SLAC Magnetic Measurements Date: 09-09-2015 Time: 15:45:22 Project: LCLS Magnet Type: Phase_Shifter Magnet Name: Phase_Shifter_Delta Serial Number: Run Number: 49 Measurement Device: 10wire Operator: sda Comment: By integral at 14 mm gap with 5.35 mm width shims whose centers are set 10 mm from centerline. Shim thickness 0.020" Test Currents for Main (A): 0.00 0.00 0.00 15:45:41 Move Stages, x_position = 0.000000 m, y_position = 0.000000 m 15:45:41 X and Y Stages Zeroed 15:45:41 MoveXY Devices: Move Device = ESP300, Read Device = ESP300 15:46:04 Ambient Measurement Device(s) Being Used, type = None 15:46:04 Magnet Voltage Measurement Device(s) Being Used, type = None 15:46:04 Temperature Measurement Device(s) Being Used, type = None 15:46:04 XMAG Device(s) Being Used, type = None 15:46:04 Ambient Measurement Device(s) Being Used, type = None 15:46:22 Move Stages, x_position = 0.000000 m, y_position = 0.000000 m 15:46:22 8` Voltage Measurement, V = 0.000000 V 15:46:22 8` Voltage Measurement, V = 0.000000 V 15:46:22 Magnet Test Parameter Measurements P0 = Main , Value = 0.00 P1 = Magnet voltage (V), Value = 0.00 15:46:22 Integrated Voltage Measurement... 15:46:37 Voltage Offset Correction: v_off_ini = 0.000001077 V, v_off_fin = -0.000000933 V 15:46:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003355 Vs 15:46:52 Voltage Offset Correction: v_off_ini = 0.000000581 V, v_off_fin = 0.000001391 V 15:46:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003150 Vs 15:47:07 Voltage Offset Correction: v_off_ini = 0.000000622 V, v_off_fin = -0.000001648 V 15:47:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003429 Vs 15:47:21 Voltage Offset Correction: v_off_ini = 0.000001652 V, v_off_fin = 0.000003649 V 15:47:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003424 Vs 15:47:36 Voltage Offset Correction: v_off_ini = 0.000001714 V, v_off_fin = 0.000000896 V 15:47:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003081 Vs 15:47:51 Voltage Offset Correction: v_off_ini = 0.000002097 V, v_off_fin = 0.000004318 V 15:47:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003716 Vs 15:48:06 Voltage Offset Correction: v_off_ini = 0.000002131 V, v_off_fin = -0.000000403 V 15:48:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003548 Vs 15:48:21 Voltage Offset Correction: v_off_ini = 0.000002487 V, v_off_fin = 0.000003717 V 15:48:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003257 Vs 15:48:21 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003370 +- 0.000000069 Vs 15:48:21 First Field Integral, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0040000 m, Nturns = 10.0 VT = 0.000003370 +- 0.000000069 Vs FI1 = 0.00008425 +- 0.00000172 Tm 15:48:28 Integrated Voltage Measurement... 15:48:38 Voltage Offset Correction: v_off_ini = 0.000002872 V, v_off_fin = 0.000000981 V 15:48:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000002839 Vs 15:48:45 Voltage Offset Correction: v_off_ini = 0.000002595 V, v_off_fin = 0.000005114 V 15:48:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000002849 Vs 15:48:52 Voltage Offset Correction: v_off_ini = 0.000002944 V, v_off_fin = -0.000000269 V 15:48:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000002959 Vs 15:48:59 Voltage Offset Correction: v_off_ini = 0.000002935 V, v_off_fin = 0.000004607 V 15:48:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000002517 Vs 15:49:06 Voltage Offset Correction: v_off_ini = 0.000002621 V, v_off_fin = -0.000000946 V 15:49:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003047 Vs 15:49:13 Voltage Offset Correction: v_off_ini = 0.000002545 V, v_off_fin = 0.000005009 V 15:49:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000002967 Vs 15:49:20 Voltage Offset Correction: v_off_ini = 0.000002649 V, v_off_fin = -0.000002563 V 15:49:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003555 Vs 15:49:27 Voltage Offset Correction: v_off_ini = 0.000002726 V, v_off_fin = 0.000003565 V 15:49:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000002645 Vs 15:49:27 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000002922 +- 0.000000140 Vs 15:49:27 Second Field Integral, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0040000 m, Nturns = 10.0 VT = 0.000002922 +- 0.000000140 Vs FI2 = 0.00004456 +- 0.00000214 Tm 15:49:34 8` Voltage Measurement, V = 0.000000 V 15:49:34 8` Voltage Measurement, V = 0.000000 V 15:49:34 Magnet Test Parameter Measurements P0 = Main , Value = 0.00 P1 = Magnet voltage (V), Value = 0.00 15:49:34 Integrated Voltage Measurement... 15:49:49 Voltage Offset Correction: v_off_ini = 0.000003226 V, v_off_fin = 0.000001242 V 15:49:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003443 Vs 15:50:04 Voltage Offset Correction: v_off_ini = 0.000002850 V, v_off_fin = 0.000004470 V 15:50:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003294 Vs 15:50:19 Voltage Offset Correction: v_off_ini = 0.000002179 V, v_off_fin = 0.000000968 V 15:50:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003157 Vs 15:50:34 Voltage Offset Correction: v_off_ini = 0.000002399 V, v_off_fin = 0.000004274 V 15:50:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003507 Vs 15:50:50 Voltage Offset Correction: v_off_ini = 0.000002155 V, v_off_fin = -0.000000489 V 15:50:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003633 Vs 15:51:05 Voltage Offset Correction: v_off_ini = 0.000002516 V, v_off_fin = 0.000004364 V 15:51:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003444 Vs 15:51:21 Voltage Offset Correction: v_off_ini = 0.000002556 V, v_off_fin = 0.000001252 V 15:51:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003390 Vs 15:51:37 Voltage Offset Correction: v_off_ini = 0.000001987 V, v_off_fin = 0.000004909 V 15:51:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003964 Vs 15:51:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003479 +- 0.000000138 Vs 15:51:37 First Field Integral, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0040000 m, Nturns = 10.0 VT = 0.000003479 +- 0.000000138 Vs FI1 = 0.00008697 +- 0.00000346 Tm 15:51:44 8` Voltage Measurement, V = 0.000000 V 15:51:44 8` Voltage Measurement, V = 0.000000 V 15:51:44 Magnet Test Parameter Measurements P0 = Main , Value = 0.00 P1 = Magnet voltage (V), Value = 0.00 15:51:45 Integrated Voltage Measurement... 15:51:59 Voltage Offset Correction: v_off_ini = 0.000001262 V, v_off_fin = -0.000000961 V 15:52:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003515 Vs 15:52:15 Voltage Offset Correction: v_off_ini = 0.000001196 V, v_off_fin = 0.000003529 V 15:52:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003446 Vs 15:52:30 Voltage Offset Correction: v_off_ini = 0.000001468 V, v_off_fin = -0.000000181 V 15:52:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003340 Vs 15:52:45 Voltage Offset Correction: v_off_ini = 0.000002316 V, v_off_fin = 0.000004691 V 15:52:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003519 Vs 15:53:03 Voltage Offset Correction: v_off_ini = 0.000001582 V, v_off_fin = -0.000001178 V 15:53:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003608 Vs 15:53:18 Voltage Offset Correction: v_off_ini = 0.000001783 V, v_off_fin = 0.000003744 V 15:53:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003093 Vs 15:53:33 Voltage Offset Correction: v_off_ini = 0.000002032 V, v_off_fin = -0.000000203 V 15:53:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003487 Vs 15:53:48 Voltage Offset Correction: v_off_ini = 0.000002241 V, v_off_fin = 0.000004647 V 15:53:48 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000003476 Vs 15:53:48 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000003436 +- 0.000000053 Vs 15:53:48 First Field Integral, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0040000 m, Nturns = 10.0 VT = 0.000003436 +- 0.000000053 Vs FI1 = 0.00008589 +- 0.00000134 Tm 15:54:14 Move Stages, x_position = 0.000000 m, y_position = 0.000000 m