV 15:13:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.358 uVs 15:13:28 I2X Average and RMS of Measurements VT = 156.142 +- 9.685 uVs I2X = -14.365 +- 0.891 uTm^2 15:13:33 Wire Position: X0 = 15.0000 mm, Y0 = 0.0000 mm 15:13:36 Integrated Voltage Measurement... 15:13:45 Voltage Offset Correction: v_off_ini = 173.755 uV, v_off_fin = 181.235 uV 15:13:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 238.933 uVs 15:13:46 Integrated Voltage Measurement... 15:13:55 Voltage Offset Correction: v_off_ini = 180.171 uV, v_off_fin = 177.037 uV 15:13:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -252.217 uVs 15:13:55 Integrated Voltage Measurement... 15:14:05 Voltage Offset Correction: v_off_ini = 193.226 uV, v_off_fin = 173.343 uV 15:14:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 209.141 uVs 15:14:05 Integrated Voltage Measurement... 15:14:14 Voltage Offset Correction: v_off_ini = 183.580 uV, v_off_fin = 194.156 uV 15:14:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -207.367 uVs 15:14:14 Integrated Voltage Measurement... 15:14:24 Voltage Offset Correction: v_off_ini = 183.395 uV, v_off_fin = 177.249 uV 15:14:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 258.070 uVs 15:14:24 Integrated Voltage Measurement... 15:14:33 Voltage Offset Correction: v_off_ini = 179.384 uV, v_off_fin = 177.347 uV 15:14:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -217.109 uVs 15:14:36 I2Y Average and RMS of Measurements VT = 230.473 +- 16.046 uVs I2Y = 21.204 +- 1.476 uTm^2 15:15:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:17:04 Wire Position: X0 = 50.0000 mm, Y0 = 0.0000 mm