SLAC Magnetic Measurements Date: 07-23-2026 Time: 13:34:35 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16336 Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 2 Measurement Device: Stretched Wire Operator: YL Comment: 13:34:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:34:45 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 13:34:47 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 13:34:48 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 13:34:50 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 13:35:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:35:08 Integrated Voltage Measurement... 13:35:17 Voltage Offset Correction: v_off_ini = 2309.227 uV, v_off_fin = 2311.267 uV 13:35:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.602 uVs 13:35:17 Integrated Voltage Measurement... 13:35:26 Voltage Offset Correction: v_off_ini = 2312.946 uV, v_off_fin = 2325.435 uV 13:35:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.709 uVs 13:35:26 Integrated Voltage Measurement... 13:35:35 Voltage Offset Correction: v_off_ini = 2312.745 uV, v_off_fin = 2329.699 uV 13:35:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -43.492 uVs 13:35:35 Integrated Voltage Measurement... 13:35:44 Voltage Offset Correction: v_off_ini = 2353.674 uV, v_off_fin = 2327.751 uV 13:35:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 31.409 uVs 13:35:44 Integrated Voltage Measurement... 13:35:53 Voltage Offset Correction: v_off_ini = 2331.559 uV, v_off_fin = 2335.870 uV 13:35:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.570 uVs 13:35:53 Integrated Voltage Measurement... 13:36:02 Voltage Offset Correction: v_off_ini = 2368.552 uV, v_off_fin = 2393.980 uV 13:36:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.576 uVs 13:36:06 I1X Average and RMS of Measurements VT = -45.893 +- 9.180 uVs I1X = 9.179 +- 1.836 uTm 13:36:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:36:13 Integrated Voltage Measurement... 13:36:22 Voltage Offset Correction: v_off_ini = 2355.834 uV, v_off_fin = 2382.384 uV 13:36:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -208.930 uVs 13:36:22 Integrated Voltage Measurement... 13:36:31 Voltage Offset Correction: v_off_ini = 2394.612 uV, v_off_fin = 2400.788 uV 13:36:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 259.566 uVs 13:36:31 Integrated Voltage Measurement... 13:36:40 Voltage Offset Correction: v_off_ini = 2383.677 uV, v_off_fin = 2421.166 uV 13:36:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -197.313 uVs 13:36:40 Integrated Voltage Measurement... 13:36:49 Voltage Offset Correction: v_off_ini = 2417.337 uV, v_off_fin = 2443.685 uV 13:36:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 253.373 uVs 13:36:49 Integrated Voltage Measurement... 13:36:58 Voltage Offset Correction: v_off_ini = 2416.864 uV, v_off_fin = 2416.799 uV 13:36:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -238.487 uVs 13:36:58 Integrated Voltage Measurement... 13:37:07 Voltage Offset Correction: v_off_ini = 2402.518 uV, v_off_fin = 2424.401 uV 13:37:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.976 uVs 13:37:11 I1Y Average and RMS of Measurements VT = -225.941 +- 6.552 uVs I1Y = -45.186 +- 1.309 uTm 13:37:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:37:19 Integrated Voltage Measurement... 13:37:28 Voltage Offset Correction: v_off_ini = 2438.564 uV, v_off_fin = 2403.048 uV 13:37:28 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -128.131 uVs 13:37:28 Integrated Voltage Measurement... 13:37:37 Voltage Offset Correction: v_off_ini = 2398.031 uV, v_off_fin = 2406.679 uV 13:37:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.432 uVs 13:37:37 Integrated Voltage Measurement... 13:37:46 Voltage Offset Correction: v_off_ini = 2354.604 uV, v_off_fin = 2378.516 uV 13:37:46 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 0.390 uVs 13:37:46 Integrated Voltage Measurement... 13:37:55 Voltage Offset Correction: v_off_ini = 2391.826 uV, v_off_fin = 2407.782 uV 13:37:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 49.214 uVs 13:37:55 Integrated Voltage Measurement... 13:38:04 Voltage Offset Correction: v_off_ini = 2405.775 uV, v_off_fin = 2397.217 uV 13:38:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -46.863 uVs 13:38:04 Integrated Voltage Measurement... 13:38:13 Voltage Offset Correction: v_off_ini = 2429.063 uV, v_off_fin = 2406.795 uV 13:38:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.314 uVs 13:38:15 I2X Average and RMS of Measurements VT = -55.427 +- 28.529 uVs I2X = 5.099 +- 2.624 uTm^2 13:38:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:38:23 Integrated Voltage Measurement... 13:38:32 Voltage Offset Correction: v_off_ini = 2424.513 uV, v_off_fin = 2420.977 uV 13:38:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -401.857 uVs 13:38:32 Integrated Voltage Measurement... 13:38:41 Voltage Offset Correction: v_off_ini = 2432.412 uV, v_off_fin = 2436.998 uV 13:38:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 382.285 uVs 13:38:41 Integrated Voltage Measurement... 13:38:50 Voltage Offset Correction: v_off_ini = 2391.871 uV, v_off_fin = 2411.317 uV 13:38:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -356.419 uVs 13:38:50 Integrated Voltage Measurement... 13:38:59 Voltage Offset Correction: v_off_ini = 2399.046 uV, v_off_fin = 2423.398 uV 13:38:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 391.225 uVs 13:38:59 Integrated Voltage Measurement... 13:39:08 Voltage Offset Correction: v_off_ini = 2392.871 uV, v_off_fin = 2434.263 uV 13:39:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -350.712 uVs 13:39:08 Integrated Voltage Measurement... 13:39:17 Voltage Offset Correction: v_off_ini = 2413.779 uV, v_off_fin = 2389.511 uV 13:39:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 414.825 uVs 13:39:19 I2Y Average and RMS of Measurements VT = -382.887 +- 7.451 uVs I2Y = -35.226 +- 0.685 uTm^2 13:39:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:39:34 Integrated Voltage Measurement... 13:39:43 Voltage Offset Correction: v_off_ini = 2444.092 uV, v_off_fin = 2419.471 uV 13:39:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -47.339 uVs 13:39:43 Integrated Voltage Measurement... 13:39:52 Voltage Offset Correction: v_off_ini = 2421.428 uV, v_off_fin = 2426.761 uV 13:39:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.486 uVs 13:39:52 Integrated Voltage Measurement... 13:40:01 Voltage Offset Correction: v_off_ini = 2455.328 uV, v_off_fin = 2450.344 uV 13:40:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.114 uVs 13:40:01 Integrated Voltage Measurement... 13:40:10 Voltage Offset Correction: v_off_ini = 2417.017 uV, v_off_fin = 2414.925 uV 13:40:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 34.793 uVs 13:40:10 Integrated Voltage Measurement... 13:40:19 Voltage Offset Correction: v_off_ini = 2439.120 uV, v_off_fin = 2428.315 uV 13:40:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.433 uVs 13:40:19 Integrated Voltage Measurement... 13:40:28 Voltage Offset Correction: v_off_ini = 2434.063 uV, v_off_fin = 2446.294 uV 13:40:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.330 uVs 13:40:33 I1X Average and RMS of Measurements VT = -51.582 +- 1.625 uVs I1X = 10.316 +- 0.325 uTm 13:40:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:40:39 Integrated Voltage Measurement... 13:40:48 Voltage Offset Correction: v_off_ini = 2423.073 uV, v_off_fin = 2411.277 uV 13:40:48 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -250.274 uVs 13:40:48 Integrated Voltage Measurement... 13:40:57 Voltage Offset Correction: v_off_ini = 2440.889 uV, v_off_fin = 2477.975 uV 13:40:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 233.450 uVs 13:40:57 Integrated Voltage Measurement... 13:41:06 Voltage Offset Correction: v_off_ini = 2457.021 uV, v_off_fin = 2428.214 uV 13:41:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.797 uVs 13:41:06 Integrated Voltage Measurement... 13:41:15 Voltage Offset Correction: v_off_ini = 2454.355 uV, v_off_fin = 2463.927 uV 13:41:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.208 uVs 13:41:15 Integrated Voltage Measurement... 13:41:24 Voltage Offset Correction: v_off_ini = 2469.031 uV, v_off_fin = 2462.130 uV 13:41:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -230.618 uVs 13:41:24 Integrated Voltage Measurement... 13:41:33 Voltage Offset Correction: v_off_ini = 2483.908 uV, v_off_fin = 2459.825 uV 13:41:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 243.212 uVs 13:41:37 I1Y Average and RMS of Measurements VT = -238.926 +- 2.122 uVs I1Y = -47.783 +- 0.425 uTm 13:41:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:41:45 Integrated Voltage Measurement... 13:41:54 Voltage Offset Correction: v_off_ini = 2493.508 uV, v_off_fin = 2496.968 uV 13:41:54 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -58.260 uVs 13:41:54 Integrated Voltage Measurement... 13:42:03 Voltage Offset Correction: v_off_ini = 2498.653 uV, v_off_fin = 2496.046 uV 13:42:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.770 uVs 13:42:03 Integrated Voltage Measurement... 13:42:12 Voltage Offset Correction: v_off_ini = 2465.277 uV, v_off_fin = 2468.717 uV 13:42:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -35.784 uVs 13:42:12 Integrated Voltage Measurement... 13:42:21 Voltage Offset Correction: v_off_ini = 2487.440 uV, v_off_fin = 2507.406 uV 13:42:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.779 uVs 13:42:21 Integrated Voltage Measurement... 13:42:30 Voltage Offset Correction: v_off_ini = 2509.947 uV, v_off_fin = 2489.513 uV 13:42:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.541 uVs 13:42:30 Integrated Voltage Measurement... 13:42:39 Voltage Offset Correction: v_off_ini = 2487.660 uV, v_off_fin = 2477.520 uV 13:42:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 24.508 uVs 13:42:42 I2X Average and RMS of Measurements VT = -38.940 +- 11.742 uVs I2X = 3.582 +- 1.080 uTm^2 13:42:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:42:49 Integrated Voltage Measurement... 13:42:58 Voltage Offset Correction: v_off_ini = 2443.400 uV, v_off_fin = 2464.233 uV 13:42:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -384.936 uVs 13:42:58 Integrated Voltage Measurement... 13:43:07 Voltage Offset Correction: v_off_ini = 2463.987 uV, v_off_fin = 2489.968 uV 13:43:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 389.835 uVs 13:43:07 Integrated Voltage Measurement... 13:43:16 Voltage Offset Correction: v_off_ini = 2474.363 uV, v_off_fin = 2460.468 uV 13:43:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -375.346 uVs 13:43:16 Integrated Voltage Measurement... 13:43:25 Voltage Offset Correction: v_off_ini = 2465.132 uV, v_off_fin = 2477.583 uV 13:43:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 369.701 uVs 13:43:25 Integrated Voltage Measurement... 13:43:34 Voltage Offset Correction: v_off_ini = 2462.718 uV, v_off_fin = 2458.241 uV 13:43:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -344.584 uVs 13:43:34 Integrated Voltage Measurement... 13:43:43 Voltage Offset Correction: v_off_ini = 2478.291 uV, v_off_fin = 2475.209 uV 13:43:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 378.889 uVs 13:43:46 I2Y Average and RMS of Measurements VT = -373.882 +- 10.515 uVs I2Y = -34.397 +- 0.967 uTm^2 13:43:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:44:01 Integrated Voltage Measurement... 13:44:10 Voltage Offset Correction: v_off_ini = 2503.792 uV, v_off_fin = 2504.721 uV 13:44:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -46.606 uVs 13:44:10 Integrated Voltage Measurement... 13:44:19 Voltage Offset Correction: v_off_ini = 2488.579 uV, v_off_fin = 2487.938 uV 13:44:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.669 uVs 13:44:19 Integrated Voltage Measurement... 13:44:28 Voltage Offset Correction: v_off_ini = 2467.656 uV, v_off_fin = 2478.250 uV 13:44:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.817 uVs 13:44:28 Integrated Voltage Measurement... 13:44:37 Voltage Offset Correction: v_off_ini = 2479.221 uV, v_off_fin = 2501.889 uV 13:44:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.238 uVs 13:44:37 Integrated Voltage Measurement... 13:44:46 Voltage Offset Correction: v_off_ini = 2496.814 uV, v_off_fin = 2474.751 uV 13:44:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.775 uVs 13:44:46 Integrated Voltage Measurement... 13:44:55 Voltage Offset Correction: v_off_ini = 2487.634 uV, v_off_fin = 2464.543 uV 13:44:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.461 uVs 13:44:59 I1X Average and RMS of Measurements VT = -55.261 +- 9.530 uVs I1X = 11.052 +- 1.906 uTm 13:45:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:45:06 Integrated Voltage Measurement... 13:45:15 Voltage Offset Correction: v_off_ini = 2485.999 uV, v_off_fin = 2463.321 uV 13:45:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.719 uVs 13:45:15 Integrated Voltage Measurement... 13:45:24 Voltage Offset Correction: v_off_ini = 2453.417 uV, v_off_fin = 2479.356 uV 13:45:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 275.460 uVs 13:45:24 Integrated Voltage Measurement... 13:45:33 Voltage Offset Correction: v_off_ini = 2473.988 uV, v_off_fin = 2458.988 uV 13:45:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -235.362 uVs 13:45:33 Integrated Voltage Measurement... 13:45:42 Voltage Offset Correction: v_off_ini = 2461.820 uV, v_off_fin = 2471.698 uV 13:45:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 225.465 uVs 13:45:42 Integrated Voltage Measurement... 13:45:51 Voltage Offset Correction: v_off_ini = 2495.444 uV, v_off_fin = 2469.539 uV 13:45:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.108 uVs 13:45:51 Integrated Voltage Measurement... 13:46:00 Voltage Offset Correction: v_off_ini = 2498.226 uV, v_off_fin = 2482.790 uV 13:46:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 219.987 uVs 13:46:04 I1Y Average and RMS of Measurements VT = -226.184 +- 9.482 uVs I1Y = -45.232 +- 1.896 uTm 13:46:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:46:12 Integrated Voltage Measurement... 13:46:20 Voltage Offset Correction: v_off_ini = 2480.338 uV, v_off_fin = 2460.878 uV 13:46:21 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -25.361 uVs 13:46:21 Integrated Voltage Measurement... 13:46:30 Voltage Offset Correction: v_off_ini = 2475.825 uV, v_off_fin = 2514.012 uV 13:46:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.512 uVs 13:46:30 Integrated Voltage Measurement... 13:46:39 Voltage Offset Correction: v_off_ini = 2507.685 uV, v_off_fin = 2502.313 uV 13:46:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.933 uVs 13:46:39 Integrated Voltage Measurement... 13:46:48 Voltage Offset Correction: v_off_ini = 2503.514 uV, v_off_fin = 2499.639 uV 13:46:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.820 uVs 13:46:48 Integrated Voltage Measurement... 13:46:57 Voltage Offset Correction: v_off_ini = 2497.441 uV, v_off_fin = 2486.001 uV 13:46:57 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -57.005 uVs 13:46:57 Integrated Voltage Measurement... 13:47:06 Voltage Offset Correction: v_off_ini = 2501.592 uV, v_off_fin = 2498.706 uV 13:47:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.926 uVs 13:47:08 I2X Average and RMS of Measurements VT = -52.259 +- 4.057 uVs I2X = 4.808 +- 0.373 uTm^2 13:47:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:47:16 Integrated Voltage Measurement... 13:47:25 Voltage Offset Correction: v_off_ini = 2486.540 uV, v_off_fin = 2487.547 uV 13:47:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -356.116 uVs 13:47:25 Integrated Voltage Measurement... 13:47:34 Voltage Offset Correction: v_off_ini = 2471.571 uV, v_off_fin = 2474.712 uV 13:47:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 353.996 uVs 13:47:34 Integrated Voltage Measurement... 13:47:43 Voltage Offset Correction: v_off_ini = 2483.715 uV, v_off_fin = 2470.384 uV 13:47:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -360.251 uVs 13:47:43 Integrated Voltage Measurement... 13:47:52 Voltage Offset Correction: v_off_ini = 2495.349 uV, v_off_fin = 2510.426 uV 13:47:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 358.020 uVs 13:47:52 Integrated Voltage Measurement... 13:48:01 Voltage Offset Correction: v_off_ini = 2468.947 uV, v_off_fin = 2511.611 uV 13:48:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -366.545 uVs 13:48:01 Integrated Voltage Measurement... 13:48:10 Voltage Offset Correction: v_off_ini = 2500.574 uV, v_off_fin = 2506.887 uV 13:48:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 377.215 uVs 13:48:12 I2Y Average and RMS of Measurements VT = -362.024 +- 7.166 uVs I2Y = -33.306 +- 0.659 uTm^2 13:48:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:48:27 Integrated Voltage Measurement... 13:48:36 Voltage Offset Correction: v_off_ini = 2507.310 uV, v_off_fin = 2504.810 uV 13:48:36 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -54.152 uVs 13:48:36 Integrated Voltage Measurement... 13:48:45 Voltage Offset Correction: v_off_ini = 2506.835 uV, v_off_fin = 2533.412 uV 13:48:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 32.298 uVs 13:48:45 Integrated Voltage Measurement... 13:48:54 Voltage Offset Correction: v_off_ini = 2523.793 uV, v_off_fin = 2516.451 uV 13:48:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -19.892 uVs 13:48:54 Integrated Voltage Measurement... 13:49:03 Voltage Offset Correction: v_off_ini = 2509.758 uV, v_off_fin = 2511.828 uV 13:49:03 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.909 uVs 13:49:03 Integrated Voltage Measurement... 13:49:12 Voltage Offset Correction: v_off_ini = 2496.367 uV, v_off_fin = 2489.579 uV 13:49:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -39.050 uVs 13:49:12 Integrated Voltage Measurement... 13:49:21 Voltage Offset Correction: v_off_ini = 2498.303 uV, v_off_fin = 2504.371 uV 13:49:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 24.010 uVs 13:49:26 I1X Average and RMS of Measurements VT = -37.385 +- 4.774 uVs I1X = 7.478 +- 0.955 uTm 13:49:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:49:32 Integrated Voltage Measurement... 13:49:41 Voltage Offset Correction: v_off_ini = 2493.118 uV, v_off_fin = 2504.516 uV 13:49:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -202.427 uVs 13:49:41 Integrated Voltage Measurement... 13:49:50 Voltage Offset Correction: v_off_ini = 2525.979 uV, v_off_fin = 2519.679 uV 13:49:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.181 uVs 13:49:50 Integrated Voltage Measurement... 13:49:59 Voltage Offset Correction: v_off_ini = 2491.591 uV, v_off_fin = 2492.102 uV 13:49:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -211.282 uVs 13:49:59 Integrated Voltage Measurement... 13:50:08 Voltage Offset Correction: v_off_ini = 2504.772 uV, v_off_fin = 2497.203 uV 13:50:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 233.276 uVs 13:50:08 Integrated Voltage Measurement... 13:50:17 Voltage Offset Correction: v_off_ini = 2495.989 uV, v_off_fin = 2500.030 uV 13:50:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.426 uVs 13:50:17 Integrated Voltage Measurement... 13:50:26 Voltage Offset Correction: v_off_ini = 2510.074 uV, v_off_fin = 2486.195 uV 13:50:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.115 uVs 13:50:30 I1Y Average and RMS of Measurements VT = -210.284 +- 8.635 uVs I1Y = -42.053 +- 1.727 uTm 13:50:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:50:38 Integrated Voltage Measurement... 13:50:47 Voltage Offset Correction: v_off_ini = 2503.413 uV, v_off_fin = 2498.043 uV 13:50:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.558 uVs 13:50:47 Integrated Voltage Measurement... 13:50:56 Voltage Offset Correction: v_off_ini = 2512.527 uV, v_off_fin = 2525.378 uV 13:50:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.306 uVs 13:50:56 Integrated Voltage Measurement... 13:51:05 Voltage Offset Correction: v_off_ini = 2504.764 uV, v_off_fin = 2490.098 uV 13:51:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.755 uVs 13:51:05 Integrated Voltage Measurement... 13:51:14 Voltage Offset Correction: v_off_ini = 2514.172 uV, v_off_fin = 2532.437 uV 13:51:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.049 uVs 13:51:14 Integrated Voltage Measurement... 13:51:23 Voltage Offset Correction: v_off_ini = 2527.985 uV, v_off_fin = 2496.765 uV 13:51:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.731 uVs 13:51:23 Integrated Voltage Measurement... 13:51:32 Voltage Offset Correction: v_off_ini = 2473.721 uV, v_off_fin = 2488.073 uV 13:51:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.247 uVs 13:51:35 I2X Average and RMS of Measurements VT = -54.441 +- 5.314 uVs I2X = 5.009 +- 0.489 uTm^2 13:51:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:51:42 Integrated Voltage Measurement... 13:51:51 Voltage Offset Correction: v_off_ini = 2488.007 uV, v_off_fin = 2464.784 uV 13:51:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -341.497 uVs 13:51:51 Integrated Voltage Measurement... 13:52:00 Voltage Offset Correction: v_off_ini = 2429.520 uV, v_off_fin = 2438.219 uV 13:52:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 346.443 uVs 13:52:00 Integrated Voltage Measurement... 13:52:09 Voltage Offset Correction: v_off_ini = 2451.139 uV, v_off_fin = 2451.672 uV 13:52:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -338.619 uVs 13:52:09 Integrated Voltage Measurement... 13:52:18 Voltage Offset Correction: v_off_ini = 2450.766 uV, v_off_fin = 2469.169 uV 13:52:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 352.137 uVs 13:52:19 Integrated Voltage Measurement... 13:52:27 Voltage Offset Correction: v_off_ini = 2467.192 uV, v_off_fin = 2467.790 uV 13:52:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.670 uVs 13:52:28 Integrated Voltage Measurement... 13:52:37 Voltage Offset Correction: v_off_ini = 2490.883 uV, v_off_fin = 2481.674 uV 13:52:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 337.896 uVs 13:52:39 I2Y Average and RMS of Measurements VT = -338.377 +- 8.924 uVs I2Y = -31.131 +- 0.821 uTm^2 13:52:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:52:54 Integrated Voltage Measurement... 13:53:03 Voltage Offset Correction: v_off_ini = 2475.869 uV, v_off_fin = 2483.221 uV 13:53:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -68.814 uVs 13:53:03 Integrated Voltage Measurement... 13:53:12 Voltage Offset Correction: v_off_ini = 2481.645 uV, v_off_fin = 2499.007 uV 13:53:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.416 uVs 13:53:12 Integrated Voltage Measurement... 13:53:21 Voltage Offset Correction: v_off_ini = 2476.544 uV, v_off_fin = 2471.205 uV 13:53:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.377 uVs 13:53:21 Integrated Voltage Measurement... 13:53:30 Voltage Offset Correction: v_off_ini = 2482.397 uV, v_off_fin = 2479.074 uV 13:53:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 21.088 uVs 13:53:30 Integrated Voltage Measurement... 13:53:39 Voltage Offset Correction: v_off_ini = 2472.198 uV, v_off_fin = 2477.112 uV 13:53:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.275 uVs 13:53:39 Integrated Voltage Measurement... 13:53:48 Voltage Offset Correction: v_off_ini = 2455.047 uV, v_off_fin = 2471.723 uV 13:53:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.083 uVs 13:53:52 I1X Average and RMS of Measurements VT = -56.342 +- 8.566 uVs I1X = 11.269 +- 1.713 uTm 13:53:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:53:59 Integrated Voltage Measurement... 13:54:08 Voltage Offset Correction: v_off_ini = 2486.644 uV, v_off_fin = 2482.102 uV 13:54:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -205.627 uVs 13:54:08 Integrated Voltage Measurement... 13:54:17 Voltage Offset Correction: v_off_ini = 2484.703 uV, v_off_fin = 2486.754 uV 13:54:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 195.717 uVs 13:54:17 Integrated Voltage Measurement... 13:54:26 Voltage Offset Correction: v_off_ini = 2478.232 uV, v_off_fin = 2487.509 uV 13:54:26 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.206 uVs 13:54:26 Integrated Voltage Measurement... 13:54:35 Voltage Offset Correction: v_off_ini = 2509.849 uV, v_off_fin = 2515.773 uV 13:54:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.363 uVs 13:54:35 Integrated Voltage Measurement... 13:54:44 Voltage Offset Correction: v_off_ini = 2506.013 uV, v_off_fin = 2492.075 uV 13:54:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.398 uVs 13:54:44 Integrated Voltage Measurement... 13:54:53 Voltage Offset Correction: v_off_ini = 2512.037 uV, v_off_fin = 2513.938 uV 13:54:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.108 uVs 13:54:57 I1Y Average and RMS of Measurements VT = -188.903 +- 10.715 uVs I1Y = -37.779 +- 2.141 uTm 13:55:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:55:05 Integrated Voltage Measurement... 13:55:14 Voltage Offset Correction: v_off_ini = 2520.532 uV, v_off_fin = 2533.626 uV 13:55:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -22.608 uVs 13:55:14 Integrated Voltage Measurement... 13:55:23 Voltage Offset Correction: v_off_ini = 2500.293 uV, v_off_fin = 2530.739 uV 13:55:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.809 uVs 13:55:23 Integrated Voltage Measurement... 13:55:32 Voltage Offset Correction: v_off_ini = 2528.310 uV, v_off_fin = 2504.731 uV 13:55:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.770 uVs 13:55:32 Integrated Voltage Measurement... 13:55:41 Voltage Offset Correction: v_off_ini = 2484.972 uV, v_off_fin = 2509.404 uV 13:55:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 54.399 uVs 13:55:41 Integrated Voltage Measurement... 13:55:50 Voltage Offset Correction: v_off_ini = 2519.026 uV, v_off_fin = 2508.018 uV 13:55:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -43.700 uVs 13:55:50 Integrated Voltage Measurement... 13:55:59 Voltage Offset Correction: v_off_ini = 2515.104 uV, v_off_fin = 2526.857 uV 13:55:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -3.734 uVs 13:56:01 I2X Average and RMS of Measurements VT = -37.759 +- 13.787 uVs I2X = 3.474 +- 1.268 uTm^2 13:56:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:56:09 Integrated Voltage Measurement... 13:56:18 Voltage Offset Correction: v_off_ini = 2503.185 uV, v_off_fin = 2505.746 uV 13:56:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.978 uVs 13:56:18 Integrated Voltage Measurement... 13:56:27 Voltage Offset Correction: v_off_ini = 2505.633 uV, v_off_fin = 2496.253 uV 13:56:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 325.254 uVs 13:56:27 Integrated Voltage Measurement... 13:56:36 Voltage Offset Correction: v_off_ini = 2497.020 uV, v_off_fin = 2506.233 uV 13:56:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.870 uVs 13:56:36 Integrated Voltage Measurement... 13:56:45 Voltage Offset Correction: v_off_ini = 2514.933 uV, v_off_fin = 2505.175 uV 13:56:45 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 316.319 uVs 13:56:45 Integrated Voltage Measurement... 13:56:54 Voltage Offset Correction: v_off_ini = 2517.186 uV, v_off_fin = 2525.420 uV 13:56:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.235 uVs 13:56:54 Integrated Voltage Measurement... 13:57:03 Voltage Offset Correction: v_off_ini = 2534.044 uV, v_off_fin = 2544.402 uV 13:57:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 335.063 uVs 13:57:06 I2Y Average and RMS of Measurements VT = -327.120 +- 3.683 uVs I2Y = -30.096 +- 0.338 uTm^2 13:57:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:57:21 Integrated Voltage Measurement... 13:57:30 Voltage Offset Correction: v_off_ini = 2523.806 uV, v_off_fin = 2505.954 uV 13:57:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.563 uVs 13:57:30 Integrated Voltage Measurement... 13:57:39 Voltage Offset Correction: v_off_ini = 2486.285 uV, v_off_fin = 2528.394 uV 13:57:39 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.551 uVs 13:57:39 Integrated Voltage Measurement... 13:57:48 Voltage Offset Correction: v_off_ini = 2468.249 uV, v_off_fin = 2511.303 uV 13:57:48 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.393 uVs 13:57:48 Integrated Voltage Measurement... 13:57:57 Voltage Offset Correction: v_off_ini = 2497.453 uV, v_off_fin = 2482.899 uV 13:57:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 41.919 uVs 13:57:57 Integrated Voltage Measurement... 13:58:06 Voltage Offset Correction: v_off_ini = 2491.419 uV, v_off_fin = 2477.267 uV 13:58:06 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.630 uVs 13:58:06 Integrated Voltage Measurement... 13:58:15 Voltage Offset Correction: v_off_ini = 2465.596 uV, v_off_fin = 2455.927 uV 13:58:15 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.756 uVs 13:58:19 I1X Average and RMS of Measurements VT = -60.802 +- 11.610 uVs I1X = 12.161 +- 2.322 uTm 13:58:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:58:26 Integrated Voltage Measurement... 13:58:35 Voltage Offset Correction: v_off_ini = 2473.859 uV, v_off_fin = 2461.272 uV 13:58:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.914 uVs 13:58:35 Integrated Voltage Measurement... 13:58:44 Voltage Offset Correction: v_off_ini = 2440.564 uV, v_off_fin = 2495.100 uV 13:58:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.347 uVs 13:58:44 Integrated Voltage Measurement... 13:58:53 Voltage Offset Correction: v_off_ini = 2468.406 uV, v_off_fin = 2470.899 uV 13:58:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -199.371 uVs 13:58:53 Integrated Voltage Measurement... 13:59:02 Voltage Offset Correction: v_off_ini = 2473.571 uV, v_off_fin = 2488.488 uV 13:59:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 190.044 uVs 13:59:02 Integrated Voltage Measurement... 13:59:11 Voltage Offset Correction: v_off_ini = 2468.056 uV, v_off_fin = 2473.876 uV 13:59:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -192.731 uVs 13:59:11 Integrated Voltage Measurement... 13:59:20 Voltage Offset Correction: v_off_ini = 2470.363 uV, v_off_fin = 2489.264 uV 13:59:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.766 uVs 13:59:24 I1Y Average and RMS of Measurements VT = -196.362 +- 6.282 uVs I1Y = -39.270 +- 1.255 uTm