SLAC Magnetic Measurements Date: 09-09-2015 Time: 15:35:27 Project: LCLS Magnet Type: Phase_Shifter Magnet Name: Phase_Shifter_Delta Serial Number: Run Number: 48 Measurement Device: 10wire Operator: sda Comment: By integral at 17 mm gap with 5.35 mm width shims whose centers are set 10 mm from centerline. Shim thickness 0.020" Test Currents for Main (A): 0.00 0.00 0.00 15:35:47 Move Stages, x_position = 0.000000 m, y_position = 0.000000 m 15:35:47 X and Y Stages Zeroed 15:35:47 MoveXY Devices: Move Device = ESP300, Read Device = ESP300 15:36:09 Ambient Measurement Device(s) Being Used, type = None 15:36:10 Magnet Voltage Measurement Device(s) Being Used, type = None 15:36:10 Temperature Measurement Device(s) Being Used, type = None 15:36:10 XMAG Device(s) Being Used, type = None 15:36:10 Ambient Measurement Device(s) Being Used, type = None 15:36:28 Move Stages, x_position = 0.000000 m, y_position = 0.000000 m 15:36:28 8` Voltage Measurement, V = 0.000000 V 15:36:28 8` Voltage Measurement, V = 0.000000 V 15:36:28 Magnet Test Parameter Measurements P0 = Main , Value = 0.00 P1 = Magnet voltage (V), Value = 0.00 15:36:28 Integrated Voltage Measurement... 15:36:43 Voltage Offset Correction: v_off_ini = 0.000002709 V, v_off_fin = 0.000000456 V 15:36:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001371 Vs 15:36:58 Voltage Offset Correction: v_off_ini = 0.000002637 V, v_off_fin = 0.000004866 V 15:36:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001376 Vs 15:37:13 Voltage Offset Correction: v_off_ini = 0.000002302 V, v_off_fin = 0.000000333 V 15:37:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001359 Vs 15:37:28 Voltage Offset Correction: v_off_ini = 0.000001943 V, v_off_fin = 0.000003149 V 15:37:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001241 Vs 15:37:43 Voltage Offset Correction: v_off_ini = 0.000002412 V, v_off_fin = -0.000000106 V 15:37:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001380 Vs 15:37:58 Voltage Offset Correction: v_off_ini = 0.000002080 V, v_off_fin = 0.000004309 V 15:37:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001413 Vs 15:38:13 Voltage Offset Correction: v_off_ini = 0.000002676 V, v_off_fin = 0.000000519 V 15:38:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001388 Vs 15:38:27 Voltage Offset Correction: v_off_ini = 0.000002544 V, v_off_fin = 0.000003018 V 15:38:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001056 Vs 15:38:27 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001323 +- 0.000000068 Vs 15:38:27 First Field Integral, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0040000 m, Nturns = 10.0 VT = 0.000001323 +- 0.000000068 Vs FI1 = 0.00003307 +- 0.00000171 Tm 15:38:35 Integrated Voltage Measurement... 15:38:44 Voltage Offset Correction: v_off_ini = 0.000001818 V, v_off_fin = -0.000000444 V 15:38:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001630 Vs 15:38:51 Voltage Offset Correction: v_off_ini = 0.000001431 V, v_off_fin = 0.000003399 V 15:38:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001615 Vs 15:38:58 Voltage Offset Correction: v_off_ini = 0.000001009 V, v_off_fin = -0.000000714 V 15:38:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001598 Vs 15:39:05 Voltage Offset Correction: v_off_ini = 0.000001057 V, v_off_fin = 0.000002799 V 15:39:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001438 Vs 15:39:12 Voltage Offset Correction: v_off_ini = 0.000001211 V, v_off_fin = -0.000001107 V 15:39:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001564 Vs 15:39:19 Voltage Offset Correction: v_off_ini = 0.000001188 V, v_off_fin = 0.000003597 V 15:39:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001891 Vs 15:39:26 Voltage Offset Correction: v_off_ini = 0.000001398 V, v_off_fin = -0.000001031 V 15:39:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001684 Vs 15:39:33 Voltage Offset Correction: v_off_ini = 0.000001491 V, v_off_fin = 0.000003287 V 15:39:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001549 Vs 15:39:33 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001621 +- 0.000000074 Vs 15:39:33 Second Field Integral, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0040000 m, Nturns = 10.0 VT = 0.000001621 +- 0.000000074 Vs FI2 = 0.00002472 +- 0.00000113 Tm 15:39:41 8` Voltage Measurement, V = 0.000000 V 15:39:41 8` Voltage Measurement, V = 0.000000 V 15:39:41 Magnet Test Parameter Measurements P0 = Main , Value = 0.00 P1 = Magnet voltage (V), Value = 0.00 15:39:41 Integrated Voltage Measurement... 15:39:56 Voltage Offset Correction: v_off_ini = 0.000002050 V, v_off_fin = -0.000001009 V 15:39:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001534 Vs 15:40:11 Voltage Offset Correction: v_off_ini = 0.000002024 V, v_off_fin = 0.000003855 V 15:40:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001104 Vs 15:40:26 Voltage Offset Correction: v_off_ini = 0.000001599 V, v_off_fin = -0.000001117 V 15:40:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001362 Vs 15:40:40 Voltage Offset Correction: v_off_ini = 0.000002686 V, v_off_fin = 0.000004288 V 15:40:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001112 Vs 15:40:55 Voltage Offset Correction: v_off_ini = 0.000002621 V, v_off_fin = 0.000001385 V 15:40:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001091 Vs 15:41:10 Voltage Offset Correction: v_off_ini = 0.000002265 V, v_off_fin = 0.000004459 V 15:41:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001436 Vs 15:41:25 Voltage Offset Correction: v_off_ini = 0.000001456 V, v_off_fin = -0.000000482 V 15:41:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001224 Vs 15:41:40 Voltage Offset Correction: v_off_ini = 0.000001267 V, v_off_fin = 0.000002482 V 15:41:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001126 Vs 15:41:40 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001249 +- 0.000000052 Vs 15:41:40 First Field Integral, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0040000 m, Nturns = 10.0 VT = 0.000001249 +- 0.000000052 Vs FI1 = 0.00003122 +- 0.00000130 Tm 15:41:47 8` Voltage Measurement, V = 0.000000 V 15:41:47 8` Voltage Measurement, V = 0.000000 V 15:41:47 Magnet Test Parameter Measurements P0 = Main , Value = 0.00 P1 = Magnet voltage (V), Value = 0.00 15:41:48 Integrated Voltage Measurement... 15:42:02 Voltage Offset Correction: v_off_ini = -0.000000282 V, v_off_fin = -0.000004052 V 15:42:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001762 Vs 15:42:17 Voltage Offset Correction: v_off_ini = -0.000000032 V, v_off_fin = 0.000002049 V 15:42:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001288 Vs 15:42:32 Voltage Offset Correction: v_off_ini = 0.000000484 V, v_off_fin = -0.000001115 V 15:42:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001343 Vs 15:42:47 Voltage Offset Correction: v_off_ini = 0.000000558 V, v_off_fin = 0.000002336 V 15:42:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001195 Vs 15:43:02 Voltage Offset Correction: v_off_ini = 0.000000303 V, v_off_fin = -0.000001791 V 15:43:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001313 Vs 15:43:17 Voltage Offset Correction: v_off_ini = 0.000000258 V, v_off_fin = 0.000002583 V 15:43:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001527 Vs 15:43:32 Voltage Offset Correction: v_off_ini = 0.000000784 V, v_off_fin = -0.000000798 V 15:43:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001105 Vs 15:43:46 Voltage Offset Correction: v_off_ini = 0.000001371 V, v_off_fin = 0.000003010 V 15:43:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0040000 m VT = -0.000001164 Vs 15:43:46 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0040000 m VT = 0.000001337 +- 0.000000148 Vs 15:43:46 First Field Integral, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0040000 m, Nturns = 10.0 VT = 0.000001337 +- 0.000000148 Vs FI1 = 0.00003343 +- 0.00000371 Tm 15:44:12 Move Stages, x_position = 0.000000 m, y_position = 0.000000 m