SLAC Magnetic Measurements Date: 01-14-2025 Time: 14:38:00 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_10086 Dataset Number: 1 Run Purpose: Final Results Run Type: Stretched Wire Run Number: 4 Measurement Device: Background Operator: YL Comment: 14:38:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:38:10 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:38:12 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:38:14 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:38:15 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:38:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:38:43 Integrated Voltage Measurement... 14:38:52 Voltage Offset Correction: v_off_ini = 139.159 uV, v_off_fin = 129.104 uV 14:38:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -119.617 uVs 14:38:52 Integrated Voltage Measurement... 14:39:01 Voltage Offset Correction: v_off_ini = 132.145 uV, v_off_fin = 143.847 uV 14:39:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.472 uVs 14:39:01 Integrated Voltage Measurement... 14:39:10 Voltage Offset Correction: v_off_ini = 127.495 uV, v_off_fin = 116.185 uV 14:39:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.124 uVs 14:39:10 Integrated Voltage Measurement... 14:39:19 Voltage Offset Correction: v_off_ini = 125.826 uV, v_off_fin = 118.626 uV 14:39:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.967 uVs 14:39:19 Integrated Voltage Measurement... 14:39:28 Voltage Offset Correction: v_off_ini = 134.905 uV, v_off_fin = 128.231 uV 14:39:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -133.940 uVs 14:39:28 Integrated Voltage Measurement... 14:39:37 Voltage Offset Correction: v_off_ini = 119.086 uV, v_off_fin = 118.056 uV 14:39:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.151 uVs 14:39:41 I1X Average and RMS of Measurements VT = -117.712 +- 5.949 uVs I1X = 23.543 +- 1.190 uTm 14:39:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:39:48 Integrated Voltage Measurement... 14:39:57 Voltage Offset Correction: v_off_ini = 126.664 uV, v_off_fin = 101.486 uV 14:39:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.023 uVs 14:39:57 Integrated Voltage Measurement... 14:40:06 Voltage Offset Correction: v_off_ini = 97.794 uV, v_off_fin = 100.623 uV 14:40:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.411 uVs 14:40:06 Integrated Voltage Measurement... 14:40:15 Voltage Offset Correction: v_off_ini = 109.451 uV, v_off_fin = 109.644 uV 14:40:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.814 uVs 14:40:15 Integrated Voltage Measurement... 14:40:24 Voltage Offset Correction: v_off_ini = 100.146 uV, v_off_fin = 117.087 uV 14:40:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 238.945 uVs 14:40:24 Integrated Voltage Measurement... 14:40:33 Voltage Offset Correction: v_off_ini = 113.005 uV, v_off_fin = 93.564 uV 14:40:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.840 uVs 14:40:33 Integrated Voltage Measurement... 14:40:42 Voltage Offset Correction: v_off_ini = 101.459 uV, v_off_fin = 110.219 uV 14:40:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.237 uVs 14:40:47 I1Y Average and RMS of Measurements VT = -225.712 +- 8.594 uVs I1Y = -45.142 +- 1.719 uTm 14:40:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:40:54 Integrated Voltage Measurement... 14:41:03 Voltage Offset Correction: v_off_ini = 96.731 uV, v_off_fin = 106.015 uV 14:41:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -144.231 uVs 14:41:03 Integrated Voltage Measurement... 14:41:12 Voltage Offset Correction: v_off_ini = 121.680 uV, v_off_fin = 120.490 uV 14:41:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 149.264 uVs 14:41:12 Integrated Voltage Measurement... 14:41:21 Voltage Offset Correction: v_off_ini = 129.414 uV, v_off_fin = 111.347 uV 14:41:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -161.458 uVs 14:41:21 Integrated Voltage Measurement... 14:41:30 Voltage Offset Correction: v_off_ini = 109.472 uV, v_off_fin = 118.224 uV 14:41:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 146.503 uVs 14:41:30 Integrated Voltage Measurement... 14:41:39 Voltage Offset Correction: v_off_ini = 130.788 uV, v_off_fin = 111.014 uV 14:41:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -143.225 uVs 14:41:39 Integrated Voltage Measurement... 14:41:48 Voltage Offset Correction: v_off_ini = 123.396 uV, v_off_fin = 135.718 uV 14:41:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.498 uVs 14:41:51 I2X Average and RMS of Measurements VT = -144.196 +- 9.208 uVs I2X = 13.266 +- 0.847 uTm^2 14:41:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:41:59 Integrated Voltage Measurement... 14:42:08 Voltage Offset Correction: v_off_ini = 130.683 uV, v_off_fin = 132.404 uV 14:42:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -232.117 uVs 14:42:08 Integrated Voltage Measurement... 14:42:17 Voltage Offset Correction: v_off_ini = 131.043 uV, v_off_fin = 140.182 uV 14:42:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 230.426 uVs 14:42:17 Integrated Voltage Measurement... 14:42:26 Voltage Offset Correction: v_off_ini = 136.242 uV, v_off_fin = 128.815 uV 14:42:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -227.895 uVs 14:42:26 Integrated Voltage Measurement... 14:42:35 Voltage Offset Correction: v_off_ini = 131.772 uV, v_off_fin = 129.457 uV 14:42:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 224.237 uVs 14:42:35 Integrated Voltage Measurement... 14:42:44 Voltage Offset Correction: v_off_ini = 147.014 uV, v_off_fin = 142.371 uV 14:42:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -236.024 uVs 14:42:44 Integrated Voltage Measurement... 14:42:53 Voltage Offset Correction: v_off_ini = 145.689 uV, v_off_fin = 139.115 uV 14:42:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.837 uVs 14:42:55 I2Y Average and RMS of Measurements VT = -226.256 +- 4.020 uVs I2Y = -20.816 +- 0.370 uTm^2 14:42:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm