Quadrupole
Magnet Checks |
|
|
|
||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Date: |
|
Magnet: |
|
|
|
Operator(s): |
|
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
|
Notes: |
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||
|
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
|
|
|
Magnetic
Fiducial Coordinates: (inches) |
|
|
|
|
|
||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Fiducial |
Z |
X |
Y |
|
|
|
|
|
|
|
-9.4561 |
11.2444 |
16.2045 |
|
|
|
|
|
|
|
|
9.4586 |
11.2517 |
16.2049 |
|
|
|
|
|
|
|
|
-9.4575 |
-11.2503 |
16.2041 |
|
|
Offset: |
|
|
|
|
|
9.4572 |
-11.2577 |
16.2041 |
|
|
0.0000 |
inches |
|
|
|
|
-8.3617 |
-15.2047 |
-4.7037 |
|
|
|
|
|
|
|
|
9.2153 |
-15.2041 |
-4.7173 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Description: |
|
|
|
|
|
|
|
|
|
|
Fiducial
values based on the measured mechanical center line. |
|
|||||||
|
|
|
|
|
|
|
|
|
|
|
Quadrupole
Mechanical Centerline Checks:
Mandrel Check (4 Positions): |
X (in) |
Y (in) |
-0.0001 |
0.0005 |
|
-0.0017 |
0.0032 |
|
-0.0010 |
0.0035 |
|
0.0010 |
0.0012 |
Mandrel Check (Position 1): |
X (in) |
Y (in) |
-0.0007 |
0.0003 |
|
-0.0016 |
0.0003 |
|
-0.0013 |
0.0003 |
|
NA |
NA |
|
NA |
NA |
|
|
Gap Measurements: |
|
Upstream (in) |
Downstream (in) |
Approval __________________
|
Quadrupole Magnet Fiducialization Traveller |
Approval must be obtained before removing the magnet from the test stand. |
Magnetic Measurements Approval by – Jack Tanabe or Tony King. |
IHEP Magnetic Measurements Operator: ____________ Date:_____________ |
Measured Magnetic Center Offset Dx =__________ mm Dy =___________ mm |
Measured at: Integrated Field: ____________ T-m @ Current_____________ Amps |
SLAC Magnetic Measurements Operator: ____________ Date:_____________ |
Measured Magnetic Center Offset Dx =__________ mm Dy =___________ mm |
Measured at: Integrated Field: ____________ T-m @ Current_____________ Amps |
Fiducialization: |
Fiducial – Measured Coordinates |
z |
x |
y |
(in) |
(in) |
(in) |
|
-9.4561 |
11.2444 |
16.2045 |
|
9.4586 |
11.2517 |
16.2049 |
|
-9.4575 |
-11.2503 |
16.2041 |
|
9.4572 |
-11.2577 |
16.2041 |
|
-8.3617 |
-15.2047 |
-4.7037 |
|
9.2153 |
-15.2041 |
-4.7173 |
(For x, y corrected coordinates, use SLAC data if available. If not, use IHEP data.)
Fiducial – Corrected Coordinates |
z |
x |
y |
(in) |
(in) |
(in) |
|
TB1 |
|
|
|
TB2 |
|
|
|
TB3 |
|
|
|
TB4 |
|
|
|
TB5 |
|
|
|
TB6 |
|
|
|
Check Measurements:
Rod |
xmeasured |
xnominal |
ymeasured |
ynominal |
10.4970 |
10.5 in |
14.8132 |
14.8 in |
|
10.4961 |
10.5 in |
14.8134 |
14.8 in |
|
-10.4973 |
-10.5 in |
14.8151 |
14.8 in |
|
-10.4978 |
-10.5 in |
14.8151 |
14.8 in |
Cylinder at Different Positions
Position 1 (x,y)=(0,0) |
x (in) |
y (in) |
-0.0017 |
0.0027 |
|
-0.0009 |
0.0030 |
|
0.0010 |
0.0007 |
Approval __________________