SLAC Magnetic Measurements Date: 01-06-2025 Time: 15:27:09 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_10086 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 1 Measurement Device: On Axis Operator: YL Comment: Check Integrals 15:28:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:28:10 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:28:11 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:28:13 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:28:14 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:34:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:34:34 Integrated Voltage Measurement... 15:34:43 Voltage Offset Correction: v_off_ini = 720.564 uV, v_off_fin = 718.900 uV 15:34:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -154.488 uVs 15:34:43 Integrated Voltage Measurement... 15:34:52 Voltage Offset Correction: v_off_ini = 730.563 uV, v_off_fin = 732.647 uV 15:34:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 116.789 uVs 15:34:52 Integrated Voltage Measurement... 15:35:01 Voltage Offset Correction: v_off_ini = 713.614 uV, v_off_fin = 715.565 uV 15:35:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -152.315 uVs 15:35:01 Integrated Voltage Measurement... 15:35:10 Voltage Offset Correction: v_off_ini = 747.606 uV, v_off_fin = 723.275 uV 15:35:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.761 uVs 15:35:10 Integrated Voltage Measurement... 15:35:19 Voltage Offset Correction: v_off_ini = 741.923 uV, v_off_fin = 745.058 uV 15:35:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.626 uVs 15:35:19 Integrated Voltage Measurement... 15:35:28 Voltage Offset Correction: v_off_ini = 745.725 uV, v_off_fin = 726.443 uV 15:35:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.989 uVs 15:35:32 I1X Average and RMS of Measurements VT = -127.495 +- 11.446 uVs I1X = 25.500 +- 2.290 uTm 15:35:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:35:39 Integrated Voltage Measurement... 15:35:48 Voltage Offset Correction: v_off_ini = 716.947 uV, v_off_fin = 713.416 uV 15:35:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -301.869 uVs 15:35:48 Integrated Voltage Measurement... 15:35:57 Voltage Offset Correction: v_off_ini = 721.141 uV, v_off_fin = 726.454 uV 15:35:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 306.600 uVs 15:35:57 Integrated Voltage Measurement... 15:36:06 Voltage Offset Correction: v_off_ini = 768.341 uV, v_off_fin = 739.655 uV 15:36:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -289.758 uVs 15:36:06 Integrated Voltage Measurement... 15:36:15 Voltage Offset Correction: v_off_ini = 744.547 uV, v_off_fin = 751.044 uV 15:36:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 310.550 uVs 15:36:15 Integrated Voltage Measurement... 15:36:24 Voltage Offset Correction: v_off_ini = 745.115 uV, v_off_fin = 725.015 uV 15:36:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -315.398 uVs 15:36:24 Integrated Voltage Measurement... 15:36:33 Voltage Offset Correction: v_off_ini = 743.214 uV, v_off_fin = 726.589 uV 15:36:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 292.828 uVs 15:36:37 I1Y Average and RMS of Measurements VT = -302.834 +- 1.896 uVs I1Y = -60.567 +- 0.379 uTm 15:36:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:36:45 Integrated Voltage Measurement... 15:36:54 Voltage Offset Correction: v_off_ini = 738.172 uV, v_off_fin = 733.464 uV 15:36:54 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -107.828 uVs 15:36:54 Integrated Voltage Measurement... 15:37:03 Voltage Offset Correction: v_off_ini = 737.964 uV, v_off_fin = 728.095 uV 15:37:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 149.180 uVs 15:37:03 Integrated Voltage Measurement... 15:37:12 Voltage Offset Correction: v_off_ini = 705.749 uV, v_off_fin = 725.323 uV 15:37:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -114.505 uVs 15:37:12 Integrated Voltage Measurement... 15:37:21 Voltage Offset Correction: v_off_ini = 732.512 uV, v_off_fin = 736.535 uV 15:37:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 136.080 uVs 15:37:21 Integrated Voltage Measurement... 15:37:30 Voltage Offset Correction: v_off_ini = 747.501 uV, v_off_fin = 743.056 uV 15:37:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -135.251 uVs 15:37:30 Integrated Voltage Measurement... 15:37:39 Voltage Offset Correction: v_off_ini = 754.739 uV, v_off_fin = 758.459 uV 15:37:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 143.408 uVs 15:37:41 I2X Average and RMS of Measurements VT = -131.042 +- 6.005 uVs I2X = 12.056 +- 0.552 uTm^2 15:37:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:37:49 Integrated Voltage Measurement... 15:37:58 Voltage Offset Correction: v_off_ini = 742.847 uV, v_off_fin = 732.703 uV 15:37:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -408.534 uVs 15:37:58 Integrated Voltage Measurement... 15:38:07 Voltage Offset Correction: v_off_ini = 727.171 uV, v_off_fin = 768.393 uV 15:38:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 423.204 uVs 15:38:07 Integrated Voltage Measurement... 15:38:16 Voltage Offset Correction: v_off_ini = 744.713 uV, v_off_fin = 745.852 uV 15:38:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -379.331 uVs 15:38:16 Integrated Voltage Measurement... 15:38:25 Voltage Offset Correction: v_off_ini = 753.204 uV, v_off_fin = 754.507 uV 15:38:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 382.297 uVs 15:38:25 Integrated Voltage Measurement... 15:38:34 Voltage Offset Correction: v_off_ini = 748.237 uV, v_off_fin = 749.295 uV 15:38:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -369.847 uVs 15:38:34 Integrated Voltage Measurement... 15:38:43 Voltage Offset Correction: v_off_ini = 754.722 uV, v_off_fin = 748.682 uV 15:38:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 372.823 uVs 15:38:45 I2Y Average and RMS of Measurements VT = -389.340 +- 19.154 uVs I2Y = -35.819 +- 1.762 uTm^2 15:38:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:38:59 Integrated Voltage Measurement... 15:39:08 Voltage Offset Correction: v_off_ini = 778.089 uV, v_off_fin = 765.891 uV 15:39:08 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -130.427 uVs 15:39:08 Integrated Voltage Measurement... 15:39:17 Voltage Offset Correction: v_off_ini = 780.735 uV, v_off_fin = 792.695 uV 15:39:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.418 uVs 15:39:17 Integrated Voltage Measurement... 15:39:26 Voltage Offset Correction: v_off_ini = 801.229 uV, v_off_fin = 780.141 uV 15:39:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.845 uVs 15:39:26 Integrated Voltage Measurement... 15:39:35 Voltage Offset Correction: v_off_ini = 789.012 uV, v_off_fin = 799.331 uV 15:39:35 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.350 uVs 15:39:35 Integrated Voltage Measurement... 15:39:44 Voltage Offset Correction: v_off_ini = 776.010 uV, v_off_fin = 774.340 uV 15:39:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.286 uVs 15:39:44 Integrated Voltage Measurement... 15:39:53 Voltage Offset Correction: v_off_ini = 779.638 uV, v_off_fin = 767.470 uV 15:39:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.356 uVs 15:39:58 I1X Average and RMS of Measurements VT = -117.447 +- 6.440 uVs I1X = 23.491 +- 1.290 uTm 15:40:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:40:04 Integrated Voltage Measurement... 15:40:13 Voltage Offset Correction: v_off_ini = 765.411 uV, v_off_fin = 751.928 uV 15:40:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -287.130 uVs 15:40:13 Integrated Voltage Measurement... 15:40:22 Voltage Offset Correction: v_off_ini = 770.838 uV, v_off_fin = 773.061 uV 15:40:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 310.358 uVs 15:40:22 Integrated Voltage Measurement... 15:40:31 Voltage Offset Correction: v_off_ini = 763.388 uV, v_off_fin = 744.493 uV 15:40:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -298.929 uVs 15:40:31 Integrated Voltage Measurement... 15:40:40 Voltage Offset Correction: v_off_ini = 734.518 uV, v_off_fin = 757.500 uV 15:40:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 314.316 uVs 15:40:40 Integrated Voltage Measurement... 15:40:49 Voltage Offset Correction: v_off_ini = 767.896 uV, v_off_fin = 760.340 uV 15:40:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -299.924 uVs 15:40:49 Integrated Voltage Measurement... 15:40:58 Voltage Offset Correction: v_off_ini = 767.799 uV, v_off_fin = 778.227 uV 15:40:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 317.441 uVs 15:41:03 I1Y Average and RMS of Measurements VT = -304.683 +- 4.283 uVs I1Y = -60.937 +- 0.857 uTm 15:41:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:41:10 Integrated Voltage Measurement... 15:41:19 Voltage Offset Correction: v_off_ini = 776.279 uV, v_off_fin = 757.400 uV 15:41:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -117.973 uVs 15:41:19 Integrated Voltage Measurement... 15:41:28 Voltage Offset Correction: v_off_ini = 759.930 uV, v_off_fin = 771.929 uV 15:41:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 135.815 uVs 15:41:28 Integrated Voltage Measurement... 15:41:37 Voltage Offset Correction: v_off_ini = 760.027 uV, v_off_fin = 763.135 uV 15:41:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.980 uVs 15:41:37 Integrated Voltage Measurement... 15:41:46 Voltage Offset Correction: v_off_ini = 786.621 uV, v_off_fin = 775.980 uV 15:41:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.500 uVs 15:41:46 Integrated Voltage Measurement... 15:41:55 Voltage Offset Correction: v_off_ini = 757.383 uV, v_off_fin = 753.446 uV 15:41:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -114.946 uVs 15:41:55 Integrated Voltage Measurement... 15:42:04 Voltage Offset Correction: v_off_ini = 764.178 uV, v_off_fin = 763.148 uV 15:42:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 166.353 uVs 15:42:07 I2X Average and RMS of Measurements VT = -128.261 +- 9.606 uVs I2X = 11.800 +- 0.884 uTm^2 15:42:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:42:14 Integrated Voltage Measurement... 15:42:23 Voltage Offset Correction: v_off_ini = 769.944 uV, v_off_fin = 768.814 uV 15:42:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -393.010 uVs 15:42:23 Integrated Voltage Measurement... 15:42:32 Voltage Offset Correction: v_off_ini = 764.963 uV, v_off_fin = 762.187 uV 15:42:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 396.653 uVs 15:42:32 Integrated Voltage Measurement... 15:42:41 Voltage Offset Correction: v_off_ini = 750.375 uV, v_off_fin = 754.381 uV 15:42:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -364.791 uVs 15:42:42 Integrated Voltage Measurement... 15:42:50 Voltage Offset Correction: v_off_ini = 741.349 uV, v_off_fin = 762.584 uV 15:42:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 385.278 uVs 15:42:51 Integrated Voltage Measurement... 15:43:00 Voltage Offset Correction: v_off_ini = 777.742 uV, v_off_fin = 772.670 uV 15:43:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -389.325 uVs 15:43:00 Integrated Voltage Measurement... 15:43:09 Voltage Offset Correction: v_off_ini = 756.487 uV, v_off_fin = 756.138 uV 15:43:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 372.616 uVs 15:43:11 I2Y Average and RMS of Measurements VT = -383.612 +- 8.295 uVs I2Y = -35.292 +- 0.763 uTm^2 15:43:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:43:25 Integrated Voltage Measurement... 15:43:34 Voltage Offset Correction: v_off_ini = 779.450 uV, v_off_fin = 789.667 uV 15:43:34 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -93.483 uVs 15:43:34 Integrated Voltage Measurement... 15:43:43 Voltage Offset Correction: v_off_ini = 764.217 uV, v_off_fin = 789.981 uV 15:43:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 110.732 uVs 15:43:43 Integrated Voltage Measurement... 15:43:52 Voltage Offset Correction: v_off_ini = 779.962 uV, v_off_fin = 779.312 uV 15:43:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.667 uVs 15:43:52 Integrated Voltage Measurement... 15:44:01 Voltage Offset Correction: v_off_ini = 790.384 uV, v_off_fin = 778.513 uV 15:44:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.215 uVs 15:44:01 Integrated Voltage Measurement... 15:44:10 Voltage Offset Correction: v_off_ini = 804.945 uV, v_off_fin = 795.681 uV 15:44:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -141.343 uVs 15:44:10 Integrated Voltage Measurement... 15:44:19 Voltage Offset Correction: v_off_ini = 808.099 uV, v_off_fin = 819.310 uV 15:44:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 137.679 uVs 15:44:23 I1X Average and RMS of Measurements VT = -116.020 +- 16.705 uVs I1X = 23.205 +- 3.341 uTm 15:44:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:44:30 Integrated Voltage Measurement... 15:44:39 Voltage Offset Correction: v_off_ini = 808.091 uV, v_off_fin = 784.001 uV 15:44:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -275.400 uVs 15:44:39 Integrated Voltage Measurement... 15:44:48 Voltage Offset Correction: v_off_ini = 798.932 uV, v_off_fin = 823.973 uV 15:44:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 297.927 uVs 15:44:48 Integrated Voltage Measurement... 15:44:57 Voltage Offset Correction: v_off_ini = 817.658 uV, v_off_fin = 798.320 uV 15:44:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -280.444 uVs 15:44:57 Integrated Voltage Measurement... 15:45:06 Voltage Offset Correction: v_off_ini = 805.826 uV, v_off_fin = 837.834 uV 15:45:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 316.247 uVs 15:45:06 Integrated Voltage Measurement... 15:45:15 Voltage Offset Correction: v_off_ini = 809.562 uV, v_off_fin = 793.393 uV 15:45:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -300.594 uVs 15:45:15 Integrated Voltage Measurement... 15:45:24 Voltage Offset Correction: v_off_ini = 811.995 uV, v_off_fin = 825.448 uV 15:45:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 291.412 uVs 15:45:28 I1Y Average and RMS of Measurements VT = -293.671 +- 5.046 uVs I1Y = -58.734 +- 1.009 uTm 15:45:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:45:36 Integrated Voltage Measurement... 15:45:45 Voltage Offset Correction: v_off_ini = 819.126 uV, v_off_fin = 834.919 uV 15:45:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.268 uVs 15:45:45 Integrated Voltage Measurement... 15:45:54 Voltage Offset Correction: v_off_ini = 851.912 uV, v_off_fin = 858.702 uV 15:45:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.252 uVs 15:45:54 Integrated Voltage Measurement... 15:46:03 Voltage Offset Correction: v_off_ini = 837.962 uV, v_off_fin = 811.427 uV 15:46:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -168.214 uVs 15:46:03 Integrated Voltage Measurement... 15:46:12 Voltage Offset Correction: v_off_ini = 854.281 uV, v_off_fin = 836.178 uV 15:46:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.288 uVs 15:46:12 Integrated Voltage Measurement... 15:46:21 Voltage Offset Correction: v_off_ini = 833.621 uV, v_off_fin = 851.387 uV 15:46:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.255 uVs 15:46:21 Integrated Voltage Measurement... 15:46:30 Voltage Offset Correction: v_off_ini = 858.153 uV, v_off_fin = 853.547 uV 15:46:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.857 uVs 15:46:32 I2X Average and RMS of Measurements VT = -127.189 +- 13.568 uVs I2X = 11.701 +- 1.248 uTm^2 15:46:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:46:40 Integrated Voltage Measurement... 15:46:49 Voltage Offset Correction: v_off_ini = 865.491 uV, v_off_fin = 883.120 uV 15:46:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -381.454 uVs 15:46:49 Integrated Voltage Measurement... 15:46:58 Voltage Offset Correction: v_off_ini = 855.474 uV, v_off_fin = 855.631 uV 15:46:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 387.919 uVs 15:46:58 Integrated Voltage Measurement... 15:47:07 Voltage Offset Correction: v_off_ini = 862.395 uV, v_off_fin = 869.893 uV 15:47:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -358.913 uVs 15:47:07 Integrated Voltage Measurement... 15:47:16 Voltage Offset Correction: v_off_ini = 865.156 uV, v_off_fin = 870.798 uV 15:47:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 367.007 uVs 15:47:16 Integrated Voltage Measurement... 15:47:25 Voltage Offset Correction: v_off_ini = 873.167 uV, v_off_fin = 864.000 uV 15:47:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -355.926 uVs 15:47:25 Integrated Voltage Measurement... 15:47:34 Voltage Offset Correction: v_off_ini = 860.831 uV, v_off_fin = 867.009 uV 15:47:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 358.600 uVs 15:47:36 I2Y Average and RMS of Measurements VT = -368.303 +- 11.816 uVs I2Y = -33.884 +- 1.087 uTm^2 15:47:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:47:50 Integrated Voltage Measurement... 15:47:59 Voltage Offset Correction: v_off_ini = 868.902 uV, v_off_fin = 848.833 uV 15:47:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.126 uVs 15:47:59 Integrated Voltage Measurement... 15:48:08 Voltage Offset Correction: v_off_ini = 893.325 uV, v_off_fin = 897.415 uV 15:48:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.521 uVs 15:48:08 Integrated Voltage Measurement... 15:48:17 Voltage Offset Correction: v_off_ini = 899.902 uV, v_off_fin = 910.439 uV 15:48:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.231 uVs 15:48:17 Integrated Voltage Measurement... 15:48:26 Voltage Offset Correction: v_off_ini = 917.055 uV, v_off_fin = 902.043 uV 15:48:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.438 uVs 15:48:26 Integrated Voltage Measurement... 15:48:35 Voltage Offset Correction: v_off_ini = 903.052 uV, v_off_fin = 906.442 uV 15:48:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.415 uVs 15:48:35 Integrated Voltage Measurement... 15:48:44 Voltage Offset Correction: v_off_ini = 908.270 uV, v_off_fin = 928.955 uV 15:48:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.188 uVs 15:48:49 I1X Average and RMS of Measurements VT = -96.320 +- 6.375 uVs I1X = 19.264 +- 1.275 uTm 15:48:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:48:55 Integrated Voltage Measurement... 15:49:04 Voltage Offset Correction: v_off_ini = 906.265 uV, v_off_fin = 894.992 uV 15:49:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -298.461 uVs 15:49:04 Integrated Voltage Measurement... 15:49:13 Voltage Offset Correction: v_off_ini = 911.704 uV, v_off_fin = 905.495 uV 15:49:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 299.902 uVs 15:49:13 Integrated Voltage Measurement... 15:49:22 Voltage Offset Correction: v_off_ini = 891.864 uV, v_off_fin = 885.947 uV 15:49:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -270.556 uVs 15:49:22 Integrated Voltage Measurement... 15:49:31 Voltage Offset Correction: v_off_ini = 886.977 uV, v_off_fin = 922.478 uV 15:49:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 292.141 uVs 15:49:31 Integrated Voltage Measurement... 15:49:40 Voltage Offset Correction: v_off_ini = 908.694 uV, v_off_fin = 884.780 uV 15:49:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -284.845 uVs 15:49:40 Integrated Voltage Measurement... 15:49:49 Voltage Offset Correction: v_off_ini = 907.254 uV, v_off_fin = 920.830 uV 15:49:49 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 287.785 uVs 15:49:53 I1Y Average and RMS of Measurements VT = -288.948 +- 7.514 uVs I1Y = -57.791 +- 1.503 uTm 15:49:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:50:01 Integrated Voltage Measurement... 15:50:10 Voltage Offset Correction: v_off_ini = 917.573 uV, v_off_fin = 905.010 uV 15:50:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.032 uVs 15:50:10 Integrated Voltage Measurement... 15:50:19 Voltage Offset Correction: v_off_ini = 898.112 uV, v_off_fin = 897.211 uV 15:50:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 110.012 uVs 15:50:19 Integrated Voltage Measurement... 15:50:28 Voltage Offset Correction: v_off_ini = 923.535 uV, v_off_fin = 931.904 uV 15:50:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.355 uVs 15:50:28 Integrated Voltage Measurement... 15:50:37 Voltage Offset Correction: v_off_ini = 923.700 uV, v_off_fin = 945.167 uV 15:50:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.316 uVs 15:50:37 Integrated Voltage Measurement... 15:50:46 Voltage Offset Correction: v_off_ini = 921.131 uV, v_off_fin = 935.456 uV 15:50:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -131.644 uVs 15:50:46 Integrated Voltage Measurement... 15:50:55 Voltage Offset Correction: v_off_ini = 952.135 uV, v_off_fin = 951.501 uV 15:50:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.395 uVs 15:50:58 I2X Average and RMS of Measurements VT = -112.459 +- 6.532 uVs I2X = 10.346 +- 0.601 uTm^2 15:51:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:51:05 Integrated Voltage Measurement... 15:51:14 Voltage Offset Correction: v_off_ini = 944.894 uV, v_off_fin = 925.200 uV 15:51:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -387.666 uVs 15:51:14 Integrated Voltage Measurement... 15:51:23 Voltage Offset Correction: v_off_ini = 952.466 uV, v_off_fin = 954.567 uV 15:51:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 381.127 uVs 15:51:24 Integrated Voltage Measurement... 15:51:32 Voltage Offset Correction: v_off_ini = 952.414 uV, v_off_fin = 958.308 uV 15:51:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -397.533 uVs 15:51:33 Integrated Voltage Measurement... 15:51:42 Voltage Offset Correction: v_off_ini = 956.954 uV, v_off_fin = 960.069 uV 15:51:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 386.337 uVs 15:51:42 Integrated Voltage Measurement... 15:51:51 Voltage Offset Correction: v_off_ini = 938.159 uV, v_off_fin = 924.209 uV 15:51:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -357.223 uVs 15:51:51 Integrated Voltage Measurement... 15:52:00 Voltage Offset Correction: v_off_ini = 915.217 uV, v_off_fin = 956.188 uV 15:52:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 350.899 uVs 15:52:02 I2Y Average and RMS of Measurements VT = -376.797 +- 16.369 uVs I2Y = -34.665 +- 1.506 uTm^2 15:52:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:52:16 Integrated Voltage Measurement... 15:52:25 Voltage Offset Correction: v_off_ini = 969.965 uV, v_off_fin = 947.492 uV 15:52:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -123.193 uVs 15:52:25 Integrated Voltage Measurement... 15:52:34 Voltage Offset Correction: v_off_ini = 969.429 uV, v_off_fin = 969.965 uV 15:52:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.002 uVs 15:52:34 Integrated Voltage Measurement... 15:52:43 Voltage Offset Correction: v_off_ini = 962.009 uV, v_off_fin = 971.663 uV 15:52:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -127.220 uVs 15:52:43 Integrated Voltage Measurement... 15:52:52 Voltage Offset Correction: v_off_ini = 969.092 uV, v_off_fin = 974.959 uV 15:52:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.584 uVs 15:52:52 Integrated Voltage Measurement... 15:53:01 Voltage Offset Correction: v_off_ini = 971.470 uV, v_off_fin = 962.067 uV 15:53:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.948 uVs 15:53:01 Integrated Voltage Measurement... 15:53:10 Voltage Offset Correction: v_off_ini = 967.957 uV, v_off_fin = 973.836 uV 15:53:10 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.043 uVs 15:53:14 I1X Average and RMS of Measurements VT = -106.332 +- 12.341 uVs I1X = 21.267 +- 2.468 uTm 15:53:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:53:21 Integrated Voltage Measurement... 15:53:30 Voltage Offset Correction: v_off_ini = 970.552 uV, v_off_fin = 974.587 uV 15:53:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -293.433 uVs 15:53:30 Integrated Voltage Measurement... 15:53:39 Voltage Offset Correction: v_off_ini = 958.211 uV, v_off_fin = 971.057 uV 15:53:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 288.187 uVs 15:53:39 Integrated Voltage Measurement... 15:53:48 Voltage Offset Correction: v_off_ini = 966.246 uV, v_off_fin = 959.201 uV 15:53:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -287.579 uVs 15:53:48 Integrated Voltage Measurement... 15:53:57 Voltage Offset Correction: v_off_ini = 955.796 uV, v_off_fin = 985.083 uV 15:53:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 282.423 uVs 15:53:57 Integrated Voltage Measurement... 15:54:06 Voltage Offset Correction: v_off_ini = 995.258 uV, v_off_fin = 976.855 uV 15:54:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -293.868 uVs 15:54:06 Integrated Voltage Measurement... 15:54:15 Voltage Offset Correction: v_off_ini = 977.148 uV, v_off_fin = 986.290 uV 15:54:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 293.433 uVs 15:54:19 I1Y Average and RMS of Measurements VT = -289.821 +- 3.600 uVs I1Y = -57.961 +- 0.719 uTm 15:54:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:54:27 Integrated Voltage Measurement... 15:54:36 Voltage Offset Correction: v_off_ini = 988.515 uV, v_off_fin = 990.304 uV 15:54:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.052 uVs 15:54:36 Integrated Voltage Measurement... 15:54:45 Voltage Offset Correction: v_off_ini = 980.435 uV, v_off_fin = 1013.609 uV 15:54:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 133.063 uVs 15:54:45 Integrated Voltage Measurement... 15:54:54 Voltage Offset Correction: v_off_ini = 1012.556 uV, v_off_fin = 1001.899 uV 15:54:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -137.073 uVs 15:54:54 Integrated Voltage Measurement... 15:55:03 Voltage Offset Correction: v_off_ini = 1024.688 uV, v_off_fin = 1016.342 uV 15:55:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 138.472 uVs 15:55:03 Integrated Voltage Measurement... 15:55:12 Voltage Offset Correction: v_off_ini = 1037.014 uV, v_off_fin = 1018.468 uV 15:55:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -121.868 uVs 15:55:12 Integrated Voltage Measurement... 15:55:21 Voltage Offset Correction: v_off_ini = 1030.469 uV, v_off_fin = 1032.875 uV 15:55:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.282 uVs 15:55:23 I2X Average and RMS of Measurements VT = -125.968 +- 10.348 uVs I2X = 11.589 +- 0.952 uTm^2 15:55:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:55:31 Integrated Voltage Measurement... 15:55:40 Voltage Offset Correction: v_off_ini = 1011.018 uV, v_off_fin = 999.384 uV 15:55:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -372.555 uVs 15:55:40 Integrated Voltage Measurement... 15:55:49 Voltage Offset Correction: v_off_ini = 982.032 uV, v_off_fin = 999.479 uV 15:55:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 378.161 uVs 15:55:49 Integrated Voltage Measurement... 15:55:58 Voltage Offset Correction: v_off_ini = 1023.357 uV, v_off_fin = 1027.297 uV 15:55:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -356.785 uVs 15:55:58 Integrated Voltage Measurement... 15:56:07 Voltage Offset Correction: v_off_ini = 1029.172 uV, v_off_fin = 1027.895 uV 15:56:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 351.396 uVs 15:56:07 Integrated Voltage Measurement... 15:56:16 Voltage Offset Correction: v_off_ini = 1029.019 uV, v_off_fin = 1016.669 uV 15:56:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -371.828 uVs 15:56:16 Integrated Voltage Measurement... 15:56:25 Voltage Offset Correction: v_off_ini = 1044.448 uV, v_off_fin = 1036.676 uV 15:56:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 356.208 uVs 15:56:28 I2Y Average and RMS of Measurements VT = -364.489 +- 8.689 uVs I2Y = -33.533 +- 0.799 uTm^2 15:56:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:56:42 Integrated Voltage Measurement... 15:56:51 Voltage Offset Correction: v_off_ini = 1012.184 uV, v_off_fin = 999.013 uV 15:56:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -96.417 uVs 15:56:51 Integrated Voltage Measurement... 15:57:00 Voltage Offset Correction: v_off_ini = 1003.387 uV, v_off_fin = 1025.809 uV 15:57:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.539 uVs 15:57:00 Integrated Voltage Measurement... 15:57:09 Voltage Offset Correction: v_off_ini = 1034.854 uV, v_off_fin = 1018.072 uV 15:57:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.853 uVs 15:57:09 Integrated Voltage Measurement... 15:57:18 Voltage Offset Correction: v_off_ini = 995.662 uV, v_off_fin = 1012.730 uV 15:57:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.227 uVs 15:57:18 Integrated Voltage Measurement... 15:57:27 Voltage Offset Correction: v_off_ini = 1007.425 uV, v_off_fin = 997.457 uV 15:57:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.882 uVs 15:57:27 Integrated Voltage Measurement... 15:57:36 Voltage Offset Correction: v_off_ini = 1016.877 uV, v_off_fin = 1006.539 uV 15:57:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.258 uVs 15:57:40 I1X Average and RMS of Measurements VT = -105.363 +- 1.818 uVs I1X = 21.073 +- 0.364 uTm 15:57:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:57:47 Integrated Voltage Measurement... 15:57:55 Voltage Offset Correction: v_off_ini = 999.994 uV, v_off_fin = 1004.160 uV 15:57:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -292.501 uVs 15:57:56 Integrated Voltage Measurement... 15:58:04 Voltage Offset Correction: v_off_ini = 1002.776 uV, v_off_fin = 1017.146 uV 15:58:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 297.813 uVs 15:58:05 Integrated Voltage Measurement... 15:58:13 Voltage Offset Correction: v_off_ini = 1006.081 uV, v_off_fin = 994.241 uV 15:58:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -302.891 uVs 15:58:14 Integrated Voltage Measurement... 15:58:23 Voltage Offset Correction: v_off_ini = 1015.747 uV, v_off_fin = 1020.931 uV 15:58:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 287.856 uVs 15:58:23 Integrated Voltage Measurement... 15:58:32 Voltage Offset Correction: v_off_ini = 999.541 uV, v_off_fin = 996.698 uV 15:58:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -302.421 uVs 15:58:32 Integrated Voltage Measurement... 15:58:40 Voltage Offset Correction: v_off_ini = 1022.532 uV, v_off_fin = 1026.899 uV 15:58:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 281.304 uVs 15:58:45 I1Y Average and RMS of Measurements VT = -294.131 +- 1.607 uVs I1Y = -58.821 +- 0.324 uTm 15:58:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:58:52 Integrated Voltage Measurement... 15:59:01 Voltage Offset Correction: v_off_ini = 1022.536 uV, v_off_fin = 1034.785 uV 15:59:01 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -126.968 uVs 15:59:01 Integrated Voltage Measurement... 15:59:10 Voltage Offset Correction: v_off_ini = 1026.023 uV, v_off_fin = 1022.660 uV 15:59:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 131.673 uVs 15:59:11 Integrated Voltage Measurement... 15:59:19 Voltage Offset Correction: v_off_ini = 1023.874 uV, v_off_fin = 1018.745 uV 15:59:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.443 uVs 15:59:20 Integrated Voltage Measurement... 15:59:29 Voltage Offset Correction: v_off_ini = 1051.233 uV, v_off_fin = 1057.223 uV 15:59:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.603 uVs 15:59:29 Integrated Voltage Measurement... 15:59:38 Voltage Offset Correction: v_off_ini = 1051.794 uV, v_off_fin = 1046.137 uV 15:59:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.564 uVs 15:59:38 Integrated Voltage Measurement... 15:59:47 Voltage Offset Correction: v_off_ini = 1046.569 uV, v_off_fin = 1046.649 uV 15:59:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.149 uVs 15:59:49 I2X Average and RMS of Measurements VT = -114.900 +- 10.885 uVs I2X = 10.571 +- 1.001 uTm^2 15:59:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:59:57 Integrated Voltage Measurement... 16:00:06 Voltage Offset Correction: v_off_ini = 1065.307 uV, v_off_fin = 1071.286 uV 16:00:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -370.887 uVs 16:00:06 Integrated Voltage Measurement... 16:00:15 Voltage Offset Correction: v_off_ini = 1087.531 uV, v_off_fin = 1082.176 uV 16:00:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 372.948 uVs 16:00:15 Integrated Voltage Measurement... 16:00:24 Voltage Offset Correction: v_off_ini = 1095.799 uV, v_off_fin = 1109.515 uV 16:00:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -396.304 uVs 16:00:24 Integrated Voltage Measurement... 16:00:33 Voltage Offset Correction: v_off_ini = 1106.338 uV, v_off_fin = 1098.510 uV 16:00:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 354.956 uVs 16:00:33 Integrated Voltage Measurement... 16:00:42 Voltage Offset Correction: v_off_ini = 1098.570 uV, v_off_fin = 1088.382 uV 16:00:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -351.575 uVs 16:00:42 Integrated Voltage Measurement... 16:00:51 Voltage Offset Correction: v_off_ini = 1108.366 uV, v_off_fin = 1110.947 uV 16:00:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 348.174 uVs 16:00:53 I2Y Average and RMS of Measurements VT = -365.807 +- 11.368 uVs I2Y = -33.654 +- 1.046 uTm^2 16:01:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:01:07 Integrated Voltage Measurement... 16:01:16 Voltage Offset Correction: v_off_ini = 1144.013 uV, v_off_fin = 1144.999 uV 16:01:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -113.021 uVs 16:01:16 Integrated Voltage Measurement... 16:01:25 Voltage Offset Correction: v_off_ini = 1135.051 uV, v_off_fin = 1137.115 uV 16:01:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.118 uVs 16:01:25 Integrated Voltage Measurement... 16:01:34 Voltage Offset Correction: v_off_ini = 1142.641 uV, v_off_fin = 1109.742 uV 16:01:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.740 uVs 16:01:34 Integrated Voltage Measurement... 16:01:43 Voltage Offset Correction: v_off_ini = 1150.795 uV, v_off_fin = 1150.341 uV 16:01:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.595 uVs 16:01:43 Integrated Voltage Measurement... 16:01:52 Voltage Offset Correction: v_off_ini = 1156.356 uV, v_off_fin = 1111.073 uV 16:01:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -133.730 uVs 16:01:52 Integrated Voltage Measurement... 16:02:01 Voltage Offset Correction: v_off_ini = 1099.667 uV, v_off_fin = 1117.543 uV 16:02:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.441 uVs 16:02:05 I1X Average and RMS of Measurements VT = -106.941 +- 16.220 uVs I1X = 21.389 +- 3.245 uTm 16:02:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:02:12 Integrated Voltage Measurement... 16:02:21 Voltage Offset Correction: v_off_ini = 1126.718 uV, v_off_fin = 1118.436 uV 16:02:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -278.140 uVs 16:02:21 Integrated Voltage Measurement... 16:02:30 Voltage Offset Correction: v_off_ini = 1123.677 uV, v_off_fin = 1121.725 uV 16:02:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 281.017 uVs 16:02:30 Integrated Voltage Measurement... 16:02:39 Voltage Offset Correction: v_off_ini = 1111.039 uV, v_off_fin = 1097.530 uV 16:02:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -311.479 uVs 16:02:39 Integrated Voltage Measurement... 16:02:48 Voltage Offset Correction: v_off_ini = 1110.904 uV, v_off_fin = 1134.245 uV 16:02:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 291.943 uVs 16:02:48 Integrated Voltage Measurement... 16:02:57 Voltage Offset Correction: v_off_ini = 1133.509 uV, v_off_fin = 1111.529 uV 16:02:57 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5002 mm dXY = 0.5000 mm VT = -300.194 uVs 16:02:57 Integrated Voltage Measurement... 16:03:06 Voltage Offset Correction: v_off_ini = 1114.861 uV, v_off_fin = 1146.921 uV 16:03:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 295.866 uVs 16:03:10 I1Y Average and RMS of Measurements VT = -293.106 +- 9.683 uVs I1Y = -58.617 +- 1.933 uTm 16:03:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:03:18 Integrated Voltage Measurement... 16:03:27 Voltage Offset Correction: v_off_ini = 1138.143 uV, v_off_fin = 1130.988 uV 16:03:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.822 uVs 16:03:27 Integrated Voltage Measurement... 16:03:36 Voltage Offset Correction: v_off_ini = 1148.412 uV, v_off_fin = 1136.729 uV 16:03:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 136.614 uVs 16:03:36 Integrated Voltage Measurement... 16:03:45 Voltage Offset Correction: v_off_ini = 1131.905 uV, v_off_fin = 1126.392 uV 16:03:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -91.557 uVs 16:03:45 Integrated Voltage Measurement... 16:03:54 Voltage Offset Correction: v_off_ini = 1128.444 uV, v_off_fin = 1132.574 uV 16:03:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 141.661 uVs 16:03:54 Integrated Voltage Measurement... 16:04:03 Voltage Offset Correction: v_off_ini = 1142.060 uV, v_off_fin = 1148.002 uV 16:04:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.488 uVs 16:04:03 Integrated Voltage Measurement... 16:04:12 Voltage Offset Correction: v_off_ini = 1162.960 uV, v_off_fin = 1157.981 uV 16:04:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.373 uVs 16:04:15 I2X Average and RMS of Measurements VT = -114.253 +- 4.142 uVs I2X = 10.511 +- 0.381 uTm^2 16:04:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:04:22 Integrated Voltage Measurement... 16:04:31 Voltage Offset Correction: v_off_ini = 1184.956 uV, v_off_fin = 1181.191 uV 16:04:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -363.222 uVs 16:04:31 Integrated Voltage Measurement... 16:04:40 Voltage Offset Correction: v_off_ini = 1171.641 uV, v_off_fin = 1176.554 uV 16:04:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 384.539 uVs 16:04:40 Integrated Voltage Measurement... 16:04:49 Voltage Offset Correction: v_off_ini = 1172.787 uV, v_off_fin = 1161.220 uV 16:04:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -358.085 uVs 16:04:49 Integrated Voltage Measurement... 16:04:58 Voltage Offset Correction: v_off_ini = 1157.123 uV, v_off_fin = 1145.712 uV 16:04:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 351.395 uVs 16:04:59 Integrated Voltage Measurement... 16:05:07 Voltage Offset Correction: v_off_ini = 1175.728 uV, v_off_fin = 1161.630 uV 16:05:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -362.767 uVs 16:05:08 Integrated Voltage Measurement... 16:05:17 Voltage Offset Correction: v_off_ini = 1170.069 uV, v_off_fin = 1177.736 uV 16:05:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 368.715 uVs 16:05:19 I2Y Average and RMS of Measurements VT = -364.787 +- 7.843 uVs I2Y = -33.560 +- 0.722 uTm^2 16:05:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:05:33 Integrated Voltage Measurement... 16:05:42 Voltage Offset Correction: v_off_ini = 1168.669 uV, v_off_fin = 1171.002 uV 16:05:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -104.430 uVs 16:05:42 Integrated Voltage Measurement... 16:05:51 Voltage Offset Correction: v_off_ini = 1177.481 uV, v_off_fin = 1185.968 uV 16:05:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.984 uVs 16:05:51 Integrated Voltage Measurement... 16:06:00 Voltage Offset Correction: v_off_ini = 1164.273 uV, v_off_fin = 1158.124 uV 16:06:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.455 uVs 16:06:00 Integrated Voltage Measurement... 16:06:09 Voltage Offset Correction: v_off_ini = 1186.781 uV, v_off_fin = 1193.089 uV 16:06:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.639 uVs 16:06:09 Integrated Voltage Measurement... 16:06:18 Voltage Offset Correction: v_off_ini = 1198.358 uV, v_off_fin = 1194.998 uV 16:06:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.008 uVs 16:06:18 Integrated Voltage Measurement... 16:06:27 Voltage Offset Correction: v_off_ini = 1200.799 uV, v_off_fin = 1202.290 uV 16:06:27 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.773 uVs 16:06:31 I1X Average and RMS of Measurements VT = -101.715 +- 11.623 uVs I1X = 20.344 +- 2.324 uTm 16:06:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:06:38 Integrated Voltage Measurement... 16:06:47 Voltage Offset Correction: v_off_ini = 1182.001 uV, v_off_fin = 1167.070 uV 16:06:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -305.060 uVs 16:06:47 Integrated Voltage Measurement... 16:06:56 Voltage Offset Correction: v_off_ini = 1212.699 uV, v_off_fin = 1206.340 uV 16:06:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 303.029 uVs 16:06:56 Integrated Voltage Measurement... 16:07:05 Voltage Offset Correction: v_off_ini = 1204.539 uV, v_off_fin = 1198.468 uV 16:07:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -288.062 uVs 16:07:05 Integrated Voltage Measurement... 16:07:14 Voltage Offset Correction: v_off_ini = 1203.558 uV, v_off_fin = 1209.526 uV 16:07:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 294.134 uVs 16:07:14 Integrated Voltage Measurement... 16:07:23 Voltage Offset Correction: v_off_ini = 1207.157 uV, v_off_fin = 1202.773 uV 16:07:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -280.453 uVs 16:07:23 Integrated Voltage Measurement... 16:07:32 Voltage Offset Correction: v_off_ini = 1213.972 uV, v_off_fin = 1216.600 uV 16:07:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 287.057 uVs 16:07:36 I1Y Average and RMS of Measurements VT = -292.966 +- 8.388 uVs I1Y = -58.588 +- 1.679 uTm 16:07:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:07:44 Integrated Voltage Measurement... 16:07:53 Voltage Offset Correction: v_off_ini = 1209.198 uV, v_off_fin = 1217.699 uV 16:07:53 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -72.960 uVs 16:07:53 Integrated Voltage Measurement... 16:08:02 Voltage Offset Correction: v_off_ini = 1220.408 uV, v_off_fin = 1232.406 uV 16:08:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.867 uVs 16:08:02 Integrated Voltage Measurement... 16:08:11 Voltage Offset Correction: v_off_ini = 1219.605 uV, v_off_fin = 1192.013 uV 16:08:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.238 uVs 16:08:11 Integrated Voltage Measurement... 16:08:20 Voltage Offset Correction: v_off_ini = 1207.470 uV, v_off_fin = 1228.742 uV 16:08:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.002 uVs 16:08:20 Integrated Voltage Measurement... 16:08:29 Voltage Offset Correction: v_off_ini = 1225.755 uV, v_off_fin = 1215.591 uV 16:08:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -117.341 uVs 16:08:29 Integrated Voltage Measurement... 16:08:38 Voltage Offset Correction: v_off_ini = 1209.444 uV, v_off_fin = 1222.253 uV 16:08:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 131.727 uVs 16:08:40 I2X Average and RMS of Measurements VT = -109.689 +- 14.843 uVs I2X = 10.091 +- 1.366 uTm^2 16:08:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:08:48 Integrated Voltage Measurement... 16:08:57 Voltage Offset Correction: v_off_ini = 1233.104 uV, v_off_fin = 1234.392 uV 16:08:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -347.888 uVs 16:08:57 Integrated Voltage Measurement... 16:09:06 Voltage Offset Correction: v_off_ini = 1224.685 uV, v_off_fin = 1231.486 uV 16:09:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 369.886 uVs 16:09:06 Integrated Voltage Measurement... 16:09:15 Voltage Offset Correction: v_off_ini = 1216.062 uV, v_off_fin = 1222.129 uV 16:09:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -347.737 uVs 16:09:15 Integrated Voltage Measurement... 16:09:24 Voltage Offset Correction: v_off_ini = 1242.852 uV, v_off_fin = 1246.004 uV 16:09:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 360.542 uVs 16:09:24 Integrated Voltage Measurement... 16:09:33 Voltage Offset Correction: v_off_ini = 1237.442 uV, v_off_fin = 1229.666 uV 16:09:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.218 uVs 16:09:33 Integrated Voltage Measurement... 16:09:42 Voltage Offset Correction: v_off_ini = 1230.866 uV, v_off_fin = 1238.036 uV 16:09:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 344.905 uVs 16:09:45 I2Y Average and RMS of Measurements VT = -348.863 +- 10.992 uVs I2Y = -32.095 +- 1.011 uTm^2 16:09:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:09:59 Integrated Voltage Measurement... 16:10:07 Voltage Offset Correction: v_off_ini = 1208.365 uV, v_off_fin = 1216.865 uV 16:10:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -103.947 uVs 16:10:08 Integrated Voltage Measurement... 16:10:16 Voltage Offset Correction: v_off_ini = 1261.092 uV, v_off_fin = 1276.666 uV 16:10:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.152 uVs 16:10:17 Integrated Voltage Measurement... 16:10:26 Voltage Offset Correction: v_off_ini = 1260.970 uV, v_off_fin = 1263.229 uV 16:10:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.634 uVs 16:10:26 Integrated Voltage Measurement... 16:10:35 Voltage Offset Correction: v_off_ini = 1266.227 uV, v_off_fin = 1281.817 uV 16:10:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.524 uVs 16:10:35 Integrated Voltage Measurement... 16:10:44 Voltage Offset Correction: v_off_ini = 1278.391 uV, v_off_fin = 1288.584 uV 16:10:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.449 uVs 16:10:44 Integrated Voltage Measurement... 16:10:53 Voltage Offset Correction: v_off_ini = 1279.725 uV, v_off_fin = 1280.037 uV 16:10:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.908 uVs 16:10:57 I1X Average and RMS of Measurements VT = -88.102 +- 8.460 uVs I1X = 17.621 +- 1.693 uTm 16:10:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:11:04 Integrated Voltage Measurement... 16:11:12 Voltage Offset Correction: v_off_ini = 1276.737 uV, v_off_fin = 1264.102 uV 16:11:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -277.832 uVs 16:11:13 Integrated Voltage Measurement... 16:11:21 Voltage Offset Correction: v_off_ini = 1292.268 uV, v_off_fin = 1292.618 uV 16:11:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 311.104 uVs 16:11:22 Integrated Voltage Measurement... 16:11:30 Voltage Offset Correction: v_off_ini = 1283.666 uV, v_off_fin = 1279.469 uV 16:11:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -283.273 uVs 16:11:31 Integrated Voltage Measurement... 16:11:39 Voltage Offset Correction: v_off_ini = 1287.937 uV, v_off_fin = 1289.699 uV 16:11:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 283.164 uVs 16:11:40 Integrated Voltage Measurement... 16:11:48 Voltage Offset Correction: v_off_ini = 1287.634 uV, v_off_fin = 1280.277 uV 16:11:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -292.823 uVs 16:11:49 Integrated Voltage Measurement... 16:11:57 Voltage Offset Correction: v_off_ini = 1281.002 uV, v_off_fin = 1302.116 uV 16:11:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 298.887 uVs 16:12:02 I1Y Average and RMS of Measurements VT = -291.180 +- 5.658 uVs I1Y = -58.238 +- 1.133 uTm 16:12:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:12:09 Integrated Voltage Measurement... 16:12:18 Voltage Offset Correction: v_off_ini = 1281.348 uV, v_off_fin = 1257.438 uV 16:12:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.944 uVs 16:12:18 Integrated Voltage Measurement... 16:12:27 Voltage Offset Correction: v_off_ini = 1267.192 uV, v_off_fin = 1290.138 uV 16:12:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.681 uVs 16:12:27 Integrated Voltage Measurement... 16:12:36 Voltage Offset Correction: v_off_ini = 1281.214 uV, v_off_fin = 1302.371 uV 16:12:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.414 uVs 16:12:37 Integrated Voltage Measurement... 16:12:45 Voltage Offset Correction: v_off_ini = 1308.666 uV, v_off_fin = 1287.442 uV 16:12:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.463 uVs 16:12:46 Integrated Voltage Measurement... 16:12:55 Voltage Offset Correction: v_off_ini = 1305.781 uV, v_off_fin = 1292.832 uV 16:12:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.845 uVs 16:12:55 Integrated Voltage Measurement... 16:13:04 Voltage Offset Correction: v_off_ini = 1306.754 uV, v_off_fin = 1314.917 uV 16:13:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.886 uVs 16:13:06 I2X Average and RMS of Measurements VT = -96.372 +- 16.087 uVs I2X = 8.866 +- 1.480 uTm^2 16:13:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:13:14 Integrated Voltage Measurement... 16:13:23 Voltage Offset Correction: v_off_ini = 1327.691 uV, v_off_fin = 1317.975 uV 16:13:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -377.919 uVs 16:13:23 Integrated Voltage Measurement... 16:13:32 Voltage Offset Correction: v_off_ini = 1336.731 uV, v_off_fin = 1343.987 uV 16:13:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 345.953 uVs 16:13:32 Integrated Voltage Measurement... 16:13:41 Voltage Offset Correction: v_off_ini = 1347.877 uV, v_off_fin = 1332.485 uV 16:13:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -364.448 uVs 16:13:41 Integrated Voltage Measurement... 16:13:50 Voltage Offset Correction: v_off_ini = 1332.194 uV, v_off_fin = 1333.954 uV 16:13:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 376.910 uVs 16:13:50 Integrated Voltage Measurement... 16:13:59 Voltage Offset Correction: v_off_ini = 1330.325 uV, v_off_fin = 1336.013 uV 16:13:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -349.651 uVs 16:13:59 Integrated Voltage Measurement... 16:14:08 Voltage Offset Correction: v_off_ini = 1342.669 uV, v_off_fin = 1325.906 uV 16:14:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 368.863 uVs 16:14:10 I2Y Average and RMS of Measurements VT = -363.957 +- 4.877 uVs I2Y = -33.485 +- 0.448 uTm^2 16:14:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:14:24 Integrated Voltage Measurement... 16:14:33 Voltage Offset Correction: v_off_ini = 1345.667 uV, v_off_fin = 1335.803 uV 16:14:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -109.765 uVs 16:14:33 Integrated Voltage Measurement... 16:14:42 Voltage Offset Correction: v_off_ini = 1345.033 uV, v_off_fin = 1342.430 uV 16:14:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.522 uVs 16:14:42 Integrated Voltage Measurement... 16:14:51 Voltage Offset Correction: v_off_ini = 1350.167 uV, v_off_fin = 1344.380 uV 16:14:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.168 uVs 16:14:51 Integrated Voltage Measurement... 16:15:00 Voltage Offset Correction: v_off_ini = 1346.059 uV, v_off_fin = 1344.840 uV 16:15:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.277 uVs 16:15:00 Integrated Voltage Measurement... 16:15:09 Voltage Offset Correction: v_off_ini = 1374.675 uV, v_off_fin = 1378.703 uV 16:15:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.839 uVs 16:15:09 Integrated Voltage Measurement... 16:15:18 Voltage Offset Correction: v_off_ini = 1376.560 uV, v_off_fin = 1380.090 uV 16:15:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.706 uVs 16:15:22 I1X Average and RMS of Measurements VT = -93.213 +- 11.648 uVs I1X = 18.643 +- 2.330 uTm 16:15:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:15:29 Integrated Voltage Measurement... 16:15:38 Voltage Offset Correction: v_off_ini = 1348.237 uV, v_off_fin = 1346.488 uV 16:15:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -269.353 uVs 16:15:38 Integrated Voltage Measurement... 16:15:47 Voltage Offset Correction: v_off_ini = 1332.631 uV, v_off_fin = 1360.001 uV 16:15:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 297.308 uVs 16:15:47 Integrated Voltage Measurement... 16:15:56 Voltage Offset Correction: v_off_ini = 1380.853 uV, v_off_fin = 1359.914 uV 16:15:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -298.534 uVs 16:15:56 Integrated Voltage Measurement... 16:16:05 Voltage Offset Correction: v_off_ini = 1385.294 uV, v_off_fin = 1392.876 uV 16:16:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 286.393 uVs 16:16:05 Integrated Voltage Measurement... 16:16:14 Voltage Offset Correction: v_off_ini = 1380.096 uV, v_off_fin = 1369.937 uV 16:16:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -283.906 uVs 16:16:14 Integrated Voltage Measurement... 16:16:23 Voltage Offset Correction: v_off_ini = 1386.515 uV, v_off_fin = 1382.007 uV 16:16:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 284.166 uVs 16:16:27 I1Y Average and RMS of Measurements VT = -286.610 +- 4.149 uVs I1Y = -57.316 +- 0.828 uTm 16:16:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:16:35 Integrated Voltage Measurement... 16:16:44 Voltage Offset Correction: v_off_ini = 1371.431 uV, v_off_fin = 1365.734 uV 16:16:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.602 uVs 16:16:44 Integrated Voltage Measurement... 16:16:53 Voltage Offset Correction: v_off_ini = 1364.833 uV, v_off_fin = 1373.360 uV 16:16:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.537 uVs 16:16:53 Integrated Voltage Measurement... 16:17:02 Voltage Offset Correction: v_off_ini = 1361.105 uV, v_off_fin = 1349.739 uV 16:17:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.826 uVs 16:17:02 Integrated Voltage Measurement... 16:17:11 Voltage Offset Correction: v_off_ini = 1330.511 uV, v_off_fin = 1366.119 uV 16:17:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.945 uVs 16:17:11 Integrated Voltage Measurement... 16:17:20 Voltage Offset Correction: v_off_ini = 1338.543 uV, v_off_fin = 1345.637 uV 16:17:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.852 uVs 16:17:20 Integrated Voltage Measurement... 16:17:29 Voltage Offset Correction: v_off_ini = 1350.213 uV, v_off_fin = 1361.203 uV 16:17:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.575 uVs 16:17:32 I2X Average and RMS of Measurements VT = -95.556 +- 2.257 uVs I2X = 8.791 +- 0.208 uTm^2 16:17:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:17:39 Integrated Voltage Measurement... 16:17:48 Voltage Offset Correction: v_off_ini = 1343.808 uV, v_off_fin = 1339.941 uV 16:17:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -343.583 uVs 16:17:48 Integrated Voltage Measurement... 16:17:57 Voltage Offset Correction: v_off_ini = 1342.251 uV, v_off_fin = 1369.016 uV 16:17:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 360.055 uVs 16:17:57 Integrated Voltage Measurement... 16:18:06 Voltage Offset Correction: v_off_ini = 1362.170 uV, v_off_fin = 1349.079 uV 16:18:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.950 uVs 16:18:06 Integrated Voltage Measurement... 16:18:15 Voltage Offset Correction: v_off_ini = 1338.602 uV, v_off_fin = 1350.402 uV 16:18:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 357.787 uVs 16:18:16 Integrated Voltage Measurement... 16:18:24 Voltage Offset Correction: v_off_ini = 1341.632 uV, v_off_fin = 1337.011 uV 16:18:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.094 uVs 16:18:25 Integrated Voltage Measurement... 16:18:34 Voltage Offset Correction: v_off_ini = 1351.472 uV, v_off_fin = 1334.463 uV 16:18:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 350.016 uVs 16:18:36 I2Y Average and RMS of Measurements VT = -348.247 +- 2.775 uVs I2Y = -32.039 +- 0.255 uTm^2 16:18:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:18:50 Integrated Voltage Measurement... 16:18:59 Voltage Offset Correction: v_off_ini = 1346.294 uV, v_off_fin = 1339.729 uV 16:18:59 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -92.316 uVs 16:18:59 Integrated Voltage Measurement... 16:19:08 Voltage Offset Correction: v_off_ini = 1347.295 uV, v_off_fin = 1350.619 uV 16:19:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.081 uVs 16:19:08 Integrated Voltage Measurement... 16:19:17 Voltage Offset Correction: v_off_ini = 1339.143 uV, v_off_fin = 1353.259 uV 16:19:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.850 uVs 16:19:17 Integrated Voltage Measurement... 16:19:26 Voltage Offset Correction: v_off_ini = 1358.425 uV, v_off_fin = 1326.625 uV 16:19:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.662 uVs 16:19:26 Integrated Voltage Measurement... 16:19:35 Voltage Offset Correction: v_off_ini = 1331.725 uV, v_off_fin = 1323.143 uV 16:19:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -84.925 uVs 16:19:35 Integrated Voltage Measurement... 16:19:44 Voltage Offset Correction: v_off_ini = 1371.824 uV, v_off_fin = 1364.167 uV 16:19:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.265 uVs 16:19:48 I1X Average and RMS of Measurements VT = -80.683 +- 5.358 uVs I1X = 16.138 +- 1.073 uTm 16:19:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:19:55 Integrated Voltage Measurement... 16:20:04 Voltage Offset Correction: v_off_ini = 1368.495 uV, v_off_fin = 1345.958 uV 16:20:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -279.730 uVs 16:20:04 Integrated Voltage Measurement... 16:20:13 Voltage Offset Correction: v_off_ini = 1376.916 uV, v_off_fin = 1387.752 uV 16:20:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 271.735 uVs 16:20:13 Integrated Voltage Measurement... 16:20:22 Voltage Offset Correction: v_off_ini = 1385.559 uV, v_off_fin = 1368.620 uV 16:20:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -288.954 uVs 16:20:22 Integrated Voltage Measurement... 16:20:31 Voltage Offset Correction: v_off_ini = 1379.532 uV, v_off_fin = 1376.570 uV 16:20:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 276.069 uVs 16:20:31 Integrated Voltage Measurement... 16:20:40 Voltage Offset Correction: v_off_ini = 1350.522 uV, v_off_fin = 1354.198 uV 16:20:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -274.618 uVs 16:20:40 Integrated Voltage Measurement... 16:20:49 Voltage Offset Correction: v_off_ini = 1388.275 uV, v_off_fin = 1375.819 uV 16:20:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 272.693 uVs 16:20:53 I1Y Average and RMS of Measurements VT = -277.300 +- 3.782 uVs I1Y = -55.452 +- 0.756 uTm 16:20:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:21:01 Integrated Voltage Measurement... 16:21:10 Voltage Offset Correction: v_off_ini = 1396.611 uV, v_off_fin = 1386.233 uV 16:21:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.224 uVs 16:21:10 Integrated Voltage Measurement... 16:21:19 Voltage Offset Correction: v_off_ini = 1390.572 uV, v_off_fin = 1405.571 uV 16:21:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.051 uVs 16:21:19 Integrated Voltage Measurement... 16:21:28 Voltage Offset Correction: v_off_ini = 1398.972 uV, v_off_fin = 1383.885 uV 16:21:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.277 uVs 16:21:28 Integrated Voltage Measurement... 16:21:37 Voltage Offset Correction: v_off_ini = 1396.822 uV, v_off_fin = 1400.139 uV 16:21:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 119.431 uVs 16:21:37 Integrated Voltage Measurement... 16:21:46 Voltage Offset Correction: v_off_ini = 1398.907 uV, v_off_fin = 1400.611 uV 16:21:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.760 uVs 16:21:46 Integrated Voltage Measurement... 16:21:55 Voltage Offset Correction: v_off_ini = 1390.627 uV, v_off_fin = 1391.990 uV 16:21:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.624 uVs 16:21:57 I2X Average and RMS of Measurements VT = -111.228 +- 6.834 uVs I2X = 10.233 +- 0.629 uTm^2 16:22:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:22:05 Integrated Voltage Measurement... 16:22:14 Voltage Offset Correction: v_off_ini = 1376.223 uV, v_off_fin = 1375.753 uV 16:22:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -364.234 uVs 16:22:14 Integrated Voltage Measurement... 16:22:23 Voltage Offset Correction: v_off_ini = 1380.489 uV, v_off_fin = 1415.164 uV 16:22:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 351.353 uVs 16:22:23 Integrated Voltage Measurement... 16:22:32 Voltage Offset Correction: v_off_ini = 1406.015 uV, v_off_fin = 1390.935 uV 16:22:32 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -333.644 uVs 16:22:32 Integrated Voltage Measurement... 16:22:41 Voltage Offset Correction: v_off_ini = 1397.869 uV, v_off_fin = 1415.455 uV 16:22:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 377.390 uVs 16:22:41 Integrated Voltage Measurement... 16:22:50 Voltage Offset Correction: v_off_ini = 1413.386 uV, v_off_fin = 1414.938 uV 16:22:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -360.249 uVs 16:22:50 Integrated Voltage Measurement... 16:22:59 Voltage Offset Correction: v_off_ini = 1417.364 uV, v_off_fin = 1422.739 uV 16:22:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 362.066 uVs 16:23:02 I2Y Average and RMS of Measurements VT = -358.156 +- 2.317 uVs I2Y = -32.951 +- 0.213 uTm^2 16:23:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:23:19 Integrated Voltage Measurement... 16:23:28 Voltage Offset Correction: v_off_ini = 1405.110 uV, v_off_fin = 1406.110 uV 16:23:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -94.320 uVs 16:23:28 Integrated Voltage Measurement... 16:23:37 Voltage Offset Correction: v_off_ini = 1429.443 uV, v_off_fin = 1437.751 uV 16:23:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 116.174 uVs 16:23:37 Integrated Voltage Measurement... 16:23:46 Voltage Offset Correction: v_off_ini = 1445.497 uV, v_off_fin = 1423.375 uV 16:23:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.836 uVs 16:23:46 Integrated Voltage Measurement... 16:23:55 Voltage Offset Correction: v_off_ini = 1425.083 uV, v_off_fin = 1428.628 uV 16:23:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.684 uVs 16:23:55 Integrated Voltage Measurement... 16:24:04 Voltage Offset Correction: v_off_ini = 1450.083 uV, v_off_fin = 1462.854 uV 16:24:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.985 uVs 16:24:04 Integrated Voltage Measurement... 16:24:13 Voltage Offset Correction: v_off_ini = 1461.639 uV, v_off_fin = 1454.217 uV 16:24:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.395 uVs 16:24:17 I1X Average and RMS of Measurements VT = -98.065 +- 5.410 uVs I1X = 19.613 +- 1.082 uTm 16:24:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:24:24 Integrated Voltage Measurement... 16:24:33 Voltage Offset Correction: v_off_ini = 1453.874 uV, v_off_fin = 1443.458 uV 16:24:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -277.252 uVs 16:24:33 Integrated Voltage Measurement... 16:24:42 Voltage Offset Correction: v_off_ini = 1456.418 uV, v_off_fin = 1460.344 uV 16:24:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 269.690 uVs 16:24:42 Integrated Voltage Measurement... 16:24:51 Voltage Offset Correction: v_off_ini = 1451.666 uV, v_off_fin = 1449.016 uV 16:24:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -287.392 uVs 16:24:51 Integrated Voltage Measurement... 16:25:00 Voltage Offset Correction: v_off_ini = 1449.060 uV, v_off_fin = 1465.278 uV 16:25:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 280.941 uVs 16:25:00 Integrated Voltage Measurement... 16:25:09 Voltage Offset Correction: v_off_ini = 1460.526 uV, v_off_fin = 1439.292 uV 16:25:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -277.835 uVs 16:25:09 Integrated Voltage Measurement... 16:25:18 Voltage Offset Correction: v_off_ini = 1467.971 uV, v_off_fin = 1479.519 uV 16:25:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 281.503 uVs 16:25:22 I1Y Average and RMS of Measurements VT = -279.102 +- 4.385 uVs I1Y = -55.816 +- 0.875 uTm 16:25:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:25:30 Integrated Voltage Measurement... 16:25:39 Voltage Offset Correction: v_off_ini = 1460.216 uV, v_off_fin = 1472.642 uV 16:25:39 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -101.735 uVs 16:25:39 Integrated Voltage Measurement... 16:25:48 Voltage Offset Correction: v_off_ini = 1452.197 uV, v_off_fin = 1468.072 uV 16:25:48 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 101.097 uVs 16:25:48 Integrated Voltage Measurement... 16:25:57 Voltage Offset Correction: v_off_ini = 1471.722 uV, v_off_fin = 1463.211 uV 16:25:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.301 uVs 16:25:57 Integrated Voltage Measurement... 16:26:06 Voltage Offset Correction: v_off_ini = 1474.416 uV, v_off_fin = 1475.587 uV 16:26:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.569 uVs 16:26:06 Integrated Voltage Measurement... 16:26:15 Voltage Offset Correction: v_off_ini = 1487.560 uV, v_off_fin = 1495.193 uV 16:26:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.394 uVs 16:26:15 Integrated Voltage Measurement... 16:26:24 Voltage Offset Correction: v_off_ini = 1483.809 uV, v_off_fin = 1493.419 uV 16:26:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.199 uVs 16:26:26 I2X Average and RMS of Measurements VT = -103.049 +- 6.287 uVs I2X = 9.481 +- 0.578 uTm^2 16:26:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:26:34 Integrated Voltage Measurement... 16:26:43 Voltage Offset Correction: v_off_ini = 1493.963 uV, v_off_fin = 1487.567 uV 16:26:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -375.293 uVs 16:26:43 Integrated Voltage Measurement... 16:26:52 Voltage Offset Correction: v_off_ini = 1476.889 uV, v_off_fin = 1483.316 uV 16:26:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0001 mm dXY = -1.0000 mm VT = 356.081 uVs 16:26:52 Integrated Voltage Measurement... 16:27:01 Voltage Offset Correction: v_off_ini = 1490.479 uV, v_off_fin = 1486.488 uV 16:27:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = -0.0001 mm dXY = 1.0000 mm VT = -366.618 uVs 16:27:01 Integrated Voltage Measurement... 16:27:10 Voltage Offset Correction: v_off_ini = 1471.291 uV, v_off_fin = 1450.462 uV 16:27:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 356.401 uVs 16:27:10 Integrated Voltage Measurement... 16:27:19 Voltage Offset Correction: v_off_ini = 1447.210 uV, v_off_fin = 1454.866 uV 16:27:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -372.738 uVs 16:27:19 Integrated Voltage Measurement... 16:27:28 Voltage Offset Correction: v_off_ini = 1465.106 uV, v_off_fin = 1467.375 uV 16:27:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 339.989 uVs 16:27:31 I2Y Average and RMS of Measurements VT = -361.187 +- 3.813 uVs I2Y = -33.229 +- 0.351 uTm^2 16:27:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:27:48 Integrated Voltage Measurement... 16:27:57 Voltage Offset Correction: v_off_ini = 1472.069 uV, v_off_fin = 1459.516 uV 16:27:57 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -115.534 uVs 16:27:57 Integrated Voltage Measurement... 16:28:06 Voltage Offset Correction: v_off_ini = 1447.927 uV, v_off_fin = 1470.755 uV 16:28:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.394 uVs 16:28:06 Integrated Voltage Measurement... 16:28:15 Voltage Offset Correction: v_off_ini = 1479.038 uV, v_off_fin = 1491.601 uV 16:28:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -88.532 uVs 16:28:15 Integrated Voltage Measurement... 16:28:24 Voltage Offset Correction: v_off_ini = 1475.914 uV, v_off_fin = 1475.010 uV 16:28:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.810 uVs 16:28:24 Integrated Voltage Measurement... 16:28:33 Voltage Offset Correction: v_off_ini = 1483.646 uV, v_off_fin = 1470.775 uV 16:28:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -97.951 uVs 16:28:33 Integrated Voltage Measurement... 16:28:42 Voltage Offset Correction: v_off_ini = 1475.503 uV, v_off_fin = 1482.132 uV 16:28:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.179 uVs 16:28:46 I1X Average and RMS of Measurements VT = -96.567 +- 3.463 uVs I1X = 19.315 +- 0.694 uTm 16:28:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:28:53 Integrated Voltage Measurement... 16:29:02 Voltage Offset Correction: v_off_ini = 1452.847 uV, v_off_fin = 1449.268 uV 16:29:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -293.271 uVs 16:29:02 Integrated Voltage Measurement... 16:29:11 Voltage Offset Correction: v_off_ini = 1464.742 uV, v_off_fin = 1488.145 uV 16:29:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 281.979 uVs 16:29:11 Integrated Voltage Measurement... 16:29:20 Voltage Offset Correction: v_off_ini = 1505.334 uV, v_off_fin = 1502.902 uV 16:29:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -263.814 uVs 16:29:20 Integrated Voltage Measurement... 16:29:29 Voltage Offset Correction: v_off_ini = 1500.388 uV, v_off_fin = 1507.277 uV 16:29:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 283.684 uVs 16:29:29 Integrated Voltage Measurement... 16:29:38 Voltage Offset Correction: v_off_ini = 1485.390 uV, v_off_fin = 1457.720 uV 16:29:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -283.470 uVs 16:29:38 Integrated Voltage Measurement... 16:29:47 Voltage Offset Correction: v_off_ini = 1469.636 uV, v_off_fin = 1507.803 uV 16:29:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 273.112 uVs 16:29:51 I1Y Average and RMS of Measurements VT = -279.888 +- 5.776 uVs I1Y = -55.970 +- 1.154 uTm 16:29:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:29:59 Integrated Voltage Measurement... 16:30:08 Voltage Offset Correction: v_off_ini = 1479.625 uV, v_off_fin = 1485.404 uV 16:30:08 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -98.990 uVs 16:30:08 Integrated Voltage Measurement... 16:30:17 Voltage Offset Correction: v_off_ini = 1495.264 uV, v_off_fin = 1512.789 uV 16:30:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.949 uVs 16:30:17 Integrated Voltage Measurement... 16:30:26 Voltage Offset Correction: v_off_ini = 1514.680 uV, v_off_fin = 1492.909 uV 16:30:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.906 uVs 16:30:26 Integrated Voltage Measurement... 16:30:35 Voltage Offset Correction: v_off_ini = 1512.005 uV, v_off_fin = 1539.768 uV 16:30:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.469 uVs 16:30:35 Integrated Voltage Measurement... 16:30:44 Voltage Offset Correction: v_off_ini = 1532.255 uV, v_off_fin = 1529.592 uV 16:30:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.031 uVs 16:30:44 Integrated Voltage Measurement... 16:30:53 Voltage Offset Correction: v_off_ini = 1542.570 uV, v_off_fin = 1543.234 uV 16:30:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.205 uVs 16:30:55 I2X Average and RMS of Measurements VT = -94.259 +- 4.199 uVs I2X = 8.671 +- 0.386 uTm^2 16:31:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:31:03 Integrated Voltage Measurement... 16:31:12 Voltage Offset Correction: v_off_ini = 1547.830 uV, v_off_fin = 1540.955 uV 16:31:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.799 uVs 16:31:12 Integrated Voltage Measurement... 16:31:21 Voltage Offset Correction: v_off_ini = 1528.427 uV, v_off_fin = 1514.420 uV 16:31:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.081 uVs 16:31:21 Integrated Voltage Measurement... 16:31:30 Voltage Offset Correction: v_off_ini = 1510.592 uV, v_off_fin = 1507.368 uV 16:31:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -349.210 uVs 16:31:30 Integrated Voltage Measurement... 16:31:39 Voltage Offset Correction: v_off_ini = 1520.077 uV, v_off_fin = 1515.986 uV 16:31:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.283 uVs 16:31:39 Integrated Voltage Measurement... 16:31:48 Voltage Offset Correction: v_off_ini = 1527.559 uV, v_off_fin = 1526.281 uV 16:31:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.832 uVs 16:31:48 Integrated Voltage Measurement... 16:31:57 Voltage Offset Correction: v_off_ini = 1513.644 uV, v_off_fin = 1517.111 uV 16:31:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 345.366 uVs 16:32:00 I2Y Average and RMS of Measurements VT = -335.095 +- 4.935 uVs I2Y = -30.829 +- 0.454 uTm^2 16:32:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:32:22 Integrated Voltage Measurement... 16:32:31 Voltage Offset Correction: v_off_ini = 1510.521 uV, v_off_fin = 1487.524 uV 16:32:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -113.424 uVs 16:32:31 Integrated Voltage Measurement... 16:32:40 Voltage Offset Correction: v_off_ini = 1496.015 uV, v_off_fin = 1496.413 uV 16:32:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.941 uVs 16:32:40 Integrated Voltage Measurement... 16:32:49 Voltage Offset Correction: v_off_ini = 1480.784 uV, v_off_fin = 1477.733 uV 16:32:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.135 uVs 16:32:49 Integrated Voltage Measurement... 16:32:58 Voltage Offset Correction: v_off_ini = 1515.480 uV, v_off_fin = 1519.456 uV 16:32:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 116.435 uVs 16:32:58 Integrated Voltage Measurement... 16:33:07 Voltage Offset Correction: v_off_ini = 1508.453 uV, v_off_fin = 1487.174 uV 16:33:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.254 uVs 16:33:07 Integrated Voltage Measurement... 16:33:16 Voltage Offset Correction: v_off_ini = 1506.620 uV, v_off_fin = 1502.772 uV 16:33:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.526 uVs 16:33:20 I1X Average and RMS of Measurements VT = -92.453 +- 11.453 uVs I1X = 18.491 +- 2.291 uTm 16:33:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:33:27 Integrated Voltage Measurement... 16:33:35 Voltage Offset Correction: v_off_ini = 1506.635 uV, v_off_fin = 1483.667 uV 16:33:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -244.103 uVs 16:33:36 Integrated Voltage Measurement... 16:33:44 Voltage Offset Correction: v_off_ini = 1482.635 uV, v_off_fin = 1504.892 uV 16:33:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 275.801 uVs 16:33:44 Integrated Voltage Measurement... 16:33:53 Voltage Offset Correction: v_off_ini = 1494.917 uV, v_off_fin = 1481.336 uV 16:33:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -260.047 uVs 16:33:53 Integrated Voltage Measurement... 16:34:02 Voltage Offset Correction: v_off_ini = 1493.576 uV, v_off_fin = 1508.883 uV 16:34:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 270.321 uVs 16:34:02 Integrated Voltage Measurement... 16:34:11 Voltage Offset Correction: v_off_ini = 1496.257 uV, v_off_fin = 1484.448 uV 16:34:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -248.153 uVs 16:34:11 Integrated Voltage Measurement... 16:34:20 Voltage Offset Correction: v_off_ini = 1505.160 uV, v_off_fin = 1512.333 uV 16:34:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 276.642 uVs 16:34:25 I1Y Average and RMS of Measurements VT = -262.511 +- 2.137 uVs I1Y = -52.496 +- 0.429 uTm 16:34:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:34:32 Integrated Voltage Measurement... 16:34:41 Voltage Offset Correction: v_off_ini = 1475.899 uV, v_off_fin = 1488.897 uV 16:34:41 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -89.760 uVs 16:34:41 Integrated Voltage Measurement... 16:34:50 Voltage Offset Correction: v_off_ini = 1512.308 uV, v_off_fin = 1523.099 uV 16:34:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.115 uVs 16:34:50 Integrated Voltage Measurement... 16:34:59 Voltage Offset Correction: v_off_ini = 1523.161 uV, v_off_fin = 1496.775 uV 16:34:59 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -101.250 uVs 16:34:59 Integrated Voltage Measurement... 16:35:08 Voltage Offset Correction: v_off_ini = 1511.192 uV, v_off_fin = 1517.456 uV 16:35:08 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 95.629 uVs 16:35:08 Integrated Voltage Measurement... 16:35:17 Voltage Offset Correction: v_off_ini = 1503.750 uV, v_off_fin = 1485.650 uV 16:35:17 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -72.955 uVs 16:35:17 Integrated Voltage Measurement... 16:35:26 Voltage Offset Correction: v_off_ini = 1491.627 uV, v_off_fin = 1501.328 uV 16:35:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 92.131 uVs 16:35:29 I2X Average and RMS of Measurements VT = -87.307 +- 7.899 uVs I2X = 8.032 +- 0.727 uTm^2 16:35:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:35:36 Integrated Voltage Measurement... 16:35:45 Voltage Offset Correction: v_off_ini = 1505.512 uV, v_off_fin = 1503.831 uV 16:35:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.367 uVs 16:35:45 Integrated Voltage Measurement... 16:35:54 Voltage Offset Correction: v_off_ini = 1506.499 uV, v_off_fin = 1487.634 uV 16:35:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.085 uVs 16:35:54 Integrated Voltage Measurement... 16:36:03 Voltage Offset Correction: v_off_ini = 1504.313 uV, v_off_fin = 1515.627 uV 16:36:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -338.315 uVs 16:36:03 Integrated Voltage Measurement... 16:36:12 Voltage Offset Correction: v_off_ini = 1525.990 uV, v_off_fin = 1527.421 uV 16:36:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 338.412 uVs 16:36:12 Integrated Voltage Measurement... 16:36:21 Voltage Offset Correction: v_off_ini = 1523.654 uV, v_off_fin = 1506.143 uV 16:36:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.191 uVs 16:36:21 Integrated Voltage Measurement... 16:36:30 Voltage Offset Correction: v_off_ini = 1481.084 uV, v_off_fin = 1501.929 uV 16:36:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 335.589 uVs 16:36:33 I2Y Average and RMS of Measurements VT = -328.326 +- 8.023 uVs I2Y = -30.206 +- 0.738 uTm^2 16:36:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:36:55 Integrated Voltage Measurement... 16:37:04 Voltage Offset Correction: v_off_ini = 1508.029 uV, v_off_fin = 1498.937 uV 16:37:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -88.266 uVs 16:37:04 Integrated Voltage Measurement... 16:37:13 Voltage Offset Correction: v_off_ini = 1507.677 uV, v_off_fin = 1499.084 uV 16:37:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.530 uVs 16:37:13 Integrated Voltage Measurement... 16:37:22 Voltage Offset Correction: v_off_ini = 1459.497 uV, v_off_fin = 1470.120 uV 16:37:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.682 uVs 16:37:22 Integrated Voltage Measurement... 16:37:31 Voltage Offset Correction: v_off_ini = 1483.254 uV, v_off_fin = 1493.980 uV 16:37:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.241 uVs 16:37:31 Integrated Voltage Measurement... 16:37:40 Voltage Offset Correction: v_off_ini = 1487.081 uV, v_off_fin = 1505.213 uV 16:37:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.200 uVs 16:37:40 Integrated Voltage Measurement... 16:37:49 Voltage Offset Correction: v_off_ini = 1501.481 uV, v_off_fin = 1507.582 uV 16:37:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.921 uVs 16:37:53 I1X Average and RMS of Measurements VT = -82.640 +- 12.391 uVs I1X = 16.528 +- 2.478 uTm 16:37:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:38:00 Integrated Voltage Measurement... 16:38:08 Voltage Offset Correction: v_off_ini = 1510.420 uV, v_off_fin = 1498.051 uV 16:38:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.879 uVs 16:38:09 Integrated Voltage Measurement... 16:38:17 Voltage Offset Correction: v_off_ini = 1518.271 uV, v_off_fin = 1541.151 uV 16:38:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 263.508 uVs 16:38:17 Integrated Voltage Measurement... 16:38:26 Voltage Offset Correction: v_off_ini = 1529.910 uV, v_off_fin = 1519.845 uV 16:38:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -243.709 uVs 16:38:26 Integrated Voltage Measurement... 16:38:35 Voltage Offset Correction: v_off_ini = 1511.999 uV, v_off_fin = 1534.522 uV 16:38:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 265.839 uVs 16:38:35 Integrated Voltage Measurement... 16:38:44 Voltage Offset Correction: v_off_ini = 1502.937 uV, v_off_fin = 1502.309 uV 16:38:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -257.504 uVs 16:38:44 Integrated Voltage Measurement... 16:38:53 Voltage Offset Correction: v_off_ini = 1542.443 uV, v_off_fin = 1540.298 uV 16:38:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.235 uVs 16:38:58 I1Y Average and RMS of Measurements VT = -256.946 +- 2.147 uVs I1Y = -51.387 +- 0.429 uTm 16:39:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:39:05 Integrated Voltage Measurement... 16:39:14 Voltage Offset Correction: v_off_ini = 1528.001 uV, v_off_fin = 1536.255 uV 16:39:14 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -96.883 uVs 16:39:14 Integrated Voltage Measurement... 16:39:23 Voltage Offset Correction: v_off_ini = 1527.303 uV, v_off_fin = 1541.870 uV 16:39:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.849 uVs 16:39:23 Integrated Voltage Measurement... 16:39:32 Voltage Offset Correction: v_off_ini = 1555.808 uV, v_off_fin = 1548.724 uV 16:39:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.609 uVs 16:39:32 Integrated Voltage Measurement... 16:39:41 Voltage Offset Correction: v_off_ini = 1537.858 uV, v_off_fin = 1532.276 uV 16:39:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.741 uVs 16:39:41 Integrated Voltage Measurement... 16:39:50 Voltage Offset Correction: v_off_ini = 1538.284 uV, v_off_fin = 1517.947 uV 16:39:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.791 uVs 16:39:50 Integrated Voltage Measurement... 16:39:59 Voltage Offset Correction: v_off_ini = 1549.170 uV, v_off_fin = 1558.113 uV 16:39:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.992 uVs 16:40:02 I2X Average and RMS of Measurements VT = -89.144 +- 2.751 uVs I2X = 8.201 +- 0.253 uTm^2 16:40:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:40:09 Integrated Voltage Measurement... 16:40:18 Voltage Offset Correction: v_off_ini = 1553.622 uV, v_off_fin = 1557.084 uV 16:40:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -302.306 uVs 16:40:18 Integrated Voltage Measurement... 16:40:27 Voltage Offset Correction: v_off_ini = 1539.610 uV, v_off_fin = 1556.532 uV 16:40:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 311.347 uVs 16:40:27 Integrated Voltage Measurement... 16:40:36 Voltage Offset Correction: v_off_ini = 1559.453 uV, v_off_fin = 1550.021 uV 16:40:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.401 uVs 16:40:36 Integrated Voltage Measurement... 16:40:45 Voltage Offset Correction: v_off_ini = 1554.009 uV, v_off_fin = 1538.909 uV 16:40:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.425 uVs 16:40:46 Integrated Voltage Measurement... 16:40:54 Voltage Offset Correction: v_off_ini = 1543.124 uV, v_off_fin = 1560.611 uV 16:40:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.687 uVs 16:40:55 Integrated Voltage Measurement... 16:41:03 Voltage Offset Correction: v_off_ini = 1555.760 uV, v_off_fin = 1555.803 uV 16:41:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 312.630 uVs 16:41:06 I2Y Average and RMS of Measurements VT = -313.466 +- 4.748 uVs I2Y = -28.839 +- 0.437 uTm^2 16:41:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:41:28 Integrated Voltage Measurement... 16:41:37 Voltage Offset Correction: v_off_ini = 1536.366 uV, v_off_fin = 1536.118 uV 16:41:37 Integrated Voltage Measurement dXYus = -0.4998 mm, dXYds = -0.4997 mm dXY = -0.4997 mm VT = -100.003 uVs 16:41:37 Integrated Voltage Measurement... 16:41:46 Voltage Offset Correction: v_off_ini = 1535.241 uV, v_off_fin = 1550.788 uV 16:41:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 98.511 uVs 16:41:46 Integrated Voltage Measurement... 16:41:55 Voltage Offset Correction: v_off_ini = 1529.721 uV, v_off_fin = 1524.014 uV 16:41:55 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.029 uVs 16:41:55 Integrated Voltage Measurement... 16:42:04 Voltage Offset Correction: v_off_ini = 1528.786 uV, v_off_fin = 1528.197 uV 16:42:04 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 114.917 uVs 16:42:04 Integrated Voltage Measurement... 16:42:13 Voltage Offset Correction: v_off_ini = 1552.770 uV, v_off_fin = 1545.552 uV 16:42:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.575 uVs 16:42:13 Integrated Voltage Measurement... 16:42:22 Voltage Offset Correction: v_off_ini = 1548.526 uV, v_off_fin = 1544.136 uV 16:42:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.139 uVs 16:42:26 I1X Average and RMS of Measurements VT = -98.529 +- 3.959 uVs I1X = 19.709 +- 0.793 uTm 16:42:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:42:33 Integrated Voltage Measurement... 16:42:41 Voltage Offset Correction: v_off_ini = 1529.309 uV, v_off_fin = 1507.929 uV 16:42:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -259.417 uVs 16:42:42 Integrated Voltage Measurement... 16:42:50 Voltage Offset Correction: v_off_ini = 1528.968 uV, v_off_fin = 1542.750 uV 16:42:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 271.497 uVs 16:42:51 Integrated Voltage Measurement... 16:42:59 Voltage Offset Correction: v_off_ini = 1546.931 uV, v_off_fin = 1531.096 uV 16:42:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -248.332 uVs 16:42:59 Integrated Voltage Measurement... 16:43:08 Voltage Offset Correction: v_off_ini = 1555.955 uV, v_off_fin = 1550.331 uV 16:43:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 242.143 uVs 16:43:08 Integrated Voltage Measurement... 16:43:17 Voltage Offset Correction: v_off_ini = 1522.978 uV, v_off_fin = 1506.591 uV 16:43:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -245.988 uVs 16:43:17 Integrated Voltage Measurement... 16:43:26 Voltage Offset Correction: v_off_ini = 1522.089 uV, v_off_fin = 1523.096 uV 16:43:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.152 uVs 16:43:31 I1Y Average and RMS of Measurements VT = -254.255 +- 8.398 uVs I1Y = -50.848 +- 1.680 uTm 16:43:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:43:38 Integrated Voltage Measurement... 16:43:47 Voltage Offset Correction: v_off_ini = 1510.286 uV, v_off_fin = 1516.839 uV 16:43:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.229 uVs 16:43:47 Integrated Voltage Measurement... 16:43:56 Voltage Offset Correction: v_off_ini = 1514.014 uV, v_off_fin = 1533.590 uV 16:43:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.132 uVs 16:43:56 Integrated Voltage Measurement... 16:44:05 Voltage Offset Correction: v_off_ini = 1548.061 uV, v_off_fin = 1539.674 uV 16:44:05 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -91.047 uVs 16:44:05 Integrated Voltage Measurement... 16:44:14 Voltage Offset Correction: v_off_ini = 1531.416 uV, v_off_fin = 1539.197 uV 16:44:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 83.276 uVs 16:44:14 Integrated Voltage Measurement... 16:44:23 Voltage Offset Correction: v_off_ini = 1524.720 uV, v_off_fin = 1515.145 uV 16:44:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.085 uVs 16:44:23 Integrated Voltage Measurement... 16:44:32 Voltage Offset Correction: v_off_ini = 1537.972 uV, v_off_fin = 1552.004 uV 16:44:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.120 uVs 16:44:35 I2X Average and RMS of Measurements VT = -94.648 +- 5.908 uVs I2X = 8.708 +- 0.543 uTm^2 16:44:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:44:42 Integrated Voltage Measurement... 16:44:51 Voltage Offset Correction: v_off_ini = 1544.978 uV, v_off_fin = 1537.169 uV 16:44:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -295.916 uVs 16:44:51 Integrated Voltage Measurement... 16:45:00 Voltage Offset Correction: v_off_ini = 1538.859 uV, v_off_fin = 1542.701 uV 16:45:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 306.340 uVs 16:45:00 Integrated Voltage Measurement... 16:45:09 Voltage Offset Correction: v_off_ini = 1540.100 uV, v_off_fin = 1539.206 uV 16:45:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.510 uVs 16:45:09 Integrated Voltage Measurement... 16:45:18 Voltage Offset Correction: v_off_ini = 1542.089 uV, v_off_fin = 1554.551 uV 16:45:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 337.386 uVs 16:45:18 Integrated Voltage Measurement... 16:45:27 Voltage Offset Correction: v_off_ini = 1542.306 uV, v_off_fin = 1544.588 uV 16:45:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.089 uVs 16:45:28 Integrated Voltage Measurement... 16:45:37 Voltage Offset Correction: v_off_ini = 1548.779 uV, v_off_fin = 1548.479 uV 16:45:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 296.607 uVs 16:45:39 I2Y Average and RMS of Measurements VT = -305.641 +- 6.942 uVs I2Y = -28.119 +- 0.639 uTm^2 16:45:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:46:01 Integrated Voltage Measurement... 16:46:10 Voltage Offset Correction: v_off_ini = 1590.618 uV, v_off_fin = 1590.940 uV 16:46:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -62.609 uVs 16:46:10 Integrated Voltage Measurement... 16:46:19 Voltage Offset Correction: v_off_ini = 1591.243 uV, v_off_fin = 1579.671 uV 16:46:19 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.932 uVs 16:46:19 Integrated Voltage Measurement... 16:46:28 Voltage Offset Correction: v_off_ini = 1586.519 uV, v_off_fin = 1594.525 uV 16:46:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.852 uVs 16:46:28 Integrated Voltage Measurement... 16:46:37 Voltage Offset Correction: v_off_ini = 1590.513 uV, v_off_fin = 1596.808 uV 16:46:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.116 uVs 16:46:37 Integrated Voltage Measurement... 16:46:46 Voltage Offset Correction: v_off_ini = 1598.633 uV, v_off_fin = 1600.373 uV 16:46:46 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -82.222 uVs 16:46:46 Integrated Voltage Measurement... 16:46:55 Voltage Offset Correction: v_off_ini = 1617.149 uV, v_off_fin = 1623.479 uV 16:46:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 95.109 uVs 16:46:59 I1X Average and RMS of Measurements VT = -85.473 +- 6.993 uVs I1X = 17.096 +- 1.398 uTm 16:47:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:47:06 Integrated Voltage Measurement... 16:47:15 Voltage Offset Correction: v_off_ini = 1614.232 uV, v_off_fin = 1596.368 uV 16:47:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -254.981 uVs 16:47:15 Integrated Voltage Measurement... 16:47:24 Voltage Offset Correction: v_off_ini = 1630.471 uV, v_off_fin = 1633.734 uV 16:47:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 234.763 uVs 16:47:24 Integrated Voltage Measurement... 16:47:33 Voltage Offset Correction: v_off_ini = 1622.793 uV, v_off_fin = 1602.595 uV 16:47:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -259.192 uVs 16:47:33 Integrated Voltage Measurement... 16:47:42 Voltage Offset Correction: v_off_ini = 1617.115 uV, v_off_fin = 1641.661 uV 16:47:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 253.861 uVs 16:47:42 Integrated Voltage Measurement... 16:47:51 Voltage Offset Correction: v_off_ini = 1615.353 uV, v_off_fin = 1591.551 uV 16:47:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -239.171 uVs 16:47:51 Integrated Voltage Measurement... 16:48:00 Voltage Offset Correction: v_off_ini = 1617.669 uV, v_off_fin = 1624.672 uV 16:48:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.210 uVs 16:48:04 I1Y Average and RMS of Measurements VT = -249.530 +- 5.037 uVs I1Y = -49.901 +- 1.005 uTm 16:48:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:48:12 Integrated Voltage Measurement... 16:48:21 Voltage Offset Correction: v_off_ini = 1604.775 uV, v_off_fin = 1619.338 uV 16:48:21 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -63.831 uVs 16:48:21 Integrated Voltage Measurement... 16:48:30 Voltage Offset Correction: v_off_ini = 1625.358 uV, v_off_fin = 1647.761 uV 16:48:30 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 96.913 uVs 16:48:30 Integrated Voltage Measurement... 16:48:39 Voltage Offset Correction: v_off_ini = 1616.499 uV, v_off_fin = 1603.117 uV 16:48:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.558 uVs 16:48:39 Integrated Voltage Measurement... 16:48:48 Voltage Offset Correction: v_off_ini = 1642.475 uV, v_off_fin = 1655.100 uV 16:48:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.033 uVs 16:48:48 Integrated Voltage Measurement... 16:48:57 Voltage Offset Correction: v_off_ini = 1638.791 uV, v_off_fin = 1661.921 uV 16:48:57 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -102.273 uVs 16:48:57 Integrated Voltage Measurement... 16:49:06 Voltage Offset Correction: v_off_ini = 1646.189 uV, v_off_fin = 1654.681 uV 16:49:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.177 uVs 16:49:08 I2X Average and RMS of Measurements VT = -84.464 +- 8.061 uVs I2X = 7.771 +- 0.742 uTm^2 16:49:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:49:16 Integrated Voltage Measurement... 16:49:25 Voltage Offset Correction: v_off_ini = 1636.128 uV, v_off_fin = 1640.571 uV 16:49:25 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -291.178 uVs 16:49:25 Integrated Voltage Measurement... 16:49:34 Voltage Offset Correction: v_off_ini = 1644.168 uV, v_off_fin = 1656.413 uV 16:49:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.330 uVs 16:49:34 Integrated Voltage Measurement... 16:49:43 Voltage Offset Correction: v_off_ini = 1640.225 uV, v_off_fin = 1632.928 uV 16:49:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.516 uVs 16:49:43 Integrated Voltage Measurement... 16:49:52 Voltage Offset Correction: v_off_ini = 1638.251 uV, v_off_fin = 1640.257 uV 16:49:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.049 uVs 16:49:52 Integrated Voltage Measurement... 16:50:01 Voltage Offset Correction: v_off_ini = 1636.801 uV, v_off_fin = 1631.621 uV 16:50:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -311.220 uVs 16:50:01 Integrated Voltage Measurement... 16:50:10 Voltage Offset Correction: v_off_ini = 1642.168 uV, v_off_fin = 1660.414 uV 16:50:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.873 uVs 16:50:12 I2Y Average and RMS of Measurements VT = -289.027 +- 11.154 uVs I2Y = -26.591 +- 1.026 uTm^2 16:50:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:50:34 Integrated Voltage Measurement... 16:50:43 Voltage Offset Correction: v_off_ini = 1654.276 uV, v_off_fin = 1650.950 uV 16:50:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -86.648 uVs 16:50:43 Integrated Voltage Measurement... 16:50:52 Voltage Offset Correction: v_off_ini = 1652.480 uV, v_off_fin = 1655.622 uV 16:50:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.501 uVs 16:50:52 Integrated Voltage Measurement... 16:51:01 Voltage Offset Correction: v_off_ini = 1648.794 uV, v_off_fin = 1652.572 uV 16:51:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -92.271 uVs 16:51:01 Integrated Voltage Measurement... 16:51:10 Voltage Offset Correction: v_off_ini = 1656.077 uV, v_off_fin = 1676.781 uV 16:51:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.862 uVs 16:51:10 Integrated Voltage Measurement... 16:51:19 Voltage Offset Correction: v_off_ini = 1684.714 uV, v_off_fin = 1661.361 uV 16:51:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -90.256 uVs 16:51:19 Integrated Voltage Measurement... 16:51:28 Voltage Offset Correction: v_off_ini = 1627.719 uV, v_off_fin = 1639.587 uV 16:51:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.202 uVs 16:51:33 I1X Average and RMS of Measurements VT = -82.623 +- 4.428 uVs I1X = 16.525 +- 0.886 uTm 16:51:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:51:39 Integrated Voltage Measurement... 16:51:48 Voltage Offset Correction: v_off_ini = 1632.255 uV, v_off_fin = 1627.084 uV 16:51:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -227.387 uVs 16:51:48 Integrated Voltage Measurement... 16:51:57 Voltage Offset Correction: v_off_ini = 1634.393 uV, v_off_fin = 1652.763 uV 16:51:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 250.230 uVs 16:51:57 Integrated Voltage Measurement... 16:52:06 Voltage Offset Correction: v_off_ini = 1626.155 uV, v_off_fin = 1622.227 uV 16:52:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -236.388 uVs 16:52:06 Integrated Voltage Measurement... 16:52:15 Voltage Offset Correction: v_off_ini = 1648.542 uV, v_off_fin = 1646.369 uV 16:52:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.373 uVs 16:52:15 Integrated Voltage Measurement... 16:52:24 Voltage Offset Correction: v_off_ini = 1639.352 uV, v_off_fin = 1631.938 uV 16:52:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -238.843 uVs 16:52:24 Integrated Voltage Measurement... 16:52:33 Voltage Offset Correction: v_off_ini = 1645.307 uV, v_off_fin = 1657.095 uV 16:52:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.503 uVs 16:52:37 I1Y Average and RMS of Measurements VT = -240.954 +- 3.710 uVs I1Y = -48.187 +- 0.743 uTm 16:52:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:52:45 Integrated Voltage Measurement... 16:52:54 Voltage Offset Correction: v_off_ini = 1648.602 uV, v_off_fin = 1649.869 uV 16:52:54 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -116.889 uVs 16:52:54 Integrated Voltage Measurement... 16:53:03 Voltage Offset Correction: v_off_ini = 1643.223 uV, v_off_fin = 1635.185 uV 16:53:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.260 uVs 16:53:03 Integrated Voltage Measurement... 16:53:12 Voltage Offset Correction: v_off_ini = 1651.872 uV, v_off_fin = 1637.033 uV 16:53:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -102.672 uVs 16:53:12 Integrated Voltage Measurement... 16:53:21 Voltage Offset Correction: v_off_ini = 1667.183 uV, v_off_fin = 1669.602 uV 16:53:21 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 107.338 uVs 16:53:21 Integrated Voltage Measurement... 16:53:30 Voltage Offset Correction: v_off_ini = 1676.405 uV, v_off_fin = 1661.097 uV 16:53:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.028 uVs 16:53:30 Integrated Voltage Measurement... 16:53:39 Voltage Offset Correction: v_off_ini = 1653.485 uV, v_off_fin = 1673.393 uV 16:53:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.609 uVs 16:53:42 I2X Average and RMS of Measurements VT = -101.799 +- 17.197 uVs I2X = 9.366 +- 1.582 uTm^2 16:53:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:53:49 Integrated Voltage Measurement... 16:53:58 Voltage Offset Correction: v_off_ini = 1661.985 uV, v_off_fin = 1677.944 uV 16:53:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -278.828 uVs 16:53:58 Integrated Voltage Measurement... 16:54:07 Voltage Offset Correction: v_off_ini = 1679.583 uV, v_off_fin = 1659.004 uV 16:54:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.949 uVs 16:54:07 Integrated Voltage Measurement... 16:54:16 Voltage Offset Correction: v_off_ini = 1682.416 uV, v_off_fin = 1675.048 uV 16:54:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -289.568 uVs 16:54:17 Integrated Voltage Measurement... 16:54:25 Voltage Offset Correction: v_off_ini = 1679.512 uV, v_off_fin = 1672.845 uV 16:54:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 291.378 uVs 16:54:26 Integrated Voltage Measurement... 16:54:35 Voltage Offset Correction: v_off_ini = 1668.340 uV, v_off_fin = 1661.610 uV 16:54:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = -0.0001 mm dXY = 1.0000 mm VT = -272.563 uVs 16:54:35 Integrated Voltage Measurement... 16:54:44 Voltage Offset Correction: v_off_ini = 1659.405 uV, v_off_fin = 1671.776 uV 16:54:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 292.425 uVs 16:54:46 I2Y Average and RMS of Measurements VT = -283.619 +- 5.199 uVs I2Y = -26.093 +- 0.478 uTm^2 16:55:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:55:08 Integrated Voltage Measurement... 16:55:17 Voltage Offset Correction: v_off_ini = 1682.155 uV, v_off_fin = 1654.352 uV 16:55:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -116.518 uVs 16:55:17 Integrated Voltage Measurement... 16:55:26 Voltage Offset Correction: v_off_ini = 1647.248 uV, v_off_fin = 1654.387 uV 16:55:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.587 uVs 16:55:26 Integrated Voltage Measurement... 16:55:35 Voltage Offset Correction: v_off_ini = 1640.539 uV, v_off_fin = 1617.633 uV 16:55:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.794 uVs 16:55:35 Integrated Voltage Measurement... 16:55:44 Voltage Offset Correction: v_off_ini = 1640.217 uV, v_off_fin = 1661.262 uV 16:55:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.844 uVs 16:55:44 Integrated Voltage Measurement... 16:55:53 Voltage Offset Correction: v_off_ini = 1644.599 uV, v_off_fin = 1652.716 uV 16:55:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -90.878 uVs 16:55:53 Integrated Voltage Measurement... 16:56:02 Voltage Offset Correction: v_off_ini = 1647.489 uV, v_off_fin = 1641.293 uV 16:56:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.412 uVs 16:56:06 I1X Average and RMS of Measurements VT = -90.339 +- 4.642 uVs I1X = 18.068 +- 0.928 uTm 16:56:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:56:13 Integrated Voltage Measurement... 16:56:22 Voltage Offset Correction: v_off_ini = 1657.850 uV, v_off_fin = 1618.402 uV 16:56:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -219.838 uVs 16:56:22 Integrated Voltage Measurement... 16:56:31 Voltage Offset Correction: v_off_ini = 1634.449 uV, v_off_fin = 1661.401 uV 16:56:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 241.762 uVs 16:56:31 Integrated Voltage Measurement... 16:56:39 Voltage Offset Correction: v_off_ini = 1654.314 uV, v_off_fin = 1649.016 uV 16:56:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -239.472 uVs 16:56:40 Integrated Voltage Measurement... 16:56:48 Voltage Offset Correction: v_off_ini = 1645.917 uV, v_off_fin = 1638.981 uV 16:56:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 231.072 uVs 16:56:49 Integrated Voltage Measurement... 16:56:57 Voltage Offset Correction: v_off_ini = 1632.874 uV, v_off_fin = 1625.477 uV 16:56:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -267.851 uVs 16:56:58 Integrated Voltage Measurement... 16:57:06 Voltage Offset Correction: v_off_ini = 1618.364 uV, v_off_fin = 1640.196 uV 16:57:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.519 uVs 16:57:11 I1Y Average and RMS of Measurements VT = -241.419 +- 11.995 uVs I1Y = -48.278 +- 2.399 uTm 16:57:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:57:18 Integrated Voltage Measurement... 16:57:27 Voltage Offset Correction: v_off_ini = 1643.494 uV, v_off_fin = 1649.111 uV 16:57:27 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -90.141 uVs 16:57:27 Integrated Voltage Measurement... 16:57:36 Voltage Offset Correction: v_off_ini = 1661.879 uV, v_off_fin = 1686.708 uV 16:57:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.021 uVs 16:57:36 Integrated Voltage Measurement... 16:57:45 Voltage Offset Correction: v_off_ini = 1672.642 uV, v_off_fin = 1633.582 uV 16:57:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.764 uVs 16:57:45 Integrated Voltage Measurement... 16:57:54 Voltage Offset Correction: v_off_ini = 1648.465 uV, v_off_fin = 1635.715 uV 16:57:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.136 uVs 16:57:54 Integrated Voltage Measurement... 16:58:03 Voltage Offset Correction: v_off_ini = 1654.058 uV, v_off_fin = 1636.873 uV 16:58:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.802 uVs 16:58:03 Integrated Voltage Measurement... 16:58:12 Voltage Offset Correction: v_off_ini = 1634.568 uV, v_off_fin = 1652.962 uV 16:58:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.681 uVs 16:58:15 I2X Average and RMS of Measurements VT = -89.091 +- 8.475 uVs I2X = 8.196 +- 0.780 uTm^2 16:58:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:58:22 Integrated Voltage Measurement... 16:58:31 Voltage Offset Correction: v_off_ini = 1640.919 uV, v_off_fin = 1644.351 uV 16:58:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -286.552 uVs 16:58:31 Integrated Voltage Measurement... 16:58:40 Voltage Offset Correction: v_off_ini = 1649.965 uV, v_off_fin = 1655.002 uV 16:58:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.702 uVs 16:58:41 Integrated Voltage Measurement... 16:58:49 Voltage Offset Correction: v_off_ini = 1644.144 uV, v_off_fin = 1638.820 uV 16:58:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -253.488 uVs 16:58:50 Integrated Voltage Measurement... 16:58:58 Voltage Offset Correction: v_off_ini = 1638.999 uV, v_off_fin = 1669.109 uV 16:58:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 295.808 uVs 16:58:59 Integrated Voltage Measurement... 16:59:08 Voltage Offset Correction: v_off_ini = 1662.039 uV, v_off_fin = 1671.933 uV 16:59:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.709 uVs 16:59:08 Integrated Voltage Measurement... 16:59:17 Voltage Offset Correction: v_off_ini = 1658.218 uV, v_off_fin = 1660.829 uV 16:59:17 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 289.223 uVs 16:59:19 I2Y Average and RMS of Measurements VT = -274.247 +- 4.973 uVs I2Y = -25.231 +- 0.457 uTm^2 16:59:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:35:03 Wire Position: X0 = 50.0000 mm, Y0 = 0.0000 mm