SLAC Magnetic Measurements Date: 07-18-2025 Time: 09:42:50 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16353 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 12 Measurement Device: Operator: YL Comment: 09:42:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:43:01 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:43:03 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:43:04 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:43:06 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:43:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:43:21 Integrated Voltage Measurement... 09:43:30 Voltage Offset Correction: v_off_ini = 283.012 uV, v_off_fin = 281.339 uV 09:43:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.454 uVs 09:43:31 Integrated Voltage Measurement... 09:43:39 Voltage Offset Correction: v_off_ini = 274.347 uV, v_off_fin = 282.266 uV 09:43:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.356 uVs 09:43:40 Integrated Voltage Measurement... 09:43:49 Voltage Offset Correction: v_off_ini = 286.562 uV, v_off_fin = 288.796 uV 09:43:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.442 uVs 09:43:49 Integrated Voltage Measurement... 09:43:58 Voltage Offset Correction: v_off_ini = 289.976 uV, v_off_fin = 297.446 uV 09:43:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.222 uVs 09:43:58 Integrated Voltage Measurement... 09:44:07 Voltage Offset Correction: v_off_ini = 290.054 uV, v_off_fin = 291.090 uV 09:44:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.169 uVs 09:44:07 Integrated Voltage Measurement... 09:44:16 Voltage Offset Correction: v_off_ini = 307.140 uV, v_off_fin = 312.701 uV 09:44:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.297 uVs 09:44:20 I1X Average and RMS of Measurements VT = -62.157 +- 4.074 uVs I1X = 12.431 +- 0.815 uTm 09:44:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:44:27 Integrated Voltage Measurement... 09:44:36 Voltage Offset Correction: v_off_ini = 321.271 uV, v_off_fin = 316.232 uV 09:44:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -185.909 uVs 09:44:36 Integrated Voltage Measurement... 09:44:45 Voltage Offset Correction: v_off_ini = 319.460 uV, v_off_fin = 338.141 uV 09:44:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 202.641 uVs 09:44:45 Integrated Voltage Measurement... 09:44:54 Voltage Offset Correction: v_off_ini = 324.581 uV, v_off_fin = 316.041 uV 09:44:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -189.368 uVs 09:44:54 Integrated Voltage Measurement... 09:45:03 Voltage Offset Correction: v_off_ini = 318.229 uV, v_off_fin = 326.308 uV 09:45:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 202.697 uVs 09:45:03 Integrated Voltage Measurement... 09:45:12 Voltage Offset Correction: v_off_ini = 323.039 uV, v_off_fin = 308.377 uV 09:45:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -190.579 uVs 09:45:12 Integrated Voltage Measurement... 09:45:21 Voltage Offset Correction: v_off_ini = 316.195 uV, v_off_fin = 342.412 uV 09:45:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 196.607 uVs 09:45:25 I1Y Average and RMS of Measurements VT = -194.634 +- 1.028 uVs I1Y = -38.921 +- 0.206 uTm 09:45:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:45:33 Integrated Voltage Measurement... 09:45:42 Voltage Offset Correction: v_off_ini = 321.487 uV, v_off_fin = 317.785 uV 09:45:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -26.484 uVs 09:45:42 Integrated Voltage Measurement... 09:45:51 Voltage Offset Correction: v_off_ini = 328.005 uV, v_off_fin = 337.598 uV 09:45:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 29.844 uVs 09:45:51 Integrated Voltage Measurement... 09:46:00 Voltage Offset Correction: v_off_ini = 342.320 uV, v_off_fin = 322.897 uV 09:46:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -46.097 uVs 09:46:00 Integrated Voltage Measurement... 09:46:09 Voltage Offset Correction: v_off_ini = 342.670 uV, v_off_fin = 344.314 uV 09:46:09 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 45.695 uVs 09:46:09 Integrated Voltage Measurement... 09:46:18 Voltage Offset Correction: v_off_ini = 336.262 uV, v_off_fin = 338.396 uV 09:46:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -26.074 uVs 09:46:18 Integrated Voltage Measurement... 09:46:27 Voltage Offset Correction: v_off_ini = 355.711 uV, v_off_fin = 380.535 uV 09:46:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.424 uVs 09:46:29 I2X Average and RMS of Measurements VT = -36.436 +- 7.287 uVs I2X = 3.352 +- 0.671 uTm^2 09:46:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:46:37 Integrated Voltage Measurement... 09:46:46 Voltage Offset Correction: v_off_ini = 363.338 uV, v_off_fin = 365.400 uV 09:46:46 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -529.971 uVs 09:46:46 Integrated Voltage Measurement... 09:46:55 Voltage Offset Correction: v_off_ini = 368.738 uV, v_off_fin = 368.983 uV 09:46:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 542.997 uVs 09:46:55 Integrated Voltage Measurement... 09:47:04 Voltage Offset Correction: v_off_ini = 370.194 uV, v_off_fin = 366.381 uV 09:47:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -523.065 uVs 09:47:04 Integrated Voltage Measurement... 09:47:13 Voltage Offset Correction: v_off_ini = 361.605 uV, v_off_fin = 379.701 uV 09:47:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 533.181 uVs 09:47:13 Integrated Voltage Measurement... 09:47:22 Voltage Offset Correction: v_off_ini = 383.377 uV, v_off_fin = 383.290 uV 09:47:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -516.440 uVs 09:47:22 Integrated Voltage Measurement... 09:47:31 Voltage Offset Correction: v_off_ini = 385.405 uV, v_off_fin = 381.456 uV 09:47:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 542.302 uVs 09:47:34 I2Y Average and RMS of Measurements VT = -531.326 +- 3.683 uVs I2Y = -48.883 +- 0.340 uTm^2 09:47:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:47:49 Integrated Voltage Measurement... 09:47:57 Voltage Offset Correction: v_off_ini = 370.824 uV, v_off_fin = 365.628 uV 09:47:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.801 uVs 09:47:58 Integrated Voltage Measurement... 09:48:07 Voltage Offset Correction: v_off_ini = 377.865 uV, v_off_fin = 387.189 uV 09:48:07 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.599 uVs 09:48:07 Integrated Voltage Measurement... 09:48:16 Voltage Offset Correction: v_off_ini = 386.603 uV, v_off_fin = 359.948 uV 09:48:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.815 uVs 09:48:16 Integrated Voltage Measurement... 09:48:25 Voltage Offset Correction: v_off_ini = 364.628 uV, v_off_fin = 359.841 uV 09:48:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.110 uVs 09:48:25 Integrated Voltage Measurement... 09:48:34 Voltage Offset Correction: v_off_ini = 361.575 uV, v_off_fin = 348.950 uV 09:48:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.600 uVs 09:48:34 Integrated Voltage Measurement... 09:48:43 Voltage Offset Correction: v_off_ini = 353.400 uV, v_off_fin = 366.308 uV 09:48:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.926 uVs 09:48:47 I1X Average and RMS of Measurements VT = -77.975 +- 5.224 uVs I1X = 15.595 +- 1.045 uTm 09:48:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:48:54 Integrated Voltage Measurement... 09:49:03 Voltage Offset Correction: v_off_ini = 349.971 uV, v_off_fin = 349.332 uV 09:49:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.136 uVs 09:49:03 Integrated Voltage Measurement... 09:49:12 Voltage Offset Correction: v_off_ini = 347.337 uV, v_off_fin = 374.129 uV 09:49:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 206.134 uVs 09:49:12 Integrated Voltage Measurement... 09:49:21 Voltage Offset Correction: v_off_ini = 367.905 uV, v_off_fin = 359.811 uV 09:49:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -192.473 uVs 09:49:21 Integrated Voltage Measurement... 09:49:30 Voltage Offset Correction: v_off_ini = 374.947 uV, v_off_fin = 381.652 uV 09:49:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 193.576 uVs 09:49:30 Integrated Voltage Measurement... 09:49:39 Voltage Offset Correction: v_off_ini = 378.950 uV, v_off_fin = 357.246 uV 09:49:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -192.640 uVs 09:49:39 Integrated Voltage Measurement... 09:49:48 Voltage Offset Correction: v_off_ini = 378.178 uV, v_off_fin = 385.944 uV 09:49:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 200.335 uVs 09:49:52 I1Y Average and RMS of Measurements VT = -198.716 +- 5.776 uVs I1Y = -39.738 +- 1.156 uTm 09:49:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:50:00 Integrated Voltage Measurement... 09:50:09 Voltage Offset Correction: v_off_ini = 375.173 uV, v_off_fin = 387.360 uV 09:50:09 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -30.780 uVs 09:50:09 Integrated Voltage Measurement... 09:50:18 Voltage Offset Correction: v_off_ini = 396.961 uV, v_off_fin = 391.108 uV 09:50:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.972 uVs 09:50:18 Integrated Voltage Measurement... 09:50:27 Voltage Offset Correction: v_off_ini = 388.912 uV, v_off_fin = 393.069 uV 09:50:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -44.985 uVs 09:50:27 Integrated Voltage Measurement... 09:50:36 Voltage Offset Correction: v_off_ini = 405.596 uV, v_off_fin = 423.060 uV 09:50:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.572 uVs 09:50:36 Integrated Voltage Measurement... 09:50:45 Voltage Offset Correction: v_off_ini = 414.592 uV, v_off_fin = 393.054 uV 09:50:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.355 uVs 09:50:45 Integrated Voltage Measurement... 09:50:54 Voltage Offset Correction: v_off_ini = 414.262 uV, v_off_fin = 419.447 uV 09:50:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.457 uVs 09:50:56 I2X Average and RMS of Measurements VT = -45.853 +- 6.568 uVs I2X = 4.218 +- 0.604 uTm^2 09:51:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:51:04 Integrated Voltage Measurement... 09:51:13 Voltage Offset Correction: v_off_ini = 412.149 uV, v_off_fin = 402.724 uV 09:51:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -516.195 uVs 09:51:13 Integrated Voltage Measurement... 09:51:22 Voltage Offset Correction: v_off_ini = 418.330 uV, v_off_fin = 410.523 uV 09:51:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 516.925 uVs 09:51:22 Integrated Voltage Measurement... 09:51:31 Voltage Offset Correction: v_off_ini = 416.366 uV, v_off_fin = 408.987 uV 09:51:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -532.703 uVs 09:51:31 Integrated Voltage Measurement... 09:51:40 Voltage Offset Correction: v_off_ini = 414.559 uV, v_off_fin = 419.722 uV 09:51:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 525.148 uVs 09:51:40 Integrated Voltage Measurement... 09:51:49 Voltage Offset Correction: v_off_ini = 429.815 uV, v_off_fin = 429.850 uV 09:51:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -536.981 uVs 09:51:49 Integrated Voltage Measurement... 09:51:58 Voltage Offset Correction: v_off_ini = 425.455 uV, v_off_fin = 415.038 uV 09:51:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 533.158 uVs 09:52:01 I2Y Average and RMS of Measurements VT = -526.852 +- 7.698 uVs I2Y = -48.470 +- 0.708 uTm^2 09:52:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:52:16 Integrated Voltage Measurement... 09:52:25 Voltage Offset Correction: v_off_ini = 430.462 uV, v_off_fin = 426.723 uV 09:52:25 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -109.119 uVs 09:52:25 Integrated Voltage Measurement... 09:52:34 Voltage Offset Correction: v_off_ini = 432.347 uV, v_off_fin = 440.676 uV 09:52:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.800 uVs 09:52:34 Integrated Voltage Measurement... 09:52:43 Voltage Offset Correction: v_off_ini = 444.558 uV, v_off_fin = 437.377 uV 09:52:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.289 uVs 09:52:43 Integrated Voltage Measurement... 09:52:52 Voltage Offset Correction: v_off_ini = 439.434 uV, v_off_fin = 438.228 uV 09:52:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.873 uVs 09:52:52 Integrated Voltage Measurement... 09:53:01 Voltage Offset Correction: v_off_ini = 440.382 uV, v_off_fin = 438.578 uV 09:53:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.810 uVs 09:53:01 Integrated Voltage Measurement... 09:53:10 Voltage Offset Correction: v_off_ini = 448.413 uV, v_off_fin = 456.271 uV 09:53:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.152 uVs 09:53:14 I1X Average and RMS of Measurements VT = -87.340 +- 9.665 uVs I1X = 17.469 +- 1.934 uTm 09:53:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:53:21 Integrated Voltage Measurement... 09:53:30 Voltage Offset Correction: v_off_ini = 436.550 uV, v_off_fin = 440.471 uV 09:53:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.587 uVs 09:53:30 Integrated Voltage Measurement... 09:53:39 Voltage Offset Correction: v_off_ini = 448.488 uV, v_off_fin = 470.181 uV 09:53:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 198.764 uVs 09:53:39 Integrated Voltage Measurement... 09:53:48 Voltage Offset Correction: v_off_ini = 454.287 uV, v_off_fin = 437.202 uV 09:53:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.106 uVs 09:53:48 Integrated Voltage Measurement... 09:53:57 Voltage Offset Correction: v_off_ini = 450.184 uV, v_off_fin = 457.677 uV 09:53:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.589 uVs 09:53:57 Integrated Voltage Measurement...