Coil Constant for 1DQB26 2 Main Coils using magnet Sext TB05SX1 New Calibrated Coil Constant = 158620.73975, File coilpar.ru (Old) Coil Constant = 151244.86090 New Constant/Old Constant = 1.0488, Old Coil Radius = 0.0127000, New Radius = 0.0123667, Number of coils = 2 (Only change radius if Number of coils = 1!) Wire Strength = 10.278539 T/m, Coil Strength = 9.800586 T/m, Magnet Type is Sextupole at Imag = 7.000 for coil. Number of Coil Turns = 10 Coil Constant Calculated from radius, turns and magnet order = 146456.9858 (Should be within 5% of Calibrated Coil Constant)