SLAC Magnetic Measurements Date: 08-01-2025 Time: 10:25:22 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16351 Dataset Number: 1 Run Purpose: Final Results Run Type: Stretched Wire Run Number: 1 Measurement Device: Operator: YL Comment: On Axis 10:25:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:25:32 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:25:33 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:25:35 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:25:36 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:25:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:25:55 Integrated Voltage Measurement... 10:26:04 Voltage Offset Correction: v_off_ini = -179.553 uV, v_off_fin = -178.564 uV 10:26:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.341 uVs 10:26:04 Integrated Voltage Measurement... 10:26:13 Voltage Offset Correction: v_off_ini = -170.947 uV, v_off_fin = -162.463 uV 10:26:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.101 uVs 10:26:13 Integrated Voltage Measurement... 10:26:22 Voltage Offset Correction: v_off_ini = -166.284 uV, v_off_fin = -173.093 uV 10:26:22 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -109.723 uVs 10:26:22 Integrated Voltage Measurement... 10:26:31 Voltage Offset Correction: v_off_ini = -166.203 uV, v_off_fin = -162.608 uV 10:26:31 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 107.191 uVs 10:26:31 Integrated Voltage Measurement... 10:26:40 Voltage Offset Correction: v_off_ini = -153.312 uV, v_off_fin = -154.658 uV 10:26:40 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -139.458 uVs 10:26:40 Integrated Voltage Measurement... 10:26:49 Voltage Offset Correction: v_off_ini = -168.527 uV, v_off_fin = -148.531 uV 10:26:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 108.348 uVs 10:26:53 I1X Average and RMS of Measurements VT = -110.694 +- 9.998 uVs I1X = 22.141 +- 2.000 uTm 10:26:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:27:00 Integrated Voltage Measurement... 10:27:09 Voltage Offset Correction: v_off_ini = -138.393 uV, v_off_fin = -146.014 uV 10:27:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -239.655 uVs 10:27:09 Integrated Voltage Measurement... 10:27:18 Voltage Offset Correction: v_off_ini = -127.385 uV, v_off_fin = -116.856 uV 10:27:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 213.002 uVs 10:27:18 Integrated Voltage Measurement... 10:27:27 Voltage Offset Correction: v_off_ini = -120.184 uV, v_off_fin = -136.787 uV 10:27:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -236.571 uVs 10:27:27 Integrated Voltage Measurement... 10:27:36 Voltage Offset Correction: v_off_ini = -124.697 uV, v_off_fin = -127.180 uV 10:27:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 227.408 uVs 10:27:36 Integrated Voltage Measurement... 10:27:45 Voltage Offset Correction: v_off_ini = -133.610 uV, v_off_fin = -132.161 uV 10:27:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.079 uVs 10:27:45 Integrated Voltage Measurement... 10:27:54 Voltage Offset Correction: v_off_ini = -107.298 uV, v_off_fin = -92.594 uV 10:27:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 221.649 uVs 10:27:58 I1Y Average and RMS of Measurements VT = -228.061 +- 2.784 uVs I1Y = -45.612 +- 0.555 uTm 10:28:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:28:06 Integrated Voltage Measurement... 10:28:15 Voltage Offset Correction: v_off_ini = -108.802 uV, v_off_fin = -110.536 uV 10:28:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.264 uVs 10:28:15 Integrated Voltage Measurement... 10:28:24 Voltage Offset Correction: v_off_ini = -90.372 uV, v_off_fin = -66.177 uV 10:28:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.962 uVs 10:28:24 Integrated Voltage Measurement... 10:28:33 Voltage Offset Correction: v_off_ini = -60.150 uV, v_off_fin = -80.028 uV 10:28:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.321 uVs 10:28:33 Integrated Voltage Measurement... 10:28:42 Voltage Offset Correction: v_off_ini = -72.468 uV, v_off_fin = -72.933 uV 10:28:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.860 uVs 10:28:42 Integrated Voltage Measurement... 10:28:51 Voltage Offset Correction: v_off_ini = -81.974 uV, v_off_fin = -85.838 uV 10:28:51 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -52.810 uVs 10:28:51 Integrated Voltage Measurement... 10:29:00 Voltage Offset Correction: v_off_ini = -69.898 uV, v_off_fin = -52.280 uV 10:29:00 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 60.490 uVs 10:29:03 I2X Average and RMS of Measurements VT = -68.618 +- 9.074 uVs I2X = 6.313 +- 0.835 uTm^2 10:29:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:29:10 Integrated Voltage Measurement... 10:29:19 Voltage Offset Correction: v_off_ini = -56.214 uV, v_off_fin = -63.841 uV 10:29:19 Integrated Voltage Measurement dXYus = 1.0001 mm, dXYds = 0.0000 mm dXY = 1.0001 mm VT = -309.154 uVs 10:29:19 Integrated Voltage Measurement... 10:29:28 Voltage Offset Correction: v_off_ini = -48.687 uV, v_off_fin = -47.139 uV 10:29:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 287.932 uVs 10:29:28 Integrated Voltage Measurement... 10:29:37 Voltage Offset Correction: v_off_ini = -46.595 uV, v_off_fin = -50.755 uV 10:29:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -303.740 uVs 10:29:37 Integrated Voltage Measurement... 10:29:46 Voltage Offset Correction: v_off_ini = -41.009 uV, v_off_fin = -46.370 uV 10:29:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 303.069 uVs 10:29:46 Integrated Voltage Measurement... 10:29:55 Voltage Offset Correction: v_off_ini = -33.800 uV, v_off_fin = -31.602 uV 10:29:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -307.863 uVs 10:29:56 Integrated Voltage Measurement... 10:30:04 Voltage Offset Correction: v_off_ini = -44.757 uV, v_off_fin = -43.109 uV 10:30:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.346 uVs 10:30:07 I2Y Average and RMS of Measurements VT = -304.684 +- 5.610 uVs I2Y = -28.030 +- 0.517 uTm^2 10:30:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:30:22 Integrated Voltage Measurement... 10:30:31 Voltage Offset Correction: v_off_ini = -29.193 uV, v_off_fin = -43.940 uV 10:30:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -97.009 uVs 10:30:31 Integrated Voltage Measurement... 10:30:40 Voltage Offset Correction: v_off_ini = -41.600 uV, v_off_fin = -34.911 uV 10:30:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.278 uVs 10:30:40 Integrated Voltage Measurement... 10:30:49 Voltage Offset Correction: v_off_ini = -19.295 uV, v_off_fin = -25.911 uV 10:30:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -133.485 uVs 10:30:49 Integrated Voltage Measurement... 10:30:58 Voltage Offset Correction: v_off_ini = -25.039 uV, v_off_fin = -39.490 uV 10:30:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 131.377 uVs 10:30:58 Integrated Voltage Measurement... 10:31:07 Voltage Offset Correction: v_off_ini = -23.599 uV, v_off_fin = -16.146 uV 10:31:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.530 uVs 10:31:07 Integrated Voltage Measurement... 10:31:16 Voltage Offset Correction: v_off_ini = 7.806 uV, v_off_fin = 11.266 uV 10:31:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.148 uVs 10:31:20 I1X Average and RMS of Measurements VT = -118.471 +- 9.896 uVs I1X = 23.695 +- 1.979 uTm 10:31:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:31:27 Integrated Voltage Measurement... 10:31:36 Voltage Offset Correction: v_off_ini = -1.366 uV, v_off_fin = 6.765 uV 10:31:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.319 uVs 10:31:36 Integrated Voltage Measurement... 10:31:45 Voltage Offset Correction: v_off_ini = 19.800 uV, v_off_fin = 26.329 uV 10:31:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 205.419 uVs 10:31:45 Integrated Voltage Measurement... 10:31:54 Voltage Offset Correction: v_off_ini = -1.458 uV, v_off_fin = -10.166 uV 10:31:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -225.912 uVs 10:31:54 Integrated Voltage Measurement... 10:32:03 Voltage Offset Correction: v_off_ini = -5.976 uV, v_off_fin = 24.091 uV 10:32:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 222.611 uVs 10:32:03 Integrated Voltage Measurement... 10:32:12 Voltage Offset Correction: v_off_ini = 5.054 uV, v_off_fin = 15.803 uV 10:32:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -218.496 uVs 10:32:12 Integrated Voltage Measurement... 10:32:21 Voltage Offset Correction: v_off_ini = 25.191 uV, v_off_fin = 28.254 uV 10:32:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.002 uVs 10:32:25 I1Y Average and RMS of Measurements VT = -216.960 +- 5.175 uVs I1Y = -43.390 +- 1.034 uTm 10:32:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:32:33 Integrated Voltage Measurement... 10:32:42 Voltage Offset Correction: v_off_ini = 29.935 uV, v_off_fin = 27.114 uV 10:32:42 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -83.281 uVs 10:32:42 Integrated Voltage Measurement... 10:32:51 Voltage Offset Correction: v_off_ini = 41.374 uV, v_off_fin = 33.863 uV 10:32:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.611 uVs 10:32:51 Integrated Voltage Measurement... 10:33:00 Voltage Offset Correction: v_off_ini = 25.199 uV, v_off_fin = 15.473 uV 10:33:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.841 uVs 10:33:00 Integrated Voltage Measurement... 10:33:09 Voltage Offset Correction: v_off_ini = 31.867 uV, v_off_fin = 43.995 uV 10:33:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.344 uVs 10:33:09 Integrated Voltage Measurement... 10:33:18 Voltage Offset Correction: v_off_ini = 34.982 uV, v_off_fin = 41.609 uV 10:33:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.250 uVs 10:33:18 Integrated Voltage Measurement... 10:33:27 Voltage Offset Correction: v_off_ini = 46.210 uV, v_off_fin = 48.444 uV 10:33:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.440 uVs 10:33:29 I2X Average and RMS of Measurements VT = -79.294 +- 7.138 uVs I2X = 7.295 +- 0.657 uTm^2 10:33:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:33:37 Integrated Voltage Measurement... 10:33:46 Voltage Offset Correction: v_off_ini = 44.097 uV, v_off_fin = 26.421 uV 10:33:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -285.556 uVs 10:33:46 Integrated Voltage Measurement... 10:33:55 Voltage Offset Correction: v_off_ini = 47.532 uV, v_off_fin = 63.604 uV 10:33:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.060 uVs 10:33:55 Integrated Voltage Measurement... 10:34:04 Voltage Offset Correction: v_off_ini = 28.648 uV, v_off_fin = 42.313 uV 10:34:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -280.517 uVs 10:34:04 Integrated Voltage Measurement... 10:34:13 Voltage Offset Correction: v_off_ini = 41.643 uV, v_off_fin = 61.391 uV 10:34:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 277.261 uVs 10:34:13 Integrated Voltage Measurement... 10:34:22 Voltage Offset Correction: v_off_ini = 63.294 uV, v_off_fin = 62.616 uV 10:34:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -307.426 uVs 10:34:22 Integrated Voltage Measurement... 10:34:31 Voltage Offset Correction: v_off_ini = 59.034 uV, v_off_fin = 54.534 uV 10:34:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.485 uVs 10:34:34 I2Y Average and RMS of Measurements VT = -285.217 +- 5.829 uVs I2Y = -26.240 +- 0.536 uTm^2 10:34:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:34:49 Integrated Voltage Measurement... 10:34:58 Voltage Offset Correction: v_off_ini = 64.944 uV, v_off_fin = 64.047 uV 10:34:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -137.802 uVs 10:34:58 Integrated Voltage Measurement... 10:35:07 Voltage Offset Correction: v_off_ini = 61.038 uV, v_off_fin = 67.512 uV 10:35:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 132.690 uVs 10:35:07 Integrated Voltage Measurement... 10:35:16 Voltage Offset Correction: v_off_ini = 66.133 uV, v_off_fin = 65.807 uV 10:35:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -139.076 uVs 10:35:16 Integrated Voltage Measurement... 10:35:25 Voltage Offset Correction: v_off_ini = 77.502 uV, v_off_fin = 80.926 uV 10:35:25 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 132.569 uVs 10:35:25 Integrated Voltage Measurement... 10:35:34 Voltage Offset Correction: v_off_ini = 75.853 uV, v_off_fin = 86.844 uV 10:35:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -115.169 uVs 10:35:34 Integrated Voltage Measurement... 10:35:43 Voltage Offset Correction: v_off_ini = 93.874 uV, v_off_fin = 106.223 uV 10:35:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.235 uVs 10:35:47 I1X Average and RMS of Measurements VT = -128.090 +- 10.530 uVs I1X = 25.620 +- 2.107 uTm 10:35:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:35:54 Integrated Voltage Measurement... 10:36:03 Voltage Offset Correction: v_off_ini = 87.041 uV, v_off_fin = 84.209 uV 10:36:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -223.707 uVs 10:36:03 Integrated Voltage Measurement... 10:36:12 Voltage Offset Correction: v_off_ini = 88.081 uV, v_off_fin = 97.461 uV 10:36:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.028 uVs 10:36:12 Integrated Voltage Measurement... 10:36:21 Voltage Offset Correction: v_off_ini = 84.205 uV, v_off_fin = 95.615 uV 10:36:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -214.506 uVs 10:36:21 Integrated Voltage Measurement... 10:36:30 Voltage Offset Correction: v_off_ini = 81.955 uV, v_off_fin = 110.431 uV 10:36:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 195.005 uVs 10:36:30 Integrated Voltage Measurement... 10:36:39 Voltage Offset Correction: v_off_ini = 114.309 uV, v_off_fin = 95.502 uV 10:36:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -223.659 uVs 10:36:39 Integrated Voltage Measurement... 10:36:48 Voltage Offset Correction: v_off_ini = 85.321 uV, v_off_fin = 94.418 uV 10:36:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 219.675 uVs 10:36:52 I1Y Average and RMS of Measurements VT = -212.263 +- 7.033 uVs I1Y = -42.450 +- 1.407 uTm 10:36:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:37:00 Integrated Voltage Measurement... 10:37:09 Voltage Offset Correction: v_off_ini = 85.136 uV, v_off_fin = 75.425 uV 10:37:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.577 uVs 10:37:09 Integrated Voltage Measurement... 10:37:18 Voltage Offset Correction: v_off_ini = 93.555 uV, v_off_fin = 115.315 uV 10:37:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.844 uVs 10:37:18 Integrated Voltage Measurement... 10:37:27 Voltage Offset Correction: v_off_ini = 87.095 uV, v_off_fin = 84.211 uV 10:37:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.800 uVs 10:37:27 Integrated Voltage Measurement... 10:37:36 Voltage Offset Correction: v_off_ini = 88.249 uV, v_off_fin = 99.848 uV 10:37:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.576 uVs 10:37:36 Integrated Voltage Measurement... 10:37:45 Voltage Offset Correction: v_off_ini = 99.698 uV, v_off_fin = 72.594 uV 10:37:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.865 uVs 10:37:45 Integrated Voltage Measurement... 10:37:54 Voltage Offset Correction: v_off_ini = 78.683 uV, v_off_fin = 92.735 uV 10:37:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.795 uVs 10:37:56 I2X Average and RMS of Measurements VT = -77.576 +- 4.443 uVs I2X = 7.137 +- 0.409 uTm^2 10:38:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:38:04 Integrated Voltage Measurement... 10:38:13 Voltage Offset Correction: v_off_ini = 90.471 uV, v_off_fin = 90.065 uV 10:38:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -264.764 uVs 10:38:13 Integrated Voltage Measurement... 10:38:22 Voltage Offset Correction: v_off_ini = 99.930 uV, v_off_fin = 106.757 uV 10:38:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.929 uVs 10:38:22 Integrated Voltage Measurement... 10:38:31 Voltage Offset Correction: v_off_ini = 93.556 uV, v_off_fin = 92.005 uV 10:38:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -273.410 uVs 10:38:31 Integrated Voltage Measurement... 10:38:40 Voltage Offset Correction: v_off_ini = 113.300 uV, v_off_fin = 126.590 uV 10:38:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 288.586 uVs 10:38:40 Integrated Voltage Measurement... 10:38:49 Voltage Offset Correction: v_off_ini = 121.507 uV, v_off_fin = 125.029 uV 10:38:49 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -273.693 uVs 10:38:49 Integrated Voltage Measurement... 10:38:58 Voltage Offset Correction: v_off_ini = 117.498 uV, v_off_fin = 107.506 uV 10:38:58 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 286.326 uVs 10:39:01 I2Y Average and RMS of Measurements VT = -275.785 +- 6.686 uVs I2Y = -25.373 +- 0.616 uTm^2 10:39:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:39:16 Integrated Voltage Measurement... 10:39:25 Voltage Offset Correction: v_off_ini = 151.200 uV, v_off_fin = 142.045 uV 10:39:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -140.111 uVs 10:39:25 Integrated Voltage Measurement... 10:39:34 Voltage Offset Correction: v_off_ini = 127.891 uV, v_off_fin = 127.959 uV 10:39:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.916 uVs 10:39:34 Integrated Voltage Measurement... 10:39:43 Voltage Offset Correction: v_off_ini = 126.041 uV, v_off_fin = 131.916 uV 10:39:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -112.614 uVs 10:39:43 Integrated Voltage Measurement... 10:39:52 Voltage Offset Correction: v_off_ini = 139.982 uV, v_off_fin = 132.802 uV 10:39:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 116.741 uVs 10:39:52 Integrated Voltage Measurement... 10:40:01 Voltage Offset Correction: v_off_ini = 129.374 uV, v_off_fin = 134.528 uV 10:40:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -133.447 uVs 10:40:01 Integrated Voltage Measurement... 10:40:10 Voltage Offset Correction: v_off_ini = 142.122 uV, v_off_fin = 119.685 uV 10:40:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.970 uVs 10:40:14 I1X Average and RMS of Measurements VT = -123.800 +- 7.287 uVs I1X = 24.760 +- 1.457 uTm 10:40:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:40:21 Integrated Voltage Measurement... 10:40:30 Voltage Offset Correction: v_off_ini = 154.278 uV, v_off_fin = 132.942 uV 10:40:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.213 uVs 10:40:30 Integrated Voltage Measurement... 10:40:39 Voltage Offset Correction: v_off_ini = 121.618 uV, v_off_fin = 137.128 uV 10:40:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.116 uVs 10:40:39 Integrated Voltage Measurement... 10:40:48 Voltage Offset Correction: v_off_ini = 119.106 uV, v_off_fin = 123.244 uV 10:40:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.596 uVs 10:40:48 Integrated Voltage Measurement... 10:40:57 Voltage Offset Correction: v_off_ini = 133.052 uV, v_off_fin = 144.673 uV 10:40:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.222 uVs 10:40:57 Integrated Voltage Measurement... 10:41:06 Voltage Offset Correction: v_off_ini = 130.297 uV, v_off_fin = 130.136 uV 10:41:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -216.229 uVs 10:41:06 Integrated Voltage Measurement... 10:41:15 Voltage Offset Correction: v_off_ini = 127.128 uV, v_off_fin = 132.997 uV 10:41:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.018 uVs 10:41:19 I1Y Average and RMS of Measurements VT = -205.232 +- 4.583 uVs I1Y = -41.045 +- 0.917 uTm 10:41:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:41:27 Integrated Voltage Measurement... 10:41:36 Voltage Offset Correction: v_off_ini = 123.212 uV, v_off_fin = 120.158 uV 10:41:36 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -46.195 uVs 10:41:36 Integrated Voltage Measurement... 10:41:45 Voltage Offset Correction: v_off_ini = 126.438 uV, v_off_fin = 139.696 uV 10:41:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.193 uVs 10:41:45 Integrated Voltage Measurement... 10:41:54 Voltage Offset Correction: v_off_ini = 124.610 uV, v_off_fin = 129.044 uV 10:41:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.881 uVs 10:41:54 Integrated Voltage Measurement... 10:42:03 Voltage Offset Correction: v_off_ini = 119.091 uV, v_off_fin = 119.585 uV 10:42:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.260 uVs 10:42:03 Integrated Voltage Measurement... 10:42:12 Voltage Offset Correction: v_off_ini = 105.573 uV, v_off_fin = 111.712 uV 10:42:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -70.404 uVs 10:42:12 Integrated Voltage Measurement... 10:42:21 Voltage Offset Correction: v_off_ini = 116.713 uV, v_off_fin = 120.146 uV 10:42:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.984 uVs 10:42:23 I2X Average and RMS of Measurements VT = -70.153 +- 10.213 uVs I2X = 6.454 +- 0.940 uTm^2 10:42:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:42:31 Integrated Voltage Measurement... 10:42:40 Voltage Offset Correction: v_off_ini = 113.465 uV, v_off_fin = 108.990 uV 10:42:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -270.325 uVs 10:42:40 Integrated Voltage Measurement... 10:42:49 Voltage Offset Correction: v_off_ini = 116.652 uV, v_off_fin = 114.154 uV 10:42:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.248 uVs 10:42:49 Integrated Voltage Measurement... 10:42:58 Voltage Offset Correction: v_off_ini = 112.448 uV, v_off_fin = 118.761 uV 10:42:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.144 uVs 10:42:58 Integrated Voltage Measurement... 10:43:07 Voltage Offset Correction: v_off_ini = 105.977 uV, v_off_fin = 116.449 uV 10:43:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 254.153 uVs 10:43:07 Integrated Voltage Measurement... 10:43:16 Voltage Offset Correction: v_off_ini = 117.380 uV, v_off_fin = 115.156 uV 10:43:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -258.689 uVs 10:43:16 Integrated Voltage Measurement... 10:43:25 Voltage Offset Correction: v_off_ini = 130.143 uV, v_off_fin = 115.366 uV 10:43:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.351 uVs 10:43:27 I2Y Average and RMS of Measurements VT = -257.318 +- 7.567 uVs I2Y = -23.673 +- 0.696 uTm^2 10:43:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:43:42 Integrated Voltage Measurement... 10:43:51 Voltage Offset Correction: v_off_ini = 133.522 uV, v_off_fin = 129.360 uV 10:43:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -136.887 uVs 10:43:51 Integrated Voltage Measurement... 10:44:00 Voltage Offset Correction: v_off_ini = 120.399 uV, v_off_fin = 126.259 uV 10:44:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 125.961 uVs 10:44:00 Integrated Voltage Measurement... 10:44:09 Voltage Offset Correction: v_off_ini = 140.107 uV, v_off_fin = 130.370 uV 10:44:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.797 uVs 10:44:09 Integrated Voltage Measurement... 10:44:18 Voltage Offset Correction: v_off_ini = 131.370 uV, v_off_fin = 136.673 uV 10:44:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.579 uVs 10:44:18 Integrated Voltage Measurement... 10:44:27 Voltage Offset Correction: v_off_ini = 134.013 uV, v_off_fin = 117.367 uV 10:44:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.962 uVs 10:44:28 Integrated Voltage Measurement... 10:44:36 Voltage Offset Correction: v_off_ini = 130.497 uV, v_off_fin = 131.289 uV 10:44:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 126.744 uVs 10:44:41 I1X Average and RMS of Measurements VT = -120.655 +- 10.077 uVs I1X = 24.131 +- 2.016 uTm 10:44:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:44:47 Integrated Voltage Measurement... 10:44:56 Voltage Offset Correction: v_off_ini = 126.109 uV, v_off_fin = 119.763 uV 10:44:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.297 uVs 10:44:57 Integrated Voltage Measurement... 10:45:05 Voltage Offset Correction: v_off_ini = 134.877 uV, v_off_fin = 149.614 uV 10:45:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 188.015 uVs 10:45:05 Integrated Voltage Measurement... 10:45:14 Voltage Offset Correction: v_off_ini = 137.183 uV, v_off_fin = 125.709 uV 10:45:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.813 uVs 10:45:14 Integrated Voltage Measurement... 10:45:23 Voltage Offset Correction: v_off_ini = 138.939 uV, v_off_fin = 153.033 uV 10:45:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.959 uVs 10:45:23 Integrated Voltage Measurement... 10:45:32 Voltage Offset Correction: v_off_ini = 143.676 uV, v_off_fin = 132.542 uV 10:45:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.657 uVs 10:45:32 Integrated Voltage Measurement... 10:45:41 Voltage Offset Correction: v_off_ini = 137.264 uV, v_off_fin = 156.316 uV 10:45:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 205.841 uVs 10:45:46 I1Y Average and RMS of Measurements VT = -194.430 +- 7.089 uVs I1Y = -38.884 +- 1.418 uTm 10:45:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:45:53 Integrated Voltage Measurement... 10:46:02 Voltage Offset Correction: v_off_ini = 160.892 uV, v_off_fin = 159.469 uV 10:46:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.567 uVs 10:46:02 Integrated Voltage Measurement... 10:46:11 Voltage Offset Correction: v_off_ini = 173.849 uV, v_off_fin = 179.161 uV 10:46:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.798 uVs 10:46:11 Integrated Voltage Measurement... 10:46:20 Voltage Offset Correction: v_off_ini = 178.273 uV, v_off_fin = 169.038 uV 10:46:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.670 uVs 10:46:20 Integrated Voltage Measurement... 10:46:29 Voltage Offset Correction: v_off_ini = 174.089 uV, v_off_fin = 180.949 uV 10:46:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.659 uVs 10:46:29 Integrated Voltage Measurement... 10:46:38 Voltage Offset Correction: v_off_ini = 175.035 uV, v_off_fin = 165.409 uV 10:46:38 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -69.557 uVs 10:46:38 Integrated Voltage Measurement... 10:46:47 Voltage Offset Correction: v_off_ini = 166.984 uV, v_off_fin = 186.008 uV 10:46:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.823 uVs 10:46:50 I2X Average and RMS of Measurements VT = -77.346 +- 2.621 uVs I2X = 7.116 +- 0.241 uTm^2 10:46:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:46:58 Integrated Voltage Measurement... 10:47:07 Voltage Offset Correction: v_off_ini = 167.908 uV, v_off_fin = 169.056 uV 10:47:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -234.303 uVs 10:47:07 Integrated Voltage Measurement... 10:47:16 Voltage Offset Correction: v_off_ini = 177.686 uV, v_off_fin = 169.945 uV 10:47:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.122 uVs 10:47:16 Integrated Voltage Measurement... 10:47:25 Voltage Offset Correction: v_off_ini = 183.245 uV, v_off_fin = 172.581 uV 10:47:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.734 uVs 10:47:25 Integrated Voltage Measurement... 10:47:34 Voltage Offset Correction: v_off_ini = 175.806 uV, v_off_fin = 183.916 uV 10:47:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.896 uVs 10:47:34 Integrated Voltage Measurement... 10:47:43 Voltage Offset Correction: v_off_ini = 183.468 uV, v_off_fin = 179.669 uV 10:47:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.187 uVs 10:47:43 Integrated Voltage Measurement... 10:47:52 Voltage Offset Correction: v_off_ini = 187.424 uV, v_off_fin = 212.258 uV 10:47:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.716 uVs 10:47:54 I2Y Average and RMS of Measurements VT = -251.159 +- 4.559 uVs I2Y = -23.107 +- 0.419 uTm^2 10:48:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:48:09 Integrated Voltage Measurement... 10:48:18 Voltage Offset Correction: v_off_ini = 193.508 uV, v_off_fin = 173.086 uV 10:48:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.580 uVs 10:48:18 Integrated Voltage Measurement... 10:48:27 Voltage Offset Correction: v_off_ini = 188.370 uV, v_off_fin = 187.183 uV 10:48:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.498 uVs 10:48:27 Integrated Voltage Measurement... 10:48:36 Voltage Offset Correction: v_off_ini = 183.488 uV, v_off_fin = 183.742 uV 10:48:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.491 uVs 10:48:36 Integrated Voltage Measurement... 10:48:45 Voltage Offset Correction: v_off_ini = 180.474 uV, v_off_fin = 185.257 uV 10:48:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 104.005 uVs 10:48:45 Integrated Voltage Measurement... 10:48:54 Voltage Offset Correction: v_off_ini = 182.511 uV, v_off_fin = 194.001 uV 10:48:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -126.141 uVs 10:48:54 Integrated Voltage Measurement... 10:49:03 Voltage Offset Correction: v_off_ini = 199.767 uV, v_off_fin = 195.129 uV 10:49:03 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.584 uVs 10:49:07 I1X Average and RMS of Measurements VT = -112.050 +- 1.322 uVs I1X = 22.410 +- 0.265 uTm 10:49:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:49:14 Integrated Voltage Measurement... 10:49:23 Voltage Offset Correction: v_off_ini = 191.424 uV, v_off_fin = 190.138 uV 10:49:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -187.509 uVs 10:49:23 Integrated Voltage Measurement... 10:49:32 Voltage Offset Correction: v_off_ini = 179.541 uV, v_off_fin = 191.731 uV 10:49:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 200.111 uVs 10:49:32 Integrated Voltage Measurement... 10:49:41 Voltage Offset Correction: v_off_ini = 181.527 uV, v_off_fin = 171.266 uV 10:49:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.299 uVs 10:49:41 Integrated Voltage Measurement... 10:49:50 Voltage Offset Correction: v_off_ini = 194.811 uV, v_off_fin = 192.184 uV 10:49:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.964 uVs 10:49:50 Integrated Voltage Measurement... 10:49:59 Voltage Offset Correction: v_off_ini = 186.689 uV, v_off_fin = 173.289 uV 10:49:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.070 uVs 10:49:59 Integrated Voltage Measurement... 10:50:08 Voltage Offset Correction: v_off_ini = 173.920 uV, v_off_fin = 193.594 uV 10:50:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 184.461 uVs 10:50:12 I1Y Average and RMS of Measurements VT = -189.902 +- 2.965 uVs I1Y = -37.979 +- 0.594 uTm 10:50:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:50:20 Integrated Voltage Measurement... 10:50:29 Voltage Offset Correction: v_off_ini = 153.051 uV, v_off_fin = 153.359 uV 10:50:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.630 uVs 10:50:29 Integrated Voltage Measurement... 10:50:38 Voltage Offset Correction: v_off_ini = 177.260 uV, v_off_fin = 199.669 uV 10:50:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.490 uVs 10:50:38 Integrated Voltage Measurement... 10:50:47 Voltage Offset Correction: v_off_ini = 192.252 uV, v_off_fin = 175.498 uV 10:50:47 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -71.997 uVs 10:50:47 Integrated Voltage Measurement... 10:50:56 Voltage Offset Correction: v_off_ini = 182.959 uV, v_off_fin = 186.939 uV 10:50:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.488 uVs 10:50:56 Integrated Voltage Measurement... 10:51:05 Voltage Offset Correction: v_off_ini = 178.462 uV, v_off_fin = 171.450 uV 10:51:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.475 uVs 10:51:05 Integrated Voltage Measurement... 10:51:14 Voltage Offset Correction: v_off_ini = 163.325 uV, v_off_fin = 177.645 uV 10:51:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.576 uVs 10:51:16 I2X Average and RMS of Measurements VT = -79.443 +- 4.680 uVs I2X = 7.309 +- 0.430 uTm^2 10:51:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:51:24 Integrated Voltage Measurement... 10:51:33 Voltage Offset Correction: v_off_ini = 162.748 uV, v_off_fin = 175.263 uV 10:51:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -215.150 uVs 10:51:33 Integrated Voltage Measurement... 10:51:42 Voltage Offset Correction: v_off_ini = 163.196 uV, v_off_fin = 172.775 uV 10:51:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 226.851 uVs 10:51:42 Integrated Voltage Measurement... 10:51:51 Voltage Offset Correction: v_off_ini = 164.039 uV, v_off_fin = 160.930 uV 10:51:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -229.665 uVs 10:51:51 Integrated Voltage Measurement... 10:52:00 Voltage Offset Correction: v_off_ini = 178.424 uV, v_off_fin = 171.178 uV 10:52:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 219.677 uVs 10:52:00 Integrated Voltage Measurement... 10:52:09 Voltage Offset Correction: v_off_ini = 167.503 uV, v_off_fin = 158.549 uV 10:52:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -227.931 uVs 10:52:09 Integrated Voltage Measurement... 10:52:18 Voltage Offset Correction: v_off_ini = 150.987 uV, v_off_fin = 154.379 uV 10:52:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 243.135 uVs 10:52:21 I2Y Average and RMS of Measurements VT = -227.068 +- 6.170 uVs I2Y = -20.890 +- 0.568 uTm^2 10:52:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:52:36 Integrated Voltage Measurement... 10:52:45 Voltage Offset Correction: v_off_ini = 156.921 uV, v_off_fin = 153.486 uV 10:52:45 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -120.021 uVs 10:52:45 Integrated Voltage Measurement... 10:52:54 Voltage Offset Correction: v_off_ini = 147.657 uV, v_off_fin = 169.307 uV 10:52:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 97.400 uVs 10:52:54 Integrated Voltage Measurement... 10:53:03 Voltage Offset Correction: v_off_ini = 169.501 uV, v_off_fin = 166.678 uV 10:53:03 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -116.890 uVs 10:53:03 Integrated Voltage Measurement... 10:53:12 Voltage Offset Correction: v_off_ini = 160.284 uV, v_off_fin = 165.223 uV 10:53:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 110.697 uVs 10:53:12 Integrated Voltage Measurement... 10:53:21 Voltage Offset Correction: v_off_ini = 172.787 uV, v_off_fin = 169.419 uV 10:53:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.245 uVs 10:53:21 Integrated Voltage Measurement... 10:53:30 Voltage Offset Correction: v_off_ini = 168.647 uV, v_off_fin = 163.602 uV 10:53:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.019 uVs 10:53:34 I1X Average and RMS of Measurements VT = -107.546 +- 5.638 uVs I1X = 21.511 +- 1.129 uTm 10:53:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:53:41 Integrated Voltage Measurement... 10:53:50 Voltage Offset Correction: v_off_ini = 156.478 uV, v_off_fin = 155.106 uV 10:53:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -172.932 uVs 10:53:50 Integrated Voltage Measurement... 10:53:59 Voltage Offset Correction: v_off_ini = 153.979 uV, v_off_fin = 162.910 uV 10:53:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 193.394 uVs 10:53:59 Integrated Voltage Measurement... 10:54:08 Voltage Offset Correction: v_off_ini = 152.913 uV, v_off_fin = 159.454 uV 10:54:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -173.795 uVs 10:54:08 Integrated Voltage Measurement... 10:54:17 Voltage Offset Correction: v_off_ini = 168.716 uV, v_off_fin = 188.055 uV 10:54:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 181.894 uVs 10:54:17 Integrated Voltage Measurement... 10:54:26 Voltage Offset Correction: v_off_ini = 167.393 uV, v_off_fin = 161.957 uV 10:54:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -186.082 uVs 10:54:26 Integrated Voltage Measurement... 10:54:35 Voltage Offset Correction: v_off_ini = 168.043 uV, v_off_fin = 183.653 uV 10:54:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.299 uVs 10:54:39 I1Y Average and RMS of Measurements VT = -182.233 +- 3.270 uVs I1Y = -36.443 +- 0.654 uTm 10:54:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:54:47 Integrated Voltage Measurement... 10:54:56 Voltage Offset Correction: v_off_ini = 175.741 uV, v_off_fin = 157.905 uV 10:54:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.096 uVs 10:54:56 Integrated Voltage Measurement... 10:55:05 Voltage Offset Correction: v_off_ini = 164.242 uV, v_off_fin = 175.263 uV 10:55:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.129 uVs 10:55:05 Integrated Voltage Measurement... 10:55:14 Voltage Offset Correction: v_off_ini = 177.727 uV, v_off_fin = 177.318 uV 10:55:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.220 uVs 10:55:14 Integrated Voltage Measurement... 10:55:23 Voltage Offset Correction: v_off_ini = 187.623 uV, v_off_fin = 196.643 uV 10:55:23 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 93.588 uVs 10:55:23 Integrated Voltage Measurement... 10:55:32 Voltage Offset Correction: v_off_ini = 191.453 uV, v_off_fin = 186.409 uV 10:55:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.655 uVs 10:55:32 Integrated Voltage Measurement... 10:55:41 Voltage Offset Correction: v_off_ini = 193.435 uV, v_off_fin = 208.589 uV 10:55:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.938 uVs 10:55:43 I2X Average and RMS of Measurements VT = -79.938 +- 4.253 uVs I2X = 7.354 +- 0.391 uTm^2 10:55:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:55:51 Integrated Voltage Measurement... 10:56:00 Voltage Offset Correction: v_off_ini = 189.496 uV, v_off_fin = 194.135 uV 10:56:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -224.008 uVs 10:56:00 Integrated Voltage Measurement... 10:56:09 Voltage Offset Correction: v_off_ini = 200.488 uV, v_off_fin = 194.612 uV 10:56:09 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 241.839 uVs 10:56:09 Integrated Voltage Measurement... 10:56:18 Voltage Offset Correction: v_off_ini = 202.209 uV, v_off_fin = 205.561 uV 10:56:18 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -231.914 uVs 10:56:18 Integrated Voltage Measurement... 10:56:27 Voltage Offset Correction: v_off_ini = 206.071 uV, v_off_fin = 206.058 uV 10:56:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 231.510 uVs 10:56:27 Integrated Voltage Measurement... 10:56:36 Voltage Offset Correction: v_off_ini = 194.617 uV, v_off_fin = 201.102 uV 10:56:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.819 uVs 10:56:36 Integrated Voltage Measurement... 10:56:45 Voltage Offset Correction: v_off_ini = 210.227 uV, v_off_fin = 212.105 uV 10:56:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.610 uVs 10:56:47 I2Y Average and RMS of Measurements VT = -230.950 +- 1.997 uVs I2Y = -21.248 +- 0.184 uTm^2 10:56:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:57:02 Integrated Voltage Measurement... 10:57:11 Voltage Offset Correction: v_off_ini = 228.568 uV, v_off_fin = 213.130 uV 10:57:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -110.734 uVs 10:57:11 Integrated Voltage Measurement... 10:57:20 Voltage Offset Correction: v_off_ini = 235.597 uV, v_off_fin = 222.829 uV 10:57:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.019 uVs 10:57:20 Integrated Voltage Measurement... 10:57:29 Voltage Offset Correction: v_off_ini = 216.672 uV, v_off_fin = 228.971 uV 10:57:29 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -92.159 uVs 10:57:29 Integrated Voltage Measurement... 10:57:38 Voltage Offset Correction: v_off_ini = 210.161 uV, v_off_fin = 220.234 uV 10:57:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 132.461 uVs 10:57:38 Integrated Voltage Measurement... 10:57:47 Voltage Offset Correction: v_off_ini = 217.564 uV, v_off_fin = 224.661 uV 10:57:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -104.292 uVs 10:57:47 Integrated Voltage Measurement... 10:57:56 Voltage Offset Correction: v_off_ini = 233.030 uV, v_off_fin = 241.128 uV 10:57:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 106.626 uVs 10:58:01 I1X Average and RMS of Measurements VT = -111.048 +- 4.146 uVs I1X = 22.213 +- 0.830 uTm 10:58:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:58:07 Integrated Voltage Measurement... 10:58:16 Voltage Offset Correction: v_off_ini = 232.378 uV, v_off_fin = 235.735 uV 10:58:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.461 uVs 10:58:16 Integrated Voltage Measurement... 10:58:25 Voltage Offset Correction: v_off_ini = 246.960 uV, v_off_fin = 250.865 uV 10:58:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 190.601 uVs 10:58:25 Integrated Voltage Measurement... 10:58:34 Voltage Offset Correction: v_off_ini = 241.150 uV, v_off_fin = 233.794 uV 10:58:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.409 uVs 10:58:34 Integrated Voltage Measurement... 10:58:43 Voltage Offset Correction: v_off_ini = 244.338 uV, v_off_fin = 264.054 uV 10:58:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.582 uVs 10:58:43 Integrated Voltage Measurement... 10:58:52 Voltage Offset Correction: v_off_ini = 244.202 uV, v_off_fin = 239.445 uV 10:58:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.470 uVs 10:58:52 Integrated Voltage Measurement... 10:59:01 Voltage Offset Correction: v_off_ini = 245.237 uV, v_off_fin = 250.489 uV 10:59:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.651 uVs 10:59:05 I1Y Average and RMS of Measurements VT = -186.029 +- 2.666 uVs I1Y = -37.203 +- 0.532 uTm 10:59:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:59:13 Integrated Voltage Measurement... 10:59:22 Voltage Offset Correction: v_off_ini = 226.138 uV, v_off_fin = 226.656 uV 10:59:22 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -69.028 uVs 10:59:22 Integrated Voltage Measurement... 10:59:31 Voltage Offset Correction: v_off_ini = 231.676 uV, v_off_fin = 255.541 uV 10:59:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.696 uVs 10:59:31 Integrated Voltage Measurement... 10:59:40 Voltage Offset Correction: v_off_ini = 249.666 uV, v_off_fin = 220.720 uV 10:59:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.858 uVs 10:59:40 Integrated Voltage Measurement... 10:59:49 Voltage Offset Correction: v_off_ini = 233.911 uV, v_off_fin = 246.233 uV 10:59:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.094 uVs 10:59:49 Integrated Voltage Measurement... 10:59:58 Voltage Offset Correction: v_off_ini = 262.915 uV, v_off_fin = 237.817 uV 10:59:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.712 uVs 10:59:58 Integrated Voltage Measurement... 11:00:07 Voltage Offset Correction: v_off_ini = 236.195 uV, v_off_fin = 237.521 uV 11:00:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.781 uVs 11:00:10 I2X Average and RMS of Measurements VT = -73.362 +- 3.313 uVs I2X = 6.749 +- 0.305 uTm^2 11:00:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:00:17 Integrated Voltage Measurement... 11:00:26 Voltage Offset Correction: v_off_ini = 235.309 uV, v_off_fin = 240.443 uV 11:00:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -219.394 uVs 11:00:26 Integrated Voltage Measurement... 11:00:35 Voltage Offset Correction: v_off_ini = 246.882 uV, v_off_fin = 251.147 uV 11:00:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.274 uVs 11:00:35 Integrated Voltage Measurement... 11:00:44 Voltage Offset Correction: v_off_ini = 244.751 uV, v_off_fin = 246.806 uV 11:00:44 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -224.682 uVs 11:00:44 Integrated Voltage Measurement... 11:00:53 Voltage Offset Correction: v_off_ini = 251.608 uV, v_off_fin = 251.622 uV 11:00:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.802 uVs 11:00:54 Integrated Voltage Measurement... 11:01:02 Voltage Offset Correction: v_off_ini = 245.185 uV, v_off_fin = 234.122 uV 11:01:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -215.335 uVs 11:01:03 Integrated Voltage Measurement... 11:01:11 Voltage Offset Correction: v_off_ini = 243.759 uV, v_off_fin = 247.741 uV 11:01:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.860 uVs 11:01:14 I2Y Average and RMS of Measurements VT = -220.224 +- 2.774 uVs I2Y = -20.261 +- 0.255 uTm^2 11:01:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:01:29 Integrated Voltage Measurement... 11:01:38 Voltage Offset Correction: v_off_ini = 262.936 uV, v_off_fin = 251.497 uV 11:01:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -127.937 uVs 11:01:38 Integrated Voltage Measurement... 11:01:47 Voltage Offset Correction: v_off_ini = 255.481 uV, v_off_fin = 257.275 uV 11:01:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.221 uVs 11:01:47 Integrated Voltage Measurement... 11:01:56 Voltage Offset Correction: v_off_ini = 266.010 uV, v_off_fin = 270.219 uV 11:01:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -124.265 uVs 11:01:56 Integrated Voltage Measurement... 11:02:05 Voltage Offset Correction: v_off_ini = 256.662 uV, v_off_fin = 257.523 uV 11:02:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 103.910 uVs 11:02:05 Integrated Voltage Measurement... 11:02:14 Voltage Offset Correction: v_off_ini = 246.590 uV, v_off_fin = 241.028 uV 11:02:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -91.010 uVs 11:02:14 Integrated Voltage Measurement... 11:02:23 Voltage Offset Correction: v_off_ini = 247.809 uV, v_off_fin = 248.830 uV 11:02:23 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.835 uVs 11:02:27 I1X Average and RMS of Measurements VT = -107.030 +- 11.053 uVs I1X = 21.407 +- 2.210 uTm 11:02:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:02:34 Integrated Voltage Measurement... 11:02:43 Voltage Offset Correction: v_off_ini = 258.993 uV, v_off_fin = 247.342 uV 11:02:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.191 uVs 11:02:43 Integrated Voltage Measurement... 11:02:52 Voltage Offset Correction: v_off_ini = 247.582 uV, v_off_fin = 261.556 uV 11:02:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 195.270 uVs 11:02:52 Integrated Voltage Measurement... 11:03:01 Voltage Offset Correction: v_off_ini = 247.889 uV, v_off_fin = 251.374 uV 11:03:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -170.836 uVs 11:03:01 Integrated Voltage Measurement... 11:03:10 Voltage Offset Correction: v_off_ini = 272.854 uV, v_off_fin = 263.077 uV 11:03:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.922 uVs 11:03:10 Integrated Voltage Measurement... 11:03:19 Voltage Offset Correction: v_off_ini = 252.754 uV, v_off_fin = 239.020 uV 11:03:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -183.698 uVs 11:03:19 Integrated Voltage Measurement... 11:03:28 Voltage Offset Correction: v_off_ini = 241.669 uV, v_off_fin = 270.960 uV 11:03:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 179.328 uVs 11:03:32 I1Y Average and RMS of Measurements VT = -178.874 +- 6.539 uVs I1Y = -35.771 +- 1.307 uTm 11:03:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:03:40 Integrated Voltage Measurement... 11:03:49 Voltage Offset Correction: v_off_ini = 247.221 uV, v_off_fin = 253.492 uV 11:03:49 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -76.396 uVs 11:03:49 Integrated Voltage Measurement... 11:03:58 Voltage Offset Correction: v_off_ini = 269.450 uV, v_off_fin = 264.689 uV 11:03:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.677 uVs 11:03:58 Integrated Voltage Measurement... 11:04:07 Voltage Offset Correction: v_off_ini = 259.263 uV, v_off_fin = 245.017 uV 11:04:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.866 uVs 11:04:07 Integrated Voltage Measurement... 11:04:16 Voltage Offset Correction: v_off_ini = 250.887 uV, v_off_fin = 253.460 uV 11:04:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.275 uVs 11:04:16 Integrated Voltage Measurement... 11:04:25 Voltage Offset Correction: v_off_ini = 233.341 uV, v_off_fin = 230.769 uV 11:04:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.525 uVs 11:04:25 Integrated Voltage Measurement... 11:04:34 Voltage Offset Correction: v_off_ini = 266.037 uV, v_off_fin = 256.774 uV 11:04:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.488 uVs 11:04:36 I2X Average and RMS of Measurements VT = -74.871 +- 9.303 uVs I2X = 6.888 +- 0.856 uTm^2 11:04:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:04:44 Integrated Voltage Measurement... 11:04:53 Voltage Offset Correction: v_off_ini = 253.459 uV, v_off_fin = 251.909 uV 11:04:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -218.186 uVs 11:04:53 Integrated Voltage Measurement... 11:05:02 Voltage Offset Correction: v_off_ini = 258.445 uV, v_off_fin = 252.178 uV 11:05:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 214.914 uVs 11:05:02 Integrated Voltage Measurement... 11:05:11 Voltage Offset Correction: v_off_ini = 256.800 uV, v_off_fin = 245.609 uV 11:05:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -229.733 uVs 11:05:11 Integrated Voltage Measurement... 11:05:20 Voltage Offset Correction: v_off_ini = 253.919 uV, v_off_fin = 240.471 uV 11:05:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.656 uVs 11:05:20 Integrated Voltage Measurement... 11:05:29 Voltage Offset Correction: v_off_ini = 237.597 uV, v_off_fin = 240.876 uV 11:05:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -212.643 uVs 11:05:29 Integrated Voltage Measurement... 11:05:38 Voltage Offset Correction: v_off_ini = 238.304 uV, v_off_fin = 243.291 uV 11:05:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 214.720 uVs 11:05:40 I2Y Average and RMS of Measurements VT = -216.809 +- 2.665 uVs I2Y = -19.946 +- 0.245 uTm^2 11:05:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:05:55 Integrated Voltage Measurement... 11:06:04 Voltage Offset Correction: v_off_ini = 233.545 uV, v_off_fin = 232.554 uV 11:06:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -132.627 uVs 11:06:04 Integrated Voltage Measurement... 11:06:13 Voltage Offset Correction: v_off_ini = 240.638 uV, v_off_fin = 230.698 uV 11:06:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 123.842 uVs 11:06:13 Integrated Voltage Measurement... 11:06:22 Voltage Offset Correction: v_off_ini = 230.900 uV, v_off_fin = 232.515 uV 11:06:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -103.162 uVs 11:06:22 Integrated Voltage Measurement... 11:06:31 Voltage Offset Correction: v_off_ini = 246.346 uV, v_off_fin = 233.861 uV 11:06:31 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.641 uVs 11:06:31 Integrated Voltage Measurement... 11:06:40 Voltage Offset Correction: v_off_ini = 238.322 uV, v_off_fin = 238.454 uV 11:06:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.495 uVs 11:06:40 Integrated Voltage Measurement... 11:06:49 Voltage Offset Correction: v_off_ini = 223.834 uV, v_off_fin = 229.263 uV 11:06:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.546 uVs 11:06:54 I1X Average and RMS of Measurements VT = -112.052 +- 11.913 uVs I1X = 22.411 +- 2.382 uTm 11:06:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:07:00 Integrated Voltage Measurement... 11:07:09 Voltage Offset Correction: v_off_ini = 246.291 uV, v_off_fin = 235.718 uV 11:07:09 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -179.753 uVs 11:07:09 Integrated Voltage Measurement... 11:07:18 Voltage Offset Correction: v_off_ini = 251.030 uV, v_off_fin = 258.138 uV 11:07:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 163.885 uVs 11:07:18 Integrated Voltage Measurement... 11:07:27 Voltage Offset Correction: v_off_ini = 252.765 uV, v_off_fin = 228.052 uV 11:07:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -175.831 uVs 11:07:27 Integrated Voltage Measurement... 11:07:36 Voltage Offset Correction: v_off_ini = 232.484 uV, v_off_fin = 250.301 uV 11:07:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 167.790 uVs 11:07:36 Integrated Voltage Measurement... 11:07:45 Voltage Offset Correction: v_off_ini = 222.883 uV, v_off_fin = 214.300 uV 11:07:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -156.744 uVs 11:07:45 Integrated Voltage Measurement... 11:07:54 Voltage Offset Correction: v_off_ini = 237.113 uV, v_off_fin = 242.700 uV 11:07:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 173.990 uVs 11:07:59 I1Y Average and RMS of Measurements VT = -169.666 +- 3.039 uVs I1Y = -33.931 +- 0.607 uTm 11:08:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:08:06 Integrated Voltage Measurement... 11:08:15 Voltage Offset Correction: v_off_ini = 225.853 uV, v_off_fin = 233.358 uV 11:08:15 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -72.957 uVs 11:08:15 Integrated Voltage Measurement... 11:08:24 Voltage Offset Correction: v_off_ini = 234.409 uV, v_off_fin = 239.050 uV 11:08:24 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 81.484 uVs 11:08:24 Integrated Voltage Measurement... 11:08:33 Voltage Offset Correction: v_off_ini = 236.593 uV, v_off_fin = 227.772 uV 11:08:33 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -66.053 uVs 11:08:33 Integrated Voltage Measurement... 11:08:42 Voltage Offset Correction: v_off_ini = 263.846 uV, v_off_fin = 270.003 uV 11:08:42 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = 60.481 uVs 11:08:42 Integrated Voltage Measurement... 11:08:51 Voltage Offset Correction: v_off_ini = 259.912 uV, v_off_fin = 226.290 uV 11:08:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.552 uVs 11:08:51 Integrated Voltage Measurement... 11:09:00 Voltage Offset Correction: v_off_ini = 222.768 uV, v_off_fin = 248.383 uV 11:09:00 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 75.340 uVs 11:09:03 I2X Average and RMS of Measurements VT = -76.478 +- 10.497 uVs I2X = 7.036 +- 0.965 uTm^2 11:09:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:09:10 Integrated Voltage Measurement... 11:09:19 Voltage Offset Correction: v_off_ini = 245.506 uV, v_off_fin = 252.767 uV 11:09:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -204.258 uVs 11:09:19 Integrated Voltage Measurement... 11:09:28 Voltage Offset Correction: v_off_ini = 253.872 uV, v_off_fin = 262.142 uV 11:09:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.476 uVs 11:09:29 Integrated Voltage Measurement... 11:09:37 Voltage Offset Correction: v_off_ini = 252.230 uV, v_off_fin = 229.956 uV 11:09:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -217.354 uVs 11:09:38 Integrated Voltage Measurement... 11:09:47 Voltage Offset Correction: v_off_ini = 252.805 uV, v_off_fin = 251.269 uV 11:09:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 228.742 uVs 11:09:47 Integrated Voltage Measurement... 11:09:56 Voltage Offset Correction: v_off_ini = 247.858 uV, v_off_fin = 250.030 uV 11:09:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -212.363 uVs 11:09:56 Integrated Voltage Measurement... 11:10:05 Voltage Offset Correction: v_off_ini = 256.004 uV, v_off_fin = 253.075 uV 11:10:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.287 uVs 11:10:07 I2Y Average and RMS of Measurements VT = -213.580 +- 8.052 uVs I2Y = -19.649 +- 0.741 uTm^2 11:10:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:10:22 Integrated Voltage Measurement... 11:10:31 Voltage Offset Correction: v_off_ini = 257.682 uV, v_off_fin = 260.597 uV 11:10:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.520 uVs 11:10:31 Integrated Voltage Measurement... 11:10:40 Voltage Offset Correction: v_off_ini = 264.612 uV, v_off_fin = 271.515 uV 11:10:40 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.612 uVs 11:10:40 Integrated Voltage Measurement... 11:10:49 Voltage Offset Correction: v_off_ini = 263.875 uV, v_off_fin = 272.388 uV 11:10:49 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.297 uVs 11:10:49 Integrated Voltage Measurement... 11:10:58 Voltage Offset Correction: v_off_ini = 252.694 uV, v_off_fin = 254.363 uV 11:10:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 112.136 uVs 11:10:58 Integrated Voltage Measurement... 11:11:07 Voltage Offset Correction: v_off_ini = 275.492 uV, v_off_fin = 262.943 uV 11:11:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.856 uVs 11:11:07 Integrated Voltage Measurement... 11:11:16 Voltage Offset Correction: v_off_ini = 264.432 uV, v_off_fin = 272.368 uV 11:11:16 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.992 uVs 11:11:20 I1X Average and RMS of Measurements VT = -98.569 +- 3.922 uVs I1X = 19.715 +- 0.785 uTm 11:11:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:11:27 Integrated Voltage Measurement... 11:11:36 Voltage Offset Correction: v_off_ini = 285.592 uV, v_off_fin = 270.502 uV 11:11:36 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -172.992 uVs 11:11:36 Integrated Voltage Measurement... 11:11:45 Voltage Offset Correction: v_off_ini = 277.332 uV, v_off_fin = 285.612 uV 11:11:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 182.268 uVs 11:11:45 Integrated Voltage Measurement... 11:11:54 Voltage Offset Correction: v_off_ini = 273.535 uV, v_off_fin = 241.827 uV 11:11:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -160.594 uVs 11:11:54 Integrated Voltage Measurement... 11:12:03 Voltage Offset Correction: v_off_ini = 263.002 uV, v_off_fin = 272.289 uV 11:12:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 182.912 uVs 11:12:03 Integrated Voltage Measurement... 11:12:12 Voltage Offset Correction: v_off_ini = 277.016 uV, v_off_fin = 266.212 uV 11:12:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.528 uVs 11:12:12 Integrated Voltage Measurement... 11:12:21 Voltage Offset Correction: v_off_ini = 271.294 uV, v_off_fin = 283.890 uV 11:12:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 167.982 uVs 11:12:25 I1Y Average and RMS of Measurements VT = -175.713 +- 2.800 uVs I1Y = -35.142 +- 0.562 uTm 11:12:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:12:33 Integrated Voltage Measurement... 11:12:42 Voltage Offset Correction: v_off_ini = 273.690 uV, v_off_fin = 277.028 uV 11:12:42 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -74.557 uVs 11:12:42 Integrated Voltage Measurement... 11:12:51 Voltage Offset Correction: v_off_ini = 286.892 uV, v_off_fin = 270.468 uV 11:12:51 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 73.264 uVs 11:12:51 Integrated Voltage Measurement... 11:13:00 Voltage Offset Correction: v_off_ini = 270.283 uV, v_off_fin = 270.414 uV 11:13:00 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -77.533 uVs 11:13:00 Integrated Voltage Measurement... 11:13:09 Voltage Offset Correction: v_off_ini = 285.121 uV, v_off_fin = 291.807 uV 11:13:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.008 uVs 11:13:09 Integrated Voltage Measurement... 11:13:18 Voltage Offset Correction: v_off_ini = 297.598 uV, v_off_fin = 273.044 uV 11:13:18 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -105.204 uVs 11:13:18 Integrated Voltage Measurement... 11:13:27 Voltage Offset Correction: v_off_ini = 297.051 uV, v_off_fin = 289.401 uV 11:13:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.975 uVs 11:13:29 I2X Average and RMS of Measurements VT = -79.424 +- 8.615 uVs I2X = 7.307 +- 0.793 uTm^2 11:13:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:13:37 Integrated Voltage Measurement... 11:13:46 Voltage Offset Correction: v_off_ini = 296.338 uV, v_off_fin = 288.393 uV 11:13:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -203.495 uVs 11:13:46 Integrated Voltage Measurement... 11:13:55 Voltage Offset Correction: v_off_ini = 278.065 uV, v_off_fin = 286.746 uV 11:13:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.511 uVs 11:13:55 Integrated Voltage Measurement... 11:14:04 Voltage Offset Correction: v_off_ini = 295.626 uV, v_off_fin = 298.046 uV 11:14:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.182 uVs 11:14:04 Integrated Voltage Measurement... 11:14:13 Voltage Offset Correction: v_off_ini = 295.981 uV, v_off_fin = 298.458 uV 11:14:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.960 uVs 11:14:13 Integrated Voltage Measurement... 11:14:22 Voltage Offset Correction: v_off_ini = 292.099 uV, v_off_fin = 295.972 uV 11:14:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -211.372 uVs 11:14:22 Integrated Voltage Measurement... 11:14:31 Voltage Offset Correction: v_off_ini = 279.659 uV, v_off_fin = 295.541 uV 11:14:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.630 uVs 11:14:34 I2Y Average and RMS of Measurements VT = -209.692 +- 4.297 uVs I2Y = -19.292 +- 0.395 uTm^2 11:14:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:14:50 Integrated Voltage Measurement... 11:14:59 Voltage Offset Correction: v_off_ini = 288.501 uV, v_off_fin = 289.103 uV 11:14:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -89.558 uVs 11:14:59 Integrated Voltage Measurement... 11:15:08 Voltage Offset Correction: v_off_ini = 299.429 uV, v_off_fin = 299.818 uV 11:15:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.767 uVs 11:15:08 Integrated Voltage Measurement... 11:15:17 Voltage Offset Correction: v_off_ini = 317.046 uV, v_off_fin = 292.210 uV 11:15:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -90.980 uVs 11:15:17 Integrated Voltage Measurement... 11:15:26 Voltage Offset Correction: v_off_ini = 277.716 uV, v_off_fin = 295.951 uV 11:15:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.640 uVs 11:15:26 Integrated Voltage Measurement... 11:15:35 Voltage Offset Correction: v_off_ini = 303.981 uV, v_off_fin = 306.724 uV 11:15:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.309 uVs 11:15:35 Integrated Voltage Measurement... 11:15:44 Voltage Offset Correction: v_off_ini = 321.961 uV, v_off_fin = 319.013 uV 11:15:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.094 uVs 11:15:48 I1X Average and RMS of Measurements VT = -99.558 +- 5.361 uVs I1X = 19.912 +- 1.072 uTm 11:15:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:15:55 Integrated Voltage Measurement... 11:16:04 Voltage Offset Correction: v_off_ini = 317.078 uV, v_off_fin = 310.181 uV 11:16:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -154.820 uVs 11:16:04 Integrated Voltage Measurement... 11:16:13 Voltage Offset Correction: v_off_ini = 307.018 uV, v_off_fin = 316.923 uV 11:16:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 172.267 uVs 11:16:13 Integrated Voltage Measurement... 11:16:22 Voltage Offset Correction: v_off_ini = 302.307 uV, v_off_fin = 292.041 uV 11:16:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -164.155 uVs 11:16:22 Integrated Voltage Measurement... 11:16:31 Voltage Offset Correction: v_off_ini = 289.232 uV, v_off_fin = 301.343 uV 11:16:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 156.025 uVs 11:16:31 Integrated Voltage Measurement... 11:16:40 Voltage Offset Correction: v_off_ini = 288.160 uV, v_off_fin = 276.298 uV 11:16:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -146.570 uVs 11:16:40 Integrated Voltage Measurement... 11:16:49 Voltage Offset Correction: v_off_ini = 294.765 uV, v_off_fin = 299.189 uV 11:16:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 153.761 uVs 11:16:53 I1Y Average and RMS of Measurements VT = -157.933 +- 5.671 uVs I1Y = -31.581 +- 1.134 uTm 11:16:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:17:01 Integrated Voltage Measurement... 11:17:10 Voltage Offset Correction: v_off_ini = 272.066 uV, v_off_fin = 291.408 uV 11:17:10 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -39.009 uVs 11:17:10 Integrated Voltage Measurement... 11:17:19 Voltage Offset Correction: v_off_ini = 301.530 uV, v_off_fin = 303.899 uV 11:17:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.585 uVs 11:17:19 Integrated Voltage Measurement... 11:17:28 Voltage Offset Correction: v_off_ini = 300.549 uV, v_off_fin = 291.790 uV 11:17:28 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -74.359 uVs 11:17:28 Integrated Voltage Measurement... 11:17:37 Voltage Offset Correction: v_off_ini = 302.937 uV, v_off_fin = 306.788 uV 11:17:37 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 72.320 uVs 11:17:37 Integrated Voltage Measurement... 11:17:46 Voltage Offset Correction: v_off_ini = 303.252 uV, v_off_fin = 290.967 uV 11:17:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.440 uVs 11:17:46 Integrated Voltage Measurement... 11:17:55 Voltage Offset Correction: v_off_ini = 309.535 uV, v_off_fin = 313.200 uV 11:17:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.164 uVs 11:17:57 I2X Average and RMS of Measurements VT = -66.313 +- 8.209 uVs I2X = 6.101 +- 0.755 uTm^2 11:18:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:18:05 Integrated Voltage Measurement... 11:18:14 Voltage Offset Correction: v_off_ini = 314.237 uV, v_off_fin = 315.256 uV 11:18:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -219.332 uVs 11:18:14 Integrated Voltage Measurement... 11:18:23 Voltage Offset Correction: v_off_ini = 311.824 uV, v_off_fin = 310.990 uV 11:18:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.899 uVs 11:18:23 Integrated Voltage Measurement... 11:18:32 Voltage Offset Correction: v_off_ini = 320.557 uV, v_off_fin = 311.217 uV 11:18:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.390 uVs 11:18:32 Integrated Voltage Measurement... 11:18:41 Voltage Offset Correction: v_off_ini = 310.617 uV, v_off_fin = 298.254 uV 11:18:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.095 uVs 11:18:41 Integrated Voltage Measurement... 11:18:50 Voltage Offset Correction: v_off_ini = 308.150 uV, v_off_fin = 302.823 uV 11:18:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -210.582 uVs 11:18:50 Integrated Voltage Measurement... 11:18:59 Voltage Offset Correction: v_off_ini = 316.581 uV, v_off_fin = 315.929 uV 11:18:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 214.237 uVs 11:19:02 I2Y Average and RMS of Measurements VT = -214.089 +- 3.236 uVs I2Y = -19.696 +- 0.298 uTm^2 11:19:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:19:18 Integrated Voltage Measurement... 11:19:27 Voltage Offset Correction: v_off_ini = 330.776 uV, v_off_fin = 338.534 uV 11:19:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -113.677 uVs 11:19:27 Integrated Voltage Measurement... 11:19:36 Voltage Offset Correction: v_off_ini = 332.346 uV, v_off_fin = 333.276 uV 11:19:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.904 uVs 11:19:36 Integrated Voltage Measurement... 11:19:45 Voltage Offset Correction: v_off_ini = 325.256 uV, v_off_fin = 327.694 uV 11:19:45 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -96.125 uVs 11:19:45 Integrated Voltage Measurement... 11:19:54 Voltage Offset Correction: v_off_ini = 314.077 uV, v_off_fin = 308.716 uV 11:19:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 91.206 uVs 11:19:54 Integrated Voltage Measurement... 11:20:03 Voltage Offset Correction: v_off_ini = 319.752 uV, v_off_fin = 323.057 uV 11:20:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.169 uVs 11:20:03 Integrated Voltage Measurement... 11:20:12 Voltage Offset Correction: v_off_ini = 322.265 uV, v_off_fin = 349.423 uV 11:20:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 96.530 uVs 11:20:16 I1X Average and RMS of Measurements VT = -105.602 +- 8.633 uVs I1X = 21.121 +- 1.725 uTm 11:20:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:20:23 Integrated Voltage Measurement... 11:20:32 Voltage Offset Correction: v_off_ini = 329.863 uV, v_off_fin = 327.559 uV 11:20:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -156.615 uVs 11:20:32 Integrated Voltage Measurement... 11:20:41 Voltage Offset Correction: v_off_ini = 340.518 uV, v_off_fin = 357.813 uV 11:20:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 172.280 uVs 11:20:41 Integrated Voltage Measurement... 11:20:50 Voltage Offset Correction: v_off_ini = 344.779 uV, v_off_fin = 339.942 uV 11:20:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -161.843 uVs 11:20:50 Integrated Voltage Measurement... 11:20:59 Voltage Offset Correction: v_off_ini = 351.959 uV, v_off_fin = 365.389 uV 11:20:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.977 uVs 11:20:59 Integrated Voltage Measurement... 11:21:08 Voltage Offset Correction: v_off_ini = 362.460 uV, v_off_fin = 344.627 uV 11:21:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.533 uVs 11:21:08 Integrated Voltage Measurement... 11:21:17 Voltage Offset Correction: v_off_ini = 363.152 uV, v_off_fin = 369.924 uV 11:21:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 169.711 uVs 11:21:21 I1Y Average and RMS of Measurements VT = -170.160 +- 4.998 uVs I1Y = -34.029 +- 1.000 uTm 11:21:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:21:29 Integrated Voltage Measurement... 11:21:38 Voltage Offset Correction: v_off_ini = 352.285 uV, v_off_fin = 348.878 uV 11:21:38 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -59.507 uVs 11:21:38 Integrated Voltage Measurement... 11:21:47 Voltage Offset Correction: v_off_ini = 362.300 uV, v_off_fin = 390.170 uV 11:21:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.854 uVs 11:21:47 Integrated Voltage Measurement... 11:21:56 Voltage Offset Correction: v_off_ini = 368.757 uV, v_off_fin = 346.135 uV 11:21:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.203 uVs 11:21:56 Integrated Voltage Measurement... 11:22:05 Voltage Offset Correction: v_off_ini = 354.081 uV, v_off_fin = 373.131 uV 11:22:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.741 uVs 11:22:05 Integrated Voltage Measurement... 11:22:14 Voltage Offset Correction: v_off_ini = 367.736 uV, v_off_fin = 348.038 uV 11:22:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.646 uVs 11:22:14 Integrated Voltage Measurement... 11:22:23 Voltage Offset Correction: v_off_ini = 357.245 uV, v_off_fin = 364.006 uV 11:22:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.853 uVs 11:22:25 I2X Average and RMS of Measurements VT = -70.301 +- 7.176 uVs I2X = 6.468 +- 0.660 uTm^2 11:22:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:22:33 Integrated Voltage Measurement... 11:22:42 Voltage Offset Correction: v_off_ini = 356.157 uV, v_off_fin = 367.736 uV 11:22:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -224.299 uVs 11:22:42 Integrated Voltage Measurement... 11:22:51 Voltage Offset Correction: v_off_ini = 361.104 uV, v_off_fin = 359.067 uV 11:22:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 199.907 uVs 11:22:51 Integrated Voltage Measurement... 11:23:00 Voltage Offset Correction: v_off_ini = 361.105 uV, v_off_fin = 369.715 uV 11:23:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.334 uVs 11:23:00 Integrated Voltage Measurement... 11:23:09 Voltage Offset Correction: v_off_ini = 363.845 uV, v_off_fin = 371.971 uV 11:23:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 207.109 uVs 11:23:09 Integrated Voltage Measurement... 11:23:18 Voltage Offset Correction: v_off_ini = 365.567 uV, v_off_fin = 360.572 uV 11:23:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.908 uVs 11:23:18 Integrated Voltage Measurement... 11:23:27 Voltage Offset Correction: v_off_ini = 358.634 uV, v_off_fin = 352.229 uV 11:23:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.822 uVs 11:23:29 I2Y Average and RMS of Measurements VT = -197.563 +- 15.636 uVs I2Y = -18.176 +- 1.439 uTm^2 11:23:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:23:48 Integrated Voltage Measurement... 11:23:57 Voltage Offset Correction: v_off_ini = 372.691 uV, v_off_fin = 383.854 uV 11:23:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.152 uVs 11:23:57 Integrated Voltage Measurement... 11:24:06 Voltage Offset Correction: v_off_ini = 359.870 uV, v_off_fin = 359.679 uV 11:24:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.933 uVs 11:24:06 Integrated Voltage Measurement... 11:24:15 Voltage Offset Correction: v_off_ini = 360.877 uV, v_off_fin = 373.594 uV 11:24:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -78.804 uVs 11:24:15 Integrated Voltage Measurement... 11:24:24 Voltage Offset Correction: v_off_ini = 382.240 uV, v_off_fin = 377.260 uV 11:24:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.665 uVs 11:24:24 Integrated Voltage Measurement... 11:24:33 Voltage Offset Correction: v_off_ini = 371.234 uV, v_off_fin = 358.524 uV 11:24:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.257 uVs 11:24:33 Integrated Voltage Measurement... 11:24:42 Voltage Offset Correction: v_off_ini = 366.910 uV, v_off_fin = 363.665 uV 11:24:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.396 uVs 11:24:46 I1X Average and RMS of Measurements VT = -82.868 +- 10.161 uVs I1X = 16.574 +- 2.032 uTm 11:24:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:24:53 Integrated Voltage Measurement... 11:25:02 Voltage Offset Correction: v_off_ini = 356.150 uV, v_off_fin = 363.658 uV 11:25:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -166.325 uVs 11:25:02 Integrated Voltage Measurement... 11:25:11 Voltage Offset Correction: v_off_ini = 358.551 uV, v_off_fin = 382.435 uV 11:25:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 190.869 uVs 11:25:11 Integrated Voltage Measurement... 11:25:20 Voltage Offset Correction: v_off_ini = 387.109 uV, v_off_fin = 371.954 uV 11:25:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -159.703 uVs 11:25:20 Integrated Voltage Measurement... 11:25:29 Voltage Offset Correction: v_off_ini = 364.038 uV, v_off_fin = 381.840 uV 11:25:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 172.950 uVs 11:25:29 Integrated Voltage Measurement... 11:25:38 Voltage Offset Correction: v_off_ini = 355.951 uV, v_off_fin = 367.942 uV 11:25:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -150.760 uVs 11:25:38 Integrated Voltage Measurement... 11:25:47 Voltage Offset Correction: v_off_ini = 372.808 uV, v_off_fin = 378.054 uV 11:25:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.103 uVs 11:25:51 I1Y Average and RMS of Measurements VT = -172.118 +- 5.033 uVs I1Y = -34.421 +- 1.006 uTm 11:25:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:25:59 Integrated Voltage Measurement... 11:26:08 Voltage Offset Correction: v_off_ini = 373.862 uV, v_off_fin = 371.486 uV 11:26:08 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -54.051 uVs 11:26:08 Integrated Voltage Measurement... 11:26:17 Voltage Offset Correction: v_off_ini = 388.950 uV, v_off_fin = 395.945 uV 11:26:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.975 uVs 11:26:17 Integrated Voltage Measurement... 11:26:26 Voltage Offset Correction: v_off_ini = 383.628 uV, v_off_fin = 387.872 uV 11:26:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.857 uVs 11:26:26 Integrated Voltage Measurement... 11:26:35 Voltage Offset Correction: v_off_ini = 396.116 uV, v_off_fin = 399.081 uV 11:26:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.335 uVs 11:26:35 Integrated Voltage Measurement... 11:26:44 Voltage Offset Correction: v_off_ini = 394.847 uV, v_off_fin = 379.859 uV 11:26:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.842 uVs 11:26:44 Integrated Voltage Measurement... 11:26:53 Voltage Offset Correction: v_off_ini = 395.322 uV, v_off_fin = 402.147 uV 11:26:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.337 uVs 11:26:55 I2X Average and RMS of Measurements VT = -54.566 +- 0.378 uVs I2X = 5.020 +- 0.035 uTm^2 11:27:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:27:03 Integrated Voltage Measurement... 11:27:12 Voltage Offset Correction: v_off_ini = 396.541 uV, v_off_fin = 397.546 uV 11:27:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -186.731 uVs 11:27:12 Integrated Voltage Measurement... 11:27:21 Voltage Offset Correction: v_off_ini = 406.726 uV, v_off_fin = 411.124 uV 11:27:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.901 uVs 11:27:21 Integrated Voltage Measurement... 11:27:30 Voltage Offset Correction: v_off_ini = 407.403 uV, v_off_fin = 407.811 uV 11:27:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -208.013 uVs 11:27:30 Integrated Voltage Measurement... 11:27:39 Voltage Offset Correction: v_off_ini = 426.525 uV, v_off_fin = 418.251 uV 11:27:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.225 uVs 11:27:39 Integrated Voltage Measurement... 11:27:48 Voltage Offset Correction: v_off_ini = 404.618 uV, v_off_fin = 389.994 uV 11:27:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.270 uVs 11:27:48 Integrated Voltage Measurement... 11:27:57 Voltage Offset Correction: v_off_ini = 397.702 uV, v_off_fin = 397.277 uV 11:27:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 187.495 uVs 11:27:59 I2Y Average and RMS of Measurements VT = -198.273 +- 6.187 uVs I2Y = -18.241 +- 0.569 uTm^2 11:28:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:28:18 Integrated Voltage Measurement... 11:28:27 Voltage Offset Correction: v_off_ini = 424.626 uV, v_off_fin = 401.851 uV 11:28:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -77.769 uVs 11:28:27 Integrated Voltage Measurement... 11:28:36 Voltage Offset Correction: v_off_ini = 394.838 uV, v_off_fin = 410.694 uV 11:28:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.667 uVs 11:28:36 Integrated Voltage Measurement... 11:28:45 Voltage Offset Correction: v_off_ini = 413.103 uV, v_off_fin = 411.717 uV 11:28:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -68.989 uVs 11:28:45 Integrated Voltage Measurement... 11:28:54 Voltage Offset Correction: v_off_ini = 419.916 uV, v_off_fin = 404.644 uV 11:28:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.002 uVs 11:28:54 Integrated Voltage Measurement... 11:29:03 Voltage Offset Correction: v_off_ini = 416.605 uV, v_off_fin = 417.426 uV 11:29:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.601 uVs 11:29:03 Integrated Voltage Measurement... 11:29:12 Voltage Offset Correction: v_off_ini = 440.274 uV, v_off_fin = 435.439 uV 11:29:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.078 uVs 11:29:16 I1X Average and RMS of Measurements VT = -69.018 +- 7.095 uVs I1X = 13.804 +- 1.419 uTm 11:29:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:29:23 Integrated Voltage Measurement... 11:29:32 Voltage Offset Correction: v_off_ini = 387.401 uV, v_off_fin = 390.603 uV 11:29:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -168.492 uVs 11:29:32 Integrated Voltage Measurement... 11:29:41 Voltage Offset Correction: v_off_ini = 398.139 uV, v_off_fin = 414.182 uV 11:29:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 180.459 uVs 11:29:41 Integrated Voltage Measurement... 11:29:50 Voltage Offset Correction: v_off_ini = 404.759 uV, v_off_fin = 392.162 uV 11:29:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -167.541 uVs 11:29:50 Integrated Voltage Measurement... 11:29:59 Voltage Offset Correction: v_off_ini = 404.294 uV, v_off_fin = 411.938 uV 11:29:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 180.102 uVs 11:29:59 Integrated Voltage Measurement... 11:30:08 Voltage Offset Correction: v_off_ini = 420.964 uV, v_off_fin = 410.345 uV 11:30:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -160.392 uVs 11:30:08 Integrated Voltage Measurement... 11:30:17 Voltage Offset Correction: v_off_ini = 412.826 uV, v_off_fin = 424.642 uV 11:30:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.280 uVs 11:30:21 I1Y Average and RMS of Measurements VT = -173.711 +- 0.674 uVs I1Y = -34.739 +- 0.135 uTm 11:30:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:30:29 Integrated Voltage Measurement... 11:30:38 Voltage Offset Correction: v_off_ini = 431.780 uV, v_off_fin = 430.495 uV 11:30:38 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -43.997 uVs 11:30:38 Integrated Voltage Measurement... 11:30:47 Voltage Offset Correction: v_off_ini = 435.257 uV, v_off_fin = 437.694 uV 11:30:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 49.658 uVs 11:30:47 Integrated Voltage Measurement... 11:30:56 Voltage Offset Correction: v_off_ini = 437.388 uV, v_off_fin = 427.968 uV 11:30:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.443 uVs 11:30:56 Integrated Voltage Measurement... 11:31:05 Voltage Offset Correction: v_off_ini = 421.044 uV, v_off_fin = 434.868 uV 11:31:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.659 uVs 11:31:05 Integrated Voltage Measurement... 11:31:14 Voltage Offset Correction: v_off_ini = 446.870 uV, v_off_fin = 431.673 uV 11:31:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -55.086 uVs 11:31:14 Integrated Voltage Measurement... 11:31:23 Voltage Offset Correction: v_off_ini = 449.919 uV, v_off_fin = 457.108 uV 11:31:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.667 uVs 11:31:25 I2X Average and RMS of Measurements VT = -48.252 +- 1.115 uVs I2X = 4.439 +- 0.103 uTm^2 11:31:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:31:33 Integrated Voltage Measurement... 11:31:42 Voltage Offset Correction: v_off_ini = 433.622 uV, v_off_fin = 419.283 uV 11:31:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -212.709 uVs 11:31:42 Integrated Voltage Measurement... 11:31:51 Voltage Offset Correction: v_off_ini = 430.847 uV, v_off_fin = 420.505 uV 11:31:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 191.826 uVs 11:31:51 Integrated Voltage Measurement... 11:32:00 Voltage Offset Correction: v_off_ini = 418.312 uV, v_off_fin = 422.477 uV 11:32:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.845 uVs 11:32:00 Integrated Voltage Measurement... 11:32:09 Voltage Offset Correction: v_off_ini = 417.868 uV, v_off_fin = 432.800 uV 11:32:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.754 uVs 11:32:09 Integrated Voltage Measurement... 11:32:18 Voltage Offset Correction: v_off_ini = 432.420 uV, v_off_fin = 431.074 uV 11:32:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -199.786 uVs 11:32:18 Integrated Voltage Measurement... 11:32:27 Voltage Offset Correction: v_off_ini = 422.755 uV, v_off_fin = 416.649 uV 11:32:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.865 uVs 11:32:29 I2Y Average and RMS of Measurements VT = -201.631 +- 5.337 uVs I2Y = -18.550 +- 0.491 uTm^2 11:32:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:32:48 Integrated Voltage Measurement... 11:32:57 Voltage Offset Correction: v_off_ini = 409.533 uV, v_off_fin = 417.716 uV 11:32:57 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.031 uVs 11:32:57 Integrated Voltage Measurement... 11:33:06 Voltage Offset Correction: v_off_ini = 414.899 uV, v_off_fin = 416.419 uV 11:33:06 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 54.960 uVs 11:33:06 Integrated Voltage Measurement... 11:33:15 Voltage Offset Correction: v_off_ini = 421.808 uV, v_off_fin = 426.078 uV 11:33:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.588 uVs 11:33:15 Integrated Voltage Measurement... 11:33:24 Voltage Offset Correction: v_off_ini = 406.826 uV, v_off_fin = 419.797 uV 11:33:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.771 uVs 11:33:24 Integrated Voltage Measurement... 11:33:33 Voltage Offset Correction: v_off_ini = 415.348 uV, v_off_fin = 419.188 uV 11:33:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.846 uVs 11:33:33 Integrated Voltage Measurement... 11:33:42 Voltage Offset Correction: v_off_ini = 413.559 uV, v_off_fin = 429.366 uV 11:33:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.249 uVs 11:33:46 I1X Average and RMS of Measurements VT = -57.574 +- 4.955 uVs I1X = 11.515 +- 0.991 uTm 11:33:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:33:53 Integrated Voltage Measurement... 11:34:02 Voltage Offset Correction: v_off_ini = 410.442 uV, v_off_fin = 400.779 uV 11:34:02 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -181.957 uVs 11:34:02 Integrated Voltage Measurement... 11:34:11 Voltage Offset Correction: v_off_ini = 416.289 uV, v_off_fin = 434.928 uV 11:34:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.648 uVs 11:34:11 Integrated Voltage Measurement... 11:34:20 Voltage Offset Correction: v_off_ini = 423.936 uV, v_off_fin = 420.708 uV 11:34:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -178.587 uVs 11:34:20 Integrated Voltage Measurement... 11:34:29 Voltage Offset Correction: v_off_ini = 427.979 uV, v_off_fin = 436.514 uV 11:34:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 181.092 uVs 11:34:29 Integrated Voltage Measurement... 11:34:38 Voltage Offset Correction: v_off_ini = 430.377 uV, v_off_fin = 423.912 uV 11:34:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -179.201 uVs 11:34:38 Integrated Voltage Measurement... 11:34:47 Voltage Offset Correction: v_off_ini = 424.865 uV, v_off_fin = 438.089 uV 11:34:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.084 uVs 11:34:51 I1Y Average and RMS of Measurements VT = -187.095 +- 5.131 uVs I1Y = -37.417 +- 1.027 uTm 11:34:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:34:59 Integrated Voltage Measurement... 11:35:08 Voltage Offset Correction: v_off_ini = 415.508 uV, v_off_fin = 396.489 uV 11:35:08 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -30.510 uVs 11:35:08 Integrated Voltage Measurement... 11:35:17 Voltage Offset Correction: v_off_ini = 421.858 uV, v_off_fin = 439.076 uV 11:35:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 24.996 uVs 11:35:17 Integrated Voltage Measurement... 11:35:26 Voltage Offset Correction: v_off_ini = 427.325 uV, v_off_fin = 411.165 uV 11:35:26 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -43.991 uVs 11:35:26 Integrated Voltage Measurement... 11:35:35 Voltage Offset Correction: v_off_ini = 419.566 uV, v_off_fin = 432.815 uV 11:35:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.193 uVs 11:35:35 Integrated Voltage Measurement... 11:35:44 Voltage Offset Correction: v_off_ini = 416.629 uV, v_off_fin = 403.790 uV 11:35:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -60.469 uVs 11:35:44 Integrated Voltage Measurement... 11:35:53 Voltage Offset Correction: v_off_ini = 426.057 uV, v_off_fin = 435.514 uV 11:35:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.427 uVs 11:35:55 I2X Average and RMS of Measurements VT = -41.431 +- 10.014 uVs I2X = 3.812 +- 0.921 uTm^2 11:36:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:36:03 Integrated Voltage Measurement... 11:36:12 Voltage Offset Correction: v_off_ini = 428.590 uV, v_off_fin = 434.019 uV 11:36:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.749 uVs 11:36:12 Integrated Voltage Measurement... 11:36:21 Voltage Offset Correction: v_off_ini = 443.033 uV, v_off_fin = 429.861 uV 11:36:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 212.407 uVs 11:36:21 Integrated Voltage Measurement... 11:36:30 Voltage Offset Correction: v_off_ini = 421.183 uV, v_off_fin = 417.357 uV 11:36:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.120 uVs 11:36:30 Integrated Voltage Measurement... 11:36:39 Voltage Offset Correction: v_off_ini = 421.924 uV, v_off_fin = 441.403 uV 11:36:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 223.051 uVs 11:36:39 Integrated Voltage Measurement... 11:36:48 Voltage Offset Correction: v_off_ini = 439.775 uV, v_off_fin = 436.426 uV 11:36:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -210.088 uVs 11:36:48 Integrated Voltage Measurement... 11:36:57 Voltage Offset Correction: v_off_ini = 453.917 uV, v_off_fin = 451.495 uV 11:36:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 214.979 uVs 11:37:00 I2Y Average and RMS of Measurements VT = -214.899 +- 5.958 uVs I2Y = -19.771 +- 0.548 uTm^2 11:37:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:37:18 Integrated Voltage Measurement... 11:37:27 Voltage Offset Correction: v_off_ini = 457.803 uV, v_off_fin = 446.007 uV 11:37:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -47.021 uVs 11:37:27 Integrated Voltage Measurement... 11:37:36 Voltage Offset Correction: v_off_ini = 446.650 uV, v_off_fin = 441.923 uV 11:37:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.488 uVs 11:37:36 Integrated Voltage Measurement... 11:37:45 Voltage Offset Correction: v_off_ini = 446.353 uV, v_off_fin = 448.415 uV 11:37:45 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.726 uVs 11:37:45 Integrated Voltage Measurement... 11:37:54 Voltage Offset Correction: v_off_ini = 444.853 uV, v_off_fin = 458.385 uV 11:37:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 78.393 uVs 11:37:54 Integrated Voltage Measurement... 11:38:03 Voltage Offset Correction: v_off_ini = 463.346 uV, v_off_fin = 462.177 uV 11:38:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.566 uVs 11:38:03 Integrated Voltage Measurement... 11:38:12 Voltage Offset Correction: v_off_ini = 465.682 uV, v_off_fin = 459.518 uV 11:38:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 57.428 uVs 11:38:16 I1X Average and RMS of Measurements VT = -59.937 +- 6.458 uVs I1X = 11.988 +- 1.292 uTm 11:38:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:38:23 Integrated Voltage Measurement... 11:38:32 Voltage Offset Correction: v_off_ini = 444.581 uV, v_off_fin = 447.156 uV 11:38:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.855 uVs 11:38:32 Integrated Voltage Measurement... 11:38:41 Voltage Offset Correction: v_off_ini = 467.738 uV, v_off_fin = 481.800 uV 11:38:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.551 uVs 11:38:41 Integrated Voltage Measurement... 11:38:50 Voltage Offset Correction: v_off_ini = 475.539 uV, v_off_fin = 451.268 uV 11:38:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -191.034 uVs 11:38:50 Integrated Voltage Measurement... 11:38:59 Voltage Offset Correction: v_off_ini = 463.146 uV, v_off_fin = 459.090 uV 11:38:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 201.136 uVs 11:38:59 Integrated Voltage Measurement... 11:39:08 Voltage Offset Correction: v_off_ini = 464.177 uV, v_off_fin = 445.743 uV 11:39:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -201.043 uVs 11:39:08 Integrated Voltage Measurement... 11:39:17 Voltage Offset Correction: v_off_ini = 452.682 uV, v_off_fin = 465.475 uV 11:39:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.511 uVs 11:39:21 I1Y Average and RMS of Measurements VT = -193.022 +- 7.494 uVs I1Y = -38.601 +- 1.498 uTm 11:39:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:39:29 Integrated Voltage Measurement... 11:39:38 Voltage Offset Correction: v_off_ini = 447.888 uV, v_off_fin = 449.134 uV 11:39:38 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -38.071 uVs 11:39:38 Integrated Voltage Measurement... 11:39:47 Voltage Offset Correction: v_off_ini = 459.938 uV, v_off_fin = 459.616 uV 11:39:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.108 uVs 11:39:47 Integrated Voltage Measurement... 11:39:56 Voltage Offset Correction: v_off_ini = 451.067 uV, v_off_fin = 451.693 uV 11:39:56 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -27.314 uVs 11:39:56 Integrated Voltage Measurement... 11:40:05 Voltage Offset Correction: v_off_ini = 456.482 uV, v_off_fin = 472.425 uV 11:40:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 38.324 uVs 11:40:05 Integrated Voltage Measurement... 11:40:14 Voltage Offset Correction: v_off_ini = 463.969 uV, v_off_fin = 455.325 uV 11:40:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -40.526 uVs 11:40:14 Integrated Voltage Measurement... 11:40:23 Voltage Offset Correction: v_off_ini = 466.877 uV, v_off_fin = 466.483 uV 11:40:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 33.222 uVs 11:40:26 I2X Average and RMS of Measurements VT = -33.428 +- 2.602 uVs I2X = 3.075 +- 0.239 uTm^2 11:40:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:40:33 Integrated Voltage Measurement... 11:40:42 Voltage Offset Correction: v_off_ini = 474.798 uV, v_off_fin = 479.345 uV 11:40:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.486 uVs 11:40:42 Integrated Voltage Measurement... 11:40:51 Voltage Offset Correction: v_off_ini = 478.356 uV, v_off_fin = 478.078 uV 11:40:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 199.692 uVs 11:40:51 Integrated Voltage Measurement... 11:41:00 Voltage Offset Correction: v_off_ini = 469.052 uV, v_off_fin = 457.268 uV 11:41:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -203.217 uVs 11:41:00 Integrated Voltage Measurement... 11:41:09 Voltage Offset Correction: v_off_ini = 472.942 uV, v_off_fin = 487.089 uV 11:41:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 212.556 uVs 11:41:09 Integrated Voltage Measurement... 11:41:18 Voltage Offset Correction: v_off_ini = 471.546 uV, v_off_fin = 472.934 uV 11:41:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.915 uVs 11:41:18 Integrated Voltage Measurement... 11:41:27 Voltage Offset Correction: v_off_ini = 476.959 uV, v_off_fin = 461.762 uV 11:41:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.371 uVs 11:41:30 I2Y Average and RMS of Measurements VT = -209.039 +- 5.391 uVs I2Y = -19.232 +- 0.496 uTm^2 11:41:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:41:48 Integrated Voltage Measurement... 11:41:57 Voltage Offset Correction: v_off_ini = 475.535 uV, v_off_fin = 471.850 uV 11:41:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.235 uVs 11:41:57 Integrated Voltage Measurement... 11:42:06 Voltage Offset Correction: v_off_ini = 463.968 uV, v_off_fin = 459.134 uV 11:42:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.503 uVs 11:42:06 Integrated Voltage Measurement... 11:42:15 Voltage Offset Correction: v_off_ini = 462.328 uV, v_off_fin = 458.698 uV 11:42:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -60.509 uVs 11:42:15 Integrated Voltage Measurement... 11:42:24 Voltage Offset Correction: v_off_ini = 462.741 uV, v_off_fin = 470.753 uV 11:42:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.356 uVs 11:42:24 Integrated Voltage Measurement... 11:42:33 Voltage Offset Correction: v_off_ini = 462.283 uV, v_off_fin = 470.833 uV 11:42:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.696 uVs 11:42:33 Integrated Voltage Measurement... 11:42:42 Voltage Offset Correction: v_off_ini = 467.931 uV, v_off_fin = 463.082 uV 11:42:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 43.494 uVs 11:42:46 I1X Average and RMS of Measurements VT = -59.466 +- 10.706 uVs I1X = 11.893 +- 2.142 uTm 11:42:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:42:53 Integrated Voltage Measurement... 11:43:02 Voltage Offset Correction: v_off_ini = 460.510 uV, v_off_fin = 454.516 uV 11:43:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.831 uVs 11:43:02 Integrated Voltage Measurement... 11:43:11 Voltage Offset Correction: v_off_ini = 458.059 uV, v_off_fin = 487.757 uV 11:43:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.151 uVs 11:43:11 Integrated Voltage Measurement... 11:43:20 Voltage Offset Correction: v_off_ini = 481.967 uV, v_off_fin = 472.638 uV 11:43:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -198.491 uVs 11:43:20 Integrated Voltage Measurement... 11:43:29 Voltage Offset Correction: v_off_ini = 474.983 uV, v_off_fin = 490.606 uV 11:43:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.410 uVs 11:43:29 Integrated Voltage Measurement... 11:43:38 Voltage Offset Correction: v_off_ini = 485.121 uV, v_off_fin = 461.750 uV 11:43:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.820 uVs 11:43:38 Integrated Voltage Measurement... 11:43:47 Voltage Offset Correction: v_off_ini = 472.843 uV, v_off_fin = 487.813 uV 11:43:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 204.604 uVs 11:43:51 I1Y Average and RMS of Measurements VT = -201.218 +- 3.031 uVs I1Y = -40.240 +- 0.607 uTm 11:43:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:43:59 Integrated Voltage Measurement... 11:44:08 Voltage Offset Correction: v_off_ini = 476.901 uV, v_off_fin = 478.955 uV 11:44:08 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -10.337 uVs 11:44:08 Integrated Voltage Measurement... 11:44:17 Voltage Offset Correction: v_off_ini = 501.038 uV, v_off_fin = 489.127 uV 11:44:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 19.612 uVs 11:44:17 Integrated Voltage Measurement... 11:44:26 Voltage Offset Correction: v_off_ini = 484.956 uV, v_off_fin = 460.910 uV 11:44:26 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -24.241 uVs 11:44:26 Integrated Voltage Measurement... 11:44:35 Voltage Offset Correction: v_off_ini = 487.637 uV, v_off_fin = 482.049 uV 11:44:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.259 uVs 11:44:35 Integrated Voltage Measurement... 11:44:44 Voltage Offset Correction: v_off_ini = 463.604 uV, v_off_fin = 470.439 uV 11:44:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -33.832 uVs 11:44:44 Integrated Voltage Measurement... 11:44:53 Voltage Offset Correction: v_off_ini = 478.342 uV, v_off_fin = 489.795 uV 11:44:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 17.983 uVs 11:44:56 I2X Average and RMS of Measurements VT = -23.544 +- 6.259 uVs I2X = 2.166 +- 0.576 uTm^2 11:45:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:45:03 Integrated Voltage Measurement... 11:45:12 Voltage Offset Correction: v_off_ini = 480.505 uV, v_off_fin = 483.586 uV 11:45:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -213.170 uVs 11:45:12 Integrated Voltage Measurement... 11:45:21 Voltage Offset Correction: v_off_ini = 495.880 uV, v_off_fin = 498.793 uV 11:45:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 216.484 uVs 11:45:21 Integrated Voltage Measurement... 11:45:30 Voltage Offset Correction: v_off_ini = 485.914 uV, v_off_fin = 485.868 uV 11:45:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.087 uVs 11:45:31 Integrated Voltage Measurement... 11:45:39 Voltage Offset Correction: v_off_ini = 512.234 uV, v_off_fin = 522.786 uV 11:45:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.667 uVs 11:45:40 Integrated Voltage Measurement... 11:45:49 Voltage Offset Correction: v_off_ini = 508.719 uV, v_off_fin = 509.876 uV 11:45:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -211.454 uVs 11:45:49 Integrated Voltage Measurement... 11:45:58 Voltage Offset Correction: v_off_ini = 512.444 uV, v_off_fin = 516.572 uV 11:45:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 218.342 uVs 11:46:00 I2Y Average and RMS of Measurements VT = -213.201 +- 2.350 uVs I2Y = -19.614 +- 0.216 uTm^2 12:03:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:03:18 Integrated Voltage Measurement... 12:03:27 Voltage Offset Correction: v_off_ini = 608.066 uV, v_off_fin = 608.650 uV 12:03:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.014 uVs 12:03:27 Integrated Voltage Measurement... 12:03:36 Voltage Offset Correction: v_off_ini = 626.785 uV, v_off_fin = 610.313 uV 12:03:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.451 uVs 12:03:36 Integrated Voltage Measurement... 12:03:45 Voltage Offset Correction: v_off_ini = 615.002 uV, v_off_fin = 609.134 uV 12:03:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -34.557 uVs 12:03:45 Integrated Voltage Measurement... 12:03:54 Voltage Offset Correction: v_off_ini = 596.392 uV, v_off_fin = 590.446 uV 12:03:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 61.358 uVs 12:03:54 Integrated Voltage Measurement... 12:04:03 Voltage Offset Correction: v_off_ini = 607.436 uV, v_off_fin = 614.656 uV 12:04:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -50.706 uVs 12:04:03 Integrated Voltage Measurement... 12:04:12 Voltage Offset Correction: v_off_ini = 623.920 uV, v_off_fin = 619.688 uV 12:04:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.735 uVs 12:04:16 I1X Average and RMS of Measurements VT = -56.304 +- 5.905 uVs I1X = 11.261 +- 1.181 uTm 12:04:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:04:23 Integrated Voltage Measurement... 12:04:32 Voltage Offset Correction: v_off_ini = 603.577 uV, v_off_fin = 597.901 uV 12:04:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -207.390 uVs 12:04:32 Integrated Voltage Measurement... 12:04:41 Voltage Offset Correction: v_off_ini = 601.335 uV, v_off_fin = 619.096 uV 12:04:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 205.998 uVs 12:04:41 Integrated Voltage Measurement... 12:04:50 Voltage Offset Correction: v_off_ini = 601.784 uV, v_off_fin = 602.480 uV 12:04:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.506 uVs 12:04:50 Integrated Voltage Measurement... 12:04:59 Voltage Offset Correction: v_off_ini = 604.478 uV, v_off_fin = 613.921 uV 12:04:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.077 uVs 12:04:59 Integrated Voltage Measurement... 12:05:08 Voltage Offset Correction: v_off_ini = 597.838 uV, v_off_fin = 598.507 uV 12:05:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.784 uVs 12:05:08 Integrated Voltage Measurement... 12:05:17 Voltage Offset Correction: v_off_ini = 605.814 uV, v_off_fin = 633.221 uV 12:05:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 220.436 uVs 12:05:21 I1Y Average and RMS of Measurements VT = -208.699 +- 3.869 uVs I1Y = -41.738 +- 0.773 uTm 12:05:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:05:29 Integrated Voltage Measurement... 12:05:38 Voltage Offset Correction: v_off_ini = 618.548 uV, v_off_fin = 628.205 uV 12:05:38 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -27.418 uVs 12:05:38 Integrated Voltage Measurement... 12:05:47 Voltage Offset Correction: v_off_ini = 630.159 uV, v_off_fin = 633.156 uV 12:05:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 11.974 uVs 12:05:47 Integrated Voltage Measurement... 12:05:56 Voltage Offset Correction: v_off_ini = 625.251 uV, v_off_fin = 615.267 uV 12:05:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -37.154 uVs 12:05:56 Integrated Voltage Measurement... 12:06:05 Voltage Offset Correction: v_off_ini = 610.291 uV, v_off_fin = 617.130 uV 12:06:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.628 uVs 12:06:05 Integrated Voltage Measurement... 12:06:14 Voltage Offset Correction: v_off_ini = 618.757 uV, v_off_fin = 613.482 uV 12:06:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -41.235 uVs 12:06:14 Integrated Voltage Measurement... 12:06:23 Voltage Offset Correction: v_off_ini = 617.688 uV, v_off_fin = 628.072 uV 12:06:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 28.225 uVs 12:06:26 I2X Average and RMS of Measurements VT = -28.272 +- 6.318 uVs I2X = 2.601 +- 0.581 uTm^2 12:06:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:06:33 Integrated Voltage Measurement... 12:06:42 Voltage Offset Correction: v_off_ini = 618.476 uV, v_off_fin = 622.288 uV 12:06:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.236 uVs 12:06:43 Integrated Voltage Measurement... 12:06:52 Voltage Offset Correction: v_off_ini = 611.559 uV, v_off_fin = 618.716 uV 12:06:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.936 uVs 12:06:52 Integrated Voltage Measurement... 12:07:01 Voltage Offset Correction: v_off_ini = 626.756 uV, v_off_fin = 622.241 uV 12:07:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.729 uVs 12:07:01 Integrated Voltage Measurement... 12:07:10 Voltage Offset Correction: v_off_ini = 624.068 uV, v_off_fin = 625.543 uV 12:07:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.559 uVs 12:07:10 Integrated Voltage Measurement... 12:07:19 Voltage Offset Correction: v_off_ini = 629.426 uV, v_off_fin = 636.123 uV 12:07:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -199.578 uVs 12:07:19 Integrated Voltage Measurement... 12:07:28 Voltage Offset Correction: v_off_ini = 654.006 uV, v_off_fin = 640.794 uV 12:07:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 231.695 uVs 12:07:30 I2Y Average and RMS of Measurements VT = -226.289 +- 7.542 uVs I2Y = -20.819 +- 0.694 uTm^2 12:07:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm