SLAC Magnetic Measurements Date: 11-13-2025 Time: 08:52:04 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16339 Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 1 Measurement Device: Stretched Wire Operator: YL Comment: 08:53:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:53:05 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:53:07 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:53:08 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:53:10 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:55:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:55:29 Integrated Voltage Measurement... 08:55:38 Voltage Offset Correction: v_off_ini = 891.626 uV, v_off_fin = 883.826 uV 08:55:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -103.110 uVs 08:55:38 Integrated Voltage Measurement... 08:55:47 Voltage Offset Correction: v_off_ini = 877.646 uV, v_off_fin = 879.678 uV 08:55:47 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.131 uVs 08:55:47 Integrated Voltage Measurement... 08:55:56 Voltage Offset Correction: v_off_ini = 866.731 uV, v_off_fin = 873.761 uV 08:55:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.240 uVs 08:55:57 Integrated Voltage Measurement... 08:56:05 Voltage Offset Correction: v_off_ini = 880.420 uV, v_off_fin = 877.892 uV 08:56:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.599 uVs 08:56:06 Integrated Voltage Measurement... 08:56:15 Voltage Offset Correction: v_off_ini = 878.565 uV, v_off_fin = 878.448 uV 08:56:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.630 uVs 08:56:15 Integrated Voltage Measurement... 08:56:24 Voltage Offset Correction: v_off_ini = 879.921 uV, v_off_fin = 874.335 uV 08:56:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.040 uVs 08:56:28 I1X Average and RMS of Measurements VT = -87.292 +- 9.105 uVs I1X = 17.459 +- 1.822 uTm 08:56:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:56:35 Integrated Voltage Measurement... 08:56:44 Voltage Offset Correction: v_off_ini = 879.438 uV, v_off_fin = 868.175 uV 08:56:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -229.784 uVs 08:56:44 Integrated Voltage Measurement... 08:56:53 Voltage Offset Correction: v_off_ini = 874.796 uV, v_off_fin = 882.885 uV 08:56:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 212.166 uVs 08:56:53 Integrated Voltage Measurement... 08:57:02 Voltage Offset Correction: v_off_ini = 870.692 uV, v_off_fin = 852.304 uV 08:57:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.949 uVs 08:57:02 Integrated Voltage Measurement... 08:57:11 Voltage Offset Correction: v_off_ini = 873.166 uV, v_off_fin = 881.625 uV 08:57:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.756 uVs 08:57:11 Integrated Voltage Measurement... 08:57:20 Voltage Offset Correction: v_off_ini = 879.937 uV, v_off_fin = 861.201 uV 08:57:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.932 uVs 08:57:20 Integrated Voltage Measurement... 08:57:29 Voltage Offset Correction: v_off_ini = 882.997 uV, v_off_fin = 878.503 uV 08:57:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.355 uVs 08:57:33 I1Y Average and RMS of Measurements VT = -216.324 +- 3.302 uVs I1Y = -43.266 +- 0.660 uTm 08:57:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:57:41 Integrated Voltage Measurement... 08:57:50 Voltage Offset Correction: v_off_ini = 872.664 uV, v_off_fin = 862.233 uV 08:57:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.456 uVs 08:57:50 Integrated Voltage Measurement... 08:57:59 Voltage Offset Correction: v_off_ini = 865.549 uV, v_off_fin = 881.850 uV 08:57:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.200 uVs 08:57:59 Integrated Voltage Measurement... 08:58:08 Voltage Offset Correction: v_off_ini = 856.561 uV, v_off_fin = 859.784 uV 08:58:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.432 uVs 08:58:08 Integrated Voltage Measurement... 08:58:17 Voltage Offset Correction: v_off_ini = 865.511 uV, v_off_fin = 867.888 uV 08:58:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 113.609 uVs 08:58:17 Integrated Voltage Measurement... 08:58:26 Voltage Offset Correction: v_off_ini = 862.142 uV, v_off_fin = 860.722 uV 08:58:26 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -115.224 uVs 08:58:26 Integrated Voltage Measurement... 08:58:35 Voltage Offset Correction: v_off_ini = 858.008 uV, v_off_fin = 873.768 uV 08:58:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.838 uVs 08:58:37 I2X Average and RMS of Measurements VT = -111.126 +- 2.193 uVs I2X = 10.224 +- 0.201 uTm^2 08:58:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:58:45 Integrated Voltage Measurement... 08:58:54 Voltage Offset Correction: v_off_ini = 857.234 uV, v_off_fin = 854.823 uV 08:58:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -400.541 uVs 08:58:54 Integrated Voltage Measurement... 08:59:03 Voltage Offset Correction: v_off_ini = 870.964 uV, v_off_fin = 868.599 uV 08:59:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 365.293 uVs 08:59:03 Integrated Voltage Measurement... 08:59:12 Voltage Offset Correction: v_off_ini = 859.385 uV, v_off_fin = 848.812 uV 08:59:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -373.920 uVs 08:59:12 Integrated Voltage Measurement... 08:59:21 Voltage Offset Correction: v_off_ini = 865.497 uV, v_off_fin = 858.213 uV 08:59:21 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 368.314 uVs 08:59:21 Integrated Voltage Measurement... 08:59:30 Voltage Offset Correction: v_off_ini = 855.620 uV, v_off_fin = 853.930 uV 08:59:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.101 uVs 08:59:30 Integrated Voltage Measurement... 08:59:39 Voltage Offset Correction: v_off_ini = 862.138 uV, v_off_fin = 866.961 uV 08:59:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 380.852 uVs 08:59:42 I2Y Average and RMS of Measurements VT = -372.504 +- 7.997 uVs I2Y = -34.271 +- 0.736 uTm^2 08:59:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:59:57 Integrated Voltage Measurement... 09:00:06 Voltage Offset Correction: v_off_ini = 864.038 uV, v_off_fin = 854.932 uV 09:00:06 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.417 uVs 09:00:06 Integrated Voltage Measurement... 09:00:15 Voltage Offset Correction: v_off_ini = 846.999 uV, v_off_fin = 855.535 uV 09:00:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 92.081 uVs 09:00:15 Integrated Voltage Measurement... 09:00:24 Voltage Offset Correction: v_off_ini = 863.958 uV, v_off_fin = 846.415 uV 09:00:24 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.935 uVs 09:00:24 Integrated Voltage Measurement... 09:00:33 Voltage Offset Correction: v_off_ini = 852.494 uV, v_off_fin = 857.336 uV 09:00:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 77.202 uVs 09:00:33 Integrated Voltage Measurement... 09:00:42 Voltage Offset Correction: v_off_ini = 860.022 uV, v_off_fin = 865.943 uV 09:00:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.279 uVs 09:00:42 Integrated Voltage Measurement... 09:00:51 Voltage Offset Correction: v_off_ini = 854.409 uV, v_off_fin = 863.662 uV 09:00:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.578 uVs 09:00:55 I1X Average and RMS of Measurements VT = -86.915 +- 1.073 uVs I1X = 17.384 +- 0.214 uTm 09:00:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:01:02 Integrated Voltage Measurement... 09:01:11 Voltage Offset Correction: v_off_ini = 851.771 uV, v_off_fin = 856.156 uV 09:01:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.466 uVs 09:01:11 Integrated Voltage Measurement... 09:01:20 Voltage Offset Correction: v_off_ini = 865.160 uV, v_off_fin = 860.885 uV 09:01:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 227.067 uVs 09:01:20 Integrated Voltage Measurement... 09:01:29 Voltage Offset Correction: v_off_ini = 863.079 uV, v_off_fin = 849.700 uV 09:01:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -211.515 uVs 09:01:29 Integrated Voltage Measurement... 09:01:38 Voltage Offset Correction: v_off_ini = 851.392 uV, v_off_fin = 868.430 uV 09:01:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 218.174 uVs 09:01:38 Integrated Voltage Measurement... 09:01:47 Voltage Offset Correction: v_off_ini = 868.933 uV, v_off_fin = 845.594 uV 09:01:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.234 uVs 09:01:47 Integrated Voltage Measurement... 09:01:56 Voltage Offset Correction: v_off_ini = 859.625 uV, v_off_fin = 867.834 uV 09:01:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.323 uVs 09:02:00 I1Y Average and RMS of Measurements VT = -217.797 +- 4.221 uVs I1Y = -43.559 +- 0.844 uTm 09:02:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:02:08 Integrated Voltage Measurement... 09:02:17 Voltage Offset Correction: v_off_ini = 849.928 uV, v_off_fin = 839.562 uV 09:02:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.027 uVs 09:02:17 Integrated Voltage Measurement... 09:02:26 Voltage Offset Correction: v_off_ini = 839.710 uV, v_off_fin = 847.649 uV 09:02:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 116.386 uVs 09:02:26 Integrated Voltage Measurement... 09:02:35 Voltage Offset Correction: v_off_ini = 844.498 uV, v_off_fin = 825.894 uV 09:02:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -118.659 uVs 09:02:35 Integrated Voltage Measurement... 09:02:44 Voltage Offset Correction: v_off_ini = 841.432 uV, v_off_fin = 862.334 uV 09:02:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.932 uVs 09:02:44 Integrated Voltage Measurement... 09:02:53 Voltage Offset Correction: v_off_ini = 844.400 uV, v_off_fin = 838.159 uV 09:02:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.238 uVs 09:02:53 Integrated Voltage Measurement... 09:03:02 Voltage Offset Correction: v_off_ini = 847.596 uV, v_off_fin = 862.500 uV 09:03:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.947 uVs 09:03:05 I2X Average and RMS of Measurements VT = -115.698 +- 4.966 uVs I2X = 10.644 +- 0.457 uTm^2 09:03:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:03:12 Integrated Voltage Measurement... 09:03:21 Voltage Offset Correction: v_off_ini = 850.423 uV, v_off_fin = 857.927 uV 09:03:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -360.162 uVs 09:03:21 Integrated Voltage Measurement... 09:03:30 Voltage Offset Correction: v_off_ini = 860.264 uV, v_off_fin = 860.213 uV 09:03:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 364.660 uVs 09:03:30 Integrated Voltage Measurement... 09:03:39 Voltage Offset Correction: v_off_ini = 845.112 uV, v_off_fin = 837.984 uV 09:03:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -352.316 uVs 09:03:40 Integrated Voltage Measurement... 09:03:49 Voltage Offset Correction: v_off_ini = 856.334 uV, v_off_fin = 855.911 uV 09:03:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 363.366 uVs 09:03:49 Integrated Voltage Measurement... 09:03:58 Voltage Offset Correction: v_off_ini = 854.768 uV, v_off_fin = 834.467 uV 09:03:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -382.193 uVs 09:03:58 Integrated Voltage Measurement... 09:04:07 Voltage Offset Correction: v_off_ini = 848.855 uV, v_off_fin = 854.436 uV 09:04:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 373.386 uVs 09:04:09 I2Y Average and RMS of Measurements VT = -366.014 +- 8.533 uVs I2Y = -33.673 +- 0.785 uTm^2 09:04:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:04:24 Integrated Voltage Measurement... 09:04:33 Voltage Offset Correction: v_off_ini = 856.138 uV, v_off_fin = 839.743 uV 09:04:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.944 uVs 09:04:33 Integrated Voltage Measurement... 09:04:42 Voltage Offset Correction: v_off_ini = 839.543 uV, v_off_fin = 840.499 uV 09:04:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.210 uVs 09:04:42 Integrated Voltage Measurement... 09:04:51 Voltage Offset Correction: v_off_ini = 845.620 uV, v_off_fin = 851.125 uV 09:04:51 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.660 uVs 09:04:51 Integrated Voltage Measurement... 09:05:00 Voltage Offset Correction: v_off_ini = 849.330 uV, v_off_fin = 864.165 uV 09:05:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 81.730 uVs 09:05:00 Integrated Voltage Measurement... 09:05:09 Voltage Offset Correction: v_off_ini = 844.783 uV, v_off_fin = 849.757 uV 09:05:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.507 uVs 09:05:09 Integrated Voltage Measurement... 09:05:18 Voltage Offset Correction: v_off_ini = 853.346 uV, v_off_fin = 863.115 uV 09:05:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.538 uVs 09:05:22 I1X Average and RMS of Measurements VT = -87.098 +- 7.013 uVs I1X = 17.420 +- 1.402 uTm 09:05:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:05:29 Integrated Voltage Measurement... 09:05:38 Voltage Offset Correction: v_off_ini = 834.192 uV, v_off_fin = 816.635 uV 09:05:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -219.595 uVs 09:05:38 Integrated Voltage Measurement... 09:05:47 Voltage Offset Correction: v_off_ini = 831.963 uV, v_off_fin = 836.332 uV 09:05:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 217.757 uVs 09:05:47 Integrated Voltage Measurement... 09:05:56 Voltage Offset Correction: v_off_ini = 828.572 uV, v_off_fin = 830.864 uV 09:05:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -207.268 uVs 09:05:56 Integrated Voltage Measurement... 09:06:05 Voltage Offset Correction: v_off_ini = 827.621 uV, v_off_fin = 843.661 uV 09:06:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 218.222 uVs 09:06:05 Integrated Voltage Measurement... 09:06:14 Voltage Offset Correction: v_off_ini = 824.758 uV, v_off_fin = 823.892 uV 09:06:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.205 uVs 09:06:14 Integrated Voltage Measurement... 09:06:23 Voltage Offset Correction: v_off_ini = 843.945 uV, v_off_fin = 855.592 uV 09:06:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 200.820 uVs 09:06:28 I1Y Average and RMS of Measurements VT = -211.811 +- 6.023 uVs I1Y = -42.360 +- 1.205 uTm 09:06:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:06:35 Integrated Voltage Measurement... 09:06:44 Voltage Offset Correction: v_off_ini = 832.612 uV, v_off_fin = 825.890 uV 09:06:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.610 uVs 09:06:44 Integrated Voltage Measurement... 09:06:53 Voltage Offset Correction: v_off_ini = 846.718 uV, v_off_fin = 850.618 uV 09:06:53 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 89.833 uVs 09:06:53 Integrated Voltage Measurement... 09:07:02 Voltage Offset Correction: v_off_ini = 849.885 uV, v_off_fin = 831.591 uV 09:07:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -131.421 uVs 09:07:02 Integrated Voltage Measurement... 09:07:11 Voltage Offset Correction: v_off_ini = 851.831 uV, v_off_fin = 853.837 uV 09:07:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.554 uVs 09:07:12 Integrated Voltage Measurement... 09:07:21 Voltage Offset Correction: v_off_ini = 852.601 uV, v_off_fin = 839.869 uV 09:07:21 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -109.696 uVs 09:07:21 Integrated Voltage Measurement... 09:07:30 Voltage Offset Correction: v_off_ini = 848.909 uV, v_off_fin = 861.141 uV 09:07:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.827 uVs 09:07:32 I2X Average and RMS of Measurements VT = -108.824 +- 4.814 uVs I2X = 10.012 +- 0.443 uTm^2 09:07:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:07:40 Integrated Voltage Measurement... 09:07:49 Voltage Offset Correction: v_off_ini = 855.015 uV, v_off_fin = 853.388 uV 09:07:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -354.742 uVs 09:07:49 Integrated Voltage Measurement... 09:07:58 Voltage Offset Correction: v_off_ini = 842.123 uV, v_off_fin = 849.299 uV 09:07:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 359.004 uVs 09:07:58 Integrated Voltage Measurement... 09:08:07 Voltage Offset Correction: v_off_ini = 855.111 uV, v_off_fin = 854.778 uV 09:08:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -356.495 uVs 09:08:07 Integrated Voltage Measurement... 09:08:16 Voltage Offset Correction: v_off_ini = 854.269 uV, v_off_fin = 856.326 uV 09:08:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 364.269 uVs 09:08:16 Integrated Voltage Measurement... 09:08:25 Voltage Offset Correction: v_off_ini = 850.468 uV, v_off_fin = 849.385 uV 09:08:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -358.619 uVs 09:08:25 Integrated Voltage Measurement... 09:08:34 Voltage Offset Correction: v_off_ini = 860.642 uV, v_off_fin = 860.410 uV 09:08:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 359.692 uVs 09:08:36 I2Y Average and RMS of Measurements VT = -358.803 +- 1.454 uVs I2Y = -33.010 +- 0.134 uTm^2 09:08:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:08:51 Integrated Voltage Measurement... 09:09:00 Voltage Offset Correction: v_off_ini = 850.164 uV, v_off_fin = 848.963 uV 09:09:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.229 uVs 09:09:00 Integrated Voltage Measurement... 09:09:09 Voltage Offset Correction: v_off_ini = 866.095 uV, v_off_fin = 868.724 uV 09:09:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 59.624 uVs 09:09:10 Integrated Voltage Measurement... 09:09:18 Voltage Offset Correction: v_off_ini = 876.213 uV, v_off_fin = 864.823 uV 09:09:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.959 uVs 09:09:19 Integrated Voltage Measurement... 09:09:28 Voltage Offset Correction: v_off_ini = 876.231 uV, v_off_fin = 882.529 uV 09:09:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.140 uVs 09:09:28 Integrated Voltage Measurement... 09:09:37 Voltage Offset Correction: v_off_ini = 876.519 uV, v_off_fin = 866.560 uV 09:09:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.063 uVs 09:09:37 Integrated Voltage Measurement... 09:09:46 Voltage Offset Correction: v_off_ini = 885.809 uV, v_off_fin = 866.715 uV 09:09:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.233 uVs 09:09:50 I1X Average and RMS of Measurements VT = -72.375 +- 2.870 uVs I1X = 14.475 +- 0.575 uTm 09:09:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:09:57 Integrated Voltage Measurement... 09:10:06 Voltage Offset Correction: v_off_ini = 884.106 uV, v_off_fin = 883.718 uV 09:10:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.446 uVs 09:10:06 Integrated Voltage Measurement... 09:10:15 Voltage Offset Correction: v_off_ini = 892.843 uV, v_off_fin = 892.619 uV 09:10:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.585 uVs 09:10:15 Integrated Voltage Measurement... 09:10:24 Voltage Offset Correction: v_off_ini = 867.752 uV, v_off_fin = 872.469 uV 09:10:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.414 uVs 09:10:24 Integrated Voltage Measurement... 09:10:33 Voltage Offset Correction: v_off_ini = 884.722 uV, v_off_fin = 899.261 uV 09:10:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.874 uVs 09:10:33 Integrated Voltage Measurement... 09:10:42 Voltage Offset Correction: v_off_ini = 882.200 uV, v_off_fin = 867.753 uV 09:10:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -220.640 uVs 09:10:42 Integrated Voltage Measurement... 09:10:51 Voltage Offset Correction: v_off_ini = 883.296 uV, v_off_fin = 888.067 uV 09:10:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.414 uVs 09:10:55 I1Y Average and RMS of Measurements VT = -216.062 +- 2.629 uVs I1Y = -43.211 +- 0.527 uTm 09:11:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:11:03 Integrated Voltage Measurement... 09:11:12 Voltage Offset Correction: v_off_ini = 887.243 uV, v_off_fin = 866.678 uV 09:11:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -114.936 uVs 09:11:12 Integrated Voltage Measurement... 09:11:21 Voltage Offset Correction: v_off_ini = 881.231 uV, v_off_fin = 879.666 uV 09:11:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.829 uVs 09:11:21 Integrated Voltage Measurement... 09:11:30 Voltage Offset Correction: v_off_ini = 890.613 uV, v_off_fin = 877.841 uV 09:11:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.968 uVs 09:11:30 Integrated Voltage Measurement... 09:11:39 Voltage Offset Correction: v_off_ini = 870.352 uV, v_off_fin = 883.678 uV 09:11:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 124.945 uVs 09:11:39 Integrated Voltage Measurement... 09:11:48 Voltage Offset Correction: v_off_ini = 881.119 uV, v_off_fin = 872.361 uV 09:11:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.064 uVs 09:11:48 Integrated Voltage Measurement... 09:11:57 Voltage Offset Correction: v_off_ini = 880.538 uV, v_off_fin = 877.484 uV 09:11:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.538 uVs 09:11:59 I2X Average and RMS of Measurements VT = -111.713 +- 11.256 uVs I2X = 10.278 +- 1.036 uTm^2 09:12:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:12:07 Integrated Voltage Measurement... 09:12:16 Voltage Offset Correction: v_off_ini = 885.256 uV, v_off_fin = 874.822 uV 09:12:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -377.574 uVs 09:12:16 Integrated Voltage Measurement... 09:12:25 Voltage Offset Correction: v_off_ini = 865.399 uV, v_off_fin = 871.861 uV 09:12:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 364.504 uVs 09:12:25 Integrated Voltage Measurement... 09:12:34 Voltage Offset Correction: v_off_ini = 882.650 uV, v_off_fin = 879.233 uV 09:12:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -363.344 uVs 09:12:34 Integrated Voltage Measurement... 09:12:43 Voltage Offset Correction: v_off_ini = 860.115 uV, v_off_fin = 861.876 uV 09:12:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 345.987 uVs 09:12:44 Integrated Voltage Measurement... 09:12:53 Voltage Offset Correction: v_off_ini = 875.241 uV, v_off_fin = 874.049 uV 09:12:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -365.088 uVs 09:12:53 Integrated Voltage Measurement... 09:13:02 Voltage Offset Correction: v_off_ini = 866.952 uV, v_off_fin = 865.157 uV 09:13:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 366.306 uVs 09:13:04 I2Y Average and RMS of Measurements VT = -363.801 +- 6.818 uVs I2Y = -33.470 +- 0.627 uTm^2 09:13:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:13:19 Integrated Voltage Measurement... 09:13:28 Voltage Offset Correction: v_off_ini = 882.326 uV, v_off_fin = 871.137 uV 09:13:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.526 uVs 09:13:28 Integrated Voltage Measurement... 09:13:37 Voltage Offset Correction: v_off_ini = 890.995 uV, v_off_fin = 886.848 uV 09:13:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.329 uVs 09:13:37 Integrated Voltage Measurement... 09:13:46 Voltage Offset Correction: v_off_ini = 886.502 uV, v_off_fin = 889.661 uV 09:13:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.973 uVs 09:13:46 Integrated Voltage Measurement... 09:13:55 Voltage Offset Correction: v_off_ini = 892.530 uV, v_off_fin = 898.975 uV 09:13:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.857 uVs 09:13:55 Integrated Voltage Measurement... 09:14:04 Voltage Offset Correction: v_off_ini = 895.708 uV, v_off_fin = 893.130 uV 09:14:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.903 uVs 09:14:04 Integrated Voltage Measurement... 09:14:13 Voltage Offset Correction: v_off_ini = 899.876 uV, v_off_fin = 912.082 uV 09:14:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.844 uVs 09:14:17 I1X Average and RMS of Measurements VT = -66.405 +- 4.942 uVs I1X = 13.281 +- 0.988 uTm 09:14:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:14:24 Integrated Voltage Measurement... 09:14:33 Voltage Offset Correction: v_off_ini = 893.554 uV, v_off_fin = 878.411 uV 09:14:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -207.555 uVs 09:14:33 Integrated Voltage Measurement... 09:14:42 Voltage Offset Correction: v_off_ini = 904.838 uV, v_off_fin = 915.979 uV 09:14:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.491 uVs 09:14:42 Integrated Voltage Measurement... 09:14:51 Voltage Offset Correction: v_off_ini = 890.988 uV, v_off_fin = 883.035 uV 09:14:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -201.949 uVs 09:14:51 Integrated Voltage Measurement... 09:15:00 Voltage Offset Correction: v_off_ini = 903.971 uV, v_off_fin = 912.926 uV 09:15:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.030 uVs 09:15:00 Integrated Voltage Measurement... 09:15:09 Voltage Offset Correction: v_off_ini = 902.080 uV, v_off_fin = 892.546 uV 09:15:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -207.796 uVs 09:15:09 Integrated Voltage Measurement... 09:15:18 Voltage Offset Correction: v_off_ini = 898.510 uV, v_off_fin = 916.095 uV 09:15:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 223.643 uVs 09:15:22 I1Y Average and RMS of Measurements VT = -212.077 +- 4.644 uVs I1Y = -42.412 +- 0.929 uTm 09:15:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:15:30 Integrated Voltage Measurement... 09:15:39 Voltage Offset Correction: v_off_ini = 898.620 uV, v_off_fin = 893.225 uV 09:15:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.269 uVs 09:15:39 Integrated Voltage Measurement... 09:15:48 Voltage Offset Correction: v_off_ini = 917.672 uV, v_off_fin = 928.254 uV 09:15:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.913 uVs 09:15:48 Integrated Voltage Measurement... 09:15:57 Voltage Offset Correction: v_off_ini = 929.098 uV, v_off_fin = 928.192 uV 09:15:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.836 uVs 09:15:57 Integrated Voltage Measurement... 09:16:06 Voltage Offset Correction: v_off_ini = 934.402 uV, v_off_fin = 937.720 uV 09:16:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.526 uVs 09:16:06 Integrated Voltage Measurement... 09:16:15 Voltage Offset Correction: v_off_ini = 933.821 uV, v_off_fin = 915.446 uV 09:16:15 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -101.683 uVs 09:16:15 Integrated Voltage Measurement... 09:16:24 Voltage Offset Correction: v_off_ini = 924.059 uV, v_off_fin = 924.028 uV 09:16:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.534 uVs 09:16:27 I2X Average and RMS of Measurements VT = -96.460 +- 4.957 uVs I2X = 8.874 +- 0.456 uTm^2 09:16:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:16:34 Integrated Voltage Measurement... 09:16:43 Voltage Offset Correction: v_off_ini = 913.235 uV, v_off_fin = 925.768 uV 09:16:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -355.850 uVs 09:16:44 Integrated Voltage Measurement... 09:16:52 Voltage Offset Correction: v_off_ini = 940.753 uV, v_off_fin = 919.148 uV 09:16:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 351.511 uVs 09:16:53 Integrated Voltage Measurement... 09:17:02 Voltage Offset Correction: v_off_ini = 920.315 uV, v_off_fin = 899.845 uV 09:17:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.021 uVs 09:17:02 Integrated Voltage Measurement... 09:17:11 Voltage Offset Correction: v_off_ini = 904.362 uV, v_off_fin = 911.898 uV 09:17:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 358.770 uVs 09:17:11 Integrated Voltage Measurement... 09:17:20 Voltage Offset Correction: v_off_ini = 910.707 uV, v_off_fin = 904.560 uV 09:17:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -380.821 uVs 09:17:20 Integrated Voltage Measurement... 09:17:29 Voltage Offset Correction: v_off_ini = 907.577 uV, v_off_fin = 899.786 uV 09:17:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 348.959 uVs 09:17:31 I2Y Average and RMS of Measurements VT = -355.322 +- 7.236 uVs I2Y = -32.690 +- 0.666 uTm^2 09:17:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:17:46 Integrated Voltage Measurement... 09:17:55 Voltage Offset Correction: v_off_ini = 901.782 uV, v_off_fin = 887.504 uV 09:17:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.086 uVs 09:17:55 Integrated Voltage Measurement... 09:18:04 Voltage Offset Correction: v_off_ini = 907.143 uV, v_off_fin = 899.379 uV 09:18:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.417 uVs 09:18:04 Integrated Voltage Measurement... 09:18:13 Voltage Offset Correction: v_off_ini = 906.445 uV, v_off_fin = 896.510 uV 09:18:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.759 uVs 09:18:13 Integrated Voltage Measurement... 09:18:22 Voltage Offset Correction: v_off_ini = 899.182 uV, v_off_fin = 903.712 uV 09:18:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.422 uVs 09:18:22 Integrated Voltage Measurement... 09:18:31 Voltage Offset Correction: v_off_ini = 895.284 uV, v_off_fin = 897.376 uV 09:18:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.011 uVs 09:18:31 Integrated Voltage Measurement... 09:18:40 Voltage Offset Correction: v_off_ini = 905.674 uV, v_off_fin = 910.498 uV 09:18:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.170 uVs 09:18:44 I1X Average and RMS of Measurements VT = -67.811 +- 2.967 uVs I1X = 13.562 +- 0.593 uTm 09:18:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:18:51 Integrated Voltage Measurement... 09:19:00 Voltage Offset Correction: v_off_ini = 894.182 uV, v_off_fin = 888.259 uV 09:19:00 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.174 uVs 09:19:00 Integrated Voltage Measurement... 09:19:09 Voltage Offset Correction: v_off_ini = 909.240 uV, v_off_fin = 914.655 uV 09:19:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 210.769 uVs 09:19:09 Integrated Voltage Measurement... 09:19:18 Voltage Offset Correction: v_off_ini = 910.221 uV, v_off_fin = 893.507 uV 09:19:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -215.888 uVs 09:19:18 Integrated Voltage Measurement... 09:19:27 Voltage Offset Correction: v_off_ini = 903.013 uV, v_off_fin = 916.118 uV 09:19:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.612 uVs 09:19:27 Integrated Voltage Measurement... 09:19:36 Voltage Offset Correction: v_off_ini = 912.533 uV, v_off_fin = 894.601 uV 09:19:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.131 uVs 09:19:36 Integrated Voltage Measurement... 09:19:45 Voltage Offset Correction: v_off_ini = 905.513 uV, v_off_fin = 915.114 uV 09:19:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 205.290 uVs 09:19:49 I1Y Average and RMS of Measurements VT = -210.477 +- 1.292 uVs I1Y = -42.093 +- 0.259 uTm 09:19:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:19:57 Integrated Voltage Measurement... 09:20:06 Voltage Offset Correction: v_off_ini = 922.300 uV, v_off_fin = 924.912 uV 09:20:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.104 uVs 09:20:06 Integrated Voltage Measurement... 09:20:15 Voltage Offset Correction: v_off_ini = 923.031 uV, v_off_fin = 934.211 uV 09:20:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.090 uVs 09:20:15 Integrated Voltage Measurement... 09:20:24 Voltage Offset Correction: v_off_ini = 925.451 uV, v_off_fin = 924.150 uV 09:20:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.459 uVs 09:20:24 Integrated Voltage Measurement... 09:20:33 Voltage Offset Correction: v_off_ini = 936.279 uV, v_off_fin = 948.724 uV 09:20:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.146 uVs 09:20:33 Integrated Voltage Measurement... 09:20:42 Voltage Offset Correction: v_off_ini = 941.456 uV, v_off_fin = 936.440 uV 09:20:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.135 uVs 09:20:42 Integrated Voltage Measurement... 09:20:51 Voltage Offset Correction: v_off_ini = 949.233 uV, v_off_fin = 950.938 uV 09:20:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 116.657 uVs 09:20:54 I2X Average and RMS of Measurements VT = -101.932 +- 6.746 uVs I2X = 9.378 +- 0.621 uTm^2 09:20:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:21:02 Integrated Voltage Measurement... 09:21:11 Voltage Offset Correction: v_off_ini = 938.369 uV, v_off_fin = 936.244 uV 09:21:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -354.182 uVs 09:21:11 Integrated Voltage Measurement... 09:21:20 Voltage Offset Correction: v_off_ini = 936.460 uV, v_off_fin = 935.965 uV 09:21:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 351.112 uVs 09:21:20 Integrated Voltage Measurement... 09:21:29 Voltage Offset Correction: v_off_ini = 938.367 uV, v_off_fin = 937.339 uV 09:21:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -363.884 uVs 09:21:29 Integrated Voltage Measurement... 09:21:38 Voltage Offset Correction: v_off_ini = 937.619 uV, v_off_fin = 942.118 uV 09:21:38 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 355.225 uVs 09:21:38 Integrated Voltage Measurement... 09:21:47 Voltage Offset Correction: v_off_ini = 927.711 uV, v_off_fin = 938.920 uV 09:21:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.629 uVs 09:21:47 Integrated Voltage Measurement... 09:21:56 Voltage Offset Correction: v_off_ini = 943.503 uV, v_off_fin = 936.997 uV 09:21:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 351.605 uVs 09:21:58 I2Y Average and RMS of Measurements VT = -350.273 +- 8.711 uVs I2Y = -32.226 +- 0.802 uTm^2 09:22:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:22:13 Integrated Voltage Measurement... 09:22:22 Voltage Offset Correction: v_off_ini = 947.480 uV, v_off_fin = 939.445 uV 09:22:22 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.859 uVs 09:22:22 Integrated Voltage Measurement... 09:22:31 Voltage Offset Correction: v_off_ini = 951.063 uV, v_off_fin = 951.881 uV 09:22:31 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 53.823 uVs 09:22:31 Integrated Voltage Measurement... 09:22:40 Voltage Offset Correction: v_off_ini = 949.170 uV, v_off_fin = 945.457 uV 09:22:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -50.109 uVs 09:22:40 Integrated Voltage Measurement... 09:22:49 Voltage Offset Correction: v_off_ini = 955.388 uV, v_off_fin = 960.614 uV 09:22:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.470 uVs 09:22:49 Integrated Voltage Measurement... 09:22:58 Voltage Offset Correction: v_off_ini = 969.850 uV, v_off_fin = 955.046 uV 09:22:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.754 uVs 09:22:58 Integrated Voltage Measurement... 09:23:07 Voltage Offset Correction: v_off_ini = 956.044 uV, v_off_fin = 971.540 uV 09:23:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.785 uVs 09:23:12 I1X Average and RMS of Measurements VT = -60.800 +- 6.851 uVs I1X = 12.161 +- 1.370 uTm 09:23:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:23:18 Integrated Voltage Measurement... 09:23:27 Voltage Offset Correction: v_off_ini = 960.037 uV, v_off_fin = 944.114 uV 09:23:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.318 uVs 09:23:27 Integrated Voltage Measurement... 09:23:36 Voltage Offset Correction: v_off_ini = 958.941 uV, v_off_fin = 975.913 uV 09:23:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 201.806 uVs 09:23:36 Integrated Voltage Measurement... 09:23:45 Voltage Offset Correction: v_off_ini = 953.588 uV, v_off_fin = 939.603 uV 09:23:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.990 uVs 09:23:45 Integrated Voltage Measurement... 09:23:54 Voltage Offset Correction: v_off_ini = 972.732 uV, v_off_fin = 985.400 uV 09:23:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 229.854 uVs 09:23:54 Integrated Voltage Measurement... 09:24:03 Voltage Offset Correction: v_off_ini = 974.356 uV, v_off_fin = 966.248 uV 09:24:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.965 uVs 09:24:03 Integrated Voltage Measurement... 09:24:12 Voltage Offset Correction: v_off_ini = 987.710 uV, v_off_fin = 998.948 uV 09:24:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 239.754 uVs 09:24:17 I1Y Average and RMS of Measurements VT = -214.781 +- 8.022 uVs I1Y = -42.953 +- 1.604 uTm 09:24:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:24:24 Integrated Voltage Measurement... 09:24:33 Voltage Offset Correction: v_off_ini = 973.524 uV, v_off_fin = 976.168 uV 09:24:33 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -93.461 uVs 09:24:33 Integrated Voltage Measurement... 09:24:42 Voltage Offset Correction: v_off_ini = 984.398 uV, v_off_fin = 997.538 uV 09:24:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.219 uVs 09:24:42 Integrated Voltage Measurement... 09:24:51 Voltage Offset Correction: v_off_ini = 995.869 uV, v_off_fin = 985.398 uV 09:24:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.233 uVs 09:24:51 Integrated Voltage Measurement... 09:25:00 Voltage Offset Correction: v_off_ini = 995.928 uV, v_off_fin = 1001.500 uV 09:25:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.559 uVs 09:25:00 Integrated Voltage Measurement... 09:25:09 Voltage Offset Correction: v_off_ini = 992.992 uV, v_off_fin = 985.500 uV 09:25:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.470 uVs 09:25:10 Integrated Voltage Measurement... 09:25:19 Voltage Offset Correction: v_off_ini = 998.011 uV, v_off_fin = 992.742 uV 09:25:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.600 uVs 09:25:21 I2X Average and RMS of Measurements VT = -90.424 +- 7.567 uVs I2X = 8.319 +- 0.696 uTm^2 09:25:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:25:29 Integrated Voltage Measurement... 09:25:38 Voltage Offset Correction: v_off_ini = 990.360 uV, v_off_fin = 992.534 uV 09:25:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -358.171 uVs 09:25:38 Integrated Voltage Measurement... 09:25:47 Voltage Offset Correction: v_off_ini = 995.910 uV, v_off_fin = 988.012 uV 09:25:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 341.702 uVs 09:25:47 Integrated Voltage Measurement... 09:25:56 Voltage Offset Correction: v_off_ini = 989.239 uV, v_off_fin = 976.948 uV 09:25:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -351.861 uVs 09:25:56 Integrated Voltage Measurement... 09:26:05 Voltage Offset Correction: v_off_ini = 972.771 uV, v_off_fin = 983.813 uV 09:26:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 364.285 uVs 09:26:05 Integrated Voltage Measurement... 09:26:14 Voltage Offset Correction: v_off_ini = 991.391 uV, v_off_fin = 986.308 uV 09:26:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -358.102 uVs 09:26:14 Integrated Voltage Measurement... 09:26:23 Voltage Offset Correction: v_off_ini = 986.189 uV, v_off_fin = 981.821 uV 09:26:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.821 uVs 09:26:25 I2Y Average and RMS of Measurements VT = -352.490 +- 3.952 uVs I2Y = -32.429 +- 0.364 uTm^2 09:26:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:26:40 Integrated Voltage Measurement... 09:26:49 Voltage Offset Correction: v_off_ini = 983.808 uV, v_off_fin = 972.437 uV 09:26:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.589 uVs 09:26:49 Integrated Voltage Measurement... 09:26:58 Voltage Offset Correction: v_off_ini = 962.647 uV, v_off_fin = 979.045 uV 09:26:58 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 69.063 uVs 09:26:58 Integrated Voltage Measurement... 09:27:07 Voltage Offset Correction: v_off_ini = 977.688 uV, v_off_fin = 968.305 uV 09:27:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.741 uVs 09:27:07 Integrated Voltage Measurement... 09:27:16 Voltage Offset Correction: v_off_ini = 973.453 uV, v_off_fin = 988.958 uV 09:27:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.777 uVs 09:27:16 Integrated Voltage Measurement... 09:27:25 Voltage Offset Correction: v_off_ini = 981.489 uV, v_off_fin = 973.196 uV 09:27:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.086 uVs 09:27:25 Integrated Voltage Measurement... 09:27:34 Voltage Offset Correction: v_off_ini = 975.434 uV, v_off_fin = 973.426 uV 09:27:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.646 uVs 09:27:39 I1X Average and RMS of Measurements VT = -68.984 +- 4.722 uVs I1X = 13.797 +- 0.944 uTm 09:27:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:27:45 Integrated Voltage Measurement... 09:27:54 Voltage Offset Correction: v_off_ini = 970.091 uV, v_off_fin = 955.951 uV 09:27:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -203.276 uVs 09:27:54 Integrated Voltage Measurement... 09:28:03 Voltage Offset Correction: v_off_ini = 966.141 uV, v_off_fin = 978.783 uV 09:28:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 213.433 uVs 09:28:03 Integrated Voltage Measurement... 09:28:12 Voltage Offset Correction: v_off_ini = 963.826 uV, v_off_fin = 941.228 uV 09:28:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.145 uVs 09:28:12 Integrated Voltage Measurement... 09:28:21 Voltage Offset Correction: v_off_ini = 958.165 uV, v_off_fin = 959.014 uV 09:28:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 214.024 uVs 09:28:21 Integrated Voltage Measurement... 09:28:30 Voltage Offset Correction: v_off_ini = 960.075 uV, v_off_fin = 944.953 uV 09:28:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.349 uVs 09:28:30 Integrated Voltage Measurement... 09:28:39 Voltage Offset Correction: v_off_ini = 948.707 uV, v_off_fin = 965.575 uV 09:28:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.959 uVs 09:28:44 I1Y Average and RMS of Measurements VT = -211.198 +- 2.572 uVs I1Y = -42.239 +- 0.514 uTm 09:28:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:28:51 Integrated Voltage Measurement... 09:29:00 Voltage Offset Correction: v_off_ini = 943.193 uV, v_off_fin = 939.978 uV 09:29:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.386 uVs 09:29:00 Integrated Voltage Measurement... 09:29:09 Voltage Offset Correction: v_off_ini = 947.485 uV, v_off_fin = 961.813 uV 09:29:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.198 uVs 09:29:09 Integrated Voltage Measurement... 09:29:18 Voltage Offset Correction: v_off_ini = 938.302 uV, v_off_fin = 922.742 uV 09:29:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.666 uVs 09:29:18 Integrated Voltage Measurement... 09:29:27 Voltage Offset Correction: v_off_ini = 947.287 uV, v_off_fin = 952.712 uV 09:29:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.412 uVs 09:29:28 Integrated Voltage Measurement... 09:29:36 Voltage Offset Correction: v_off_ini = 944.886 uV, v_off_fin = 929.069 uV 09:29:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.365 uVs 09:29:37 Integrated Voltage Measurement... 09:29:46 Voltage Offset Correction: v_off_ini = 945.583 uV, v_off_fin = 948.535 uV 09:29:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.659 uVs 09:29:48 I2X Average and RMS of Measurements VT = -83.948 +- 15.080 uVs I2X = 7.723 +- 1.387 uTm^2 09:29:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:29:56 Integrated Voltage Measurement... 09:30:05 Voltage Offset Correction: v_off_ini = 950.573 uV, v_off_fin = 947.299 uV 09:30:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.954 uVs 09:30:05 Integrated Voltage Measurement... 09:30:14 Voltage Offset Correction: v_off_ini = 955.210 uV, v_off_fin = 949.904 uV 09:30:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 335.197 uVs 09:30:14 Integrated Voltage Measurement... 09:30:23 Voltage Offset Correction: v_off_ini = 959.956 uV, v_off_fin = 957.648 uV 09:30:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -349.761 uVs 09:30:23 Integrated Voltage Measurement... 09:30:32 Voltage Offset Correction: v_off_ini = 955.724 uV, v_off_fin = 957.265 uV 09:30:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 337.204 uVs 09:30:32 Integrated Voltage Measurement... 09:30:41 Voltage Offset Correction: v_off_ini = 944.804 uV, v_off_fin = 939.183 uV 09:30:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -335.330 uVs 09:30:41 Integrated Voltage Measurement... 09:30:50 Voltage Offset Correction: v_off_ini = 950.118 uV, v_off_fin = 956.918 uV 09:30:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 344.500 uVs 09:30:52 I2Y Average and RMS of Measurements VT = -339.824 +- 3.024 uVs I2Y = -31.264 +- 0.278 uTm^2 09:31:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:31:07 Integrated Voltage Measurement... 09:31:16 Voltage Offset Correction: v_off_ini = 951.597 uV, v_off_fin = 948.449 uV 09:31:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.047 uVs 09:31:16 Integrated Voltage Measurement... 09:31:25 Voltage Offset Correction: v_off_ini = 935.087 uV, v_off_fin = 934.992 uV 09:31:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.749 uVs 09:31:26 Integrated Voltage Measurement... 09:31:34 Voltage Offset Correction: v_off_ini = 909.329 uV, v_off_fin = 905.919 uV 09:31:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.958 uVs 09:31:35 Integrated Voltage Measurement... 09:31:44 Voltage Offset Correction: v_off_ini = 926.770 uV, v_off_fin = 918.783 uV 09:31:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.339 uVs 09:31:44 Integrated Voltage Measurement... 09:31:53 Voltage Offset Correction: v_off_ini = 927.562 uV, v_off_fin = 935.939 uV 09:31:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.961 uVs 09:31:53 Integrated Voltage Measurement... 09:32:02 Voltage Offset Correction: v_off_ini = 946.992 uV, v_off_fin = 947.360 uV 09:32:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.355 uVs 09:32:06 I1X Average and RMS of Measurements VT = -65.402 +- 8.320 uVs I1X = 13.080 +- 1.664 uTm 09:32:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:32:13 Integrated Voltage Measurement... 09:32:22 Voltage Offset Correction: v_off_ini = 945.816 uV, v_off_fin = 938.279 uV 09:32:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -224.513 uVs 09:32:22 Integrated Voltage Measurement... 09:32:31 Voltage Offset Correction: v_off_ini = 930.915 uV, v_off_fin = 947.203 uV 09:32:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 231.167 uVs 09:32:31 Integrated Voltage Measurement... 09:32:40 Voltage Offset Correction: v_off_ini = 936.205 uV, v_off_fin = 921.296 uV 09:32:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -217.254 uVs 09:32:40 Integrated Voltage Measurement... 09:32:49 Voltage Offset Correction: v_off_ini = 926.654 uV, v_off_fin = 945.654 uV 09:32:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 225.050 uVs 09:32:49 Integrated Voltage Measurement... 09:32:58 Voltage Offset Correction: v_off_ini = 938.014 uV, v_off_fin = 923.689 uV 09:32:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -214.994 uVs 09:32:58 Integrated Voltage Measurement... 09:33:07 Voltage Offset Correction: v_off_ini = 934.164 uV, v_off_fin = 950.196 uV 09:33:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.248 uVs 09:33:11 I1Y Average and RMS of Measurements VT = -220.371 +- 6.441 uVs I1Y = -44.071 +- 1.288 uTm 09:33:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:33:19 Integrated Voltage Measurement... 09:33:28 Voltage Offset Correction: v_off_ini = 917.480 uV, v_off_fin = 922.514 uV 09:33:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.306 uVs 09:33:28 Integrated Voltage Measurement... 09:33:37 Voltage Offset Correction: v_off_ini = 925.876 uV, v_off_fin = 931.942 uV 09:33:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.286 uVs 09:33:37 Integrated Voltage Measurement... 09:33:46 Voltage Offset Correction: v_off_ini = 936.883 uV, v_off_fin = 943.522 uV 09:33:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.217 uVs 09:33:46 Integrated Voltage Measurement... 09:33:55 Voltage Offset Correction: v_off_ini = 955.449 uV, v_off_fin = 945.054 uV 09:33:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.693 uVs 09:33:55 Integrated Voltage Measurement... 09:34:04 Voltage Offset Correction: v_off_ini = 939.127 uV, v_off_fin = 916.578 uV 09:34:04 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -100.917 uVs 09:34:04 Integrated Voltage Measurement... 09:34:13 Voltage Offset Correction: v_off_ini = 944.141 uV, v_off_fin = 928.199 uV 09:34:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.652 uVs 09:34:15 I2X Average and RMS of Measurements VT = -83.178 +- 8.031 uVs I2X = 7.653 +- 0.739 uTm^2 09:34:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:34:23 Integrated Voltage Measurement... 09:34:32 Voltage Offset Correction: v_off_ini = 916.205 uV, v_off_fin = 919.991 uV 09:34:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.503 uVs 09:34:32 Integrated Voltage Measurement... 09:34:41 Voltage Offset Correction: v_off_ini = 929.531 uV, v_off_fin = 915.591 uV 09:34:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 348.383 uVs 09:34:41 Integrated Voltage Measurement... 09:34:50 Voltage Offset Correction: v_off_ini = 908.489 uV, v_off_fin = 913.899 uV 09:34:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.144 uVs 09:34:50 Integrated Voltage Measurement... 09:34:59 Voltage Offset Correction: v_off_ini = 922.763 uV, v_off_fin = 913.902 uV 09:34:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 328.793 uVs 09:34:59 Integrated Voltage Measurement... 09:35:08 Voltage Offset Correction: v_off_ini = 919.760 uV, v_off_fin = 911.540 uV 09:35:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -332.425 uVs 09:35:08 Integrated Voltage Measurement... 09:35:17 Voltage Offset Correction: v_off_ini = 902.554 uV, v_off_fin = 909.799 uV 09:35:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 352.740 uVs 09:35:20 I2Y Average and RMS of Measurements VT = -335.831 +- 6.018 uVs I2Y = -30.896 +- 0.554 uTm^2 09:35:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:35:35 Integrated Voltage Measurement... 09:35:44 Voltage Offset Correction: v_off_ini = 921.652 uV, v_off_fin = 903.643 uV 09:35:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -52.419 uVs 09:35:44 Integrated Voltage Measurement... 09:35:53 Voltage Offset Correction: v_off_ini = 927.827 uV, v_off_fin = 932.351 uV 09:35:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.590 uVs 09:35:53 Integrated Voltage Measurement... 09:36:02 Voltage Offset Correction: v_off_ini = 932.409 uV, v_off_fin = 925.079 uV 09:36:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.465 uVs 09:36:02 Integrated Voltage Measurement... 09:36:11 Voltage Offset Correction: v_off_ini = 927.967 uV, v_off_fin = 931.086 uV 09:36:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.016 uVs 09:36:11 Integrated Voltage Measurement... 09:36:20 Voltage Offset Correction: v_off_ini = 913.222 uV, v_off_fin = 925.517 uV 09:36:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.725 uVs 09:36:20 Integrated Voltage Measurement... 09:36:29 Voltage Offset Correction: v_off_ini = 925.617 uV, v_off_fin = 933.986 uV 09:36:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.593 uVs 09:36:33 I1X Average and RMS of Measurements VT = -57.802 +- 8.894 uVs I1X = 11.560 +- 1.779 uTm 09:36:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:36:40 Integrated Voltage Measurement... 09:36:49 Voltage Offset Correction: v_off_ini = 938.336 uV, v_off_fin = 911.472 uV 09:36:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.538 uVs 09:36:49 Integrated Voltage Measurement... 09:36:58 Voltage Offset Correction: v_off_ini = 914.590 uV, v_off_fin = 940.851 uV 09:36:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 212.306 uVs 09:36:58 Integrated Voltage Measurement... 09:37:07 Voltage Offset Correction: v_off_ini = 936.480 uV, v_off_fin = 922.685 uV 09:37:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.745 uVs 09:37:07 Integrated Voltage Measurement... 09:37:16 Voltage Offset Correction: v_off_ini = 932.303 uV, v_off_fin = 946.961 uV 09:37:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 215.364 uVs 09:37:16 Integrated Voltage Measurement... 09:37:25 Voltage Offset Correction: v_off_ini = 935.527 uV, v_off_fin = 924.519 uV 09:37:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.934 uVs 09:37:25 Integrated Voltage Measurement... 09:37:34 Voltage Offset Correction: v_off_ini = 940.962 uV, v_off_fin = 948.278 uV 09:37:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.693 uVs 09:37:38 I1Y Average and RMS of Measurements VT = -208.597 +- 5.542 uVs I1Y = -41.719 +- 1.109 uTm 09:37:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:37:46 Integrated Voltage Measurement... 09:37:55 Voltage Offset Correction: v_off_ini = 944.445 uV, v_off_fin = 932.848 uV 09:37:55 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -67.568 uVs 09:37:55 Integrated Voltage Measurement... 09:38:04 Voltage Offset Correction: v_off_ini = 943.245 uV, v_off_fin = 952.777 uV 09:38:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.735 uVs 09:38:04 Integrated Voltage Measurement... 09:38:13 Voltage Offset Correction: v_off_ini = 940.359 uV, v_off_fin = 921.219 uV 09:38:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.247 uVs 09:38:13 Integrated Voltage Measurement... 09:38:22 Voltage Offset Correction: v_off_ini = 949.851 uV, v_off_fin = 971.843 uV 09:38:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.215 uVs 09:38:22 Integrated Voltage Measurement... 09:38:31 Voltage Offset Correction: v_off_ini = 981.340 uV, v_off_fin = 967.024 uV 09:38:31 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -95.691 uVs 09:38:31 Integrated Voltage Measurement... 09:38:40 Voltage Offset Correction: v_off_ini = 958.051 uV, v_off_fin = 968.810 uV 09:38:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.462 uVs 09:38:42 I2X Average and RMS of Measurements VT = -89.486 +- 9.272 uVs I2X = 8.233 +- 0.853 uTm^2 09:38:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:38:50 Integrated Voltage Measurement... 09:38:59 Voltage Offset Correction: v_off_ini = 979.098 uV, v_off_fin = 977.540 uV 09:38:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.019 uVs 09:38:59 Integrated Voltage Measurement... 09:39:08 Voltage Offset Correction: v_off_ini = 973.215 uV, v_off_fin = 981.272 uV 09:39:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 321.779 uVs 09:39:08 Integrated Voltage Measurement... 09:39:17 Voltage Offset Correction: v_off_ini = 976.080 uV, v_off_fin = 976.364 uV 09:39:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.821 uVs 09:39:17 Integrated Voltage Measurement... 09:39:26 Voltage Offset Correction: v_off_ini = 983.535 uV, v_off_fin = 952.148 uV 09:39:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 333.632 uVs 09:39:26 Integrated Voltage Measurement... 09:39:35 Voltage Offset Correction: v_off_ini = 949.752 uV, v_off_fin = 956.671 uV 09:39:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.375 uVs 09:39:35 Integrated Voltage Measurement... 09:39:44 Voltage Offset Correction: v_off_ini = 959.756 uV, v_off_fin = 976.043 uV 09:39:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.868 uVs 09:39:47 I2Y Average and RMS of Measurements VT = -327.749 +- 3.558 uVs I2Y = -30.153 +- 0.327 uTm^2 09:39:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:40:02 Integrated Voltage Measurement... 09:40:11 Voltage Offset Correction: v_off_ini = 1000.133 uV, v_off_fin = 983.997 uV 09:40:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -61.228 uVs 09:40:11 Integrated Voltage Measurement... 09:40:20 Voltage Offset Correction: v_off_ini = 981.487 uV, v_off_fin = 984.645 uV 09:40:20 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 67.012 uVs 09:40:20 Integrated Voltage Measurement... 09:40:29 Voltage Offset Correction: v_off_ini = 986.350 uV, v_off_fin = 986.620 uV 09:40:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.774 uVs 09:40:29 Integrated Voltage Measurement... 09:40:38 Voltage Offset Correction: v_off_ini = 987.806 uV, v_off_fin = 997.461 uV 09:40:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.677 uVs 09:40:38 Integrated Voltage Measurement... 09:40:47 Voltage Offset Correction: v_off_ini = 986.619 uV, v_off_fin = 988.106 uV 09:40:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.495 uVs 09:40:47 Integrated Voltage Measurement... 09:40:56 Voltage Offset Correction: v_off_ini = 1000.910 uV, v_off_fin = 1005.833 uV 09:40:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.641 uVs 09:41:00 I1X Average and RMS of Measurements VT = -61.471 +- 2.528 uVs I1X = 12.295 +- 0.506 uTm 09:41:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:41:07 Integrated Voltage Measurement... 09:41:16 Voltage Offset Correction: v_off_ini = 1004.143 uV, v_off_fin = 1002.963 uV 09:41:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -212.521 uVs 09:41:16 Integrated Voltage Measurement... 09:41:25 Voltage Offset Correction: v_off_ini = 996.205 uV, v_off_fin = 1006.132 uV 09:41:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.785 uVs 09:41:25 Integrated Voltage Measurement... 09:41:34 Voltage Offset Correction: v_off_ini = 1003.868 uV, v_off_fin = 990.722 uV 09:41:34 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -216.162 uVs 09:41:34 Integrated Voltage Measurement... 09:41:43 Voltage Offset Correction: v_off_ini = 985.290 uV, v_off_fin = 1001.736 uV 09:41:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 233.029 uVs 09:41:43 Integrated Voltage Measurement... 09:41:52 Voltage Offset Correction: v_off_ini = 998.188 uV, v_off_fin = 978.156 uV 09:41:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -212.361 uVs 09:41:52 Integrated Voltage Measurement... 09:42:01 Voltage Offset Correction: v_off_ini = 1001.721 uV, v_off_fin = 1008.676 uV 09:42:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 219.774 uVs 09:42:05 I1Y Average and RMS of Measurements VT = -216.105 +- 6.917 uVs I1Y = -43.219 +- 1.385 uTm 09:42:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:42:13 Integrated Voltage Measurement... 09:42:22 Voltage Offset Correction: v_off_ini = 1004.734 uV, v_off_fin = 1007.605 uV 09:42:22 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -75.040 uVs 09:42:22 Integrated Voltage Measurement... 09:42:31 Voltage Offset Correction: v_off_ini = 1016.987 uV, v_off_fin = 1032.466 uV 09:42:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.540 uVs 09:42:31 Integrated Voltage Measurement... 09:42:40 Voltage Offset Correction: v_off_ini = 1022.836 uV, v_off_fin = 1018.491 uV 09:42:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.592 uVs 09:42:40 Integrated Voltage Measurement... 09:42:49 Voltage Offset Correction: v_off_ini = 1022.485 uV, v_off_fin = 1029.991 uV 09:42:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.034 uVs 09:42:49 Integrated Voltage Measurement... 09:42:58 Voltage Offset Correction: v_off_ini = 1027.006 uV, v_off_fin = 1002.306 uV 09:42:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.881 uVs 09:42:58 Integrated Voltage Measurement... 09:43:07 Voltage Offset Correction: v_off_ini = 1005.940 uV, v_off_fin = 1016.551 uV 09:43:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.544 uVs 09:43:10 I2X Average and RMS of Measurements VT = -73.605 +- 8.582 uVs I2X = 6.772 +- 0.790 uTm^2 09:43:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:43:17 Integrated Voltage Measurement... 09:43:26 Voltage Offset Correction: v_off_ini = 1004.635 uV, v_off_fin = 1014.322 uV 09:43:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -344.974 uVs 09:43:26 Integrated Voltage Measurement... 09:43:35 Voltage Offset Correction: v_off_ini = 1009.616 uV, v_off_fin = 1001.701 uV 09:43:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.510 uVs 09:43:35 Integrated Voltage Measurement... 09:43:44 Voltage Offset Correction: v_off_ini = 1001.713 uV, v_off_fin = 996.265 uV 09:43:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.482 uVs 09:43:45 Integrated Voltage Measurement... 09:43:53 Voltage Offset Correction: v_off_ini = 1003.212 uV, v_off_fin = 1016.126 uV 09:43:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 325.502 uVs 09:43:54 Integrated Voltage Measurement... 09:44:03 Voltage Offset Correction: v_off_ini = 1010.054 uV, v_off_fin = 1012.455 uV 09:44:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -326.648 uVs 09:44:03 Integrated Voltage Measurement... 09:44:12 Voltage Offset Correction: v_off_ini = 1004.138 uV, v_off_fin = 1009.462 uV 09:44:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.762 uVs 09:44:14 I2Y Average and RMS of Measurements VT = -326.646 +- 4.311 uVs I2Y = -30.051 +- 0.397 uTm^2 09:44:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:44:30 Integrated Voltage Measurement... 09:44:39 Voltage Offset Correction: v_off_ini = 1020.735 uV, v_off_fin = 999.177 uV 09:44:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.814 uVs 09:44:39 Integrated Voltage Measurement... 09:44:48 Voltage Offset Correction: v_off_ini = 1019.155 uV, v_off_fin = 1016.120 uV 09:44:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 36.927 uVs 09:44:48 Integrated Voltage Measurement... 09:44:57 Voltage Offset Correction: v_off_ini = 1007.853 uV, v_off_fin = 996.265 uV 09:44:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.890 uVs 09:44:57 Integrated Voltage Measurement... 09:45:06 Voltage Offset Correction: v_off_ini = 1004.817 uV, v_off_fin = 996.476 uV 09:45:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.232 uVs 09:45:06 Integrated Voltage Measurement... 09:45:15 Voltage Offset Correction: v_off_ini = 998.029 uV, v_off_fin = 996.552 uV 09:45:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.142 uVs 09:45:15 Integrated Voltage Measurement... 09:45:24 Voltage Offset Correction: v_off_ini = 1004.738 uV, v_off_fin = 999.482 uV 09:45:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 30.635 uVs 09:45:29 I1X Average and RMS of Measurements VT = -44.940 +- 4.997 uVs I1X = 8.988 +- 0.999 uTm 09:45:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:45:35 Integrated Voltage Measurement... 09:45:44 Voltage Offset Correction: v_off_ini = 998.078 uV, v_off_fin = 975.200 uV 09:45:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -227.876 uVs 09:45:44 Integrated Voltage Measurement... 09:45:53 Voltage Offset Correction: v_off_ini = 1000.940 uV, v_off_fin = 1017.443 uV 09:45:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.002 uVs 09:45:53 Integrated Voltage Measurement... 09:46:02 Voltage Offset Correction: v_off_ini = 985.578 uV, v_off_fin = 1003.152 uV 09:46:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.248 uVs 09:46:02 Integrated Voltage Measurement... 09:46:11 Voltage Offset Correction: v_off_ini = 999.674 uV, v_off_fin = 1009.639 uV 09:46:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.864 uVs 09:46:11 Integrated Voltage Measurement... 09:46:20 Voltage Offset Correction: v_off_ini = 999.774 uV, v_off_fin = 979.106 uV 09:46:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -214.586 uVs 09:46:20 Integrated Voltage Measurement... 09:46:29 Voltage Offset Correction: v_off_ini = 1003.888 uV, v_off_fin = 1001.404 uV 09:46:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 238.762 uVs 09:46:34 I1Y Average and RMS of Measurements VT = -210.056 +- 14.885 uVs I1Y = -42.009 +- 2.977 uTm 09:46:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:46:41 Integrated Voltage Measurement... 09:46:50 Voltage Offset Correction: v_off_ini = 992.874 uV, v_off_fin = 990.031 uV 09:46:50 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -65.702 uVs 09:46:50 Integrated Voltage Measurement... 09:46:59 Voltage Offset Correction: v_off_ini = 1001.620 uV, v_off_fin = 1017.704 uV 09:46:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.039 uVs 09:46:59 Integrated Voltage Measurement... 09:47:08 Voltage Offset Correction: v_off_ini = 1002.033 uV, v_off_fin = 997.463 uV 09:47:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.776 uVs 09:47:08 Integrated Voltage Measurement... 09:47:17 Voltage Offset Correction: v_off_ini = 1006.766 uV, v_off_fin = 1012.164 uV 09:47:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.188 uVs 09:47:18 Integrated Voltage Measurement... 09:47:27 Voltage Offset Correction: v_off_ini = 999.567 uV, v_off_fin = 983.474 uV 09:47:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.164 uVs 09:47:27 Integrated Voltage Measurement... 09:47:36 Voltage Offset Correction: v_off_ini = 1003.744 uV, v_off_fin = 1009.906 uV 09:47:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.600 uVs 09:47:38 I2X Average and RMS of Measurements VT = -69.245 +- 1.434 uVs I2X = 6.370 +- 0.132 uTm^2 09:47:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:47:46 Integrated Voltage Measurement... 09:47:55 Voltage Offset Correction: v_off_ini = 999.672 uV, v_off_fin = 995.539 uV 09:47:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -314.970 uVs 09:47:55 Integrated Voltage Measurement... 09:48:04 Voltage Offset Correction: v_off_ini = 994.485 uV, v_off_fin = 995.050 uV 09:48:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 310.352 uVs 09:48:04 Integrated Voltage Measurement... 09:48:13 Voltage Offset Correction: v_off_ini = 995.513 uV, v_off_fin = 992.346 uV 09:48:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.839 uVs 09:48:13 Integrated Voltage Measurement... 09:48:22 Voltage Offset Correction: v_off_ini = 991.537 uV, v_off_fin = 991.901 uV 09:48:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 325.655 uVs 09:48:22 Integrated Voltage Measurement... 09:48:31 Voltage Offset Correction: v_off_ini = 998.086 uV, v_off_fin = 986.082 uV 09:48:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.774 uVs 09:48:31 Integrated Voltage Measurement... 09:48:40 Voltage Offset Correction: v_off_ini = 991.913 uV, v_off_fin = 987.882 uV 09:48:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 327.075 uVs 09:48:42 I2Y Average and RMS of Measurements VT = -321.278 +- 6.242 uVs I2Y = -29.558 +- 0.574 uTm^2 09:48:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:48:59 Integrated Voltage Measurement... 09:49:07 Voltage Offset Correction: v_off_ini = 990.480 uV, v_off_fin = 993.749 uV 09:49:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -40.180 uVs 09:49:08 Integrated Voltage Measurement... 09:49:17 Voltage Offset Correction: v_off_ini = 986.798 uV, v_off_fin = 997.669 uV 09:49:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.202 uVs 09:49:17 Integrated Voltage Measurement... 09:49:26 Voltage Offset Correction: v_off_ini = 986.867 uV, v_off_fin = 985.159 uV 09:49:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.749 uVs 09:49:26 Integrated Voltage Measurement... 09:49:35 Voltage Offset Correction: v_off_ini = 986.868 uV, v_off_fin = 994.800 uV 09:49:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.184 uVs 09:49:35 Integrated Voltage Measurement... 09:49:44 Voltage Offset Correction: v_off_ini = 986.796 uV, v_off_fin = 989.948 uV 09:49:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.570 uVs 09:49:44 Integrated Voltage Measurement... 09:49:53 Voltage Offset Correction: v_off_ini = 982.826 uV, v_off_fin = 993.148 uV 09:49:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.441 uVs 09:49:57 I1X Average and RMS of Measurements VT = -54.388 +- 2.968 uVs I1X = 10.878 +- 0.594 uTm 09:49:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:50:04 Integrated Voltage Measurement... 09:50:12 Voltage Offset Correction: v_off_ini = 988.266 uV, v_off_fin = 991.373 uV 09:50:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.671 uVs 09:50:13 Integrated Voltage Measurement... 09:50:21 Voltage Offset Correction: v_off_ini = 988.372 uV, v_off_fin = 1006.089 uV 09:50:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 213.222 uVs 09:50:22 Integrated Voltage Measurement... 09:50:30 Voltage Offset Correction: v_off_ini = 991.003 uV, v_off_fin = 976.055 uV 09:50:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -211.834 uVs 09:50:31 Integrated Voltage Measurement... 09:50:39 Voltage Offset Correction: v_off_ini = 992.805 uV, v_off_fin = 1006.900 uV 09:50:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.665 uVs 09:50:40 Integrated Voltage Measurement... 09:50:48 Voltage Offset Correction: v_off_ini = 994.958 uV, v_off_fin = 983.892 uV 09:50:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.576 uVs 09:50:49 Integrated Voltage Measurement... 09:50:57 Voltage Offset Correction: v_off_ini = 1004.686 uV, v_off_fin = 1003.441 uV 09:50:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 206.019 uVs 09:51:02 I1Y Average and RMS of Measurements VT = -207.165 +- 5.012 uVs I1Y = -41.430 +- 1.001 uTm 09:51:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:51:09 Integrated Voltage Measurement... 09:51:18 Voltage Offset Correction: v_off_ini = 1003.738 uV, v_off_fin = 996.617 uV 09:51:18 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -89.071 uVs 09:51:18 Integrated Voltage Measurement... 09:51:27 Voltage Offset Correction: v_off_ini = 1003.006 uV, v_off_fin = 1005.276 uV 09:51:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.460 uVs 09:51:28 Integrated Voltage Measurement... 09:51:37 Voltage Offset Correction: v_off_ini = 1006.934 uV, v_off_fin = 1001.900 uV 09:51:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.657 uVs 09:51:37 Integrated Voltage Measurement... 09:51:46 Voltage Offset Correction: v_off_ini = 1010.345 uV, v_off_fin = 1011.684 uV 09:51:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.271 uVs 09:51:46 Integrated Voltage Measurement... 09:51:55 Voltage Offset Correction: v_off_ini = 1001.878 uV, v_off_fin = 1002.914 uV 09:51:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.988 uVs 09:51:55 Integrated Voltage Measurement... 09:52:04 Voltage Offset Correction: v_off_ini = 1013.710 uV, v_off_fin = 1024.750 uV 09:52:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.982 uVs 09:52:06 I2X Average and RMS of Measurements VT = -77.905 +- 10.221 uVs I2X = 7.167 +- 0.940 uTm^2 09:52:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:52:14 Integrated Voltage Measurement... 09:52:23 Voltage Offset Correction: v_off_ini = 1015.689 uV, v_off_fin = 1014.671 uV 09:52:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -298.815 uVs 09:52:23 Integrated Voltage Measurement... 09:52:32 Voltage Offset Correction: v_off_ini = 1018.731 uV, v_off_fin = 1022.950 uV 09:52:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.077 uVs 09:52:32 Integrated Voltage Measurement... 09:52:41 Voltage Offset Correction: v_off_ini = 1027.373 uV, v_off_fin = 1026.404 uV 09:52:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.860 uVs 09:52:41 Integrated Voltage Measurement... 09:52:50 Voltage Offset Correction: v_off_ini = 1024.244 uV, v_off_fin = 1022.178 uV 09:52:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 302.843 uVs 09:52:50 Integrated Voltage Measurement... 09:52:59 Voltage Offset Correction: v_off_ini = 1023.351 uV, v_off_fin = 1032.054 uV 09:52:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -286.498 uVs 09:52:59 Integrated Voltage Measurement... 09:53:08 Voltage Offset Correction: v_off_ini = 1037.716 uV, v_off_fin = 1038.778 uV 09:53:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 306.993 uVs 09:53:10 I2Y Average and RMS of Measurements VT = -302.348 +- 4.747 uVs I2Y = -27.816 +- 0.437 uTm^2 09:53:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:53:29 Integrated Voltage Measurement... 09:53:38 Voltage Offset Correction: v_off_ini = 1027.578 uV, v_off_fin = 1013.137 uV 09:53:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -57.068 uVs 09:53:38 Integrated Voltage Measurement... 09:53:47 Voltage Offset Correction: v_off_ini = 1019.310 uV, v_off_fin = 1024.542 uV 09:53:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.526 uVs 09:53:47 Integrated Voltage Measurement... 09:53:56 Voltage Offset Correction: v_off_ini = 1013.752 uV, v_off_fin = 1011.672 uV 09:53:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.669 uVs 09:53:56 Integrated Voltage Measurement... 09:54:05 Voltage Offset Correction: v_off_ini = 1007.060 uV, v_off_fin = 1016.305 uV 09:54:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.810 uVs 09:54:05 Integrated Voltage Measurement... 09:54:14 Voltage Offset Correction: v_off_ini = 1025.248 uV, v_off_fin = 1015.107 uV 09:54:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -37.422 uVs 09:54:14 Integrated Voltage Measurement... 09:54:23 Voltage Offset Correction: v_off_ini = 1023.429 uV, v_off_fin = 1033.979 uV 09:54:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.746 uVs 09:54:27 I1X Average and RMS of Measurements VT = -50.874 +- 5.896 uVs I1X = 10.175 +- 1.179 uTm 09:54:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:54:34 Integrated Voltage Measurement... 09:54:43 Voltage Offset Correction: v_off_ini = 1039.843 uV, v_off_fin = 1034.227 uV 09:54:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -188.967 uVs 09:54:43 Integrated Voltage Measurement... 09:54:52 Voltage Offset Correction: v_off_ini = 1035.845 uV, v_off_fin = 1044.877 uV 09:54:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.964 uVs 09:54:52 Integrated Voltage Measurement... 09:55:01 Voltage Offset Correction: v_off_ini = 1034.635 uV, v_off_fin = 1021.493 uV 09:55:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.512 uVs 09:55:01 Integrated Voltage Measurement... 09:55:10 Voltage Offset Correction: v_off_ini = 1035.952 uV, v_off_fin = 1047.020 uV 09:55:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 218.839 uVs 09:55:10 Integrated Voltage Measurement... 09:55:19 Voltage Offset Correction: v_off_ini = 1032.132 uV, v_off_fin = 1014.368 uV 09:55:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.836 uVs 09:55:19 Integrated Voltage Measurement... 09:55:28 Voltage Offset Correction: v_off_ini = 1038.092 uV, v_off_fin = 1038.360 uV 09:55:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.170 uVs 09:55:32 I1Y Average and RMS of Measurements VT = -203.215 +- 6.727 uVs I1Y = -40.644 +- 1.345 uTm 09:55:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:55:40 Integrated Voltage Measurement... 09:55:49 Voltage Offset Correction: v_off_ini = 1033.797 uV, v_off_fin = 1034.593 uV 09:55:49 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -72.456 uVs 09:55:49 Integrated Voltage Measurement... 09:55:58 Voltage Offset Correction: v_off_ini = 1047.547 uV, v_off_fin = 1048.635 uV 09:55:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.447 uVs 09:55:58 Integrated Voltage Measurement... 09:56:07 Voltage Offset Correction: v_off_ini = 1047.703 uV, v_off_fin = 1041.941 uV 09:56:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.144 uVs 09:56:07 Integrated Voltage Measurement... 09:56:16 Voltage Offset Correction: v_off_ini = 1050.936 uV, v_off_fin = 1053.895 uV 09:56:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.900 uVs 09:56:16 Integrated Voltage Measurement... 09:56:25 Voltage Offset Correction: v_off_ini = 1054.141 uV, v_off_fin = 1036.409 uV 09:56:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.135 uVs 09:56:25 Integrated Voltage Measurement... 09:56:34 Voltage Offset Correction: v_off_ini = 1035.702 uV, v_off_fin = 1038.521 uV 09:56:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.610 uVs 09:56:37 I2X Average and RMS of Measurements VT = -79.449 +- 6.758 uVs I2X = 7.309 +- 0.622 uTm^2 09:56:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:56:44 Integrated Voltage Measurement... 09:56:53 Voltage Offset Correction: v_off_ini = 1044.424 uV, v_off_fin = 1058.360 uV 09:56:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -275.114 uVs 09:56:53 Integrated Voltage Measurement... 09:57:02 Voltage Offset Correction: v_off_ini = 1055.656 uV, v_off_fin = 1038.941 uV 09:57:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 285.745 uVs 09:57:02 Integrated Voltage Measurement... 09:57:11 Voltage Offset Correction: v_off_ini = 1037.342 uV, v_off_fin = 1039.394 uV 09:57:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -292.514 uVs 09:57:12 Integrated Voltage Measurement... 09:57:20 Voltage Offset Correction: v_off_ini = 1044.826 uV, v_off_fin = 1047.374 uV 09:57:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 288.204 uVs 09:57:21 Integrated Voltage Measurement... 09:57:30 Voltage Offset Correction: v_off_ini = 1048.577 uV, v_off_fin = 1048.353 uV 09:57:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -287.788 uVs 09:57:30 Integrated Voltage Measurement... 09:57:39 Voltage Offset Correction: v_off_ini = 1051.529 uV, v_off_fin = 1052.646 uV 09:57:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 297.007 uVs 09:57:41 I2Y Average and RMS of Measurements VT = -287.728 +- 5.228 uVs I2Y = -26.471 +- 0.481 uTm^2 09:57:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:57:59 Integrated Voltage Measurement... 09:58:08 Voltage Offset Correction: v_off_ini = 1045.176 uV, v_off_fin = 1048.106 uV 09:58:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.399 uVs 09:58:08 Integrated Voltage Measurement... 09:58:17 Voltage Offset Correction: v_off_ini = 1045.834 uV, v_off_fin = 1035.701 uV 09:58:17 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.613 uVs 09:58:17 Integrated Voltage Measurement... 09:58:26 Voltage Offset Correction: v_off_ini = 1044.903 uV, v_off_fin = 1045.093 uV 09:58:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.068 uVs 09:58:26 Integrated Voltage Measurement... 09:58:35 Voltage Offset Correction: v_off_ini = 1061.788 uV, v_off_fin = 1050.799 uV 09:58:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.928 uVs 09:58:35 Integrated Voltage Measurement... 09:58:44 Voltage Offset Correction: v_off_ini = 1042.847 uV, v_off_fin = 1033.374 uV 09:58:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.493 uVs 09:58:44 Integrated Voltage Measurement... 09:58:53 Voltage Offset Correction: v_off_ini = 1049.135 uV, v_off_fin = 1060.302 uV 09:58:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.807 uVs 09:58:58 I1X Average and RMS of Measurements VT = -66.218 +- 9.408 uVs I1X = 13.244 +- 1.882 uTm 09:59:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:59:05 Integrated Voltage Measurement... 09:59:13 Voltage Offset Correction: v_off_ini = 1058.553 uV, v_off_fin = 1043.418 uV 09:59:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -197.490 uVs 09:59:14 Integrated Voltage Measurement... 09:59:22 Voltage Offset Correction: v_off_ini = 1049.611 uV, v_off_fin = 1068.497 uV 09:59:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.907 uVs 09:59:23 Integrated Voltage Measurement... 09:59:31 Voltage Offset Correction: v_off_ini = 1064.250 uV, v_off_fin = 1053.696 uV 09:59:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.468 uVs 09:59:32 Integrated Voltage Measurement... 09:59:40 Voltage Offset Correction: v_off_ini = 1063.630 uV, v_off_fin = 1064.035 uV 09:59:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.919 uVs 09:59:41 Integrated Voltage Measurement... 09:59:50 Voltage Offset Correction: v_off_ini = 1072.167 uV, v_off_fin = 1060.125 uV 09:59:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.237 uVs 09:59:50 Integrated Voltage Measurement... 09:59:59 Voltage Offset Correction: v_off_ini = 1061.445 uV, v_off_fin = 1087.822 uV 09:59:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 206.146 uVs 10:00:03 I1Y Average and RMS of Measurements VT = -203.028 +- 2.319 uVs I1Y = -40.606 +- 0.464 uTm 10:00:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:00:10 Integrated Voltage Measurement... 10:00:19 Voltage Offset Correction: v_off_ini = 1047.677 uV, v_off_fin = 1048.179 uV 10:00:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.026 uVs 10:00:20 Integrated Voltage Measurement... 10:00:29 Voltage Offset Correction: v_off_ini = 1054.135 uV, v_off_fin = 1073.970 uV 10:00:29 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 105.822 uVs 10:00:29 Integrated Voltage Measurement... 10:00:38 Voltage Offset Correction: v_off_ini = 1066.219 uV, v_off_fin = 1056.123 uV 10:00:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.762 uVs 10:00:38 Integrated Voltage Measurement... 10:00:47 Voltage Offset Correction: v_off_ini = 1065.814 uV, v_off_fin = 1068.947 uV 10:00:47 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 78.478 uVs 10:00:47 Integrated Voltage Measurement... 10:00:56 Voltage Offset Correction: v_off_ini = 1057.839 uV, v_off_fin = 1050.644 uV 10:00:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.129 uVs 10:00:56 Integrated Voltage Measurement... 10:01:05 Voltage Offset Correction: v_off_ini = 1074.709 uV, v_off_fin = 1078.920 uV 10:01:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.317 uVs 10:01:07 I2X Average and RMS of Measurements VT = -84.756 +- 5.794 uVs I2X = 7.798 +- 0.533 uTm^2 10:01:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:01:15 Integrated Voltage Measurement... 10:01:24 Voltage Offset Correction: v_off_ini = 1078.210 uV, v_off_fin = 1084.154 uV 10:01:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -283.506 uVs 10:01:24 Integrated Voltage Measurement... 10:01:33 Voltage Offset Correction: v_off_ini = 1091.234 uV, v_off_fin = 1095.604 uV 10:01:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.114 uVs 10:01:33 Integrated Voltage Measurement... 10:01:42 Voltage Offset Correction: v_off_ini = 1089.665 uV, v_off_fin = 1094.420 uV 10:01:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -259.845 uVs 10:01:42 Integrated Voltage Measurement... 10:01:51 Voltage Offset Correction: v_off_ini = 1110.067 uV, v_off_fin = 1104.471 uV 10:01:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.010 uVs 10:01:51 Integrated Voltage Measurement... 10:02:00 Voltage Offset Correction: v_off_ini = 1089.137 uV, v_off_fin = 1089.454 uV 10:02:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.156 uVs 10:02:00 Integrated Voltage Measurement... 10:02:09 Voltage Offset Correction: v_off_ini = 1089.373 uV, v_off_fin = 1085.778 uV 10:02:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 306.105 uVs 10:02:12 I2Y Average and RMS of Measurements VT = -271.623 +- 6.822 uVs I2Y = -24.989 +- 0.628 uTm^2 10:02:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:02:30 Integrated Voltage Measurement... 10:02:39 Voltage Offset Correction: v_off_ini = 1086.212 uV, v_off_fin = 1085.424 uV 10:02:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -74.162 uVs 10:02:39 Integrated Voltage Measurement... 10:02:48 Voltage Offset Correction: v_off_ini = 1084.065 uV, v_off_fin = 1084.115 uV 10:02:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.412 uVs 10:02:48 Integrated Voltage Measurement... 10:02:57 Voltage Offset Correction: v_off_ini = 1074.370 uV, v_off_fin = 1070.420 uV 10:02:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.093 uVs 10:02:57 Integrated Voltage Measurement... 10:03:06 Voltage Offset Correction: v_off_ini = 1081.991 uV, v_off_fin = 1090.576 uV 10:03:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.133 uVs 10:03:06 Integrated Voltage Measurement... 10:03:15 Voltage Offset Correction: v_off_ini = 1080.219 uV, v_off_fin = 1072.497 uV 10:03:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.940 uVs 10:03:15 Integrated Voltage Measurement... 10:03:24 Voltage Offset Correction: v_off_ini = 1074.825 uV, v_off_fin = 1080.762 uV 10:03:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.451 uVs 10:03:28 I1X Average and RMS of Measurements VT = -81.198 +- 2.927 uVs I1X = 16.240 +- 0.585 uTm 10:03:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:03:35 Integrated Voltage Measurement... 10:03:44 Voltage Offset Correction: v_off_ini = 1079.108 uV, v_off_fin = 1066.643 uV 10:03:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.879 uVs 10:03:44 Integrated Voltage Measurement... 10:03:53 Voltage Offset Correction: v_off_ini = 1074.817 uV, v_off_fin = 1096.380 uV 10:03:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.594 uVs 10:03:53 Integrated Voltage Measurement... 10:04:02 Voltage Offset Correction: v_off_ini = 1080.144 uV, v_off_fin = 1064.502 uV 10:04:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.970 uVs 10:04:02 Integrated Voltage Measurement... 10:04:11 Voltage Offset Correction: v_off_ini = 1080.191 uV, v_off_fin = 1087.232 uV 10:04:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.019 uVs 10:04:11 Integrated Voltage Measurement... 10:04:20 Voltage Offset Correction: v_off_ini = 1068.943 uV, v_off_fin = 1076.642 uV 10:04:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.868 uVs 10:04:20 Integrated Voltage Measurement... 10:04:29 Voltage Offset Correction: v_off_ini = 1081.125 uV, v_off_fin = 1096.400 uV 10:04:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 198.582 uVs 10:04:33 I1Y Average and RMS of Measurements VT = -197.819 +- 5.975 uVs I1Y = -39.563 +- 1.195 uTm 10:04:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:04:41 Integrated Voltage Measurement... 10:04:50 Voltage Offset Correction: v_off_ini = 1079.560 uV, v_off_fin = 1055.994 uV 10:04:50 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -85.861 uVs 10:04:50 Integrated Voltage Measurement... 10:04:59 Voltage Offset Correction: v_off_ini = 1058.699 uV, v_off_fin = 1079.815 uV 10:04:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 88.989 uVs 10:04:59 Integrated Voltage Measurement... 10:05:08 Voltage Offset Correction: v_off_ini = 1067.558 uV, v_off_fin = 1060.370 uV 10:05:08 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -98.517 uVs 10:05:08 Integrated Voltage Measurement... 10:05:17 Voltage Offset Correction: v_off_ini = 1069.802 uV, v_off_fin = 1076.404 uV 10:05:17 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 74.674 uVs 10:05:17 Integrated Voltage Measurement... 10:05:26 Voltage Offset Correction: v_off_ini = 1060.628 uV, v_off_fin = 1069.301 uV 10:05:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.276 uVs 10:05:26 Integrated Voltage Measurement... 10:05:35 Voltage Offset Correction: v_off_ini = 1074.373 uV, v_off_fin = 1081.367 uV 10:05:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.435 uVs 10:05:38 I2X Average and RMS of Measurements VT = -90.292 +- 4.653 uVs I2X = 8.307 +- 0.428 uTm^2 10:05:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:05:45 Integrated Voltage Measurement... 10:05:54 Voltage Offset Correction: v_off_ini = 1070.363 uV, v_off_fin = 1066.314 uV 10:05:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.482 uVs 10:05:55 Integrated Voltage Measurement... 10:06:04 Voltage Offset Correction: v_off_ini = 1070.974 uV, v_off_fin = 1075.310 uV 10:06:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.250 uVs 10:06:04 Integrated Voltage Measurement... 10:06:13 Voltage Offset Correction: v_off_ini = 1080.888 uV, v_off_fin = 1073.998 uV 10:06:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.398 uVs 10:06:13 Integrated Voltage Measurement... 10:06:22 Voltage Offset Correction: v_off_ini = 1076.163 uV, v_off_fin = 1086.871 uV 10:06:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.184 uVs 10:06:22 Integrated Voltage Measurement... 10:06:31 Voltage Offset Correction: v_off_ini = 1085.503 uV, v_off_fin = 1085.360 uV 10:06:31 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -266.576 uVs 10:06:31 Integrated Voltage Measurement... 10:06:40 Voltage Offset Correction: v_off_ini = 1080.884 uV, v_off_fin = 1098.148 uV 10:06:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.947 uVs 10:06:42 I2Y Average and RMS of Measurements VT = -253.639 +- 7.263 uVs I2Y = -23.335 +- 0.669 uTm^2 10:06:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:07:01 Integrated Voltage Measurement... 10:07:10 Voltage Offset Correction: v_off_ini = 1091.464 uV, v_off_fin = 1087.996 uV 10:07:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.003 uVs 10:07:10 Integrated Voltage Measurement... 10:07:19 Voltage Offset Correction: v_off_ini = 1082.622 uV, v_off_fin = 1088.597 uV 10:07:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.282 uVs 10:07:19 Integrated Voltage Measurement... 10:07:28 Voltage Offset Correction: v_off_ini = 1085.756 uV, v_off_fin = 1073.729 uV 10:07:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.476 uVs 10:07:28 Integrated Voltage Measurement... 10:07:37 Voltage Offset Correction: v_off_ini = 1073.031 uV, v_off_fin = 1078.101 uV 10:07:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.366 uVs 10:07:37 Integrated Voltage Measurement... 10:07:46 Voltage Offset Correction: v_off_ini = 1068.534 uV, v_off_fin = 1081.552 uV 10:07:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.797 uVs 10:07:46 Integrated Voltage Measurement... 10:07:55 Voltage Offset Correction: v_off_ini = 1087.090 uV, v_off_fin = 1073.374 uV 10:07:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.731 uVs 10:07:59 I1X Average and RMS of Measurements VT = -91.776 +- 6.727 uVs I1X = 18.355 +- 1.345 uTm 10:08:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:08:06 Integrated Voltage Measurement... 10:08:15 Voltage Offset Correction: v_off_ini = 1069.261 uV, v_off_fin = 1070.562 uV 10:08:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.635 uVs 10:08:15 Integrated Voltage Measurement... 10:08:24 Voltage Offset Correction: v_off_ini = 1080.865 uV, v_off_fin = 1097.070 uV 10:08:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.818 uVs 10:08:24 Integrated Voltage Measurement... 10:08:33 Voltage Offset Correction: v_off_ini = 1080.239 uV, v_off_fin = 1068.383 uV 10:08:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.396 uVs 10:08:33 Integrated Voltage Measurement... 10:08:42 Voltage Offset Correction: v_off_ini = 1091.578 uV, v_off_fin = 1103.575 uV 10:08:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.828 uVs 10:08:42 Integrated Voltage Measurement... 10:08:51 Voltage Offset Correction: v_off_ini = 1080.573 uV, v_off_fin = 1064.831 uV 10:08:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.953 uVs 10:08:51 Integrated Voltage Measurement... 10:09:00 Voltage Offset Correction: v_off_ini = 1069.336 uV, v_off_fin = 1081.060 uV 10:09:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.224 uVs 10:09:04 I1Y Average and RMS of Measurements VT = -188.976 +- 4.679 uVs I1Y = -37.795 +- 0.936 uTm 10:09:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:09:12 Integrated Voltage Measurement... 10:09:21 Voltage Offset Correction: v_off_ini = 1069.170 uV, v_off_fin = 1076.164 uV 10:09:21 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -76.633 uVs 10:09:21 Integrated Voltage Measurement... 10:09:30 Voltage Offset Correction: v_off_ini = 1085.094 uV, v_off_fin = 1096.033 uV 10:09:30 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 96.649 uVs 10:09:30 Integrated Voltage Measurement... 10:09:39 Voltage Offset Correction: v_off_ini = 1100.778 uV, v_off_fin = 1082.341 uV 10:09:39 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -101.733 uVs 10:09:39 Integrated Voltage Measurement... 10:09:48 Voltage Offset Correction: v_off_ini = 1086.549 uV, v_off_fin = 1091.448 uV 10:09:48 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 98.797 uVs 10:09:48 Integrated Voltage Measurement... 10:09:57 Voltage Offset Correction: v_off_ini = 1081.664 uV, v_off_fin = 1066.427 uV 10:09:57 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -93.868 uVs 10:09:57 Integrated Voltage Measurement... 10:10:06 Voltage Offset Correction: v_off_ini = 1087.744 uV, v_off_fin = 1093.827 uV 10:10:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.288 uVs 10:10:08 I2X Average and RMS of Measurements VT = -92.995 +- 5.600 uVs I2X = 8.556 +- 0.515 uTm^2 10:10:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:10:16 Integrated Voltage Measurement... 10:10:25 Voltage Offset Correction: v_off_ini = 1082.018 uV, v_off_fin = 1071.907 uV 10:10:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -237.310 uVs 10:10:25 Integrated Voltage Measurement... 10:10:34 Voltage Offset Correction: v_off_ini = 1076.646 uV, v_off_fin = 1080.902 uV 10:10:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.117 uVs 10:10:34 Integrated Voltage Measurement... 10:10:43 Voltage Offset Correction: v_off_ini = 1069.542 uV, v_off_fin = 1063.946 uV 10:10:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.224 uVs 10:10:43 Integrated Voltage Measurement... 10:10:52 Voltage Offset Correction: v_off_ini = 1083.503 uV, v_off_fin = 1065.976 uV 10:10:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.219 uVs 10:10:52 Integrated Voltage Measurement... 10:11:01 Voltage Offset Correction: v_off_ini = 1075.790 uV, v_off_fin = 1082.300 uV 10:11:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.802 uVs 10:11:01 Integrated Voltage Measurement... 10:11:10 Voltage Offset Correction: v_off_ini = 1093.093 uV, v_off_fin = 1101.645 uV 10:11:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.830 uVs 10:11:13 I2Y Average and RMS of Measurements VT = -251.584 +- 3.513 uVs I2Y = -23.146 +- 0.323 uTm^2 10:11:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:11:31 Integrated Voltage Measurement... 10:11:40 Voltage Offset Correction: v_off_ini = 1084.073 uV, v_off_fin = 1097.577 uV 10:11:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.860 uVs 10:11:40 Integrated Voltage Measurement... 10:11:49 Voltage Offset Correction: v_off_ini = 1109.295 uV, v_off_fin = 1115.047 uV 10:11:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.295 uVs 10:11:49 Integrated Voltage Measurement... 10:11:58 Voltage Offset Correction: v_off_ini = 1115.008 uV, v_off_fin = 1121.550 uV 10:11:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.836 uVs 10:11:58 Integrated Voltage Measurement... 10:12:07 Voltage Offset Correction: v_off_ini = 1105.426 uV, v_off_fin = 1119.963 uV 10:12:07 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 84.603 uVs 10:12:07 Integrated Voltage Measurement... 10:12:16 Voltage Offset Correction: v_off_ini = 1115.576 uV, v_off_fin = 1108.296 uV 10:12:16 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.882 uVs 10:12:16 Integrated Voltage Measurement... 10:12:25 Voltage Offset Correction: v_off_ini = 1118.888 uV, v_off_fin = 1117.770 uV 10:12:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 87.233 uVs 10:12:29 I1X Average and RMS of Measurements VT = -92.285 +- 1.902 uVs I1X = 18.458 +- 0.381 uTm 10:12:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:12:36 Integrated Voltage Measurement... 10:12:45 Voltage Offset Correction: v_off_ini = 1107.061 uV, v_off_fin = 1102.348 uV 10:12:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.100 uVs 10:12:45 Integrated Voltage Measurement... 10:12:54 Voltage Offset Correction: v_off_ini = 1111.079 uV, v_off_fin = 1119.417 uV 10:12:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 207.913 uVs 10:12:54 Integrated Voltage Measurement... 10:13:03 Voltage Offset Correction: v_off_ini = 1087.085 uV, v_off_fin = 1085.214 uV 10:13:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.355 uVs 10:13:03 Integrated Voltage Measurement... 10:13:12 Voltage Offset Correction: v_off_ini = 1100.900 uV, v_off_fin = 1113.797 uV 10:13:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.073 uVs 10:13:12 Integrated Voltage Measurement... 10:13:21 Voltage Offset Correction: v_off_ini = 1103.958 uV, v_off_fin = 1086.079 uV 10:13:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.341 uVs 10:13:21 Integrated Voltage Measurement... 10:13:30 Voltage Offset Correction: v_off_ini = 1089.034 uV, v_off_fin = 1114.099 uV 10:13:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.818 uVs 10:13:35 I1Y Average and RMS of Measurements VT = -193.266 +- 5.475 uVs I1Y = -38.652 +- 1.093 uTm 10:13:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:13:42 Integrated Voltage Measurement... 10:13:51 Voltage Offset Correction: v_off_ini = 1096.894 uV, v_off_fin = 1102.553 uV 10:13:51 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -103.505 uVs 10:13:51 Integrated Voltage Measurement... 10:14:00 Voltage Offset Correction: v_off_ini = 1110.653 uV, v_off_fin = 1103.858 uV 10:14:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.306 uVs 10:14:00 Integrated Voltage Measurement... 10:14:09 Voltage Offset Correction: v_off_ini = 1099.800 uV, v_off_fin = 1095.023 uV 10:14:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.112 uVs 10:14:09 Integrated Voltage Measurement... 10:14:18 Voltage Offset Correction: v_off_ini = 1110.576 uV, v_off_fin = 1119.780 uV 10:14:18 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 98.657 uVs 10:14:18 Integrated Voltage Measurement... 10:14:27 Voltage Offset Correction: v_off_ini = 1112.484 uV, v_off_fin = 1091.723 uV 10:14:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.079 uVs 10:14:27 Integrated Voltage Measurement... 10:14:36 Voltage Offset Correction: v_off_ini = 1102.201 uV, v_off_fin = 1113.220 uV 10:14:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.810 uVs 10:14:39 I2X Average and RMS of Measurements VT = -96.911 +- 3.937 uVs I2X = 8.916 +- 0.362 uTm^2 10:14:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:14:47 Integrated Voltage Measurement... 10:14:56 Voltage Offset Correction: v_off_ini = 1109.363 uV, v_off_fin = 1101.204 uV 10:14:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.575 uVs 10:14:56 Integrated Voltage Measurement... 10:15:05 Voltage Offset Correction: v_off_ini = 1101.174 uV, v_off_fin = 1104.069 uV 10:15:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.912 uVs 10:15:05 Integrated Voltage Measurement... 10:15:14 Voltage Offset Correction: v_off_ini = 1090.565 uV, v_off_fin = 1077.956 uV 10:15:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.374 uVs 10:15:14 Integrated Voltage Measurement... 10:15:23 Voltage Offset Correction: v_off_ini = 1090.181 uV, v_off_fin = 1078.537 uV 10:15:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.373 uVs 10:15:23 Integrated Voltage Measurement... 10:15:32 Voltage Offset Correction: v_off_ini = 1070.698 uV, v_off_fin = 1077.017 uV 10:15:32 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -228.576 uVs 10:15:32 Integrated Voltage Measurement... 10:15:41 Voltage Offset Correction: v_off_ini = 1071.185 uV, v_off_fin = 1066.078 uV 10:15:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 239.404 uVs 10:15:43 I2Y Average and RMS of Measurements VT = -241.036 +- 7.170 uVs I2Y = -22.176 +- 0.659 uTm^2 10:15:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:16:02 Integrated Voltage Measurement... 10:16:11 Voltage Offset Correction: v_off_ini = 1071.398 uV, v_off_fin = 1055.684 uV 10:16:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -109.017 uVs 10:16:11 Integrated Voltage Measurement... 10:16:20 Voltage Offset Correction: v_off_ini = 1056.014 uV, v_off_fin = 1056.981 uV 10:16:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.993 uVs 10:16:20 Integrated Voltage Measurement... 10:16:29 Voltage Offset Correction: v_off_ini = 1053.165 uV, v_off_fin = 1054.285 uV 10:16:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.901 uVs 10:16:29 Integrated Voltage Measurement... 10:16:38 Voltage Offset Correction: v_off_ini = 1057.224 uV, v_off_fin = 1045.650 uV 10:16:38 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.372 uVs 10:16:38 Integrated Voltage Measurement... 10:16:47 Voltage Offset Correction: v_off_ini = 1041.635 uV, v_off_fin = 1031.248 uV 10:16:47 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.738 uVs 10:16:47 Integrated Voltage Measurement... 10:16:56 Voltage Offset Correction: v_off_ini = 1019.823 uV, v_off_fin = 1044.863 uV 10:16:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.978 uVs 10:17:00 I1X Average and RMS of Measurements VT = -95.833 +- 9.358 uVs I1X = 19.168 +- 1.872 uTm 10:17:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:17:07 Integrated Voltage Measurement... 10:17:16 Voltage Offset Correction: v_off_ini = 1036.068 uV, v_off_fin = 1027.040 uV 10:17:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.302 uVs 10:17:16 Integrated Voltage Measurement... 10:17:25 Voltage Offset Correction: v_off_ini = 1041.034 uV, v_off_fin = 1040.951 uV 10:17:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.316 uVs 10:17:25 Integrated Voltage Measurement... 10:17:34 Voltage Offset Correction: v_off_ini = 1040.902 uV, v_off_fin = 1021.726 uV 10:17:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.610 uVs 10:17:34 Integrated Voltage Measurement... 10:17:43 Voltage Offset Correction: v_off_ini = 1042.125 uV, v_off_fin = 1055.298 uV 10:17:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 194.310 uVs 10:17:43 Integrated Voltage Measurement... 10:17:52 Voltage Offset Correction: v_off_ini = 1038.791 uV, v_off_fin = 1025.953 uV 10:17:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -186.413 uVs 10:17:52 Integrated Voltage Measurement... 10:18:01 Voltage Offset Correction: v_off_ini = 1038.662 uV, v_off_fin = 1036.001 uV 10:18:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.539 uVs 10:18:05 I1Y Average and RMS of Measurements VT = -189.915 +- 2.518 uVs I1Y = -37.980 +- 0.502 uTm 10:18:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:18:13 Integrated Voltage Measurement... 10:18:22 Voltage Offset Correction: v_off_ini = 1032.430 uV, v_off_fin = 1015.056 uV 10:18:22 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -123.005 uVs 10:18:22 Integrated Voltage Measurement... 10:18:31 Voltage Offset Correction: v_off_ini = 1036.115 uV, v_off_fin = 1026.332 uV 10:18:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.238 uVs 10:18:31 Integrated Voltage Measurement... 10:18:40 Voltage Offset Correction: v_off_ini = 1019.664 uV, v_off_fin = 1008.063 uV 10:18:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.705 uVs 10:18:40 Integrated Voltage Measurement... 10:18:49 Voltage Offset Correction: v_off_ini = 1004.746 uV, v_off_fin = 1016.892 uV 10:18:49 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = 110.834 uVs 10:18:49 Integrated Voltage Measurement... 10:18:58 Voltage Offset Correction: v_off_ini = 1023.779 uV, v_off_fin = 1008.099 uV 10:18:58 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -76.401 uVs 10:18:58 Integrated Voltage Measurement... 10:19:07 Voltage Offset Correction: v_off_ini = 1025.884 uV, v_off_fin = 1028.283 uV 10:19:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.659 uVs 10:19:09 I2X Average and RMS of Measurements VT = -100.474 +- 8.806 uVs I2X = 9.244 +- 0.810 uTm^2 10:19:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:19:17 Integrated Voltage Measurement... 10:19:26 Voltage Offset Correction: v_off_ini = 1010.388 uV, v_off_fin = 1028.013 uV 10:19:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -206.788 uVs 10:19:26 Integrated Voltage Measurement... 10:19:35 Voltage Offset Correction: v_off_ini = 1012.306 uV, v_off_fin = 1025.540 uV 10:19:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.759 uVs 10:19:35 Integrated Voltage Measurement... 10:19:44 Voltage Offset Correction: v_off_ini = 1031.743 uV, v_off_fin = 995.764 uV 10:19:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.426 uVs 10:19:44 Integrated Voltage Measurement... 10:19:53 Voltage Offset Correction: v_off_ini = 1015.213 uV, v_off_fin = 1001.709 uV 10:19:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.469 uVs 10:19:53 Integrated Voltage Measurement... 10:20:02 Voltage Offset Correction: v_off_ini = 997.342 uV, v_off_fin = 1002.077 uV 10:20:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -215.209 uVs 10:20:02 Integrated Voltage Measurement... 10:20:11 Voltage Offset Correction: v_off_ini = 992.257 uV, v_off_fin = 1016.137 uV 10:20:11 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = 252.532 uVs 10:20:14 I2Y Average and RMS of Measurements VT = -222.697 +- 8.448 uVs I2Y = -20.488 +- 0.777 uTm^2 10:20:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm