SLAC Magnetic Measurements Date: 01-15-2025 Time: 11:57:27 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_10086 Dataset Number: 1 Run Purpose: Final Results Run Type: Stretched Wire Run Number: 5 Measurement Device: Background Operator: YL Comment: 11:57:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:57:37 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:57:39 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:57:40 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:57:42 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:57:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:57:52 Integrated Voltage Measurement... 11:58:01 Voltage Offset Correction: v_off_ini = 186.101 uV, v_off_fin = 188.934 uV 11:58:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -120.091 uVs 11:58:02 Integrated Voltage Measurement... 11:58:10 Voltage Offset Correction: v_off_ini = 191.576 uV, v_off_fin = 189.203 uV 11:58:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 139.900 uVs 11:58:11 Integrated Voltage Measurement... 11:58:20 Voltage Offset Correction: v_off_ini = 202.202 uV, v_off_fin = 197.544 uV 11:58:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.228 uVs 11:58:20 Integrated Voltage Measurement... 11:58:29 Voltage Offset Correction: v_off_ini = 196.520 uV, v_off_fin = 196.134 uV 11:58:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.201 uVs 11:58:29 Integrated Voltage Measurement... 11:58:38 Voltage Offset Correction: v_off_ini = 194.189 uV, v_off_fin = 205.747 uV 11:58:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -114.075 uVs 11:58:38 Integrated Voltage Measurement... 11:58:47 Voltage Offset Correction: v_off_ini = 195.427 uV, v_off_fin = 190.220 uV 11:58:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 130.043 uVs 11:58:51 I1X Average and RMS of Measurements VT = -122.423 +- 6.040 uVs I1X = 24.485 +- 1.208 uTm 11:58:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:58:58 Integrated Voltage Measurement... 11:59:07 Voltage Offset Correction: v_off_ini = 176.902 uV, v_off_fin = 180.562 uV 11:59:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.735 uVs 11:59:07 Integrated Voltage Measurement... 11:59:16 Voltage Offset Correction: v_off_ini = 208.616 uV, v_off_fin = 210.727 uV 11:59:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.918 uVs 11:59:16 Integrated Voltage Measurement... 11:59:25 Voltage Offset Correction: v_off_ini = 198.115 uV, v_off_fin = 186.676 uV 11:59:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.976 uVs 11:59:25 Integrated Voltage Measurement... 11:59:34 Voltage Offset Correction: v_off_ini = 197.643 uV, v_off_fin = 206.683 uV 11:59:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 216.038 uVs 11:59:34 Integrated Voltage Measurement... 11:59:43 Voltage Offset Correction: v_off_ini = 197.228 uV, v_off_fin = 183.972 uV 11:59:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.488 uVs 11:59:43 Integrated Voltage Measurement... 11:59:52 Voltage Offset Correction: v_off_ini = 204.502 uV, v_off_fin = 221.031 uV 11:59:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 221.009 uVs 11:59:56 I1Y Average and RMS of Measurements VT = -220.527 +- 4.751 uVs I1Y = -44.105 +- 0.950 uTm 12:00:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:00:04 Integrated Voltage Measurement... 12:00:13 Voltage Offset Correction: v_off_ini = 200.887 uV, v_off_fin = 188.032 uV 12:00:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -154.885 uVs 12:00:13 Integrated Voltage Measurement... 12:00:22 Voltage Offset Correction: v_off_ini = 197.066 uV, v_off_fin = 208.145 uV 12:00:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 134.693 uVs 12:00:22 Integrated Voltage Measurement... 12:00:31 Voltage Offset Correction: v_off_ini = 200.752 uV, v_off_fin = 195.565 uV 12:00:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -144.206 uVs 12:00:31 Integrated Voltage Measurement... 12:00:40 Voltage Offset Correction: v_off_ini = 193.222 uV, v_off_fin = 207.674 uV 12:00:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.171 uVs 12:00:40 Integrated Voltage Measurement... 12:00:49 Voltage Offset Correction: v_off_ini = 189.897 uV, v_off_fin = 181.876 uV 12:00:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.906 uVs 12:00:49 Integrated Voltage Measurement... 12:00:58 Voltage Offset Correction: v_off_ini = 198.542 uV, v_off_fin = 201.683 uV 12:00:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 179.646 uVs 12:01:00 I2X Average and RMS of Measurements VT = -143.584 +- 5.612 uVs I2X = 13.210 +- 0.516 uTm^2 12:01:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:01:08 Integrated Voltage Measurement... 12:01:17 Voltage Offset Correction: v_off_ini = 199.191 uV, v_off_fin = 200.409 uV 12:01:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.595 uVs 12:01:17 Integrated Voltage Measurement... 12:01:26 Voltage Offset Correction: v_off_ini = 208.321 uV, v_off_fin = 203.906 uV 12:01:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 226.630 uVs 12:01:26 Integrated Voltage Measurement... 12:01:35 Voltage Offset Correction: v_off_ini = 197.695 uV, v_off_fin = 206.439 uV 12:01:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.531 uVs 12:01:35 Integrated Voltage Measurement... 12:01:44 Voltage Offset Correction: v_off_ini = 201.862 uV, v_off_fin = 198.362 uV 12:01:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 242.908 uVs 12:01:44 Integrated Voltage Measurement... 12:01:53 Voltage Offset Correction: v_off_ini = 219.026 uV, v_off_fin = 204.041 uV 12:01:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -216.799 uVs 12:01:53 Integrated Voltage Measurement... 12:02:02 Voltage Offset Correction: v_off_ini = 198.702 uV, v_off_fin = 198.004 uV 12:02:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 219.903 uVs 12:02:05 I2Y Average and RMS of Measurements VT = -226.061 +- 6.083 uVs I2Y = -20.798 +- 0.560 uTm^2 12:02:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:05:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:05:26 Integrated Voltage Measurement... 13:05:35 Voltage Offset Correction: v_off_ini = 344.457 uV, v_off_fin = 336.756 uV 13:05:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.869 uVs 13:05:35 Integrated Voltage Measurement... 13:05:44 Voltage Offset Correction: v_off_ini = 353.899 uV, v_off_fin = 340.844 uV 13:05:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.916 uVs 13:05:44 Integrated Voltage Measurement... 13:05:53 Voltage Offset Correction: v_off_ini = 341.054 uV, v_off_fin = 343.454 uV 13:05:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.139 uVs 13:05:53 Integrated Voltage Measurement... 13:06:02 Voltage Offset Correction: v_off_ini = 308.588 uV, v_off_fin = 328.677 uV 13:06:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 131.899 uVs 13:06:02 Integrated Voltage Measurement... 13:06:11 Voltage Offset Correction: v_off_ini = 327.342 uV, v_off_fin = 323.739 uV 13:06:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.160 uVs 13:06:11 Integrated Voltage Measurement... 13:06:20 Voltage Offset Correction: v_off_ini = 330.153 uV, v_off_fin = 329.262 uV 13:06:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.066 uVs 13:06:24 I1X Average and RMS of Measurements VT = -103.841 +- 6.698 uVs I1X = 20.768 +- 1.340 uTm 13:06:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:06:31 Integrated Voltage Measurement... 13:06:40 Voltage Offset Correction: v_off_ini = 318.523 uV, v_off_fin = 319.293 uV 13:06:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 420.610 uVs 13:06:40 Integrated Voltage Measurement... 13:06:49 Voltage Offset Correction: v_off_ini = 318.878 uV, v_off_fin = 302.011 uV 13:06:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -425.497 uVs 13:06:49 Integrated Voltage Measurement... 13:06:58 Voltage Offset Correction: v_off_ini = 315.508 uV, v_off_fin = 334.483 uV 13:06:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 413.547 uVs 13:06:58 Integrated Voltage Measurement... 13:07:07 Voltage Offset Correction: v_off_ini = 319.322 uV, v_off_fin = 299.348 uV 13:07:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -403.533 uVs 13:07:07 Integrated Voltage Measurement... 13:07:16 Voltage Offset Correction: v_off_ini = 318.383 uV, v_off_fin = 332.749 uV 13:07:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 403.854 uVs 13:07:16 Integrated Voltage Measurement... 13:07:25 Voltage Offset Correction: v_off_ini = 314.619 uV, v_off_fin = 287.838 uV 13:07:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -402.475 uVs 13:07:29 I1Y Average and RMS of Measurements VT = 411.586 +- 8.400 uVs I1Y = 82.317 +- 1.680 uTm 13:07:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:07:37 Integrated Voltage Measurement... 13:07:46 Voltage Offset Correction: v_off_ini = 290.258 uV, v_off_fin = 249.618 uV 13:07:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -142.134 uVs 13:07:46 Integrated Voltage Measurement... 13:07:55 Voltage Offset Correction: v_off_ini = 287.464 uV, v_off_fin = 288.920 uV 13:07:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.374 uVs 13:07:55 Integrated Voltage Measurement... 13:08:04 Voltage Offset Correction: v_off_ini = 297.229 uV, v_off_fin = 268.438 uV 13:08:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -117.176 uVs 13:08:04 Integrated Voltage Measurement... 13:08:13 Voltage Offset Correction: v_off_ini = 264.304 uV, v_off_fin = 276.441 uV 13:08:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.958 uVs 13:08:13 Integrated Voltage Measurement... 13:08:22 Voltage Offset Correction: v_off_ini = 261.151 uV, v_off_fin = 253.715 uV 13:08:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.467 uVs 13:08:22 Integrated Voltage Measurement... 13:08:31 Voltage Offset Correction: v_off_ini = 285.548 uV, v_off_fin = 278.351 uV 13:08:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.659 uVs 13:08:34 I2X Average and RMS of Measurements VT = -117.461 +- 11.323 uVs I2X = 10.806 +- 1.042 uTm^2 13:08:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:08:42 Integrated Voltage Measurement... 13:08:51 Voltage Offset Correction: v_off_ini = 270.910 uV, v_off_fin = 276.532 uV 13:08:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 465.769 uVs 13:08:51 Integrated Voltage Measurement... 13:09:00 Voltage Offset Correction: v_off_ini = 287.349 uV, v_off_fin = 281.437 uV 13:09:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -476.880 uVs 13:09:00 Integrated Voltage Measurement... 13:09:09 Voltage Offset Correction: v_off_ini = 262.726 uV, v_off_fin = 273.275 uV 13:09:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 481.567 uVs 13:09:09 Integrated Voltage Measurement... 13:09:18 Voltage Offset Correction: v_off_ini = 288.389 uV, v_off_fin = 287.906 uV 13:09:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -485.160 uVs 13:09:18 Integrated Voltage Measurement... 13:09:27 Voltage Offset Correction: v_off_ini = 271.446 uV, v_off_fin = 296.849 uV 13:09:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 484.607 uVs 13:09:27 Integrated Voltage Measurement... 13:09:36 Voltage Offset Correction: v_off_ini = 302.350 uV, v_off_fin = 299.808 uV 13:09:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -494.676 uVs 13:09:38 I2Y Average and RMS of Measurements VT = 481.443 +- 7.600 uVs I2Y = 44.293 +- 0.699 uTm^2 13:09:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:10:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:10:39 Integrated Voltage Measurement... 13:10:48 Voltage Offset Correction: v_off_ini = 321.395 uV, v_off_fin = 303.023 uV 13:10:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.778 uVs 13:10:48 Integrated Voltage Measurement... 13:10:57 Voltage Offset Correction: v_off_ini = 314.441 uV, v_off_fin = 316.688 uV 13:10:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.659 uVs 13:10:57 Integrated Voltage Measurement... 13:11:06 Voltage Offset Correction: v_off_ini = 325.420 uV, v_off_fin = 335.486 uV 13:11:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.992 uVs 13:11:06 Integrated Voltage Measurement... 13:11:15 Voltage Offset Correction: v_off_ini = 335.427 uV, v_off_fin = 314.791 uV 13:11:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.377 uVs 13:11:15 Integrated Voltage Measurement... 13:11:24 Voltage Offset Correction: v_off_ini = 322.322 uV, v_off_fin = 327.634 uV 13:11:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -129.156 uVs 13:11:24 Integrated Voltage Measurement... 13:11:33 Voltage Offset Correction: v_off_ini = 316.552 uV, v_off_fin = 302.657 uV 13:11:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.560 uVs 13:11:37 I1X Average and RMS of Measurements VT = -107.754 +- 6.682 uVs I1X = 21.551 +- 1.336 uTm 13:11:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:11:44 Integrated Voltage Measurement... 13:11:53 Voltage Offset Correction: v_off_ini = 323.670 uV, v_off_fin = 313.226 uV 13:11:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -852.070 uVs 13:11:53 Integrated Voltage Measurement... 13:12:02 Voltage Offset Correction: v_off_ini = 336.593 uV, v_off_fin = 354.317 uV 13:12:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 847.449 uVs 13:12:02 Integrated Voltage Measurement... 13:12:11 Voltage Offset Correction: v_off_ini = 345.896 uV, v_off_fin = 315.040 uV 13:12:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -844.586 uVs 13:12:11 Integrated Voltage Measurement... 13:12:20 Voltage Offset Correction: v_off_ini = 330.437 uV, v_off_fin = 370.315 uV 13:12:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 869.888 uVs 13:12:20 Integrated Voltage Measurement... 13:12:29 Voltage Offset Correction: v_off_ini = 327.177 uV, v_off_fin = 302.280 uV 13:12:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -837.085 uVs 13:12:29 Integrated Voltage Measurement... 13:12:38 Voltage Offset Correction: v_off_ini = 332.139 uV, v_off_fin = 355.970 uV 13:12:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 854.102 uVs 13:12:42 I1Y Average and RMS of Measurements VT = -850.863 +- 4.817 uVs I1Y = -170.175 +- 0.963 uTm 13:12:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:12:50 Integrated Voltage Measurement... 13:12:59 Voltage Offset Correction: v_off_ini = 316.563 uV, v_off_fin = 325.261 uV 13:12:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.873 uVs 13:12:59 Integrated Voltage Measurement... 13:13:08 Voltage Offset Correction: v_off_ini = 329.053 uV, v_off_fin = 331.674 uV 13:13:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 130.760 uVs 13:13:08 Integrated Voltage Measurement... 13:13:17 Voltage Offset Correction: v_off_ini = 331.005 uV, v_off_fin = 328.336 uV 13:13:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.739 uVs 13:13:17 Integrated Voltage Measurement... 13:13:26 Voltage Offset Correction: v_off_ini = 343.242 uV, v_off_fin = 343.906 uV 13:13:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 142.201 uVs 13:13:26 Integrated Voltage Measurement... 13:13:35 Voltage Offset Correction: v_off_ini = 351.042 uV, v_off_fin = 338.893 uV 13:13:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.460 uVs 13:13:35 Integrated Voltage Measurement... 13:13:44 Voltage Offset Correction: v_off_ini = 351.490 uV, v_off_fin = 348.642 uV 13:13:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.966 uVs 13:13:47 I2X Average and RMS of Measurements VT = -119.000 +- 3.276 uVs I2X = 10.948 +- 0.301 uTm^2 13:13:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:13:54 Integrated Voltage Measurement... 13:14:03 Voltage Offset Correction: v_off_ini = 368.576 uV, v_off_fin = 356.481 uV 13:14:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -944.745 uVs 13:14:03 Integrated Voltage Measurement... 13:14:12 Voltage Offset Correction: v_off_ini = 355.960 uV, v_off_fin = 359.114 uV 13:14:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 923.808 uVs 13:14:12 Integrated Voltage Measurement... 13:14:21 Voltage Offset Correction: v_off_ini = 346.420 uV, v_off_fin = 365.778 uV 13:14:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -929.588 uVs 13:14:21 Integrated Voltage Measurement... 13:14:30 Voltage Offset Correction: v_off_ini = 370.920 uV, v_off_fin = 356.457 uV 13:14:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 944.811 uVs 13:14:31 Integrated Voltage Measurement... 13:14:40 Voltage Offset Correction: v_off_ini = 355.459 uV, v_off_fin = 341.084 uV 13:14:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -915.730 uVs 13:14:40 Integrated Voltage Measurement... 13:14:49 Voltage Offset Correction: v_off_ini = 356.921 uV, v_off_fin = 350.735 uV 13:14:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 930.112 uVs 13:14:51 I2Y Average and RMS of Measurements VT = -931.466 +- 6.159 uVs I2Y = -85.695 +- 0.567 uTm^2 13:14:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm