SLAC Magnetic Measurements Date: 04-23-2014 Time: 15:55:19 Project: LCLS Device Type: Hall probe Data Type: Serial Number: SENTRON_441 Run Number: 105 Measurement Type: Reference magnet Operator: YL Comment: Run 89; Gain drift, run 41; old calibration 5.754023 -0.369567 0.067707 -0.369569 0.000182 -0.369571 5.753870 -0.369571 0.067626 -0.369571 0.000270 -0.369569 Field = -0.368917 Not adjusted for T