SLAC Magnetic Measurements Date: 12-09-2024 Time: 18:16:57 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_10082 Dataset Number: 1 Run Purpose: Final Results Run Type: Stretched Wire Run Number: 6 Measurement Device: Stretched Wire Operator: YL Comment: On Axis 18:17:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:17:07 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 18:17:08 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 18:17:10 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 18:17:11 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 18:17:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:18:00 Integrated Voltage Measurement... 18:18:09 Voltage Offset Correction: v_off_ini = 577.738 uV, v_off_fin = 564.965 uV 18:18:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.271 uVs 18:18:09 Integrated Voltage Measurement... 18:18:18 Voltage Offset Correction: v_off_ini = 565.051 uV, v_off_fin = 564.068 uV 18:18:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.706 uVs 18:18:18 Integrated Voltage Measurement... 18:18:27 Voltage Offset Correction: v_off_ini = 580.926 uV, v_off_fin = 555.179 uV 18:18:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.630 uVs 18:18:27 Integrated Voltage Measurement... 18:18:36 Voltage Offset Correction: v_off_ini = 552.250 uV, v_off_fin = 553.804 uV 18:18:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.975 uVs 18:18:36 Integrated Voltage Measurement... 18:18:45 Voltage Offset Correction: v_off_ini = 551.983 uV, v_off_fin = 555.544 uV 18:18:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.972 uVs 18:18:45 Integrated Voltage Measurement... 18:18:54 Voltage Offset Correction: v_off_ini = 557.324 uV, v_off_fin = 563.807 uV 18:18:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.522 uVs 18:18:59 I1X Average and RMS of Measurements VT = -79.179 +- 2.077 uVs I1X = 15.836 +- 0.415 uTm 18:19:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:19:05 Integrated Voltage Measurement... 18:19:14 Voltage Offset Correction: v_off_ini = 547.023 uV, v_off_fin = 551.954 uV 18:19:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -193.935 uVs 18:19:14 Integrated Voltage Measurement... 18:19:23 Voltage Offset Correction: v_off_ini = 573.131 uV, v_off_fin = 577.583 uV 18:19:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 180.571 uVs 18:19:23 Integrated Voltage Measurement... 18:19:32 Voltage Offset Correction: v_off_ini = 555.513 uV, v_off_fin = 550.619 uV 18:19:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -211.690 uVs 18:19:32 Integrated Voltage Measurement... 18:19:41 Voltage Offset Correction: v_off_ini = 542.578 uV, v_off_fin = 555.574 uV 18:19:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.430 uVs 18:19:41 Integrated Voltage Measurement... 18:19:50 Voltage Offset Correction: v_off_ini = 543.247 uV, v_off_fin = 547.950 uV 18:19:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.043 uVs 18:19:50 Integrated Voltage Measurement... 18:19:59 Voltage Offset Correction: v_off_ini = 543.469 uV, v_off_fin = 563.544 uV 18:19:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.045 uVs 18:20:03 I1Y Average and RMS of Measurements VT = -192.619 +- 3.800 uVs I1Y = -38.524 +- 0.759 uTm 18:20:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:20:11 Integrated Voltage Measurement... 18:20:20 Voltage Offset Correction: v_off_ini = 559.236 uV, v_off_fin = 560.126 uV 18:20:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.934 uVs 18:20:20 Integrated Voltage Measurement... 18:20:29 Voltage Offset Correction: v_off_ini = 555.376 uV, v_off_fin = 550.524 uV 18:20:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.495 uVs 18:20:29 Integrated Voltage Measurement... 18:20:38 Voltage Offset Correction: v_off_ini = 543.862 uV, v_off_fin = 544.604 uV 18:20:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.091 uVs 18:20:38 Integrated Voltage Measurement... 18:20:47 Voltage Offset Correction: v_off_ini = 566.856 uV, v_off_fin = 560.295 uV 18:20:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.301 uVs 18:20:47 Integrated Voltage Measurement... 18:20:56 Voltage Offset Correction: v_off_ini = 540.037 uV, v_off_fin = 548.100 uV 18:20:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.688 uVs 18:20:56 Integrated Voltage Measurement... 18:21:05 Voltage Offset Correction: v_off_ini = 552.487 uV, v_off_fin = 561.091 uV 18:21:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.954 uVs 18:21:08 I2X Average and RMS of Measurements VT = -98.077 +- 4.013 uVs I2X = 9.023 +- 0.369 uTm^2 18:21:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:21:15 Integrated Voltage Measurement... 18:21:24 Voltage Offset Correction: v_off_ini = 532.751 uV, v_off_fin = 544.031 uV 18:21:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.120 uVs 18:21:24 Integrated Voltage Measurement... 18:21:33 Voltage Offset Correction: v_off_ini = 526.577 uV, v_off_fin = 540.661 uV 18:21:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.143 uVs 18:21:33 Integrated Voltage Measurement... 18:21:42 Voltage Offset Correction: v_off_ini = 539.699 uV, v_off_fin = 545.107 uV 18:21:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -296.219 uVs 18:21:42 Integrated Voltage Measurement... 18:21:51 Voltage Offset Correction: v_off_ini = 539.851 uV, v_off_fin = 550.709 uV 18:21:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.821 uVs 18:21:52 Integrated Voltage Measurement... 18:22:01 Voltage Offset Correction: v_off_ini = 537.148 uV, v_off_fin = 534.374 uV 18:22:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -274.537 uVs 18:22:01 Integrated Voltage Measurement... 18:22:10 Voltage Offset Correction: v_off_ini = 546.367 uV, v_off_fin = 528.324 uV 18:22:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 291.166 uVs 18:22:12 I2Y Average and RMS of Measurements VT = -280.834 +- 4.475 uVs I2Y = -25.837 +- 0.412 uTm^2 18:22:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:22:26 Integrated Voltage Measurement... 18:22:35 Voltage Offset Correction: v_off_ini = 547.296 uV, v_off_fin = 539.478 uV 18:22:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -83.588 uVs 18:22:35 Integrated Voltage Measurement... 18:22:44 Voltage Offset Correction: v_off_ini = 524.506 uV, v_off_fin = 544.951 uV 18:22:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.986 uVs 18:22:44 Integrated Voltage Measurement... 18:22:53 Voltage Offset Correction: v_off_ini = 536.968 uV, v_off_fin = 542.187 uV 18:22:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.496 uVs 18:22:53 Integrated Voltage Measurement... 18:23:02 Voltage Offset Correction: v_off_ini = 547.251 uV, v_off_fin = 538.304 uV 18:23:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.991 uVs 18:23:02 Integrated Voltage Measurement... 18:23:11 Voltage Offset Correction: v_off_ini = 541.839 uV, v_off_fin = 537.572 uV 18:23:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.831 uVs 18:23:11 Integrated Voltage Measurement... 18:23:20 Voltage Offset Correction: v_off_ini = 538.181 uV, v_off_fin = 528.861 uV 18:23:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.006 uVs 18:23:24 I1X Average and RMS of Measurements VT = -77.816 +- 5.617 uVs I1X = 15.564 +- 1.124 uTm 18:23:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:23:31 Integrated Voltage Measurement... 18:23:40 Voltage Offset Correction: v_off_ini = 504.738 uV, v_off_fin = 502.695 uV 18:23:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -173.875 uVs 18:23:40 Integrated Voltage Measurement... 18:23:49 Voltage Offset Correction: v_off_ini = 520.350 uV, v_off_fin = 524.072 uV 18:23:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 194.565 uVs 18:23:49 Integrated Voltage Measurement... 18:23:58 Voltage Offset Correction: v_off_ini = 520.253 uV, v_off_fin = 513.302 uV 18:23:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -200.472 uVs 18:23:58 Integrated Voltage Measurement... 18:24:07 Voltage Offset Correction: v_off_ini = 502.055 uV, v_off_fin = 512.712 uV 18:24:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 181.184 uVs 18:24:07 Integrated Voltage Measurement... 18:24:16 Voltage Offset Correction: v_off_ini = 511.221 uV, v_off_fin = 506.924 uV 18:24:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.384 uVs 18:24:16 Integrated Voltage Measurement... 18:24:25 Voltage Offset Correction: v_off_ini = 506.238 uV, v_off_fin = 524.679 uV 18:24:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.300 uVs 18:24:29 I1Y Average and RMS of Measurements VT = -185.297 +- 4.147 uVs I1Y = -37.057 +- 0.829 uTm 18:24:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:24:37 Integrated Voltage Measurement... 18:24:46 Voltage Offset Correction: v_off_ini = 495.424 uV, v_off_fin = 504.814 uV 18:24:46 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -93.951 uVs 18:24:46 Integrated Voltage Measurement... 18:24:55 Voltage Offset Correction: v_off_ini = 521.982 uV, v_off_fin = 520.222 uV 18:24:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.405 uVs 18:24:55 Integrated Voltage Measurement... 18:25:04 Voltage Offset Correction: v_off_ini = 511.106 uV, v_off_fin = 514.924 uV 18:25:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.315 uVs 18:25:04 Integrated Voltage Measurement... 18:25:13 Voltage Offset Correction: v_off_ini = 537.443 uV, v_off_fin = 532.765 uV 18:25:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.380 uVs 18:25:13 Integrated Voltage Measurement... 18:25:22 Voltage Offset Correction: v_off_ini = 540.954 uV, v_off_fin = 523.916 uV 18:25:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.180 uVs 18:25:22 Integrated Voltage Measurement... 18:25:31 Voltage Offset Correction: v_off_ini = 518.876 uV, v_off_fin = 528.293 uV 18:25:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.077 uVs 18:25:33 I2X Average and RMS of Measurements VT = -90.218 +- 6.406 uVs I2X = 8.300 +- 0.589 uTm^2 18:25:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:25:41 Integrated Voltage Measurement... 18:25:50 Voltage Offset Correction: v_off_ini = 516.182 uV, v_off_fin = 522.993 uV 18:25:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.821 uVs 18:25:50 Integrated Voltage Measurement... 18:25:59 Voltage Offset Correction: v_off_ini = 523.113 uV, v_off_fin = 517.768 uV 18:25:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.189 uVs 18:25:59 Integrated Voltage Measurement... 18:26:08 Voltage Offset Correction: v_off_ini = 518.254 uV, v_off_fin = 521.207 uV 18:26:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.191 uVs 18:26:08 Integrated Voltage Measurement... 18:26:17 Voltage Offset Correction: v_off_ini = 525.452 uV, v_off_fin = 529.470 uV 18:26:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.765 uVs 18:26:17 Integrated Voltage Measurement... 18:26:26 Voltage Offset Correction: v_off_ini = 529.994 uV, v_off_fin = 520.989 uV 18:26:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -276.690 uVs 18:26:26 Integrated Voltage Measurement... 18:26:35 Voltage Offset Correction: v_off_ini = 515.365 uV, v_off_fin = 507.742 uV 18:26:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 281.952 uVs 18:26:37 I2Y Average and RMS of Measurements VT = -269.101 +- 7.775 uVs I2Y = -24.757 +- 0.715 uTm^2 18:26:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:27:00 Integrated Voltage Measurement... 18:27:09 Voltage Offset Correction: v_off_ini = 533.302 uV, v_off_fin = 540.487 uV 18:27:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -70.731 uVs 18:27:09 Integrated Voltage Measurement... 18:27:18 Voltage Offset Correction: v_off_ini = 521.868 uV, v_off_fin = 527.784 uV 18:27:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.474 uVs 18:27:18 Integrated Voltage Measurement... 18:27:27 Voltage Offset Correction: v_off_ini = 535.085 uV, v_off_fin = 529.476 uV 18:27:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.675 uVs 18:27:27 Integrated Voltage Measurement... 18:27:36 Voltage Offset Correction: v_off_ini = 522.177 uV, v_off_fin = 546.240 uV 18:27:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.580 uVs 18:27:36 Integrated Voltage Measurement... 18:27:45 Voltage Offset Correction: v_off_ini = 537.330 uV, v_off_fin = 532.966 uV 18:27:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.083 uVs 18:27:45 Integrated Voltage Measurement... 18:27:54 Voltage Offset Correction: v_off_ini = 544.825 uV, v_off_fin = 538.233 uV 18:27:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.081 uVs 18:27:59 I1X Average and RMS of Measurements VT = -89.937 +- 7.874 uVs I1X = 17.988 +- 1.574 uTm 18:28:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:28:05 Integrated Voltage Measurement... 18:28:14 Voltage Offset Correction: v_off_ini = 530.264 uV, v_off_fin = 531.584 uV 18:28:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.791 uVs 18:28:14 Integrated Voltage Measurement... 18:28:23 Voltage Offset Correction: v_off_ini = 536.303 uV, v_off_fin = 550.199 uV 18:28:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 179.674 uVs 18:28:23 Integrated Voltage Measurement... 18:28:32 Voltage Offset Correction: v_off_ini = 531.476 uV, v_off_fin = 535.635 uV 18:28:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.528 uVs 18:28:32 Integrated Voltage Measurement... 18:28:41 Voltage Offset Correction: v_off_ini = 544.960 uV, v_off_fin = 560.613 uV 18:28:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 184.894 uVs 18:28:41 Integrated Voltage Measurement... 18:28:50 Voltage Offset Correction: v_off_ini = 555.361 uV, v_off_fin = 541.954 uV 18:28:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -197.696 uVs 18:28:50 Integrated Voltage Measurement... 18:28:59 Voltage Offset Correction: v_off_ini = 538.978 uV, v_off_fin = 546.827 uV 18:28:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 179.864 uVs 18:29:03 I1Y Average and RMS of Measurements VT = -183.241 +- 3.966 uVs I1Y = -36.646 +- 0.792 uTm 18:29:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:29:11 Integrated Voltage Measurement... 18:29:20 Voltage Offset Correction: v_off_ini = 514.804 uV, v_off_fin = 534.156 uV 18:29:20 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -112.986 uVs 18:29:20 Integrated Voltage Measurement... 18:29:29 Voltage Offset Correction: v_off_ini = 521.664 uV, v_off_fin = 527.124 uV 18:29:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.644 uVs 18:29:29 Integrated Voltage Measurement... 18:29:38 Voltage Offset Correction: v_off_ini = 538.496 uV, v_off_fin = 517.555 uV 18:29:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.214 uVs 18:29:38 Integrated Voltage Measurement... 18:29:47 Voltage Offset Correction: v_off_ini = 516.315 uV, v_off_fin = 530.191 uV 18:29:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.516 uVs 18:29:47 Integrated Voltage Measurement... 18:29:56 Voltage Offset Correction: v_off_ini = 513.342 uV, v_off_fin = 515.360 uV 18:29:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.768 uVs 18:29:56 Integrated Voltage Measurement... 18:30:05 Voltage Offset Correction: v_off_ini = 534.189 uV, v_off_fin = 538.664 uV 18:30:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.040 uVs 18:30:08 I2X Average and RMS of Measurements VT = -94.695 +- 7.522 uVs I2X = 8.712 +- 0.692 uTm^2 18:30:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:30:15 Integrated Voltage Measurement... 18:30:24 Voltage Offset Correction: v_off_ini = 535.469 uV, v_off_fin = 551.970 uV 18:30:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -261.020 uVs 18:30:24 Integrated Voltage Measurement... 18:30:33 Voltage Offset Correction: v_off_ini = 542.798 uV, v_off_fin = 546.050 uV 18:30:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.954 uVs 18:30:33 Integrated Voltage Measurement... 18:30:42 Voltage Offset Correction: v_off_ini = 547.964 uV, v_off_fin = 539.655 uV 18:30:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.517 uVs 18:30:42 Integrated Voltage Measurement... 18:30:51 Voltage Offset Correction: v_off_ini = 550.434 uV, v_off_fin = 545.637 uV 18:30:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.059 uVs 18:30:51 Integrated Voltage Measurement... 18:31:00 Voltage Offset Correction: v_off_ini = 540.751 uV, v_off_fin = 544.452 uV 18:31:00 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -259.905 uVs 18:31:00 Integrated Voltage Measurement... 18:31:09 Voltage Offset Correction: v_off_ini = 549.626 uV, v_off_fin = 541.670 uV 18:31:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 280.303 uVs 18:31:12 I2Y Average and RMS of Measurements VT = -262.460 +- 5.517 uVs I2Y = -24.147 +- 0.508 uTm^2 18:31:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:31:26 Integrated Voltage Measurement... 18:31:35 Voltage Offset Correction: v_off_ini = 562.670 uV, v_off_fin = 561.982 uV 18:31:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.648 uVs 18:31:35 Integrated Voltage Measurement... 18:31:44 Voltage Offset Correction: v_off_ini = 546.055 uV, v_off_fin = 539.841 uV 18:31:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.149 uVs 18:31:44 Integrated Voltage Measurement... 18:31:53 Voltage Offset Correction: v_off_ini = 538.848 uV, v_off_fin = 530.526 uV 18:31:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.542 uVs 18:31:53 Integrated Voltage Measurement... 18:32:02 Voltage Offset Correction: v_off_ini = 515.987 uV, v_off_fin = 546.947 uV 18:32:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.374 uVs 18:32:02 Integrated Voltage Measurement... 18:32:11 Voltage Offset Correction: v_off_ini = 565.774 uV, v_off_fin = 560.905 uV 18:32:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.975 uVs 18:32:11 Integrated Voltage Measurement... 18:32:20 Voltage Offset Correction: v_off_ini = 560.058 uV, v_off_fin = 552.536 uV 18:32:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.301 uVs 18:32:24 I1X Average and RMS of Measurements VT = -88.332 +- 1.860 uVs I1X = 17.667 +- 0.372 uTm 18:32:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:32:31 Integrated Voltage Measurement... 18:32:40 Voltage Offset Correction: v_off_ini = 527.748 uV, v_off_fin = 501.892 uV 18:32:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -192.127 uVs 18:32:40 Integrated Voltage Measurement... 18:32:49 Voltage Offset Correction: v_off_ini = 517.443 uV, v_off_fin = 530.474 uV 18:32:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 178.201 uVs 18:32:49 Integrated Voltage Measurement... 18:32:58 Voltage Offset Correction: v_off_ini = 520.480 uV, v_off_fin = 513.464 uV 18:32:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -165.155 uVs 18:32:58 Integrated Voltage Measurement... 18:33:07 Voltage Offset Correction: v_off_ini = 516.189 uV, v_off_fin = 527.331 uV 18:33:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 194.895 uVs 18:33:07 Integrated Voltage Measurement... 18:33:16 Voltage Offset Correction: v_off_ini = 520.704 uV, v_off_fin = 514.374 uV 18:33:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.254 uVs 18:33:16 Integrated Voltage Measurement... 18:33:24 Voltage Offset Correction: v_off_ini = 503.325 uV, v_off_fin = 526.994 uV 18:33:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.686 uVs 18:33:29 I1Y Average and RMS of Measurements VT = -184.720 +- 3.665 uVs I1Y = -36.942 +- 0.733 uTm 18:33:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:33:36 Integrated Voltage Measurement... 18:33:45 Voltage Offset Correction: v_off_ini = 520.626 uV, v_off_fin = 521.203 uV 18:33:45 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -81.749 uVs 18:33:45 Integrated Voltage Measurement... 18:33:54 Voltage Offset Correction: v_off_ini = 518.423 uV, v_off_fin = 540.380 uV 18:33:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.547 uVs 18:33:54 Integrated Voltage Measurement... 18:34:03 Voltage Offset Correction: v_off_ini = 516.370 uV, v_off_fin = 513.907 uV 18:34:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.552 uVs 18:34:03 Integrated Voltage Measurement... 18:34:12 Voltage Offset Correction: v_off_ini = 512.636 uV, v_off_fin = 526.425 uV 18:34:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 119.766 uVs 18:34:13 Integrated Voltage Measurement... 18:34:21 Voltage Offset Correction: v_off_ini = 521.958 uV, v_off_fin = 522.023 uV 18:34:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.968 uVs 18:34:22 Integrated Voltage Measurement... 18:34:30 Voltage Offset Correction: v_off_ini = 534.485 uV, v_off_fin = 539.213 uV 18:34:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.540 uVs 18:34:33 I2X Average and RMS of Measurements VT = -97.353 +- 5.299 uVs I2X = 8.956 +- 0.488 uTm^2 18:34:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:34:41 Integrated Voltage Measurement... 18:34:50 Voltage Offset Correction: v_off_ini = 533.421 uV, v_off_fin = 534.526 uV 18:34:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.103 uVs 18:34:50 Integrated Voltage Measurement... 18:34:59 Voltage Offset Correction: v_off_ini = 531.330 uV, v_off_fin = 535.559 uV 18:34:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.890 uVs 18:34:59 Integrated Voltage Measurement... 18:35:08 Voltage Offset Correction: v_off_ini = 539.001 uV, v_off_fin = 545.744 uV 18:35:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.127 uVs 18:35:08 Integrated Voltage Measurement... 18:35:17 Voltage Offset Correction: v_off_ini = 549.121 uV, v_off_fin = 542.802 uV 18:35:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.185 uVs 18:35:17 Integrated Voltage Measurement... 18:35:26 Voltage Offset Correction: v_off_ini = 535.171 uV, v_off_fin = 544.599 uV 18:35:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -231.984 uVs 18:35:26 Integrated Voltage Measurement... 18:35:35 Voltage Offset Correction: v_off_ini = 540.682 uV, v_off_fin = 560.387 uV 18:35:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.032 uVs 18:35:37 I2Y Average and RMS of Measurements VT = -250.053 +- 13.957 uVs I2Y = -23.005 +- 1.284 uTm^2 18:35:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:35:57 Integrated Voltage Measurement... 18:36:06 Voltage Offset Correction: v_off_ini = 560.949 uV, v_off_fin = 552.743 uV 18:36:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -71.540 uVs 18:36:06 Integrated Voltage Measurement... 18:36:15 Voltage Offset Correction: v_off_ini = 538.470 uV, v_off_fin = 552.153 uV 18:36:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.751 uVs 18:36:15 Integrated Voltage Measurement... 18:36:24 Voltage Offset Correction: v_off_ini = 550.677 uV, v_off_fin = 546.878 uV 18:36:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.925 uVs 18:36:24 Integrated Voltage Measurement... 18:36:33 Voltage Offset Correction: v_off_ini = 562.410 uV, v_off_fin = 554.275 uV 18:36:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.843 uVs 18:36:33 Integrated Voltage Measurement... 18:36:42 Voltage Offset Correction: v_off_ini = 558.355 uV, v_off_fin = 531.200 uV 18:36:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.279 uVs 18:36:42 Integrated Voltage Measurement... 18:36:51 Voltage Offset Correction: v_off_ini = 536.976 uV, v_off_fin = 531.477 uV 18:36:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.386 uVs 18:36:55 I1X Average and RMS of Measurements VT = -79.454 +- 5.595 uVs I1X = 15.891 +- 1.119 uTm 18:36:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:37:02 Integrated Voltage Measurement... 18:37:11 Voltage Offset Correction: v_off_ini = 532.937 uV, v_off_fin = 522.743 uV 18:37:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.801 uVs 18:37:11 Integrated Voltage Measurement... 18:37:20 Voltage Offset Correction: v_off_ini = 529.111 uV, v_off_fin = 537.269 uV 18:37:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 178.743 uVs 18:37:20 Integrated Voltage Measurement... 18:37:29 Voltage Offset Correction: v_off_ini = 534.513 uV, v_off_fin = 518.381 uV 18:37:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -171.938 uVs 18:37:29 Integrated Voltage Measurement... 18:37:37 Voltage Offset Correction: v_off_ini = 527.946 uV, v_off_fin = 532.530 uV 18:37:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 196.196 uVs 18:37:38 Integrated Voltage Measurement... 18:37:46 Voltage Offset Correction: v_off_ini = 544.526 uV, v_off_fin = 528.646 uV 18:37:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.347 uVs 18:37:47 Integrated Voltage Measurement... 18:37:55 Voltage Offset Correction: v_off_ini = 546.515 uV, v_off_fin = 537.267 uV 18:37:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.882 uVs 18:38:00 I1Y Average and RMS of Measurements VT = -185.151 +- 1.863 uVs I1Y = -37.027 +- 0.373 uTm 18:38:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:38:07 Integrated Voltage Measurement... 18:38:16 Voltage Offset Correction: v_off_ini = 513.372 uV, v_off_fin = 524.851 uV 18:38:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.622 uVs 18:38:16 Integrated Voltage Measurement... 18:38:25 Voltage Offset Correction: v_off_ini = 533.272 uV, v_off_fin = 548.887 uV 18:38:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.839 uVs 18:38:25 Integrated Voltage Measurement... 18:38:34 Voltage Offset Correction: v_off_ini = 528.157 uV, v_off_fin = 522.841 uV 18:38:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.918 uVs 18:38:34 Integrated Voltage Measurement... 18:38:43 Voltage Offset Correction: v_off_ini = 511.249 uV, v_off_fin = 522.983 uV 18:38:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.560 uVs 18:38:43 Integrated Voltage Measurement... 18:38:52 Voltage Offset Correction: v_off_ini = 528.224 uV, v_off_fin = 502.204 uV 18:38:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.571 uVs 18:38:52 Integrated Voltage Measurement... 18:39:01 Voltage Offset Correction: v_off_ini = 515.995 uV, v_off_fin = 535.773 uV 18:39:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.116 uVs 18:39:04 I2X Average and RMS of Measurements VT = -96.438 +- 6.625 uVs I2X = 8.872 +- 0.609 uTm^2 18:39:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:39:12 Integrated Voltage Measurement... 18:39:21 Voltage Offset Correction: v_off_ini = 528.650 uV, v_off_fin = 532.227 uV 18:39:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.767 uVs 18:39:21 Integrated Voltage Measurement... 18:39:30 Voltage Offset Correction: v_off_ini = 527.889 uV, v_off_fin = 541.498 uV 18:39:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.834 uVs 18:39:30 Integrated Voltage Measurement... 18:39:39 Voltage Offset Correction: v_off_ini = 511.114 uV, v_off_fin = 508.180 uV 18:39:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.464 uVs 18:39:39 Integrated Voltage Measurement... 18:39:48 Voltage Offset Correction: v_off_ini = 505.070 uV, v_off_fin = 503.572 uV 18:39:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.501 uVs 18:39:48 Integrated Voltage Measurement... 18:39:57 Voltage Offset Correction: v_off_ini = 526.057 uV, v_off_fin = 528.597 uV 18:39:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.073 uVs 18:39:57 Integrated Voltage Measurement... 18:40:06 Voltage Offset Correction: v_off_ini = 518.847 uV, v_off_fin = 509.312 uV 18:40:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.323 uVs 18:40:08 I2Y Average and RMS of Measurements VT = -244.494 +- 2.541 uVs I2Y = -22.493 +- 0.234 uTm^2 18:40:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:40:22 Integrated Voltage Measurement... 18:40:31 Voltage Offset Correction: v_off_ini = 513.333 uV, v_off_fin = 494.251 uV 18:40:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -95.491 uVs 18:40:31 Integrated Voltage Measurement... 18:40:40 Voltage Offset Correction: v_off_ini = 495.641 uV, v_off_fin = 506.861 uV 18:40:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.404 uVs 18:40:40 Integrated Voltage Measurement... 18:40:49 Voltage Offset Correction: v_off_ini = 499.464 uV, v_off_fin = 489.219 uV 18:40:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -93.983 uVs 18:40:49 Integrated Voltage Measurement... 18:40:58 Voltage Offset Correction: v_off_ini = 486.598 uV, v_off_fin = 500.184 uV 18:40:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.843 uVs 18:40:58 Integrated Voltage Measurement... 18:41:07 Voltage Offset Correction: v_off_ini = 506.539 uV, v_off_fin = 496.430 uV 18:41:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.874 uVs 18:41:07 Integrated Voltage Measurement... 18:41:16 Voltage Offset Correction: v_off_ini = 510.935 uV, v_off_fin = 509.311 uV 18:41:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.549 uVs 18:41:20 I1X Average and RMS of Measurements VT = -86.524 +- 6.384 uVs I1X = 17.305 +- 1.277 uTm 18:41:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:41:27 Integrated Voltage Measurement... 18:41:36 Voltage Offset Correction: v_off_ini = 516.163 uV, v_off_fin = 490.966 uV 18:41:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -182.362 uVs 18:41:36 Integrated Voltage Measurement... 18:41:45 Voltage Offset Correction: v_off_ini = 496.550 uV, v_off_fin = 514.538 uV 18:41:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 181.782 uVs 18:41:45 Integrated Voltage Measurement... 18:41:54 Voltage Offset Correction: v_off_ini = 514.136 uV, v_off_fin = 507.799 uV 18:41:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.886 uVs 18:41:54 Integrated Voltage Measurement... 18:42:03 Voltage Offset Correction: v_off_ini = 514.741 uV, v_off_fin = 526.489 uV 18:42:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.405 uVs 18:42:03 Integrated Voltage Measurement... 18:42:12 Voltage Offset Correction: v_off_ini = 507.029 uV, v_off_fin = 506.329 uV 18:42:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.742 uVs 18:42:12 Integrated Voltage Measurement... 18:42:21 Voltage Offset Correction: v_off_ini = 478.350 uV, v_off_fin = 506.590 uV 18:42:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 186.299 uVs 18:42:25 I1Y Average and RMS of Measurements VT = -184.746 +- 3.469 uVs I1Y = -36.949 +- 0.694 uTm 18:42:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:42:33 Integrated Voltage Measurement... 18:42:42 Voltage Offset Correction: v_off_ini = 486.764 uV, v_off_fin = 485.754 uV 18:42:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.286 uVs 18:42:42 Integrated Voltage Measurement... 18:42:51 Voltage Offset Correction: v_off_ini = 501.364 uV, v_off_fin = 520.869 uV 18:42:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.894 uVs 18:42:51 Integrated Voltage Measurement... 18:43:00 Voltage Offset Correction: v_off_ini = 491.059 uV, v_off_fin = 481.918 uV 18:43:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.561 uVs 18:43:00 Integrated Voltage Measurement... 18:43:09 Voltage Offset Correction: v_off_ini = 506.595 uV, v_off_fin = 512.809 uV 18:43:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.719 uVs 18:43:09 Integrated Voltage Measurement... 18:43:18 Voltage Offset Correction: v_off_ini = 501.846 uV, v_off_fin = 486.135 uV 18:43:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.541 uVs 18:43:18 Integrated Voltage Measurement... 18:43:27 Voltage Offset Correction: v_off_ini = 478.432 uV, v_off_fin = 500.849 uV 18:43:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.213 uVs 18:43:29 I2X Average and RMS of Measurements VT = -99.202 +- 5.617 uVs I2X = 9.127 +- 0.517 uTm^2 18:43:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:43:37 Integrated Voltage Measurement... 18:43:46 Voltage Offset Correction: v_off_ini = 526.468 uV, v_off_fin = 518.344 uV 18:43:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.927 uVs 18:43:46 Integrated Voltage Measurement... 18:43:55 Voltage Offset Correction: v_off_ini = 504.768 uV, v_off_fin = 502.910 uV 18:43:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 238.911 uVs 18:43:55 Integrated Voltage Measurement... 18:44:04 Voltage Offset Correction: v_off_ini = 506.595 uV, v_off_fin = 506.865 uV 18:44:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.441 uVs 18:44:04 Integrated Voltage Measurement... 18:44:13 Voltage Offset Correction: v_off_ini = 500.692 uV, v_off_fin = 503.003 uV 18:44:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 229.998 uVs 18:44:13 Integrated Voltage Measurement... 18:44:22 Voltage Offset Correction: v_off_ini = 504.475 uV, v_off_fin = 486.992 uV 18:44:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.160 uVs 18:44:22 Integrated Voltage Measurement... 18:44:31 Voltage Offset Correction: v_off_ini = 496.978 uV, v_off_fin = 481.422 uV 18:44:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0001 mm dXY = -1.0000 mm VT = 228.427 uVs 18:44:33 I2Y Average and RMS of Measurements VT = -240.311 +- 4.777 uVs I2Y = -22.109 +- 0.439 uTm^2 18:44:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:44:47 Integrated Voltage Measurement... 18:44:56 Voltage Offset Correction: v_off_ini = 501.800 uV, v_off_fin = 500.355 uV 18:44:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -90.642 uVs 18:44:56 Integrated Voltage Measurement... 18:45:05 Voltage Offset Correction: v_off_ini = 488.295 uV, v_off_fin = 485.582 uV 18:45:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.474 uVs 18:45:05 Integrated Voltage Measurement... 18:45:14 Voltage Offset Correction: v_off_ini = 474.148 uV, v_off_fin = 495.131 uV 18:45:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.253 uVs 18:45:14 Integrated Voltage Measurement... 18:45:23 Voltage Offset Correction: v_off_ini = 471.158 uV, v_off_fin = 473.575 uV 18:45:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.028 uVs 18:45:23 Integrated Voltage Measurement... 18:45:32 Voltage Offset Correction: v_off_ini = 488.293 uV, v_off_fin = 499.058 uV 18:45:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -102.068 uVs 18:45:32 Integrated Voltage Measurement... 18:45:41 Voltage Offset Correction: v_off_ini = 493.015 uV, v_off_fin = 480.175 uV 18:45:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.054 uVs 18:45:46 I1X Average and RMS of Measurements VT = -86.920 +- 0.823 uVs I1X = 17.384 +- 0.165 uTm 18:45:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:45:52 Integrated Voltage Measurement... 18:46:01 Voltage Offset Correction: v_off_ini = 464.337 uV, v_off_fin = 449.709 uV 18:46:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -184.383 uVs 18:46:01 Integrated Voltage Measurement... 18:46:10 Voltage Offset Correction: v_off_ini = 472.722 uV, v_off_fin = 485.303 uV 18:46:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 189.435 uVs 18:46:10 Integrated Voltage Measurement... 18:46:19 Voltage Offset Correction: v_off_ini = 483.590 uV, v_off_fin = 495.612 uV 18:46:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.939 uVs 18:46:19 Integrated Voltage Measurement... 18:46:28 Voltage Offset Correction: v_off_ini = 490.066 uV, v_off_fin = 495.917 uV 18:46:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.098 uVs 18:46:28 Integrated Voltage Measurement... 18:46:37 Voltage Offset Correction: v_off_ini = 433.086 uV, v_off_fin = 440.244 uV 18:46:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.269 uVs 18:46:37 Integrated Voltage Measurement... 18:46:46 Voltage Offset Correction: v_off_ini = 467.953 uV, v_off_fin = 465.569 uV 18:46:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.264 uVs 18:46:50 I1Y Average and RMS of Measurements VT = -180.565 +- 4.624 uVs I1Y = -36.112 +- 0.925 uTm 18:46:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:46:58 Integrated Voltage Measurement... 18:47:07 Voltage Offset Correction: v_off_ini = 456.650 uV, v_off_fin = 462.415 uV 18:47:07 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -85.067 uVs 18:47:07 Integrated Voltage Measurement... 18:47:16 Voltage Offset Correction: v_off_ini = 467.441 uV, v_off_fin = 487.052 uV 18:47:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 110.543 uVs 18:47:16 Integrated Voltage Measurement... 18:47:25 Voltage Offset Correction: v_off_ini = 486.819 uV, v_off_fin = 485.690 uV 18:47:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -127.331 uVs 18:47:25 Integrated Voltage Measurement... 18:47:34 Voltage Offset Correction: v_off_ini = 486.819 uV, v_off_fin = 481.498 uV 18:47:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.301 uVs 18:47:34 Integrated Voltage Measurement... 18:47:43 Voltage Offset Correction: v_off_ini = 484.706 uV, v_off_fin = 481.449 uV 18:47:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.204 uVs 18:47:43 Integrated Voltage Measurement... 18:47:52 Voltage Offset Correction: v_off_ini = 499.800 uV, v_off_fin = 509.511 uV 18:47:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.514 uVs 18:47:55 I2X Average and RMS of Measurements VT = -99.327 +- 12.481 uVs I2X = 9.138 +- 1.148 uTm^2 18:48:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:48:02 Integrated Voltage Measurement... 18:48:11 Voltage Offset Correction: v_off_ini = 487.571 uV, v_off_fin = 480.764 uV 18:48:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.647 uVs 18:48:11 Integrated Voltage Measurement... 18:48:20 Voltage Offset Correction: v_off_ini = 473.806 uV, v_off_fin = 479.285 uV 18:48:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.193 uVs 18:48:20 Integrated Voltage Measurement... 18:48:29 Voltage Offset Correction: v_off_ini = 472.975 uV, v_off_fin = 474.287 uV 18:48:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.398 uVs 18:48:29 Integrated Voltage Measurement... 18:48:38 Voltage Offset Correction: v_off_ini = 486.039 uV, v_off_fin = 467.725 uV 18:48:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 247.075 uVs 18:48:38 Integrated Voltage Measurement... 18:48:47 Voltage Offset Correction: v_off_ini = 476.464 uV, v_off_fin = 481.560 uV 18:48:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.509 uVs 18:48:47 Integrated Voltage Measurement... 18:48:56 Voltage Offset Correction: v_off_ini = 473.983 uV, v_off_fin = 481.396 uV 18:48:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.114 uVs 18:48:59 I2Y Average and RMS of Measurements VT = -244.323 +- 1.346 uVs I2Y = -22.478 +- 0.124 uTm^2 18:49:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:49:13 Integrated Voltage Measurement... 18:49:22 Voltage Offset Correction: v_off_ini = 447.790 uV, v_off_fin = 461.430 uV 18:49:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -79.725 uVs 18:49:22 Integrated Voltage Measurement... 18:49:31 Voltage Offset Correction: v_off_ini = 471.492 uV, v_off_fin = 467.271 uV 18:49:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.114 uVs 18:49:31 Integrated Voltage Measurement... 18:49:40 Voltage Offset Correction: v_off_ini = 466.280 uV, v_off_fin = 469.166 uV 18:49:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.986 uVs 18:49:40 Integrated Voltage Measurement... 18:49:49 Voltage Offset Correction: v_off_ini = 457.237 uV, v_off_fin = 461.129 uV 18:49:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.841 uVs 18:49:49 Integrated Voltage Measurement... 18:49:58 Voltage Offset Correction: v_off_ini = 448.715 uV, v_off_fin = 469.798 uV 18:49:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.733 uVs 18:49:58 Integrated Voltage Measurement... 18:50:07 Voltage Offset Correction: v_off_ini = 462.160 uV, v_off_fin = 464.299 uV 18:50:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.196 uVs 18:50:11 I1X Average and RMS of Measurements VT = -88.766 +- 4.646 uVs I1X = 17.753 +- 0.929 uTm 18:50:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:50:18 Integrated Voltage Measurement... 18:50:27 Voltage Offset Correction: v_off_ini = 452.455 uV, v_off_fin = 439.463 uV 18:50:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -185.340 uVs 18:50:27 Integrated Voltage Measurement... 18:50:36 Voltage Offset Correction: v_off_ini = 441.649 uV, v_off_fin = 449.991 uV 18:50:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 168.767 uVs 18:50:36 Integrated Voltage Measurement... 18:50:45 Voltage Offset Correction: v_off_ini = 435.924 uV, v_off_fin = 441.648 uV 18:50:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.394 uVs 18:50:45 Integrated Voltage Measurement... 18:50:54 Voltage Offset Correction: v_off_ini = 468.227 uV, v_off_fin = 467.120 uV 18:50:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.880 uVs 18:50:54 Integrated Voltage Measurement... 18:51:03 Voltage Offset Correction: v_off_ini = 487.823 uV, v_off_fin = 453.942 uV 18:51:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.524 uVs 18:51:03 Integrated Voltage Measurement... 18:51:12 Voltage Offset Correction: v_off_ini = 465.032 uV, v_off_fin = 462.254 uV 18:51:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 179.185 uVs 18:51:16 I1Y Average and RMS of Measurements VT = -187.348 +- 8.614 uVs I1Y = -37.469 +- 1.723 uTm 18:51:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:51:23 Integrated Voltage Measurement... 18:51:32 Voltage Offset Correction: v_off_ini = 445.569 uV, v_off_fin = 443.992 uV 18:51:32 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -89.987 uVs 18:51:33 Integrated Voltage Measurement... 18:51:41 Voltage Offset Correction: v_off_ini = 443.684 uV, v_off_fin = 455.984 uV 18:51:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.997 uVs 18:51:42 Integrated Voltage Measurement... 18:51:50 Voltage Offset Correction: v_off_ini = 461.372 uV, v_off_fin = 451.486 uV 18:51:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -121.992 uVs 18:51:51 Integrated Voltage Measurement... 18:52:00 Voltage Offset Correction: v_off_ini = 444.902 uV, v_off_fin = 467.848 uV 18:52:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.343 uVs 18:52:00 Integrated Voltage Measurement... 18:52:09 Voltage Offset Correction: v_off_ini = 456.221 uV, v_off_fin = 446.618 uV 18:52:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.196 uVs 18:52:09 Integrated Voltage Measurement... 18:52:18 Voltage Offset Correction: v_off_ini = 453.382 uV, v_off_fin = 464.304 uV 18:52:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.366 uVs 18:52:20 I2X Average and RMS of Measurements VT = -97.980 +- 12.422 uVs I2X = 9.014 +- 1.143 uTm^2 18:52:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:52:28 Integrated Voltage Measurement... 18:52:37 Voltage Offset Correction: v_off_ini = 455.890 uV, v_off_fin = 456.734 uV 18:52:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.704 uVs 18:52:37 Integrated Voltage Measurement... 18:52:46 Voltage Offset Correction: v_off_ini = 466.827 uV, v_off_fin = 450.404 uV 18:52:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 232.546 uVs 18:52:46 Integrated Voltage Measurement... 18:52:55 Voltage Offset Correction: v_off_ini = 445.937 uV, v_off_fin = 458.319 uV 18:52:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.878 uVs 18:52:55 Integrated Voltage Measurement... 18:53:04 Voltage Offset Correction: v_off_ini = 451.212 uV, v_off_fin = 443.498 uV 18:53:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.499 uVs 18:53:04 Integrated Voltage Measurement... 18:53:13 Voltage Offset Correction: v_off_ini = 459.671 uV, v_off_fin = 456.811 uV 18:53:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.023 uVs 18:53:13 Integrated Voltage Measurement... 18:53:22 Voltage Offset Correction: v_off_ini = 451.690 uV, v_off_fin = 451.159 uV 18:53:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.965 uVs 18:53:24 I2Y Average and RMS of Measurements VT = -241.602 +- 6.001 uVs I2Y = -22.227 +- 0.552 uTm^2 18:53:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:53:38 Integrated Voltage Measurement... 18:53:47 Voltage Offset Correction: v_off_ini = 449.716 uV, v_off_fin = 436.136 uV 18:53:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -91.102 uVs 18:53:47 Integrated Voltage Measurement... 18:53:56 Voltage Offset Correction: v_off_ini = 440.224 uV, v_off_fin = 453.721 uV 18:53:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.480 uVs 18:53:56 Integrated Voltage Measurement... 18:54:05 Voltage Offset Correction: v_off_ini = 437.537 uV, v_off_fin = 448.262 uV 18:54:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.046 uVs 18:54:05 Integrated Voltage Measurement... 18:54:14 Voltage Offset Correction: v_off_ini = 449.048 uV, v_off_fin = 459.588 uV 18:54:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.480 uVs 18:54:14 Integrated Voltage Measurement... 18:54:23 Voltage Offset Correction: v_off_ini = 444.571 uV, v_off_fin = 459.770 uV 18:54:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.012 uVs 18:54:23 Integrated Voltage Measurement... 18:54:32 Voltage Offset Correction: v_off_ini = 461.568 uV, v_off_fin = 470.738 uV 18:54:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.065 uVs 18:54:36 I1X Average and RMS of Measurements VT = -83.364 +- 6.840 uVs I1X = 16.673 +- 1.368 uTm 18:54:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:54:43 Integrated Voltage Measurement... 18:54:52 Voltage Offset Correction: v_off_ini = 424.562 uV, v_off_fin = 434.579 uV 18:54:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -176.324 uVs 18:54:52 Integrated Voltage Measurement... 18:55:01 Voltage Offset Correction: v_off_ini = 446.085 uV, v_off_fin = 450.320 uV 18:55:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 172.957 uVs 18:55:01 Integrated Voltage Measurement... 18:55:10 Voltage Offset Correction: v_off_ini = 441.206 uV, v_off_fin = 434.389 uV 18:55:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.041 uVs 18:55:10 Integrated Voltage Measurement... 18:55:19 Voltage Offset Correction: v_off_ini = 441.133 uV, v_off_fin = 439.490 uV 18:55:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 175.846 uVs 18:55:19 Integrated Voltage Measurement... 18:55:28 Voltage Offset Correction: v_off_ini = 437.565 uV, v_off_fin = 432.922 uV 18:55:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -166.509 uVs 18:55:28 Integrated Voltage Measurement... 18:55:37 Voltage Offset Correction: v_off_ini = 447.720 uV, v_off_fin = 458.355 uV 18:55:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 175.798 uVs 18:55:41 I1Y Average and RMS of Measurements VT = -175.413 +- 3.831 uVs I1Y = -35.080 +- 0.766 uTm 18:55:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:55:49 Integrated Voltage Measurement... 18:55:58 Voltage Offset Correction: v_off_ini = 448.524 uV, v_off_fin = 426.341 uV 18:55:58 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -103.085 uVs 18:55:58 Integrated Voltage Measurement... 18:56:07 Voltage Offset Correction: v_off_ini = 449.882 uV, v_off_fin = 445.437 uV 18:56:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.426 uVs 18:56:07 Integrated Voltage Measurement... 18:56:16 Voltage Offset Correction: v_off_ini = 456.454 uV, v_off_fin = 443.275 uV 18:56:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.844 uVs 18:56:16 Integrated Voltage Measurement... 18:56:25 Voltage Offset Correction: v_off_ini = 455.895 uV, v_off_fin = 453.768 uV 18:56:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.067 uVs 18:56:25 Integrated Voltage Measurement... 18:56:34 Voltage Offset Correction: v_off_ini = 440.024 uV, v_off_fin = 456.944 uV 18:56:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.018 uVs 18:56:34 Integrated Voltage Measurement... 18:56:43 Voltage Offset Correction: v_off_ini = 442.920 uV, v_off_fin = 443.763 uV 18:56:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.724 uVs 18:56:45 I2X Average and RMS of Measurements VT = -98.361 +- 4.401 uVs I2X = 9.049 +- 0.405 uTm^2 18:56:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:56:53 Integrated Voltage Measurement... 18:57:02 Voltage Offset Correction: v_off_ini = 445.219 uV, v_off_fin = 438.166 uV 18:57:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.190 uVs 18:57:02 Integrated Voltage Measurement... 18:57:11 Voltage Offset Correction: v_off_ini = 452.580 uV, v_off_fin = 450.313 uV 18:57:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.980 uVs 18:57:11 Integrated Voltage Measurement... 18:57:20 Voltage Offset Correction: v_off_ini = 443.418 uV, v_off_fin = 445.190 uV 18:57:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -227.831 uVs 18:57:20 Integrated Voltage Measurement... 18:57:29 Voltage Offset Correction: v_off_ini = 457.786 uV, v_off_fin = 463.101 uV 18:57:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.096 uVs 18:57:29 Integrated Voltage Measurement... 18:57:38 Voltage Offset Correction: v_off_ini = 463.098 uV, v_off_fin = 475.123 uV 18:57:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -258.456 uVs 18:57:38 Integrated Voltage Measurement... 18:57:47 Voltage Offset Correction: v_off_ini = 473.474 uV, v_off_fin = 447.480 uV 18:57:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 238.715 uVs 18:57:49 I2Y Average and RMS of Measurements VT = -245.878 +- 4.187 uVs I2Y = -22.621 +- 0.385 uTm^2 18:58:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:58:04 Integrated Voltage Measurement... 18:58:13 Voltage Offset Correction: v_off_ini = 489.663 uV, v_off_fin = 462.468 uV 18:58:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -77.005 uVs 18:58:13 Integrated Voltage Measurement... 18:58:22 Voltage Offset Correction: v_off_ini = 466.210 uV, v_off_fin = 479.042 uV 18:58:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.867 uVs 18:58:22 Integrated Voltage Measurement... 18:58:31 Voltage Offset Correction: v_off_ini = 471.796 uV, v_off_fin = 445.920 uV 18:58:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.827 uVs 18:58:31 Integrated Voltage Measurement... 18:58:40 Voltage Offset Correction: v_off_ini = 451.567 uV, v_off_fin = 445.977 uV 18:58:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.150 uVs 18:58:40 Integrated Voltage Measurement... 18:58:49 Voltage Offset Correction: v_off_ini = 454.992 uV, v_off_fin = 457.460 uV 18:58:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.437 uVs 18:58:49 Integrated Voltage Measurement... 18:58:58 Voltage Offset Correction: v_off_ini = 463.567 uV, v_off_fin = 466.708 uV 18:58:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.098 uVs 18:59:03 I1X Average and RMS of Measurements VT = -93.397 +- 5.694 uVs I1X = 18.680 +- 1.139 uTm 18:59:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:59:09 Integrated Voltage Measurement... 18:59:18 Voltage Offset Correction: v_off_ini = 436.037 uV, v_off_fin = 430.831 uV 18:59:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -172.498 uVs 18:59:18 Integrated Voltage Measurement... 18:59:27 Voltage Offset Correction: v_off_ini = 433.486 uV, v_off_fin = 456.365 uV 18:59:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 167.006 uVs 18:59:27 Integrated Voltage Measurement... 18:59:36 Voltage Offset Correction: v_off_ini = 449.063 uV, v_off_fin = 443.696 uV 18:59:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -204.421 uVs 18:59:36 Integrated Voltage Measurement... 18:59:45 Voltage Offset Correction: v_off_ini = 436.111 uV, v_off_fin = 450.184 uV 18:59:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 167.542 uVs 18:59:45 Integrated Voltage Measurement... 18:59:54 Voltage Offset Correction: v_off_ini = 459.567 uV, v_off_fin = 460.225 uV 18:59:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -178.909 uVs 18:59:54 Integrated Voltage Measurement... 19:00:03 Voltage Offset Correction: v_off_ini = 457.460 uV, v_off_fin = 465.273 uV 19:00:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 161.529 uVs 19:00:07 I1Y Average and RMS of Measurements VT = -175.317 +- 7.543 uVs I1Y = -35.061 +- 1.508 uTm 19:00:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:00:15 Integrated Voltage Measurement... 19:00:24 Voltage Offset Correction: v_off_ini = 463.138 uV, v_off_fin = 441.837 uV 19:00:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.685 uVs 19:00:24 Integrated Voltage Measurement... 19:00:33 Voltage Offset Correction: v_off_ini = 447.573 uV, v_off_fin = 455.083 uV 19:00:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.395 uVs 19:00:33 Integrated Voltage Measurement... 19:00:42 Voltage Offset Correction: v_off_ini = 466.235 uV, v_off_fin = 444.603 uV 19:00:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.252 uVs 19:00:42 Integrated Voltage Measurement... 19:00:51 Voltage Offset Correction: v_off_ini = 452.421 uV, v_off_fin = 467.259 uV 19:00:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.425 uVs 19:00:51 Integrated Voltage Measurement... 19:01:00 Voltage Offset Correction: v_off_ini = 451.096 uV, v_off_fin = 450.923 uV 19:01:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.045 uVs 19:01:00 Integrated Voltage Measurement... 19:01:09 Voltage Offset Correction: v_off_ini = 467.118 uV, v_off_fin = 465.479 uV 19:01:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.405 uVs 19:01:12 I2X Average and RMS of Measurements VT = -106.368 +- 4.099 uVs I2X = 9.786 +- 0.377 uTm^2 19:01:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:01:19 Integrated Voltage Measurement... 19:01:28 Voltage Offset Correction: v_off_ini = 460.427 uV, v_off_fin = 433.704 uV 19:01:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -236.839 uVs 19:01:28 Integrated Voltage Measurement... 19:01:37 Voltage Offset Correction: v_off_ini = 429.218 uV, v_off_fin = 451.638 uV 19:01:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.214 uVs 19:01:37 Integrated Voltage Measurement... 19:01:46 Voltage Offset Correction: v_off_ini = 461.592 uV, v_off_fin = 480.331 uV 19:01:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.950 uVs 19:01:46 Integrated Voltage Measurement... 19:01:55 Voltage Offset Correction: v_off_ini = 468.758 uV, v_off_fin = 479.986 uV 19:01:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 228.634 uVs 19:01:55 Integrated Voltage Measurement... 19:02:04 Voltage Offset Correction: v_off_ini = 489.905 uV, v_off_fin = 473.738 uV 19:02:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.592 uVs 19:02:04 Integrated Voltage Measurement... 19:02:13 Voltage Offset Correction: v_off_ini = 480.106 uV, v_off_fin = 474.133 uV 19:02:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.026 uVs 19:02:16 I2Y Average and RMS of Measurements VT = -242.876 +- 2.778 uVs I2Y = -22.345 +- 0.256 uTm^2 19:02:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:02:30 Integrated Voltage Measurement... 19:02:39 Voltage Offset Correction: v_off_ini = 462.279 uV, v_off_fin = 454.539 uV 19:02:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -91.216 uVs 19:02:39 Integrated Voltage Measurement... 19:02:48 Voltage Offset Correction: v_off_ini = 448.271 uV, v_off_fin = 443.910 uV 19:02:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.703 uVs 19:02:48 Integrated Voltage Measurement... 19:02:57 Voltage Offset Correction: v_off_ini = 435.998 uV, v_off_fin = 449.966 uV 19:02:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.804 uVs 19:02:57 Integrated Voltage Measurement... 19:03:06 Voltage Offset Correction: v_off_ini = 454.980 uV, v_off_fin = 448.839 uV 19:03:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.549 uVs 19:03:06 Integrated Voltage Measurement... 19:03:15 Voltage Offset Correction: v_off_ini = 448.956 uV, v_off_fin = 446.213 uV 19:03:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.124 uVs 19:03:15 Integrated Voltage Measurement... 19:03:24 Voltage Offset Correction: v_off_ini = 456.398 uV, v_off_fin = 462.427 uV 19:03:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.730 uVs 19:03:28 I1X Average and RMS of Measurements VT = -95.854 +- 3.780 uVs I1X = 19.171 +- 0.756 uTm 19:03:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:03:35 Integrated Voltage Measurement... 19:03:44 Voltage Offset Correction: v_off_ini = 458.855 uV, v_off_fin = 449.087 uV 19:03:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.326 uVs 19:03:44 Integrated Voltage Measurement... 19:03:53 Voltage Offset Correction: v_off_ini = 438.799 uV, v_off_fin = 453.813 uV 19:03:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 186.127 uVs 19:03:53 Integrated Voltage Measurement... 19:04:02 Voltage Offset Correction: v_off_ini = 452.325 uV, v_off_fin = 442.601 uV 19:04:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -175.045 uVs 19:04:02 Integrated Voltage Measurement... 19:04:10 Voltage Offset Correction: v_off_ini = 461.900 uV, v_off_fin = 482.724 uV 19:04:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 177.209 uVs 19:04:11 Integrated Voltage Measurement... 19:04:19 Voltage Offset Correction: v_off_ini = 471.502 uV, v_off_fin = 479.022 uV 19:04:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -189.171 uVs 19:04:20 Integrated Voltage Measurement... 19:04:28 Voltage Offset Correction: v_off_ini = 475.807 uV, v_off_fin = 485.961 uV 19:04:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 174.381 uVs 19:04:33 I1Y Average and RMS of Measurements VT = -182.877 +- 6.011 uVs I1Y = -36.572 +- 1.203 uTm 19:04:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:04:40 Integrated Voltage Measurement... 19:04:49 Voltage Offset Correction: v_off_ini = 458.985 uV, v_off_fin = 438.995 uV 19:04:49 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -97.925 uVs 19:04:49 Integrated Voltage Measurement... 19:04:58 Voltage Offset Correction: v_off_ini = 467.946 uV, v_off_fin = 470.920 uV 19:04:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 113.781 uVs 19:04:58 Integrated Voltage Measurement... 19:05:07 Voltage Offset Correction: v_off_ini = 476.240 uV, v_off_fin = 465.557 uV 19:05:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.138 uVs 19:05:07 Integrated Voltage Measurement... 19:05:16 Voltage Offset Correction: v_off_ini = 483.214 uV, v_off_fin = 501.757 uV 19:05:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.688 uVs 19:05:16 Integrated Voltage Measurement... 19:05:25 Voltage Offset Correction: v_off_ini = 493.282 uV, v_off_fin = 482.064 uV 19:05:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.459 uVs 19:05:25 Integrated Voltage Measurement... 19:05:34 Voltage Offset Correction: v_off_ini = 501.522 uV, v_off_fin = 502.090 uV 19:05:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.324 uVs 19:05:37 I2X Average and RMS of Measurements VT = -105.219 +- 1.296 uVs I2X = 9.680 +- 0.119 uTm^2 19:05:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:05:45 Integrated Voltage Measurement... 19:05:54 Voltage Offset Correction: v_off_ini = 501.467 uV, v_off_fin = 463.703 uV 19:05:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.642 uVs 19:05:54 Integrated Voltage Measurement... 19:06:03 Voltage Offset Correction: v_off_ini = 476.268 uV, v_off_fin = 479.731 uV 19:06:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.680 uVs 19:06:03 Integrated Voltage Measurement... 19:06:12 Voltage Offset Correction: v_off_ini = 457.941 uV, v_off_fin = 474.337 uV 19:06:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = -0.0001 mm dXY = 1.0000 mm VT = -257.562 uVs 19:06:12 Integrated Voltage Measurement... 19:06:21 Voltage Offset Correction: v_off_ini = 488.513 uV, v_off_fin = 499.460 uV 19:06:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.482 uVs 19:06:21 Integrated Voltage Measurement... 19:06:30 Voltage Offset Correction: v_off_ini = 483.758 uV, v_off_fin = 479.493 uV 19:06:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.663 uVs 19:06:30 Integrated Voltage Measurement... 19:06:39 Voltage Offset Correction: v_off_ini = 488.919 uV, v_off_fin = 478.625 uV 19:06:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.486 uVs 19:06:41 I2Y Average and RMS of Measurements VT = -240.419 +- 11.528 uVs I2Y = -22.119 +- 1.061 uTm^2 19:06:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:06:56 Integrated Voltage Measurement... 19:07:05 Voltage Offset Correction: v_off_ini = 493.216 uV, v_off_fin = 484.449 uV 19:07:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -114.646 uVs 19:07:05 Integrated Voltage Measurement... 19:07:14 Voltage Offset Correction: v_off_ini = 465.427 uV, v_off_fin = 459.669 uV 19:07:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.671 uVs 19:07:14 Integrated Voltage Measurement... 19:07:23 Voltage Offset Correction: v_off_ini = 451.498 uV, v_off_fin = 457.350 uV 19:07:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.562 uVs 19:07:23 Integrated Voltage Measurement... 19:07:32 Voltage Offset Correction: v_off_ini = 476.946 uV, v_off_fin = 468.014 uV 19:07:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.771 uVs 19:07:32 Integrated Voltage Measurement... 19:07:41 Voltage Offset Correction: v_off_ini = 451.557 uV, v_off_fin = 454.696 uV 19:07:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.235 uVs 19:07:41 Integrated Voltage Measurement... 19:07:50 Voltage Offset Correction: v_off_ini = 441.606 uV, v_off_fin = 455.719 uV 19:07:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.594 uVs 19:07:54 I1X Average and RMS of Measurements VT = -96.913 +- 11.081 uVs I1X = 19.383 +- 2.216 uTm 19:07:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:08:01 Integrated Voltage Measurement... 19:08:10 Voltage Offset Correction: v_off_ini = 457.011 uV, v_off_fin = 438.267 uV 19:08:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.827 uVs 19:08:10 Integrated Voltage Measurement... 19:08:19 Voltage Offset Correction: v_off_ini = 445.287 uV, v_off_fin = 463.442 uV 19:08:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 182.383 uVs 19:08:19 Integrated Voltage Measurement... 19:08:28 Voltage Offset Correction: v_off_ini = 465.174 uV, v_off_fin = 453.973 uV 19:08:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -188.373 uVs 19:08:28 Integrated Voltage Measurement... 19:08:37 Voltage Offset Correction: v_off_ini = 457.791 uV, v_off_fin = 466.951 uV 19:08:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 181.549 uVs 19:08:37 Integrated Voltage Measurement... 19:08:46 Voltage Offset Correction: v_off_ini = 474.262 uV, v_off_fin = 454.980 uV 19:08:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -185.309 uVs 19:08:46 Integrated Voltage Measurement... 19:08:55 Voltage Offset Correction: v_off_ini = 455.234 uV, v_off_fin = 467.021 uV 19:08:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 185.386 uVs 19:08:59 I1Y Average and RMS of Measurements VT = -184.805 +- 0.519 uVs I1Y = -36.957 +- 0.102 uTm 19:09:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:09:07 Integrated Voltage Measurement... 19:09:16 Voltage Offset Correction: v_off_ini = 450.222 uV, v_off_fin = 461.874 uV 19:09:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.090 uVs 19:09:16 Integrated Voltage Measurement... 19:09:25 Voltage Offset Correction: v_off_ini = 463.282 uV, v_off_fin = 474.544 uV 19:09:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.296 uVs 19:09:25 Integrated Voltage Measurement... 19:09:34 Voltage Offset Correction: v_off_ini = 462.511 uV, v_off_fin = 451.435 uV 19:09:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.559 uVs 19:09:34 Integrated Voltage Measurement... 19:09:43 Voltage Offset Correction: v_off_ini = 457.967 uV, v_off_fin = 455.427 uV 19:09:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.024 uVs 19:09:43 Integrated Voltage Measurement... 19:09:52 Voltage Offset Correction: v_off_ini = 455.527 uV, v_off_fin = 447.877 uV 19:09:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.536 uVs 19:09:52 Integrated Voltage Measurement... 19:10:01 Voltage Offset Correction: v_off_ini = 460.435 uV, v_off_fin = 467.072 uV 19:10:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.352 uVs 19:10:03 I2X Average and RMS of Measurements VT = -100.476 +- 7.064 uVs I2X = 9.244 +- 0.650 uTm^2 19:10:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:10:11 Integrated Voltage Measurement... 19:10:20 Voltage Offset Correction: v_off_ini = 453.884 uV, v_off_fin = 456.446 uV 19:10:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -238.148 uVs 19:10:20 Integrated Voltage Measurement... 19:10:29 Voltage Offset Correction: v_off_ini = 446.236 uV, v_off_fin = 458.745 uV 19:10:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.046 uVs 19:10:29 Integrated Voltage Measurement... 19:10:38 Voltage Offset Correction: v_off_ini = 449.451 uV, v_off_fin = 463.200 uV 19:10:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -242.120 uVs 19:10:38 Integrated Voltage Measurement... 19:10:47 Voltage Offset Correction: v_off_ini = 457.234 uV, v_off_fin = 457.322 uV 19:10:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 251.079 uVs 19:10:47 Integrated Voltage Measurement... 19:10:56 Voltage Offset Correction: v_off_ini = 466.053 uV, v_off_fin = 471.561 uV 19:10:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.401 uVs 19:10:56 Integrated Voltage Measurement... 19:11:05 Voltage Offset Correction: v_off_ini = 455.776 uV, v_off_fin = 429.747 uV 19:11:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 222.222 uVs 19:11:07 I2Y Average and RMS of Measurements VT = -243.336 +- 8.401 uVs I2Y = -22.387 +- 0.773 uTm^2 19:11:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:11:25 Integrated Voltage Measurement... 19:11:34 Voltage Offset Correction: v_off_ini = 461.312 uV, v_off_fin = 459.288 uV 19:11:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -89.824 uVs 19:11:34 Integrated Voltage Measurement... 19:11:43 Voltage Offset Correction: v_off_ini = 459.985 uV, v_off_fin = 445.912 uV 19:11:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.277 uVs 19:11:43 Integrated Voltage Measurement... 19:11:52 Voltage Offset Correction: v_off_ini = 445.694 uV, v_off_fin = 434.741 uV 19:11:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.473 uVs 19:11:52 Integrated Voltage Measurement... 19:12:01 Voltage Offset Correction: v_off_ini = 439.102 uV, v_off_fin = 453.523 uV 19:12:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.485 uVs 19:12:01 Integrated Voltage Measurement... 19:12:10 Voltage Offset Correction: v_off_ini = 458.356 uV, v_off_fin = 452.279 uV 19:12:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.163 uVs 19:12:10 Integrated Voltage Measurement... 19:12:19 Voltage Offset Correction: v_off_ini = 445.772 uV, v_off_fin = 446.132 uV 19:12:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.790 uVs 19:12:23 I1X Average and RMS of Measurements VT = -82.835 +- 13.086 uVs I1X = 16.567 +- 2.617 uTm 19:12:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:12:30 Integrated Voltage Measurement... 19:12:39 Voltage Offset Correction: v_off_ini = 456.069 uV, v_off_fin = 428.662 uV 19:12:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.419 uVs 19:12:39 Integrated Voltage Measurement... 19:12:48 Voltage Offset Correction: v_off_ini = 427.562 uV, v_off_fin = 455.994 uV 19:12:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 186.244 uVs 19:12:48 Integrated Voltage Measurement... 19:12:57 Voltage Offset Correction: v_off_ini = 428.915 uV, v_off_fin = 431.299 uV 19:12:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -174.228 uVs 19:12:57 Integrated Voltage Measurement... 19:13:06 Voltage Offset Correction: v_off_ini = 423.037 uV, v_off_fin = 432.186 uV 19:13:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 183.798 uVs 19:13:06 Integrated Voltage Measurement... 19:13:15 Voltage Offset Correction: v_off_ini = 422.760 uV, v_off_fin = 421.674 uV 19:13:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -177.422 uVs 19:13:15 Integrated Voltage Measurement... 19:13:24 Voltage Offset Correction: v_off_ini = 424.309 uV, v_off_fin = 439.137 uV 19:13:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.402 uVs 19:13:28 I1Y Average and RMS of Measurements VT = -183.752 +- 3.513 uVs I1Y = -36.746 +- 0.703 uTm 19:13:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:13:36 Integrated Voltage Measurement... 19:13:45 Voltage Offset Correction: v_off_ini = 402.839 uV, v_off_fin = 387.558 uV 19:13:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.878 uVs 19:13:45 Integrated Voltage Measurement... 19:13:54 Voltage Offset Correction: v_off_ini = 388.869 uV, v_off_fin = 403.551 uV 19:13:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.090 uVs 19:13:54 Integrated Voltage Measurement... 19:14:03 Voltage Offset Correction: v_off_ini = 386.868 uV, v_off_fin = 372.155 uV 19:14:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -91.342 uVs 19:14:03 Integrated Voltage Measurement... 19:14:12 Voltage Offset Correction: v_off_ini = 402.565 uV, v_off_fin = 409.869 uV 19:14:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.653 uVs 19:14:12 Integrated Voltage Measurement... 19:14:21 Voltage Offset Correction: v_off_ini = 401.972 uV, v_off_fin = 399.918 uV 19:14:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.169 uVs 19:14:21 Integrated Voltage Measurement... 19:14:30 Voltage Offset Correction: v_off_ini = 408.427 uV, v_off_fin = 398.919 uV 19:14:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.240 uVs 19:14:32 I2X Average and RMS of Measurements VT = -103.062 +- 6.244 uVs I2X = 9.482 +- 0.574 uTm^2 19:14:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:14:40 Integrated Voltage Measurement... 19:14:49 Voltage Offset Correction: v_off_ini = 395.936 uV, v_off_fin = 397.688 uV 19:14:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.520 uVs 19:14:49 Integrated Voltage Measurement... 19:14:58 Voltage Offset Correction: v_off_ini = 377.770 uV, v_off_fin = 377.756 uV 19:14:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.213 uVs 19:14:58 Integrated Voltage Measurement... 19:15:07 Voltage Offset Correction: v_off_ini = 388.569 uV, v_off_fin = 404.849 uV 19:15:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.396 uVs 19:15:07 Integrated Voltage Measurement... 19:15:16 Voltage Offset Correction: v_off_ini = 394.853 uV, v_off_fin = 389.652 uV 19:15:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 214.720 uVs 19:15:16 Integrated Voltage Measurement... 19:15:25 Voltage Offset Correction: v_off_ini = 361.366 uV, v_off_fin = 375.174 uV 19:15:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -221.100 uVs 19:15:25 Integrated Voltage Measurement... 19:15:34 Voltage Offset Correction: v_off_ini = 378.078 uV, v_off_fin = 387.203 uV 19:15:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 230.905 uVs 19:15:36 I2Y Average and RMS of Measurements VT = -232.643 +- 8.315 uVs I2Y = -21.403 +- 0.765 uTm^2 19:15:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:16:01 Integrated Voltage Measurement... 19:16:10 Voltage Offset Correction: v_off_ini = 383.130 uV, v_off_fin = 384.105 uV 19:16:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -109.651 uVs 19:16:10 Integrated Voltage Measurement... 19:16:19 Voltage Offset Correction: v_off_ini = 380.059 uV, v_off_fin = 385.286 uV 19:16:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.848 uVs 19:16:19 Integrated Voltage Measurement... 19:16:28 Voltage Offset Correction: v_off_ini = 371.681 uV, v_off_fin = 380.738 uV 19:16:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.464 uVs 19:16:28 Integrated Voltage Measurement... 19:16:37 Voltage Offset Correction: v_off_ini = 387.189 uV, v_off_fin = 377.401 uV 19:16:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.637 uVs 19:16:37 Integrated Voltage Measurement... 19:16:46 Voltage Offset Correction: v_off_ini = 371.227 uV, v_off_fin = 382.251 uV 19:16:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.950 uVs 19:16:46 Integrated Voltage Measurement... 19:16:55 Voltage Offset Correction: v_off_ini = 379.223 uV, v_off_fin = 393.346 uV 19:16:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.217 uVs 19:17:00 I1X Average and RMS of Measurements VT = -97.628 +- 7.856 uVs I1X = 19.526 +- 1.571 uTm 19:17:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:17:06 Integrated Voltage Measurement... 19:17:15 Voltage Offset Correction: v_off_ini = 380.999 uV, v_off_fin = 368.565 uV 19:17:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -189.778 uVs 19:17:15 Integrated Voltage Measurement... 19:17:24 Voltage Offset Correction: v_off_ini = 381.239 uV, v_off_fin = 394.805 uV 19:17:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 190.775 uVs 19:17:24 Integrated Voltage Measurement... 19:17:33 Voltage Offset Correction: v_off_ini = 384.649 uV, v_off_fin = 379.308 uV 19:17:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -184.587 uVs 19:17:33 Integrated Voltage Measurement... 19:17:42 Voltage Offset Correction: v_off_ini = 382.368 uV, v_off_fin = 401.255 uV 19:17:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 190.723 uVs 19:17:42 Integrated Voltage Measurement... 19:17:51 Voltage Offset Correction: v_off_ini = 402.116 uV, v_off_fin = 394.733 uV 19:17:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -195.765 uVs 19:17:51 Integrated Voltage Measurement... 19:18:00 Voltage Offset Correction: v_off_ini = 405.310 uV, v_off_fin = 409.562 uV 19:18:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 211.689 uVs 19:18:04 I1Y Average and RMS of Measurements VT = -193.886 +- 7.041 uVs I1Y = -38.773 +- 1.407 uTm 19:18:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:18:12 Integrated Voltage Measurement... 19:18:21 Voltage Offset Correction: v_off_ini = 414.533 uV, v_off_fin = 418.534 uV 19:18:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -121.940 uVs 19:18:21 Integrated Voltage Measurement... 19:18:30 Voltage Offset Correction: v_off_ini = 398.157 uV, v_off_fin = 414.152 uV 19:18:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 135.509 uVs 19:18:30 Integrated Voltage Measurement... 19:18:39 Voltage Offset Correction: v_off_ini = 391.420 uV, v_off_fin = 391.368 uV 19:18:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -139.123 uVs 19:18:39 Integrated Voltage Measurement... 19:18:48 Voltage Offset Correction: v_off_ini = 407.437 uV, v_off_fin = 418.841 uV 19:18:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.082 uVs 19:18:48 Integrated Voltage Measurement... 19:18:57 Voltage Offset Correction: v_off_ini = 411.652 uV, v_off_fin = 413.096 uV 19:18:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.521 uVs 19:18:57 Integrated Voltage Measurement... 19:19:06 Voltage Offset Correction: v_off_ini = 425.850 uV, v_off_fin = 412.897 uV 19:19:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.269 uVs 19:19:09 I2X Average and RMS of Measurements VT = -120.574 +- 5.833 uVs I2X = 11.093 +- 0.537 uTm^2 19:19:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:19:16 Integrated Voltage Measurement... 19:19:25 Voltage Offset Correction: v_off_ini = 378.167 uV, v_off_fin = 398.805 uV 19:19:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -217.698 uVs 19:19:25 Integrated Voltage Measurement... 19:19:34 Voltage Offset Correction: v_off_ini = 410.430 uV, v_off_fin = 406.933 uV 19:19:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.130 uVs 19:19:34 Integrated Voltage Measurement... 19:19:43 Voltage Offset Correction: v_off_ini = 403.779 uV, v_off_fin = 393.031 uV 19:19:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.380 uVs 19:19:43 Integrated Voltage Measurement... 19:19:52 Voltage Offset Correction: v_off_ini = 397.812 uV, v_off_fin = 407.839 uV 19:19:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 243.618 uVs 19:19:52 Integrated Voltage Measurement... 19:20:01 Voltage Offset Correction: v_off_ini = 410.687 uV, v_off_fin = 402.382 uV 19:20:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -242.724 uVs 19:20:01 Integrated Voltage Measurement... 19:20:10 Voltage Offset Correction: v_off_ini = 385.121 uV, v_off_fin = 407.147 uV 19:20:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.642 uVs 19:20:13 I2Y Average and RMS of Measurements VT = -242.865 +- 2.979 uVs I2Y = -22.344 +- 0.274 uTm^2 19:20:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:20:30 Integrated Voltage Measurement... 19:20:39 Voltage Offset Correction: v_off_ini = 435.647 uV, v_off_fin = 398.333 uV 19:20:39 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -110.631 uVs 19:20:39 Integrated Voltage Measurement... 19:20:48 Voltage Offset Correction: v_off_ini = 404.362 uV, v_off_fin = 415.028 uV 19:20:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.071 uVs 19:20:48 Integrated Voltage Measurement... 19:20:57 Voltage Offset Correction: v_off_ini = 418.827 uV, v_off_fin = 415.330 uV 19:20:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.257 uVs 19:20:57 Integrated Voltage Measurement... 19:21:06 Voltage Offset Correction: v_off_ini = 402.686 uV, v_off_fin = 396.130 uV 19:21:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.045 uVs 19:21:06 Integrated Voltage Measurement... 19:21:15 Voltage Offset Correction: v_off_ini = 384.700 uV, v_off_fin = 382.727 uV 19:21:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -91.102 uVs 19:21:15 Integrated Voltage Measurement... 19:21:24 Voltage Offset Correction: v_off_ini = 400.465 uV, v_off_fin = 400.900 uV 19:21:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.774 uVs 19:21:28 I1X Average and RMS of Measurements VT = -92.647 +- 8.281 uVs I1X = 18.531 +- 1.658 uTm 19:21:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:21:35 Integrated Voltage Measurement... 19:21:44 Voltage Offset Correction: v_off_ini = 399.311 uV, v_off_fin = 379.295 uV 19:21:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.598 uVs 19:21:44 Integrated Voltage Measurement... 19:21:53 Voltage Offset Correction: v_off_ini = 398.082 uV, v_off_fin = 407.771 uV 19:21:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 202.196 uVs 19:21:53 Integrated Voltage Measurement... 19:22:02 Voltage Offset Correction: v_off_ini = 395.662 uV, v_off_fin = 381.491 uV 19:22:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -193.136 uVs 19:22:02 Integrated Voltage Measurement... 19:22:11 Voltage Offset Correction: v_off_ini = 406.663 uV, v_off_fin = 405.671 uV 19:22:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 215.587 uVs 19:22:11 Integrated Voltage Measurement... 19:22:20 Voltage Offset Correction: v_off_ini = 404.736 uV, v_off_fin = 401.337 uV 19:22:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -196.360 uVs 19:22:20 Integrated Voltage Measurement... 19:22:29 Voltage Offset Correction: v_off_ini = 399.818 uV, v_off_fin = 397.774 uV 19:22:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 187.644 uVs 19:22:33 I1Y Average and RMS of Measurements VT = -197.920 +- 5.059 uVs I1Y = -39.579 +- 1.012 uTm 19:22:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:22:41 Integrated Voltage Measurement... 19:22:50 Voltage Offset Correction: v_off_ini = 396.289 uV, v_off_fin = 365.174 uV 19:22:50 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -129.026 uVs 19:22:50 Integrated Voltage Measurement... 19:22:59 Voltage Offset Correction: v_off_ini = 369.200 uV, v_off_fin = 391.977 uV 19:22:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.388 uVs 19:22:59 Integrated Voltage Measurement... 19:23:08 Voltage Offset Correction: v_off_ini = 396.894 uV, v_off_fin = 390.980 uV 19:23:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.444 uVs 19:23:08 Integrated Voltage Measurement... 19:23:17 Voltage Offset Correction: v_off_ini = 404.104 uV, v_off_fin = 379.686 uV 19:23:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 141.872 uVs 19:23:17 Integrated Voltage Measurement... 19:23:26 Voltage Offset Correction: v_off_ini = 371.347 uV, v_off_fin = 382.944 uV 19:23:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.833 uVs 19:23:26 Integrated Voltage Measurement... 19:23:35 Voltage Offset Correction: v_off_ini = 404.250 uV, v_off_fin = 418.316 uV 19:23:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 132.201 uVs 19:23:37 I2X Average and RMS of Measurements VT = -121.961 +- 1.559 uVs I2X = 11.220 +- 0.144 uTm^2 19:23:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:23:45 Integrated Voltage Measurement... 19:23:54 Voltage Offset Correction: v_off_ini = 419.301 uV, v_off_fin = 413.198 uV 19:23:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.028 uVs 19:23:54 Integrated Voltage Measurement... 19:24:03 Voltage Offset Correction: v_off_ini = 407.664 uV, v_off_fin = 380.148 uV 19:24:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.653 uVs 19:24:03 Integrated Voltage Measurement... 19:24:12 Voltage Offset Correction: v_off_ini = 383.705 uV, v_off_fin = 391.493 uV 19:24:12 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -244.312 uVs 19:24:12 Integrated Voltage Measurement... 19:24:21 Voltage Offset Correction: v_off_ini = 389.135 uV, v_off_fin = 398.947 uV 19:24:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.017 uVs 19:24:21 Integrated Voltage Measurement... 19:24:30 Voltage Offset Correction: v_off_ini = 412.437 uV, v_off_fin = 383.658 uV 19:24:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -224.697 uVs 19:24:30 Integrated Voltage Measurement... 19:24:39 Voltage Offset Correction: v_off_ini = 386.468 uV, v_off_fin = 386.337 uV 19:24:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.149 uVs 19:24:41 I2Y Average and RMS of Measurements VT = -245.143 +- 5.069 uVs I2Y = -22.554 +- 0.466 uTm^2 19:24:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:24:58 Integrated Voltage Measurement... 19:25:07 Voltage Offset Correction: v_off_ini = 378.468 uV, v_off_fin = 363.809 uV 19:25:07 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -96.463 uVs 19:25:07 Integrated Voltage Measurement... 19:25:16 Voltage Offset Correction: v_off_ini = 397.852 uV, v_off_fin = 385.848 uV 19:25:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.124 uVs 19:25:16 Integrated Voltage Measurement... 19:25:25 Voltage Offset Correction: v_off_ini = 365.384 uV, v_off_fin = 401.520 uV 19:25:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.283 uVs 19:25:25 Integrated Voltage Measurement... 19:25:34 Voltage Offset Correction: v_off_ini = 405.201 uV, v_off_fin = 406.265 uV 19:25:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.111 uVs 19:25:34 Integrated Voltage Measurement... 19:25:43 Voltage Offset Correction: v_off_ini = 389.880 uV, v_off_fin = 386.202 uV 19:25:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.122 uVs 19:25:43 Integrated Voltage Measurement... 19:25:52 Voltage Offset Correction: v_off_ini = 373.888 uV, v_off_fin = 367.676 uV 19:25:52 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.795 uVs 19:25:57 I1X Average and RMS of Measurements VT = -97.816 +- 3.790 uVs I1X = 19.565 +- 0.757 uTm 19:25:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:26:03 Integrated Voltage Measurement... 19:26:12 Voltage Offset Correction: v_off_ini = 359.829 uV, v_off_fin = 368.497 uV 19:26:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.509 uVs 19:26:12 Integrated Voltage Measurement... 19:26:21 Voltage Offset Correction: v_off_ini = 380.507 uV, v_off_fin = 386.524 uV 19:26:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.590 uVs 19:26:21 Integrated Voltage Measurement... 19:26:30 Voltage Offset Correction: v_off_ini = 382.289 uV, v_off_fin = 371.701 uV 19:26:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -208.813 uVs 19:26:30 Integrated Voltage Measurement... 19:26:39 Voltage Offset Correction: v_off_ini = 390.145 uV, v_off_fin = 392.267 uV 19:26:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 198.919 uVs 19:26:39 Integrated Voltage Measurement... 19:26:48 Voltage Offset Correction: v_off_ini = 373.942 uV, v_off_fin = 361.533 uV 19:26:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -202.449 uVs 19:26:48 Integrated Voltage Measurement... 19:26:57 Voltage Offset Correction: v_off_ini = 367.468 uV, v_off_fin = 387.371 uV 19:26:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 205.050 uVs 19:27:01 I1Y Average and RMS of Measurements VT = -203.888 +- 0.124 uVs I1Y = -40.772 +- 0.027 uTm 19:27:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:27:09 Integrated Voltage Measurement... 19:27:18 Voltage Offset Correction: v_off_ini = 392.142 uV, v_off_fin = 396.974 uV 19:27:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -127.176 uVs 19:27:18 Integrated Voltage Measurement... 19:27:27 Voltage Offset Correction: v_off_ini = 392.258 uV, v_off_fin = 395.268 uV 19:27:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.824 uVs 19:27:27 Integrated Voltage Measurement... 19:27:36 Voltage Offset Correction: v_off_ini = 387.150 uV, v_off_fin = 389.107 uV 19:27:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.055 uVs 19:27:36 Integrated Voltage Measurement... 19:27:45 Voltage Offset Correction: v_off_ini = 395.779 uV, v_off_fin = 412.011 uV 19:27:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.536 uVs 19:27:45 Integrated Voltage Measurement... 19:27:54 Voltage Offset Correction: v_off_ini = 406.674 uV, v_off_fin = 403.129 uV 19:27:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.073 uVs 19:27:54 Integrated Voltage Measurement... 19:28:03 Voltage Offset Correction: v_off_ini = 389.013 uV, v_off_fin = 411.010 uV 19:28:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.018 uVs 19:28:05 I2X Average and RMS of Measurements VT = -116.447 +- 7.115 uVs I2X = 10.713 +- 0.655 uTm^2 19:28:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:28:13 Integrated Voltage Measurement... 19:28:22 Voltage Offset Correction: v_off_ini = 396.732 uV, v_off_fin = 408.497 uV 19:28:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.306 uVs 19:28:22 Integrated Voltage Measurement... 19:28:31 Voltage Offset Correction: v_off_ini = 396.474 uV, v_off_fin = 404.746 uV 19:28:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 232.552 uVs 19:28:31 Integrated Voltage Measurement... 19:28:40 Voltage Offset Correction: v_off_ini = 437.554 uV, v_off_fin = 415.929 uV 19:28:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -238.263 uVs 19:28:40 Integrated Voltage Measurement... 19:28:49 Voltage Offset Correction: v_off_ini = 422.992 uV, v_off_fin = 415.007 uV 19:28:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 243.426 uVs 19:28:49 Integrated Voltage Measurement... 19:28:58 Voltage Offset Correction: v_off_ini = 406.291 uV, v_off_fin = 401.215 uV 19:28:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -238.842 uVs 19:28:58 Integrated Voltage Measurement... 19:29:07 Voltage Offset Correction: v_off_ini = 391.942 uV, v_off_fin = 387.261 uV 19:29:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 254.174 uVs 19:29:09 I2Y Average and RMS of Measurements VT = -240.094 +- 5.569 uVs I2Y = -22.089 +- 0.512 uTm^2 19:29:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:29:27 Integrated Voltage Measurement... 19:29:36 Voltage Offset Correction: v_off_ini = 403.390 uV, v_off_fin = 402.778 uV 19:29:36 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -91.802 uVs 19:29:36 Integrated Voltage Measurement... 19:29:45 Voltage Offset Correction: v_off_ini = 397.854 uV, v_off_fin = 396.085 uV 19:29:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.454 uVs 19:29:45 Integrated Voltage Measurement... 19:29:54 Voltage Offset Correction: v_off_ini = 400.414 uV, v_off_fin = 396.422 uV 19:29:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -127.922 uVs 19:29:54 Integrated Voltage Measurement... 19:30:03 Voltage Offset Correction: v_off_ini = 416.600 uV, v_off_fin = 402.401 uV 19:30:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.228 uVs 19:30:03 Integrated Voltage Measurement... 19:30:12 Voltage Offset Correction: v_off_ini = 418.681 uV, v_off_fin = 414.180 uV 19:30:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -124.749 uVs 19:30:12 Integrated Voltage Measurement... 19:30:21 Voltage Offset Correction: v_off_ini = 415.919 uV, v_off_fin = 405.910 uV 19:30:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.044 uVs 19:30:25 I1X Average and RMS of Measurements VT = -108.867 +- 6.951 uVs I1X = 21.774 +- 1.389 uTm 19:30:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:30:32 Integrated Voltage Measurement... 19:30:41 Voltage Offset Correction: v_off_ini = 396.472 uV, v_off_fin = 387.798 uV 19:30:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -189.621 uVs 19:30:41 Integrated Voltage Measurement... 19:30:49 Voltage Offset Correction: v_off_ini = 401.598 uV, v_off_fin = 409.429 uV 19:30:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.359 uVs 19:30:50 Integrated Voltage Measurement... 19:30:58 Voltage Offset Correction: v_off_ini = 409.790 uV, v_off_fin = 386.949 uV 19:30:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.022 uVs 19:30:58 Integrated Voltage Measurement... 19:31:07 Voltage Offset Correction: v_off_ini = 393.001 uV, v_off_fin = 392.889 uV 19:31:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.023 uVs 19:31:07 Integrated Voltage Measurement... 19:31:16 Voltage Offset Correction: v_off_ini = 384.936 uV, v_off_fin = 382.473 uV 19:31:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -191.352 uVs 19:31:16 Integrated Voltage Measurement... 19:31:25 Voltage Offset Correction: v_off_ini = 376.096 uV, v_off_fin = 398.099 uV 19:31:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 195.219 uVs 19:31:30 I1Y Average and RMS of Measurements VT = -195.766 +- 1.804 uVs I1Y = -39.149 +- 0.361 uTm 19:31:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:31:37 Integrated Voltage Measurement... 19:31:46 Voltage Offset Correction: v_off_ini = 395.669 uV, v_off_fin = 374.477 uV 19:31:46 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -103.331 uVs 19:31:46 Integrated Voltage Measurement... 19:31:55 Voltage Offset Correction: v_off_ini = 392.940 uV, v_off_fin = 389.551 uV 19:31:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 129.653 uVs 19:31:55 Integrated Voltage Measurement... 19:32:04 Voltage Offset Correction: v_off_ini = 388.980 uV, v_off_fin = 359.657 uV 19:32:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -148.179 uVs 19:32:04 Integrated Voltage Measurement... 19:32:13 Voltage Offset Correction: v_off_ini = 376.407 uV, v_off_fin = 387.883 uV 19:32:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.461 uVs 19:32:13 Integrated Voltage Measurement... 19:32:22 Voltage Offset Correction: v_off_ini = 388.376 uV, v_off_fin = 375.460 uV 19:32:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.145 uVs 19:32:22 Integrated Voltage Measurement... 19:32:31 Voltage Offset Correction: v_off_ini = 380.977 uV, v_off_fin = 418.916 uV 19:32:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.637 uVs 19:32:34 I2X Average and RMS of Measurements VT = -124.901 +- 9.590 uVs I2X = 11.491 +- 0.883 uTm^2 19:32:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:32:41 Integrated Voltage Measurement... 19:32:50 Voltage Offset Correction: v_off_ini = 391.729 uV, v_off_fin = 397.798 uV 19:32:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -244.237 uVs 19:32:50 Integrated Voltage Measurement... 19:32:59 Voltage Offset Correction: v_off_ini = 400.033 uV, v_off_fin = 393.466 uV 19:32:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 249.166 uVs 19:32:59 Integrated Voltage Measurement... 19:33:08 Voltage Offset Correction: v_off_ini = 394.458 uV, v_off_fin = 397.117 uV 19:33:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = -0.0001 mm dXY = 1.0000 mm VT = -241.486 uVs 19:33:08 Integrated Voltage Measurement... 19:33:17 Voltage Offset Correction: v_off_ini = 390.362 uV, v_off_fin = 369.448 uV 19:33:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0001 mm dXY = -1.0000 mm VT = 243.775 uVs 19:33:17 Integrated Voltage Measurement... 19:33:26 Voltage Offset Correction: v_off_ini = 378.143 uV, v_off_fin = 374.759 uV 19:33:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.293 uVs 19:33:26 Integrated Voltage Measurement... 19:33:35 Voltage Offset Correction: v_off_ini = 364.008 uV, v_off_fin = 383.403 uV 19:33:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 232.751 uVs 19:33:38 I2Y Average and RMS of Measurements VT = -240.785 +- 5.735 uVs I2Y = -22.152 +- 0.528 uTm^2 19:33:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:33:55 Integrated Voltage Measurement... 19:34:04 Voltage Offset Correction: v_off_ini = 416.112 uV, v_off_fin = 385.923 uV 19:34:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -124.272 uVs 19:34:04 Integrated Voltage Measurement... 19:34:13 Voltage Offset Correction: v_off_ini = 390.451 uV, v_off_fin = 391.644 uV 19:34:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.542 uVs 19:34:13 Integrated Voltage Measurement... 19:34:22 Voltage Offset Correction: v_off_ini = 382.554 uV, v_off_fin = 388.876 uV 19:34:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.969 uVs 19:34:22 Integrated Voltage Measurement... 19:34:31 Voltage Offset Correction: v_off_ini = 401.901 uV, v_off_fin = 398.881 uV 19:34:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.755 uVs 19:34:31 Integrated Voltage Measurement... 19:34:40 Voltage Offset Correction: v_off_ini = 398.734 uV, v_off_fin = 391.184 uV 19:34:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.502 uVs 19:34:40 Integrated Voltage Measurement... 19:34:49 Voltage Offset Correction: v_off_ini = 396.321 uV, v_off_fin = 408.438 uV 19:34:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.838 uVs 19:34:53 I1X Average and RMS of Measurements VT = -97.980 +- 7.381 uVs I1X = 19.596 +- 1.477 uTm 19:34:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:35:00 Integrated Voltage Measurement... 19:35:09 Voltage Offset Correction: v_off_ini = 402.057 uV, v_off_fin = 391.686 uV 19:35:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -178.943 uVs 19:35:09 Integrated Voltage Measurement... 19:35:18 Voltage Offset Correction: v_off_ini = 407.370 uV, v_off_fin = 423.547 uV 19:35:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 211.590 uVs 19:35:18 Integrated Voltage Measurement... 19:35:27 Voltage Offset Correction: v_off_ini = 405.206 uV, v_off_fin = 388.344 uV 19:35:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -204.651 uVs 19:35:27 Integrated Voltage Measurement... 19:35:36 Voltage Offset Correction: v_off_ini = 415.467 uV, v_off_fin = 435.148 uV 19:35:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 197.623 uVs 19:35:36 Integrated Voltage Measurement... 19:35:45 Voltage Offset Correction: v_off_ini = 402.022 uV, v_off_fin = 392.583 uV 19:35:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.791 uVs 19:35:45 Integrated Voltage Measurement... 19:35:54 Voltage Offset Correction: v_off_ini = 427.525 uV, v_off_fin = 450.291 uV 19:35:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 192.891 uVs 19:35:58 I1Y Average and RMS of Measurements VT = -196.748 +- 3.157 uVs I1Y = -39.343 +- 0.630 uTm 19:36:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:36:06 Integrated Voltage Measurement... 19:36:15 Voltage Offset Correction: v_off_ini = 408.411 uV, v_off_fin = 420.306 uV 19:36:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.744 uVs 19:36:15 Integrated Voltage Measurement... 19:36:24 Voltage Offset Correction: v_off_ini = 419.866 uV, v_off_fin = 428.048 uV 19:36:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 145.341 uVs 19:36:24 Integrated Voltage Measurement... 19:36:33 Voltage Offset Correction: v_off_ini = 416.662 uV, v_off_fin = 417.406 uV 19:36:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.038 uVs 19:36:33 Integrated Voltage Measurement... 19:36:42 Voltage Offset Correction: v_off_ini = 416.454 uV, v_off_fin = 419.299 uV 19:36:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.959 uVs 19:36:42 Integrated Voltage Measurement... 19:36:51 Voltage Offset Correction: v_off_ini = 421.446 uV, v_off_fin = 401.651 uV 19:36:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.150 uVs 19:36:51 Integrated Voltage Measurement... 19:37:00 Voltage Offset Correction: v_off_ini = 415.125 uV, v_off_fin = 419.804 uV 19:37:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 119.275 uVs 19:37:02 I2X Average and RMS of Measurements VT = -113.085 +- 7.954 uVs I2X = 10.404 +- 0.732 uTm^2 19:37:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:37:10 Integrated Voltage Measurement... 19:37:19 Voltage Offset Correction: v_off_ini = 399.038 uV, v_off_fin = 407.684 uV 19:37:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -236.260 uVs 19:37:19 Integrated Voltage Measurement... 19:37:28 Voltage Offset Correction: v_off_ini = 414.314 uV, v_off_fin = 393.910 uV 19:37:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.102 uVs 19:37:28 Integrated Voltage Measurement... 19:37:37 Voltage Offset Correction: v_off_ini = 403.295 uV, v_off_fin = 405.925 uV 19:37:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -252.983 uVs 19:37:37 Integrated Voltage Measurement... 19:37:46 Voltage Offset Correction: v_off_ini = 417.646 uV, v_off_fin = 409.856 uV 19:37:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.457 uVs 19:37:46 Integrated Voltage Measurement... 19:37:55 Voltage Offset Correction: v_off_ini = 392.299 uV, v_off_fin = 405.784 uV 19:37:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -260.104 uVs 19:37:55 Integrated Voltage Measurement... 19:38:04 Voltage Offset Correction: v_off_ini = 394.691 uV, v_off_fin = 412.023 uV 19:38:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.889 uVs 19:38:06 I2Y Average and RMS of Measurements VT = -247.132 +- 7.822 uVs I2Y = -22.736 +- 0.720 uTm^2 19:38:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:38:24 Integrated Voltage Measurement... 19:38:33 Voltage Offset Correction: v_off_ini = 379.154 uV, v_off_fin = 389.804 uV 19:38:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.989 uVs 19:38:33 Integrated Voltage Measurement... 19:38:42 Voltage Offset Correction: v_off_ini = 405.713 uV, v_off_fin = 385.795 uV 19:38:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.829 uVs 19:38:42 Integrated Voltage Measurement... 19:38:51 Voltage Offset Correction: v_off_ini = 395.133 uV, v_off_fin = 380.930 uV 19:38:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.479 uVs 19:38:51 Integrated Voltage Measurement... 19:39:00 Voltage Offset Correction: v_off_ini = 407.563 uV, v_off_fin = 384.730 uV 19:39:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.168 uVs 19:39:00 Integrated Voltage Measurement... 19:39:09 Voltage Offset Correction: v_off_ini = 382.959 uV, v_off_fin = 367.996 uV 19:39:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.565 uVs 19:39:09 Integrated Voltage Measurement... 19:39:18 Voltage Offset Correction: v_off_ini = 382.510 uV, v_off_fin = 383.565 uV 19:39:18 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.943 uVs 19:39:22 I1X Average and RMS of Measurements VT = -95.329 +- 5.173 uVs I1X = 19.066 +- 1.035 uTm 19:39:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:39:29 Integrated Voltage Measurement... 19:39:38 Voltage Offset Correction: v_off_ini = 405.735 uV, v_off_fin = 390.028 uV 19:39:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.767 uVs 19:39:38 Integrated Voltage Measurement... 19:39:47 Voltage Offset Correction: v_off_ini = 386.823 uV, v_off_fin = 406.194 uV 19:39:47 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5002 mm dXY = -0.5000 mm VT = 216.962 uVs 19:39:47 Integrated Voltage Measurement... 19:39:56 Voltage Offset Correction: v_off_ini = 412.153 uV, v_off_fin = 418.630 uV 19:39:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -214.293 uVs 19:39:56 Integrated Voltage Measurement... 19:40:05 Voltage Offset Correction: v_off_ini = 417.202 uV, v_off_fin = 432.918 uV 19:40:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.083 uVs 19:40:05 Integrated Voltage Measurement... 19:40:14 Voltage Offset Correction: v_off_ini = 419.130 uV, v_off_fin = 401.821 uV 19:40:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.014 uVs 19:40:14 Integrated Voltage Measurement... 19:40:23 Voltage Offset Correction: v_off_ini = 421.765 uV, v_off_fin = 447.045 uV 19:40:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 194.413 uVs 19:40:27 I1Y Average and RMS of Measurements VT = -209.755 +- 2.859 uVs I1Y = -41.947 +- 0.572 uTm 19:40:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:40:34 Integrated Voltage Measurement... 19:40:43 Voltage Offset Correction: v_off_ini = 413.180 uV, v_off_fin = 397.263 uV 19:40:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -121.321 uVs 19:40:44 Integrated Voltage Measurement... 19:40:52 Voltage Offset Correction: v_off_ini = 409.224 uV, v_off_fin = 422.183 uV 19:40:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.254 uVs 19:40:53 Integrated Voltage Measurement... 19:41:01 Voltage Offset Correction: v_off_ini = 411.203 uV, v_off_fin = 401.438 uV 19:41:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -125.475 uVs 19:41:02 Integrated Voltage Measurement... 19:41:11 Voltage Offset Correction: v_off_ini = 413.957 uV, v_off_fin = 419.195 uV 19:41:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.069 uVs 19:41:11 Integrated Voltage Measurement... 19:41:20 Voltage Offset Correction: v_off_ini = 410.353 uV, v_off_fin = 389.577 uV 19:41:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -124.337 uVs 19:41:20 Integrated Voltage Measurement... 19:41:29 Voltage Offset Correction: v_off_ini = 402.886 uV, v_off_fin = 417.974 uV 19:41:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.499 uVs 19:41:31 I2X Average and RMS of Measurements VT = -113.159 +- 4.966 uVs I2X = 10.411 +- 0.457 uTm^2 19:41:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:41:39 Integrated Voltage Measurement... 19:41:48 Voltage Offset Correction: v_off_ini = 407.190 uV, v_off_fin = 409.551 uV 19:41:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.565 uVs 19:41:48 Integrated Voltage Measurement... 19:41:57 Voltage Offset Correction: v_off_ini = 401.909 uV, v_off_fin = 398.782 uV 19:41:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.957 uVs 19:41:57 Integrated Voltage Measurement... 19:42:06 Voltage Offset Correction: v_off_ini = 403.382 uV, v_off_fin = 394.545 uV 19:42:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.920 uVs 19:42:06 Integrated Voltage Measurement... 19:42:15 Voltage Offset Correction: v_off_ini = 402.151 uV, v_off_fin = 416.181 uV 19:42:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 242.856 uVs 19:42:15 Integrated Voltage Measurement... 19:42:24 Voltage Offset Correction: v_off_ini = 404.494 uV, v_off_fin = 389.453 uV 19:42:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -229.497 uVs 19:42:24 Integrated Voltage Measurement... 19:42:33 Voltage Offset Correction: v_off_ini = 385.258 uV, v_off_fin = 404.912 uV 19:42:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.826 uVs 19:42:35 I2Y Average and RMS of Measurements VT = -238.103 +- 4.740 uVs I2Y = -21.906 +- 0.436 uTm^2 19:42:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm